Detection of SARS-CoV-2 IgG Targeting Nucleocapsid or Spike Protein by Four High-Throughput Immunoassays Authorized for Emergency Use.

Harry E Prince1, Tara S Givens2, Mary Lapé-Nixon1, Nigel J Clarke1, Dale A Schwab1, Hollis J Batterman1, Robert S Jones2, William A Meyer3, Hema Kapoor4, Charles M Rowland1, Farnoosh Haji-Sheikhi1, Elizabeth M Marlowe5.   

Abstract

A total of 1,200 serum samples that were tested for SARS-CoV-2 IgG antibody using the Abbott Architect immunoassay targeting the nucleocapsid protein were run in 3 SARS-CoV-2 IgG immunoassays targeting spike proteins (DiaSorin Liaison, Ortho Vitros, and Euroimmun). Consensus-positive and consensus-negative interpretations were defined as qualitative agreement in at least 3 of the 4 assays. Agreement of the 4 individual assays with a consensus-negative interpretation (n = 610) ranged from 96.7% to 100%, and agreement with a consensus-positive interpretation (n = 584) ranged from 94.3% to 100%. Laboratory-developed inhibition assays were utilized to evaluate 49 consensus-negative samples that were positive in only one assay; true-positive reactivity was confirmed in only 2 of these 49 (4%) samples. These findings demonstrate very high levels of agreement among 4 SARS-CoV-2 IgG assays authorized for emergency use, regardless of antigen target or assay format. Although false-positive reactivity was identified, its occurrence was rare (no more than 1.7% of samples for a given assay).
Copyright © 2020 American Society for Microbiology.

Entities:  

Keywords:  SARS-CoV-2 IgG; coronavirus; immunoassay

Mesh:

Substances:

Year:  2020        PMID: 32817144      PMCID: PMC7587121          DOI: 10.1128/JCM.01742-20

Source DB:  PubMed          Journal:  J Clin Microbiol        ISSN: 0095-1137            Impact factor:   5.948


INTRODUCTION

As of 14 August 2020, over 21 million cases of coronavirus disease 2019 (COVID-19) caused by severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) have been identified worldwide, with more than 750,000 infection-related deaths (https://coronavirus.jhu.edu). The primary laboratory tool for diagnosing SARS-CoV-2 infection is detection of viral RNA by nucleic acid amplification testing (NAAT) (1). While antibody testing is not currently used to diagnose acute COVID-19 infections, assays for SARS-CoV-2 antibodies are proving important for (i) identifying some acutely infected individuals, particularly those presenting more than 1 to 2 weeks after symptom onset with negative RNA results, (ii) identifying persons who were exposed to the virus (including those with asymptomatic infection) and mounted a measurable antibody response, (iii) evaluating recovered individuals for donation of convalescent-phase plasma, and (iv) assessing prevalence of infection within community settings (https://cdc.gov/coronavirus/2019-ncov/hcp/testing-overview.html). In the future, quantitative SARS-CoV-2 IgG assays may play an important role in assessing vaccine efficacy (2, 3). Whether an IgG antibody response following infection or vaccination translates to protective immunity is still unknown (3). Many SARS-CoV-2 antibody assays have recently received emergency use authorization (EUA) from the U.S. Food and Drug Administration (https://www.fda.gov/medical-devices/emergency-situations-medical-devices/eua-authorized-serology-test-performance). However, the extent of concordance among the various EUA assays is still being defined; in particular, it is unclear how assays measuring antibodies to the SARS-CoV-2 spike protein compare to assays measuring antibodies to the SARS-CoV-2 nucleocapsid protein (3). Such information is crucial to assisting clinicians and laboratory scientists in evaluating and interpreting results generated by different laboratories that utilize multiple EUA platforms. Here, we compare results obtained using 4 high-throughput SARS-CoV-2 IgG EUA immunoassays; samples positive or negative in an assay measuring IgG to viral nucleocapsid were tested in 3 assays measuring IgG recognizing viral spike protein subunits S1 or both S1 and S2 (S1+S2).

MATERIALS AND METHODS

Samples.

