| Literature DB >> 32818097 |
Ann L Webber1, Thomas R Mandall1, Darcy T Molloy1, Lucas J Lister1, Eileen E Birch2,3.
Abstract
Purpose: To describe and evaluate an iOS application suppression test, Worth 4 Dot App (W4DApp), which was designed and developed to assess size and depth of suppression.Entities:
Keywords: Worth 4 Dot; binocular vision; suppression
Mesh:
Year: 2020 PMID: 32818097 PMCID: PMC7396169 DOI: 10.1167/tvst.9.4.9
Source DB: PubMed Journal: Transl Vis Sci Technol ISSN: 2164-2591 Impact factor: 3.283
Figure 1.Screen shot example of W4DApp display.
Color and Luminance of W4D Stimulus
| Luminance (cd/m2) | |||||||
|---|---|---|---|---|---|---|---|
| W4DApp Dots | Flashlight Worth 4 Dot | ||||||
| Hue, Saturation, Brightness | Unfiltered | With Red Filter | With Green Filter | Unfiltered | With Red Filter | With Green Filter | |
| White | 0,0,0 | 425 | 63 | 69 | 620 | 12 | 55 |
| Green | 120,100,100 | 325 | 0.2 | 65 | 170 | 0.03 | 46 |
| Red | 360,100,100 | 85 | 8.9 | 0.3 | 30 | 5 | 0.06 |
| Black surround | 0,0,0 | 0.8 | 0.7 | ||||
Figure 2.Flashlight W4D Dot parameters (Richmond Products).
Classification of Fusion by iOS W4DApp and by Flashlight W4D (Number of Participants)
| Flashlight W4D at Different Viewing Distances | ||||
|---|---|---|---|---|
| Bifoveal | Macular | Peripheral | No Fusion | |
| W4DApp @ 33 cm | ||||
| Bifoveal | 11 | 3 | 0 | 0 |
| Macular | 0 | 4 | 2 | 1 |
| Peripheral | 0 | 0 | 3 | 0 |
| No fusion | 0 | 0 | 0 | 1 |
Reliability: Test-retest ICC Values
| Test | Single Measures ICC (2-way Random) | 95% Confidence Intervals |
|---|---|---|
| Size Flashlight W4D | 0.946 | 0.855 – 0.985 |
| Size W4DApp (33 cm) | 0.968 | 0.907 – 0.992 |
| Size W4DApp (90 cm) | 0.978 | 0.933 – 0.995 |
| Depth Flashlight W4D | 0.433 | −0.40 – 0.829 |
| Depth W4DApp | 0.901 | 0.749 – 0.972 |
Figure 3.Scatterplots showing suppression zone size determined by both Flashlight W4D and W4DApp for individual participants at both 90 cm (a) and 33 cm (b). The dashed line represents absolute agreement and the solid line is the best-fit linear regression line.
Figure 4.Scatterplots showing suppression depth determined by W4DApp, Flashlight W4D + NDF and Contrast Balance Index for individual participants. The dashed line represents absolute agreement between methods and the solid line is the best-fit linear regression line for relationship between methods.
| Classification | Age (y) | Right eye visual acuity (logMAR) | Left eye visual acuity (logMAR) | Interocular difference in visual acuity (logMAR) | Randot Stereoacuity(arcsec) | Flashlight W4D at 6 m | Amblyopic treatment |
|---|---|---|---|---|---|---|---|
| Anisometropic | 21 | 0.00 | 0.00 | 0.00 | 60 | RE Suppression | |
| Anisometropic | 37 | −0.18 | −0.10 | 0.08 | 40 | Normal | |
| Anisometropic amblyope | 24 | 0.00 | 0.40 | 0.40 | 100 | RE Suppression | |
| Anisometropic amblyope | 13 | 0.70 | 0.00 | 0.70 | 100 | RE suppression | 3D computer training |
| Strabismic | 20 | −0.02 | −0.02 | 0.00 | 40 | Normal | |
| Strabismic | 46 | 0.00 | 0.00 | 0.00 | 400 | Normal | |
| Strabismic | 21 | −0.10 | −0.08 | 0.02 | 40 | Normal | |
| Strabismic | 61 | 0.00 | −0.02 | 0.02 | Nil | RE suppression | |
| Strabismic | 47 | 0.08 | 0.12 | 0.04 | Nil | LE suppression | |
| Strabismic | 24 | −0.02 | 0.04 | 0.06 | 400 | LE suppression | EOM surgery, patching |
| Strabismic | 21 | −0.08 | 0.00 | 0.08 | Nil | LE suppression | EOM surgery |
| Strabismic | 48 | 0.04 | 0.14 | 0.10 | Nil | LE suppression | Patching |
| Strabismic | 20 | 0.02 | −0.10 | 0.12 | 60 | Normal | EOM surgery, patching |
| Strabismic amblyope | 20 | 0.20 | 0.00 | 0.20 | Nil | RE suppression | EOM surgery, patching |
| Strabismic amblyope | 20 | 0.00 | 0.20 | 0.20 | Nil | LE suppression | |
| Strabismic amblyope | 23 | 0.00 | 0.40 | 0.40 | Nil | LE suppression | EOM Surgery, Patching |
| Strabismic and anisometropic | 15 | 0.00 | −0.08 | 0.08 | Nil | RE suppression | Patching, atropine drops |
| Strabismic and anisometropic amblyope | 12 | 0.40 | 0.12 | 0.28 | Nil | RE suppression | Patching |
| Strabismic and anisometropic amblyope | 47 | 0.84 | −0.10 | 0.94 | Nil | RE suppression | Patching |
Participants are grouped by classification and sorted by increasing interocular difference in visual acuity.