Development and evaluation of a Novel RT-PCR system for reliable and rapid SARS-CoV-2 screening of blood donations.

Meng Li1, Yuwei Zhao1, Ying Li1, Xue Chen1, Dongxia Luo2, Mei Luo3, Jue Hou1, Juan Liu4, Humin Liu1, Huan Wang1, Yufang Dong1, Lanjiang Zhang1, Maosheng Ji1, Xin Zhao1, Caibing Wei1, Wen Li1, Jialiang Gao1, Hua Shan5, Xuemei Fu1.   

Abstract

BACKGROUND: The ongoing outbreak of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) has caused great global concerns. In contrast to SARS, some SARS-CoV-2-infected people can be asymptomatic or have only mild nonspecific symptoms. Furthermore, there is evidence that SARS-CoV-2 may be infectious during an asymptomatic incubation period. With the discovery that SARS-CoV-2 can be detected in plasma or serum, blood safety is worthy of consideration. STUDY DESIGN AND METHODS: We developed a nucleic acid test (NAT) screening system for SARS-CoV-2 targeting nucleocapsid protein (N) and open reading frame 1ab (ORF 1ab) gene that could screen 5076 samples every 24 hours. The 2019 novel coronavirus RNA standard was used to evaluate linearity of standard curves. Diagnostic sensitivity and reproducibility were evaluated using artificial SARS-CoV-2. Specificity was evaluated with 61 other respiratory pathogens. Diagnostic performance was evaluated by testing two sputum and nine oropharyngeal swab specimens. The reverse transcription polymerase chain reaction (RT-PCR) assay was used to screen SARS-CoV-2 RNA in blood donor specimens collected during the outbreak of SARS-CoV-2 in Chengdu.
RESULTS: Limits of detection of the SARS-CoV-2 RT-PCR assay for N and ORF 1ab gene were 12.5 and 27.58 copies/mL, respectively. Intra-assay and interassay for the SARS-CoV-2 RT-PCR assay based on cycle threshold were acceptably low. No cross-reactivity was observed with other respiratory virus and bacterial isolates. The overall agreement value between the SARS-CoV-2 RT-PCR assay and clinical diagnostic results was 100%. A total of 16 287 blood specimens collected from blood donors during SARS-CoV-2 surveillance were tested negative.
CONCLUSIONS: A high-throughput NAT screening system was developed for SARS-CoV-2 screening of blood donations during the outbreak of SARS-CoV-2.
© 2020 AABB.

Entities:  

Mesh:

Substances:

Year:  2020        PMID: 32798248      PMCID: PMC7461364          DOI: 10.1111/trf.16049

Source DB:  PubMed          Journal:  Transfusion        ISSN: 0041-1132            Impact factor:   3.337


2019 novel coronavirus cycle threshold coefficient of variation limit of detection nucleocapsid protein nucleic acid test open reading frame 1ab polymerase chain reaction reverse transcription polymerase chain reaction severe acute respiratory syndrome coronavirus 2 World Health Organization