Serum samples with positive (n = 600) or negative (n = 600) SARS-CoV-2 IgG results from the Abbott chemiluminescent microparticle immunoassay (CMIA) utilizing nucleocapsid protein as the antigen target (referred to here as the nucleocapsid-based assay) were selected for further analysis in 3 SARS-CoV-2 IgG assays utilizing spike protein(s) as the antigen target (referred to here as spike-based assays). The 600 deidentified Abbott-positive samples were tested in the Abbott assay at Quest Diagnostics Nichols Institute, Chantilly, VA, and consisted of consecutive positive samples with enough remaining volume for further testing. After testing in the Abbott assay, these samples were frozen and shipped on dry ice to Quest Diagnostics Infections Disease (QDID), San Juan Capistrano, CA; there they were thawed and tested in the other 3 SARS-CoV-2 IgG assays within 24 h of thawing. The 600 deidentified Abbott-negative samples were tested in the Abbott assay at QDID and consisted of consecutive negative samples with enough remaining volume for further testing; the samples were refrigerated a maximum of 48 h before being tested in the other 3 assays. The median age of patients contributing the Abbott-positive samples was 47 years, and 48% were male; the median age of the patients contributing the Abbott-negative samples was 52 years, and 42% were male. Clinical findings and SARS-CoV-2 RNA results were not available for any of the 1,200 patients.

Immunoassays.

In addition to the Abbott CMIA, three SARS-CoV-2 IgG immunoassays (2 chemiluminescent immunoassays [CIAs] and 1 enzyme-linked immunosorbent assay [ELISA]) were evaluated (Table 1). All 4 assays were performed following the instructions for use supplied by the respective manufacturers. For the 3 chemiluminescent assays, undiluted serum samples are placed inside the respective instruments (Table 1), and all assay steps are performed inside the instrument. In contrast, for the Euroimmun ELISA, samples are diluted 1:101 in kit-supplied sample diluent, and the diluted sample is added to reaction wells. Results were classified as positive or negative following the interpretive criteria supplied in the manufacturer’s instructions, with one exception; samples with a Euroimmun equivocal index (0.8 to 1.0) were interpreted as positive for purposes of this study.
TABLE 1

Information for the 4 SARS-CoV-2 IgG immunoassays used to test study samples

Assay namea ManufacturerMethodProtein (antigen) targetNegative indexPositive index
Architect SARS-CoV-2 IgGAbbott Laboratories Inc., Abbott Park, ILCMIANucleocapsid<1.4≥1.4
Liaison SARS-CoV-2 S1/S2 IgGDiaSorin Inc., Stillwater, MNCIASpike 1 + spike 2<15.0≥15.0
Vitros anti-SARS-CoV-2 IgGOrtho Clinical Diagnostics, Raritan, NJCIASpike 1<1.0≥1.0
Anti-SARS-CoV-2 ELISA (IgG)Euroimmun Inc., Mountain Lakes, NJELISASpike 1<0.8≥0.8b

For the CMIA and CIA methods, the assay name begins with the instrument utilized for test performance.

Per the Euroimmun instructions for use, index values of 0.8, 0.9, and 1.0 are considered equivocal; however, for the purposes of this study, equivocal results were interpreted as positive.

Information for the 4 SARS-CoV-2 IgG immunoassays used to test study samples For the CMIA and CIA methods, the assay name begins with the instrument utilized for test performance. Per the Euroimmun instructions for use, index values of 0.8, 0.9, and 1.0 are considered equivocal; however, for the purposes of this study, equivocal results were interpreted as positive.

Inhibition assays.

Three assays (Abbott, DiaSorin, and Euroimmun) were modified to distinguish true-positive (TP) from false-positive (FP) reactivity based on inhibition of reactivity by soluble SARS-CoV-2 proteins. Preliminary experiments using a sample positive in all 4 assays (presumed to be TP) established the optimal concentrations of inhibitory proteins required for each assay. For the Abbott inhibition assay, serum samples were diluted 1:2 with phosphate-buffered saline (PBS), PBS containing soluble recombinant S1 protein (6 μg/ml; GenScript, Piscataway, NJ), or PBS containing soluble nucleocapsid protein (6 μg/ml; GenScript); thus, the final concentration of S1 protein or nucleocapsid protein in analyzed samples was 3 μg/ml. Similarly, for the DiaSorin inhibition assay, sera were diluted 1:2 with PBS, PBS containing S1 protein (6 μg/ml) and S2 protein (6 μg/ml; RayBiotech, Peachtree Corners, GA), or PBS containing nucleocapsid (12 μg/ml, to match the total S1+S2 concentration of 12 μg/ml). Thus, the final concentrations of analyzed samples for the DiaSorin inhibition assay were 3 μg/ml each for S1 and S2 and 6 μg/ml for nucleocapsid. For the Euroimmun inhibition assay, samples were diluted 1:101 in sample diluent, sample diluent containing S1 protein (3 μg/ml), or sample diluent containing nucleocapsid protein (3 μg/ml). For each inhibition assay, the samples tested included the samples positive only by the comparable routine assay, plus a minimum of 9 samples that were positive in all 4 assays. Inhibition was calculated as (PBS or diluent index − nucleocapsid or spike index)/PBS or diluent index and expressed as a percentage. The cutoff for distinguishing TP from FP reactivity was defined as the mean percent inhibition value minus 2 standard deviations for the samples positive in all 4 assays that were tested in the inhibition assay; values below this cutoff were considered FP. Cutoff values were 29% for the Abbott inhibition assay, 42% for the DiaSorin inhibition assay, and 61% for the Euroimmun inhibition assay.

