Literature DB >> 32810015

Feasibility and Efficacy of Vibrant Soundbridge Short Process Coupler in Patients With Aural Atresia.

Farid Alzhrani1, Roa Halawani2, Medhat Yousef1,3.   

Abstract

OBJECTIVES: Different options are available for hearing restoration in patients with aural atresia. Middle ear implantation with the Vibrant Soundbridge (VSB) has been used successfully in many patients with aural atresia with conductive or mixed hearing loss. The aim of the current study was to assess the safety and the efficiency of VSBs coupled to the short process of the incus in patients with aural atresia with conductive or mixed hearing loss.
METHODS: The study was a retrospective chart review conducted at a tertiary university hospital. Audiological and surgical data from six ears with aural atresia implanted with VSBs coupled to the short process of the incus were analyzed. The audiological results were compared with previously reported results obtained after the utilization of different coupling techniques.
RESULTS: All six atretic ears were successfully implanted using the short process coupler, and there were no major complications. Postoperative bone conduction thresholds did not differ substantially from preoperative thresholds. The respective mean preoperative pure-tone average 4, speech reception threshold, and word recognition score values were 56.7 dB, 61.7 dB, and 44%. Postoperative pure-tone average 4 thresholds (25 ± 3.5 dB) and speech reception thresholds (20.8 ± 8 dB) were significantly improved. The maximum postoperative word recognition score achieved was 100%. The audiological outcomes obtained were similar to previously reported outcomes after classic stapes placement.
CONCLUSION: Coupling of the floating mass transducer to the short process of the incus in patients with aural atresia resulted in significant audiological improvements that were comparable to previously reported improvements after placement via other methods.

Entities:  

Mesh:

Year:  2020        PMID: 32810015     DOI: 10.1097/MAO.0000000000002801

Source DB:  PubMed          Journal:  Otol Neurotol        ISSN: 1531-7129            Impact factor:   2.311


  2 in total

1.  Speech performance and subjective satisfaction of middle ear implant in congenital aural atresia.

Authors:  Farid Alzhrani; Salman F Alhabib; Medhat Yousef
Journal:  Acta Otorhinolaryngol Ital       Date:  2022-04       Impact factor: 2.618

Review 2.  [Hearing rehabilitation with the Vibrant Soundbridge in patients with congenital middle ear malformation].

Authors:  J M Hempel; A Epp; V Volgger
Journal:  HNO       Date:  2021-02-18       Impact factor: 1.284

  2 in total