INTRODUCTION

With a cluster of cases with pneumonia of unknown etiology in Wuhan City, Hubei Province, China, in December 2019, a new coronavirus was identified as the causative pathogen, which was subsequently named severe acute respiratory syndrome coronavirus 2 (SARS‐CoV‐2). , Due to its rapid spread, high infectivity, and high fatality, , , on 31 January 2020, the World Health Organization (WHO) declared the outbreak of coronavirus 2019 (COVID‐19) in China as a public health emergency of international concern. On 11 March, the WHO further declared COVID‐19 as a pandemic. According to data provided by the WHO, by 29 July 2020, there were 16 558 289 confirmed cases of COVID‐19 since the beginning of the epidemic. At that time, a total of 656 093 patients had died of the SARS‐CoV‐2 infection. SARS‐CoV‐2 was first isolated from human airway epithelial cells and characterized by next‐generation sequencing. , The genome sequence was released on 10 January (Wuhan‐Hu‐1, GenBank accession number MN908947). , SARS‐CoV‐2 belongs to beta coronaviruses that are a variety of enveloped, positive‐stranded RNA viruses with a nucleocapsid. , The genome sequence of SARS‐CoV‐2 had about 80% nucleotide identity with that of human SARS‐CoV. , In contrast to SARS, the clinical spectrum of COVID‐19 varies from asymptomatic or limited mild nonspecific flulike symptoms to respiratory failure and multiple organ dysfunctions. , , It was established that there were about 81% of SARS‐CoV‐2 cases with nonpneumonia and mild pneumonia. , Chan et al reported an asymptomatic carrier in a family cluster of COVID‐19. SARS‐CoV‐2 might be infectious during an asymptomatic incubation period. , , , , According to the investigations conducted by the Chinese Center for Disease Control, the incubation time is generally 3 to 7 days, with a mean incubation period of 5.2 days (95% confidence interval [CI], 4.1‐7.0) and a 95th percentile of the distribution of 12.5 days. , The incubation period of SARS‐CoV‐2 is longer than that of SARS‐CoV. , , These factors increase the challenge for prevention and control of SARS‐CoV‐2 infections. Although coronaviruses usually infect the upper and lower respiratory tract, viral shedding in blood is common. , , It has been demonstrated that 15% to 40% of SARS‐CoV‐2–infected individuals have detectable RNAemia. , , Furthermore, there were no differences between intensive care unit patients and patients with mild symptoms on median polymerase chain reaction (PCR) cycle threshold (Ct) value. Data are limited about SARS‐CoV‐2 viremia at different stages of the infection. Screening of blood donors from Wuhan during the height of the epidemic found four asymptomatic blood donors with positive SARS‐CoV‐2 RNA in blood samples. SARS‐CoV‐2 can infect a variety of tissue culture cells and organoids. With limited available information, a theoretical risk of SARS‐CoV‐2 transfusion‐transmitted infection could not be excluded. , To protect transfusion recipients from transfusion‐transmitted infection of SARS‐CoV‐2 during the epidemic period, Chinese blood collection facilities implemented additional predonation screening measures, including checking body temperature before blood donation and new deferral criteria. The new deferral criteria defer all blood donors for 28 days after the resolution of any COVID‐19 symptoms, after a history of residence or travel to Hubei Province and close contact with individuals diagnosed with or suspected of having COVID‐19. Even with these enhanced predonation donor screening measures, some asymptomatically infected donors may still donate blood. To prevent potential transfusion‐transmitted SARS‐CoV‐2 infection and address public concerns about the safety of the blood supply during the COVID‐19 epidemic, we developed a SARS‐CoV‐2 RNA donor screening assay and conducted donor screening testing with use of blood specimens from blood donors. Based on the genome sequence of SARS‐CoV‐2, real‐time reverse transcription polymerase chain reaction (RT‐PCR) detection assays have been quickly developed to detect SARS‐CoV‐2 in sputum, throat swabs, and nasal swabs. However, there are no validated accurate NAT screening systems for detecting SARS‐CoV‐2 RNA in blood samples to date. Furthermore, available assays have limited sensitivity and are low throughput, precluding rapid screening of large numbers of samples. In this study, we developed a high‐throughput (5076 tests per 24 hours), high‐sensitivity, and accurate NAT screening system for the SARS‐CoV‐2 RNA detection of blood donations in collaboration with Sansure Biotechnology Company. After evaluating its analytical performance and clinical applicability, the assay was used for screening SARS‐CoV‐2 RNA in blood specimens collected from blood donors during the height of SARS‐CoV‐2 epidemic in Chengdu.