Analysis and statistics.

The various reactivity patterns across all 4 SARS-CoV-2 IgG assays were combined into 3 interpretation groups. Consensus-negative was defined as a negative result in at least 3 of 4 assays, and consensus-positive was defined as a positive result in at least 3 of 4 assays; the nonconsensus group included samples positive in 2 of 4 assays. All samples were classified into consensus-negative, consensus-positive, or nonconsensus categories based on the cutoffs listed in Table 1. For each of the 4 assays, positive agreement and negative agreement (and 95% confidence intervals) were calculated to compare the assay’s performance to positive consensus or negative consensus. Samples with a nonconsensus interpretation were not included in positive-agreement/negative-agreement calculations, since there was no majority positive or negative reactivity (i.e., positive in 2 assays and negative in 2 assays). Statistical analyses were conducted using R version 3.6.3, R Foundation for Statistical Computing (https://www.R-project.org).

RESULTS

Consensus interpretations.

Summarized in Table 2 are the results for the 1,200 samples across all 4 SARS-CoV-2 IgG assays. The vast majority of samples in the consensus-negative and consensus-positive groups showed complete agreement across all 4 assays (1,109/1,194 = 93% [95% confidence interval, 91.3% to 94.2%]).
TABLE 2

Distribution of results across all 4 assays and consensus interpretations

Interpretation (n)Individual assay resulta
No. of samples
AbbottDiaSorinOrthoEuroimmun
Consensus-negative (610)NegativeNegativeNegativeNegative561
NegativePositiveNegativeNegative15
NegativeNegativeNegativePositive20
PositiveNegativeNegativeNegative14
Consensus-positive (584)PositivePositivePositivePositive548
PositiveNegativePositivePositive33
PositivePositivePositiveNegative1
NegativePositivePositivePositive2
Nonconsensus (6)NegativePositiveNegativePositive2
PositiveNegativePositiveNegative2
PositiveNegativeNegativePositive2

Discrepant results are in boldface.

Distribution of results across all 4 assays and consensus interpretations Discrepant results are in boldface.

Agreement of individual assay results with consensus-negative or -positive interpretations.

Table 3 shows the agreement of results for the 4 individual assays with a consensus-negative or consensus-positive interpretation. High levels of agreement were observed; agreement with a consensus-negative interpretation ranged from 96.7% to 100%, and agreement with a consensus-positive interpretation ranged from 94.3% to 100%. Note that the Ortho assay exhibited the highest agreement (100%) with both the consensus-negative and consensus-positive interpretations.
TABLE 3

Agreement of individual SARS-CoV-2 IgG assays with negative or positive consensus interpretations

AssayConsensus negative (n = 610)
Consensus positive (n = 584)
No. agreeing% Agreement (95% confidence interval)No. agreeingAgreement (95% confidence interval)
Abbott59697.7 (96.2, 98.6)58299.7 (98.8, 99.9)
DiaSorin59597.5 (96, 98.5)55194.3 (92.2, 95.9)
Ortho610100 (99.4, 100)584100 (99.3, 100)
Euroimmun59096.7 (95, 97.9)58399.8 (99, 100)
Agreement of individual SARS-CoV-2 IgG assays with negative or positive consensus interpretations

Results of inhibition assays.

