Harry E Prince1, Tara S Givens2, Mary Lapé-Nixon1, Nigel J Clarke1, Dale A Schwab1, Hollis J Batterman1, Robert S Jones2, William A Meyer3, Hema Kapoor4, Charles M Rowland1, Farnoosh Haji-Sheikhi1, Elizabeth M Marlowe5.
Abstract
A total of 1,200 serum samples that were tested for SARS-CoV-2 IgG antibody using the Abbott Architect immunoassay targeting the nucleocapsid protein were run in 3 SARS-CoV-2 IgG immunoassays targeting spike proteins (DiaSorin Liaison, Ortho Vitros, and Euroimmun). Consensus-positive and consensus-negative interpretations were defined as qualitative agreement in at least 3 of the 4 assays. Agreement of the 4 individual assays with a consensus-negative interpretation (n = 610) ranged from 96.7% to 100%, and agreement with a consensus-positive interpretation (n = 584) ranged from 94.3% to 100%. Laboratory-developed inhibition assays were utilized to evaluate 49 consensus-negative samples that were positive in only one assay; true-positive reactivity was confirmed in only 2 of these 49 (4%) samples. These findings demonstrate very high levels of agreement among 4 SARS-CoV-2 IgG assays authorized for emergency use, regardless of antigen target or assay format. Although false-positive reactivity was identified, its occurrence was rare (no more than 1.7% of samples for a given assay).Entities:
Keywords: SARS-CoV-2 IgG; coronavirus; immunoassay
Mesh:
Substances:
Year: 2020 PMID: 32817144 PMCID: PMC7587121 DOI: 10.1128/JCM.01742-20
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948
Information for the 4 SARS-CoV-2 IgG immunoassays used to test study samples
| Assay name | Manufacturer | Method | Protein (antigen) target | Negative index | Positive index |
|---|---|---|---|---|---|
| Architect SARS-CoV-2 IgG | Abbott Laboratories Inc., Abbott Park, IL | CMIA | Nucleocapsid | <1.4 | ≥1.4 |
| Liaison SARS-CoV-2 S1/S2 IgG | DiaSorin Inc., Stillwater, MN | CIA | Spike 1 + spike 2 | <15.0 | ≥15.0 |
| Vitros anti-SARS-CoV-2 IgG | Ortho Clinical Diagnostics, Raritan, NJ | CIA | Spike 1 | <1.0 | ≥1.0 |
| Anti-SARS-CoV-2 ELISA (IgG) | Euroimmun Inc., Mountain Lakes, NJ | ELISA | Spike 1 | <0.8 | ≥0.8 |
For the CMIA and CIA methods, the assay name begins with the instrument utilized for test performance.
Per the Euroimmun instructions for use, index values of 0.8, 0.9, and 1.0 are considered equivocal; however, for the purposes of this study, equivocal results were interpreted as positive.
Distribution of results across all 4 assays and consensus interpretations
| Interpretation ( | Individual assay result | No. of samples | |||
|---|---|---|---|---|---|
| Abbott | DiaSorin | Ortho | Euroimmun | ||
| Consensus-negative (610) | Negative | Negative | Negative | Negative | 561 |
| Negative | Negative | Negative | 15 | ||
| Negative | Negative | Negative | 20 | ||
| Negative | Negative | Negative | 14 | ||
| Consensus-positive (584) | Positive | Positive | Positive | Positive | 548 |
| Positive | Positive | Positive | 33 | ||
| Positive | Positive | Positive | 1 | ||
| Positive | Positive | Positive | 2 | ||
| Nonconsensus (6) | Negative | Positive | Negative | Positive | 2 |
| Positive | Negative | Positive | Negative | 2 | |
| Positive | Negative | Negative | Positive | 2 | |
Discrepant results are in boldface.
Agreement of individual SARS-CoV-2 IgG assays with negative or positive consensus interpretations
| Assay | Consensus negative ( | Consensus positive ( | ||
|---|---|---|---|---|
| No. agreeing | % Agreement (95% confidence interval) | No. agreeing | Agreement (95% confidence interval) | |
| Abbott | 596 | 97.7 (96.2, 98.6) | 582 | 99.7 (98.8, 99.9) |
| DiaSorin | 595 | 97.5 (96, 98.5) | 551 | 94.3 (92.2, 95.9) |
| Ortho | 610 | 100 (99.4, 100) | 584 | 100 (99.3, 100) |
| Euroimmun | 590 | 96.7 (95, 97.9) | 583 | 99.8 (99, 100) |
FIG 1Results from inhibition assays. Blue dots represent samples positive in all 4 assays (TP samples); red dots represent samples that were positive only in the indicated assay. Values for the Abbott assay represent percent inhibition by nucleocapsid, values for the DiaSorin assay represent percent inhibition by S1+S2 proteins, and values for the Euroimmun assay represent percent inhibition by S1 protein. The horizonal lines indicate the assay-specific inhibition value below which reactivity is considered FP.