MATERIALS AND METHODS

Virus and clinical specimens

Two sputum and nine oropharyngeal swab specimens were used for clinical evaluation of the SARS‐CoV‐2 assay, which were collected from 10 patients with confirmed/suspected COVID‐19 in the Public Health and Clinical Center of Chengdu. The specimens were deactivated by heating at 65°C for 30 minutes and previously tested with a nucleic acid isolation kit (Da'an Gene Corporation, Guangzhou, China; Cat: DA0630) and SARS‐CoV‐2 RT‐PCR reagent (Da'an Gene Corporation; Cat: DA0930). The specimens were stored at 2 to 8°C for no more than 24 hours before detecting. Total nucleic acid of specimens was simultaneously extracted individually with a nucleic acid extraction and purification kit (Sansure Biotechnology Company, Hunan, China) and then immediately delivered to Chengdu Blood Center on dry ice. The nucleic acid was diluted 6‐fold by Sansure eluting buffer, and then 20 μL of diluted nucleic acid was used as template for amplification. A total of 16 287 blood specimens collected from blood donors between 27 January 2020 and 29 February 2020 were tested at Chengdu Blood Center.

Primers and probes

Primer sets for 2019 novel coronavirus (2019‐nCoV) were designed using Primer 5 based on the sequence (Wuhan‐Hu‐1, GenBank accession number MN908947) targeting both the nucleocapsid protein (N) and open reading frame 1ab (ORF 1ab). The β‐globin gene was amplified simultaneously as a heterologous endogenous internal control. Primer/probe sets were predicted on Basic Local Alignment Search Tool analysis to specifically amplify 2019‐nCoV with no major combined homologies with other coronaviruses or human microflora. All primers and probes were produced by standard phosphoramidite chemical techniques at a biotechnology facility.

Pooling, extraction, and amplification

Plasma specimens were pooled and total nucleic acids were extracted on an automatic workstation. The total input volume of each pool was 240 μL consisting of 40‐μL equal aliquots from each donation (6 × 40). Samples were added to 96‐deep‐well plate (Axygen Biotech, Hangzhou, China; Cat: P‐DW‐20‐C‐S). Extraction of total nucleic acids was performed with a nucleic acid extraction and purification kit (Sansure Biotechnology Company, Hunan, China) according to the manufacturer's instructions. The nucleic acid was eluted with 50 μL elution buffer. Each PCR reaction was performed in a total volume of 50 μL (20 μL of template nucleic acid, 26 μL of RT‐PCR buffer and 4 μL of enzyme mix). Primers and probes at 240 nM were used for one reaction. Reverse transcription was performed at 50°C for 30 minutes. PCR cycling was done at 95°C for 1 minute and 45 cycles of 95°C for 15 seconds, 60°C for 30 seconds and 25°C for 10 seconds. All RT‐PCR reactions were performed with a real‐time PCR system (7500 Fast Real Time PCR System, ABI7500, Applied Biosystems, Foster City, CA). Positive and negative controls were included in each run. A positive test result was defined as a well‐defined exponential fluorescence curve that crossed the threshold of 40 cycles or less for the ORF 1ab, N, and β‐globin genes.