Within the consensus-negative group were 49 samples (4% of all samples tested) that were positive in only one assay (14 for Abbott, 15 for DiaSorin, and 20 for Euroimmun; no samples were positive by Ortho only). These samples, plus a minimum of 9 samples that were positive in all 4 assays (referred to here as TP samples), were tested in platform-specific inhibition assays designed to discriminate FP from TP reactive. The results are shown in Fig. 1. Of 14 samples positive only by Abbott, 12 exhibited FP reactivity, defined as an inhibition value of <29%. Similarly, all 9 evaluable samples positive only by DiaSorin exhibited FP reactivity (inhibition of <42%), and all 16 evaluable samples positive only by Euroimmun exhibited FP reactivity (inhibition of <61%); 6 DiaSorin-positive only samples and 4 Euroimmun-positive only samples could not be evaluated using the inhibition assay because they were negative when repeated in the routine assay as part of the inhibition protocol. Inhibition by the SARS-CoV-2 protein that was not the target protein in the comparable routine assay (i.e., S1 in the Abbott inhibition assay, nucleocapsid in the DiaSorin inhibition assay, and nucleocapsid in the Euroimmun inhibition assay) was ≤4%, ≤10%, and ≤16%, respectively, for all samples (data not shown). Taken together, these findings demonstrate that, based on inhibition assay results, only 2 of 49 (4%) samples positive in just one assay exhibited TP reactivity (both samples were positive only in the Abbott assay); 96% of samples initially positive in only one of the 4 assays exhibited FP reactivity.
FIG 1

Results from inhibition assays. Blue dots represent samples positive in all 4 assays (TP samples); red dots represent samples that were positive only in the indicated assay. Values for the Abbott assay represent percent inhibition by nucleocapsid, values for the DiaSorin assay represent percent inhibition by S1+S2 proteins, and values for the Euroimmun assay represent percent inhibition by S1 protein. The horizonal lines indicate the assay-specific inhibition value below which reactivity is considered FP.

Results from inhibition assays. Blue dots represent samples positive in all 4 assays (TP samples); red dots represent samples that were positive only in the indicated assay. Values for the Abbott assay represent percent inhibition by nucleocapsid, values for the DiaSorin assay represent percent inhibition by S1+S2 proteins, and values for the Euroimmun assay represent percent inhibition by S1 protein. The horizonal lines indicate the assay-specific inhibition value below which reactivity is considered FP.

DISCUSSION

The appropriate interpretation of results from SARS-CoV-2 IgG assays depends on a clear understanding of their performance characteristics and limitations. Robust IgG responses to both the spike protein found on the surface of virus particles and the nucleocapsid protein found inside the virus particle occur following SARS-CoV-2 infection (4–7). However, there are conflicting reports as to the relative sensitivities of spike-based versus nucleocapsid-based IgG assays, particularly during the first 14 days after disease onset. Liu et al. (using ELISAs) and Tang et al. (using the Euroimmun ELISA and the Abbott CIA) found that spike-based IgG assays were slightly more sensitive during this time frame (8, 9), whereas Burbelo et al. (using luciferase immunoprecipitation systems) found that nucleocapsid-based assays were more sensitive (10). By 2 weeks after symptom onset, however, comparable sensitivities are observed (6, 7, 11–14). Based on these published contradictory results for sensitivity (8–10), plus additional studies presenting contradictory data on the correlation of virus-neutralizing activity with spike antibody levels versus nucleocapsid antibody levels (15, 16), we sought to better understand how spike-based SARS-CoV-2 IgG assays perform compared to nucleocapsid-based assays. We thus selected samples previously tested using the Abbott assay targeting nucleocapsid protein to conduct our comparison study. Our results show that, when consensus interpretations were defined based on agreement in at least 3 of the 4 assays evaluated, individual assay agreements with the consensus interpretation were very high, ranging from 94% to 100%, regardless of antigen target and assay platform. These findings thus demonstrate highly comparable performance among these 4 SARS-CoV-2 IgG assays. Our findings are consistent with those of other investigators (17–20), who demonstrated good correlation when comparing multiple assays using panels of up to approximately 500 samples. To our knowledge, ours is the first study to assess IgG reactivity in these 4 high-throughput assays using well over 1,000 samples. A small number of samples within the consensus-negative group (4% of all samples) were positive in only one of the 4 assays. This pattern may represent increased assay sensitivity, decreased specificity, or a combination of the two. To discriminate among these possibilities, inhibition assays were developed and performed. The results showed that all samples positive only in the DiaSorin or Euroimmun assay exhibited FP reactivity; similarly, 12 of 14 samples positive only in the Abbott assay showed FP reactivity. Thus, the vast majority (96%) of samples that tested positive in only 1 of the 4 assays represented decreased assay specificity, rather than increased assay sensitivity. Of note, 13 of the 20 samples with FP reactivity in the Euroimmun assay exhibited index values within the manufacturer’s equivocal range (interpreted as positive for the purpose of this evaluation); however, 12 other Euroimmun-equivocal samples fell within the consensus positive group (also positive by Abbott and Ortho), indicating that some Euroimmun-equivocal samples show TP reactivity. These findings suggest that it may be appropriate to test Euroimmun-equivocal samples using a different SARS-CoV-2 IgG assay, in order to distinguish TP from FP reactivity. Other investigators (17) evaluating the Euroimmun SARS-CoV-2 IgG assay interpreted Euroimmun-equivocal results as negative rather than as positive, prompting us to question how using that interpretation option might alter our findings. Had samples with an index in the Euroimmun equivocal range been interpreted as negative, the consensus-negative group would increase by 1 sample (from 610 to 611), and the consensus-positive group would decrease by 11 samples (from 584 to 573); most notably, the size of the nonconsensus group would nearly triple, increasing from 6 to 16 samples. The impact of these shifts on consensus-negative percent agreement and consensus-positive percent agreement would be minimal; the Euroimmun consensus-negative agreement would increase from 97% to 99%, and the DiaSorin consensus-positive agreement would increase from 94% to 96%. Thus, when a binary (positive/negative) interpretation was used for Euroimmun assay results, either interpretation of Euroimmun-equivocal results led to the same conclusion: all 4 SARS-COV-2IgG assays evaluated were highly comparable, exhibiting >90% agreement with consensus results. A recognized limitation of our study is the lack of clinical information for the patients whose serum samples were evaluated. Information on viral RNA results, disease onset, and clinical course would likely help determine if the 2 patients positive only by Abbott but with TP reactivity simply mounted an antibody response to nucleocapsid earlier than a spike-based response, or if these patients for some reason failed to mount an antibody response to spike protein. Likewise, the same questions apply to the 2 patients positive in all 3 spike-based assays but negative in the nucleocapsid-based assay (see Table 2, consensus-positive results). Of note, Kohmer et al. (20) also identified a small number of samples with IgG recognizing only nucleocapsid or spike protein and speculated that there may be individual differences in the antibody response to SARS-CoV-2 proteins. Last, we assume that samples negative in all 4 assays were from uninfected persons and that samples positive in all 4 assays were from infected patients, but without information on viral RNA results, time since infection, and clinical course, we cannot know with certainty. An additional limitation to the study is possible sample selection bias. As indicated, samples were selected based on the qualitative result in the Abbott SARS-CoV-2 IgG assay. Had we selected samples based on results in a spike-based assay, the levels of agreement among assays might have been different. However, the large number of samples included in the study provided the needed statistical power to demonstrate that the 4 assays yield comparable results. Last, although data for the inhibition assays showed a clear distinction between TP and FP reactivity, it is unclear why the inhibition values in the Abbott inhibition assay were noticeably lower than the inhibition values in the DiaSorin and Euroimmun assays. The most obvious difference is the target antigen of the assay; the Abbott assay targets nucleocapsid, whereas the DiaSorin and Euroimmun assays target spike protein(s). For all 3 inhibition assays, the soluble inhibitory protein employed was not sourced from the assay manufacturer but rather purchased from a different vendor. Thus, subtle differences in glycosylation and tertiary structure may account for differences in the ability of the soluble protein to inhibit binding of antibodies to the immobilized antigen. Further studies are needed to characterize the relationship between soluble antigen structure and inhibitory activity in SARS-CoV-2 IgG inhibition assays. In summary, our findings show these 4 SARS-CoV-2 IgG EUA assays exhibit excellent agreement, regardless of the target antigen used or assay format (CMIA/CIA versus ELISA). While variability can occur with any assay, health professionals receiving SARS-CoV-2 IgG results obtained using any of these 4 assays, whether from different laboratories or different platforms within a given laboratory, can be assured that the results are comparable and likely equivalent as a diagnostic adjunct to NAAT.
  20 in total