Analytical performance evaluation

For validation, linearity of standard curves was verified by 2019‐nCoV RNA standard (National Institute of Metrology, Beijing, China; Cat: GBW[E]091099), which is purified RNA containing 6.89 × 102 copies/μL ORF 1ab gene and 1.36 × 103 copies/μL N gene. Sensitivity and reliability were verified with use of artificial SARS‐CoV‐2 virus (MS2 bacteriophage‐like particles packing SARS‐CoV‐2 ORF 1ab and N gene, Zeesan Biotech, Xiamen, China; Cat: JBD 279), which was obtained by armored RNA technology and quantified by 2019‐nCoV RNA standard. To determine linearity of the standard curve, 2019‐nCoV RNA standards were diluted to six concentrations by tenfold serial dilutions with N gene from 1.36 × 105 to 1.36 copies/mL and ORF 1ab gene from 6.89 × 104 to 0.689 copies/mL using RNase free water, with 10 replicates per dilution. The PCR efficiency (E) was calculated using the standard curve slope according the following formula: E = 10‐1/slope‐1. The correlation coefficient (R 2) was also calculated. The limit of detection (LOD) and reliability was determined by testing individually using artificial SARS‐CoV‐2 virus by twofold serial dilutions with N gene from 800 copies/mL to 1.56 copies/mL and ORF 1ab gene from 882.48 to 1.72 copies/mL using SARS‐CoV‐2–negative plasma. Twenty replicates per dilution. The LOD was defined as the lowest concentration at which a fluorescent signal could be detected in all reactions. Subsequently, diagnostic sensitivity of the assay was verified using 20 replicates of mock specimens with artificial SARS‐CoV‐2 virus containing 75 copies/mL N gene and 82.73 copies/mL ORF 1ab gene tested with equal volumes of five SARS‐CoV‐2–negative plasma in pools of six. Reproducibility was evaluated with artificial SARS‐CoV‐2 virus of low, moderate, or high concentrations (N gene 25, 62.5 and 123 copies/mL; ORF 1ab gene 27.58, 68.94 and 137.89 copies/mL). For intra‐assay variability evaluated, 20 replicates of three different concentrations were tested within the same run. Interassay variability was evaluated by testing three dilutions on 5 consecutive days in 20 replicates. Reproducibility (inter‐ and intra‐assay variance) was assessed using the coefficient value calculated on the basis of Ct values. Analytical specificity was evaluated using 61 other respiratory virus and bacterial isolates. Artificial Middle East respiratory syndrome and severe acute respiratory syndrome virus were obtained from GenePharma Co., Ltd. (Shanghai, China). Influenza virus A, Klebsiella pneumonia, Haemophilus influenzae, Staphylococcus aureus, Pseudomonas aeruginosa, Legionella pneumophila, and Streptococcus pneumoniae were purchased from National Institutes for Food and Drug Control of China. Other isolates were obtained from Sanway Clinical Laboratories Inc (Changsha, China).

Interference study

The influence of hemoglobin and lipids on COVID‐19 was separately evaluated in SARS‐CoV‐2–negative plasma sample spiked with increasing concentrations of hemoglobin or lipids. The concentrations of hemoglobin were 50 000, 40 000, 30 000, 20 000, 10 000, 5000, 2500, and 1250 mg/L. The triglyceride concentrations were 8000, 4000, and 2000 mg/L. The simulated plasma samples were tested for SARS‐CoV‐2 with or without about 3 × LOD artificial SARS‐CoV‐2 virus (ORF 1ab 41.37 copies/mL, N 37.5 copies/mL), with 10 replicates per dilution.

Diagnostic performance

The diagnostic performance of the assay was evaluated by testing 11 clinical specimens (two sputum and nine oropharyngeal swab specimens) from 10 patients with confirmed/suspected COVID‐19. The positive percent agreement and negative percent agreement values and kappa coefficients were calculated.

Statistical analysis

Values were log10‐transformed before analysis with computer software (Excel, Microsoft, Redmond, WA). The agreement between the Sansure SARS‐CoV‐2 RT‐PCR assay and previously clinical SARS‐CoV‐2 RT‐PCR assay was assessed by the kappa coefficient. Statistical analyses were performed by using computer software (SPSS 11, SPSS, Chicago, IL). A P value of <.05 was considered statistically significant.

Ethical statement

All specimens were collected for diagnostic evaluation, and diagnostic purposes added no extra burden to patients and blood donors. This study was approved by the Ethics Committee of Chengdu Blood Center.