1.  Antibody Detection and Dynamic Characteristics in Patients With Coronavirus Disease 2019.

Authors:  Fei Xiang; Xiaorong Wang; Xinliang He; Zhenghong Peng; Bohan Yang; Jianchu Zhang; Qiong Zhou; Hong Ye; Yanling Ma; Hui Li; Xiaoshan Wei; Pengcheng Cai; Wan-Li Ma
Journal:  Clin Infect Dis       Date:  2020-11-05       Impact factor: 9.079

2.  Profile of Immunoglobulin G and IgM Antibodies Against Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2).

Authors:  Jiuxin Qu; Chi Wu; Xiaoyong Li; Guobin Zhang; Zhaofang Jiang; Xiaohe Li; Qing Zhu; Lei Liu
Journal:  Clin Infect Dis       Date:  2020-11-19       Impact factor: 9.079

Review 3.  Current Status of Epidemiology, Diagnosis, Therapeutics, and Vaccines for Novel Coronavirus Disease 2019 (COVID-19).

Authors:  Dae-Gyun Ahn; Hye-Jin Shin; Mi-Hwa Kim; Sunhee Lee; Hae-Soo Kim; Jinjong Myoung; Bum-Tae Kim; Seong-Jun Kim
Journal:  J Microbiol Biotechnol       Date:  2020-03-28       Impact factor: 2.351

4.  Performance Characteristics of the Abbott Architect SARS-CoV-2 IgG Assay and Seroprevalence in Boise, Idaho.

Authors:  Andrew Bryan; Gregory Pepper; Mark H Wener; Susan L Fink; Chihiro Morishima; Anu Chaudhary; Keith R Jerome; Patrick C Mathias; Alexander L Greninger
Journal:  J Clin Microbiol       Date:  2020-07-23       Impact factor: 5.948

5.  The COVID-19 Serology Studies Workshop: Recommendations and Challenges.

Authors:  Andrea M Lerner; Robert W Eisinger; Douglas R Lowy; Lyle R Petersen; Rosemary Humes; Matthew Hepburn; M Cristina Cassetti
Journal:  Immunity       Date:  2020-06-23       Impact factor: 31.745

6.  Brief clinical evaluation of six high-throughput SARS-CoV-2 IgG antibody assays.

Authors:  Niko Kohmer; Sandra Westhaus; Cornelia Rühl; Sandra Ciesek; Holger F Rabenau
Journal:  J Clin Virol       Date:  2020-06-01       Impact factor: 3.168

7.  Kinetics of SARS-CoV-2 specific IgM and IgG responses in COVID-19 patients.

Authors:  Baoqing Sun; Ying Feng; Xiaoneng Mo; Peiyan Zheng; Qian Wang; Pingchao Li; Ping Peng; Xiaoqing Liu; Zhilong Chen; Huimin Huang; Fan Zhang; Wenting Luo; Xuefeng Niu; Peiyu Hu; Longyu Wang; Hui Peng; Zhifeng Huang; Liqiang Feng; Feng Li; Fuchun Zhang; Fang Li; Nanshan Zhong; Ling Chen
Journal:  Emerg Microbes Infect       Date:  2020-12       Impact factor: 7.163

8.  Clinical Performance of Two SARS-CoV-2 Serologic Assays.

Authors:  Mei San Tang; Karl G Hock; Nicole M Logsdon; Jennifer E Hayes; Ann M Gronowski; Neil W Anderson; Christopher W Farnsworth
Journal:  Clin Chem       Date:  2020-08-01       Impact factor: 8.327

9.  Evaluation of the EUROIMMUN Anti-SARS-CoV-2 ELISA Assay for detection of IgA and IgG antibodies.

Authors:  Kathleen G Beavis; Scott M Matushek; Ana Precy F Abeleda; Cindy Bethel; Carlissa Hunt; Stephanie Gillen; Angelica Moran; Vera Tesic
Journal:  J Clin Virol       Date:  2020-05-23       Impact factor: 3.168

10.  Performance Characteristics of Four High-Throughput Immunoassays for Detection of IgG Antibodies against SARS-CoV-2.

Authors:  Elitza S Theel; Julie Harring; Heather Hilgart; Dane Granger
Journal:  J Clin Microbiol       Date:  2020-07-23       Impact factor: 5.948

View more
  13 in total

Review 1.  Practical guidance for clinical laboratories for SARS-CoV-2 serology testing.

Authors:  Carmen Charlton; Jamil Kanji; Vanessa Tran; Julianne Kus; Jonathan Gubbay; Carla Osiowy; Jason Robinson; Inna Sekirov; Michael Drebot; Todd Hatchette; Derek Stein; Nadia El-Gabalawy; Amanda Lang; Lei Jiao; Paul Levett; Heidi Wood; Christian Therrien; L Robbin Lindsay; Muhammad Morshed; Jessica Forbes; Antonia Dibernardo
Journal:  Can Commun Dis Rep       Date:  2021-05-07

2.  Seroprevalence of COVID-19 IgG Antibody in Resident and Fellow Physicians in Milwaukee, Wisconsin: Analysis of a Cross-Sectional Survey.

Authors:  Dennis J Baumgardner; Alexander Schwank; Jessica J F Kram; Wilhelm Lehmann; Jacob L Bidwell; Tricia La Fratta; Kenneth Copeland
Journal:  J Patient Cent Res Rev       Date:  2022-01-17

3.  Estimated SARS-CoV-2 Seroprevalence in the US as of September 2020.

Authors:  Kristina L Bajema; Ryan E Wiegand; Kendra Cuffe; Sadhna V Patel; Ronaldo Iachan; Travis Lim; Adam Lee; Davia Moyse; Fiona P Havers; Lee Harding; Alicia M Fry; Aron J Hall; Kelly Martin; Marjorie Biel; Yangyang Deng; William A Meyer; Mohit Mathur; Tonja Kyle; Adi V Gundlapalli; Natalie J Thornburg; Lyle R Petersen; Chris Edens
Journal:  JAMA Intern Med       Date:  2021-04-01       Impact factor: 21.873

Review 4.  SARS-CoV-2 serology testing: Progress and challenges.

Authors:  Aria C Shi; Ping Ren
Journal:  J Immunol Methods       Date:  2021-04-26       Impact factor: 2.303

5.  Negative anti-SARS-CoV-2 S antibody response following Pfizer SARS-CoV-2 vaccination in a patient on ocrelizumab.

Authors:  Mahsa Khayat-Khoei; Sarah Conway; Douglas A Rubinson; Petr Jarolim; Maria K Houtchens
Journal:  J Neurol       Date:  2021-02-27       Impact factor: 4.849

6.  Virological Correlates of IgM-IgG Patterns of Response to SARS-CoV-2 Infection According to Targeted Antigens.

Authors:  Pablo Barreiro; Francisco Javier Candel; Juan Carlos Sanz; Jesús San Román; María Del Mar Carretero; Marta Pérez-Abeledo; Belén Ramos; José Manuel Viñuela-Prieto; Jesús Canora; Francisco Javier Martínez-Peromingo; Antonio Zapatero
Journal:  Viruses       Date:  2021-05-10       Impact factor: 5.048

7.  Evaluation of High-Throughput Serological Tests for SARS-CoV-2.

Authors:  Michael J Loeffelholz
Journal:  J Clin Microbiol       Date:  2020-10-21       Impact factor: 5.948

8.  Meta-analysis of diagnostic performance of serology tests for COVID-19: impact of assay design and post-symptom-onset intervals.

Authors:  Hongyu Wang; Jingwen Ai; Michael J Loeffelholz; Yi-Wei Tang; Wenhong Zhang
Journal:  Emerg Microbes Infect       Date:  2020-12       Impact factor: 7.163

9.  Case-Control Study of Individuals with Discrepant Nucleocapsid and Spike Protein SARS-CoV-2 IgG Results.

Authors:  Hannah Wang; Danica Wiredja; Lu Yang; Philip L Bulterys; Cristina Costales; Katharina Röltgen; Justin Manalac; Jennifer Yee; James Zehnder; Run Zhang Shi; Scott D Boyd; Benjamin A Pinsky
Journal:  Clin Chem       Date:  2021-07-06       Impact factor: 8.327

10.  Dimerization of SARS-CoV-2 nucleocapsid protein affects sensitivity of ELISA based diagnostics of COVID-19.

Authors:  Wajihul Hasan Khan; Nida Khan; Avinash Mishra; Surbhi Gupta; Vikrant Bansode; Deepa Mehta; Rahul Bhambure; M Ahmad Ansari; Shukla Das; Anurag S Rathore
Journal:  Int J Biol Macromol       Date:  2022-01-17       Impact factor: 6.953

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.