RESULTS

Pooling samples and extraction of total nucleic acids were performed on an automated workstation (Sansure Biotechnology Company), which reduced the hands‐on time and technical errors. This high‐throughput, efficient, and accurate NAT system is useful for screening SARS‐CoV‐2 in large numbers of samples from blood donors. We evaluated the NAT system using mock specimens and clinical specimens from COVID‐19 patients. The calculated analytical detection limit of the assay (12.5 copies/mL) appears to be low. The low percentage of coefficient value of the cycle number demonstrates the low intra‐assay variation. Stable reaction properties are also reflected in good quantification results (N gene linear range between 2.72 × 103 and 2.72 copies/reaction; ORF 1ab gene linear range between 1.378 × 103 and 1.378 copies/reaction). No cross reactivity was obtained with other respiratory isolates demonstrating high specificity for SARS‐CoV‐2. The NAT system demonstrates the feasibility of screening blood donors for SARS‐CoV‐2 in the setting of routine blood center operations.

Optimization of PCR reaction condition

Optimization of the PCR reaction components and cycling conditions were undertaken using the RNA standard including N and ORF 1ab gene of the SARS‐CoV‐2 virus. The optimum concentrations of primers, annealing, and data acquisition temperatures were evaluated. The optimum conditions were defined as those that gave the maximum fluorescence and lowest Ct values without the presence of primer dimers or nonspecific amplification. The standard curves of N and ORF 1ab were plotted with slopes of −3.0579 and −3.0848, respectively.

Linearity of standard curves

The dynamic range and efficiency of the assay were evaluated using duplicate 10‐fold serial dilutions of RNA standard including N and ORF 1ab gene. As shown in Figure 1, standard curves were constructed in terms of mean Ct value and log10(copy numbers). There was a strong linear correlation (R 2 = 99.91% for N gene and 99.76% for ORF 1ab gene) between Ct values and the corresponding amount of copy numbers. The PCR efficiency of N gene and ORF 1ab gene was 112.34% and 110.95%, respectively.
FIGURE 1

Standard curves of the SARS‐CoV‐2 RT‐PCR assay. Plots of serial 10‐fold dilutions of 2019‐nCoV RNA standard analyzed by the SARS‐CoV‐2 RT‐PCR assay with N gene from 1.36 × 105 to 1.36 copies/mL and ORF 1ab gene from 6.89 × 104 to 0.689 copies/mL using RNase free water. Linear correlation coefficients (R 2) and amplification efficiency for each assay were indicated. A, Plot inserts show mean Ct values of N gene vs Log10 (Lg) input copies/reaction; B, plot inserts show mean Ct values of ORF 1ab gene vsLg input copies/reaction

Standard curves of the SARS‐CoV‐2 RT‐PCR assay. Plots of serial 10‐fold dilutions of 2019‐nCoV RNA standard analyzed by the SARS‐CoV‐2 RT‐PCR assay with N gene from 1.36 × 105 to 1.36 copies/mL and ORF 1ab gene from 6.89 × 104 to 0.689 copies/mL using RNase free water. Linear correlation coefficients (R 2) and amplification efficiency for each assay were indicated. A, Plot inserts show mean Ct values of N gene vs Log10 (Lg) input copies/reaction; B, plot inserts show mean Ct values of ORF 1ab gene vsLg input copies/reaction

Analytical sensitivity

The LOD of the assay was determined using artificial SARS‐CoV‐2 virus. Serial twofold dilutions (N gene from 800 copies/mL to 1.56 copies/mL and ORF 1ab gene from 882.48 to 1.72 copies/mL) of artificial SARS‐CoV‐2 virus were prepared in SARS‐CoV‐2–negative plasma and were tested with each assay signature in 20 replicates for positive amplification; 240 μL of each sample was subjected to nucleic acids extraction with an elution volume of 50 μL. Then 20 μL of each elution was added as the template of amplification assay. The highest dilution of transcript at which all replicates were positive was defined as the LOD for each assay. As shown in Table 1, the LOD value of the assay was 12.5 copies/mL. The LODs for N and ORF 1ab gene were 12.5 and 27.58 copies/mL. Furthermore, 20 replicates of mock specimens using artificial SARS‐CoV‐2 virus containing 75 copies/mL N gene and 82.73 copies/mL ORF 1ab gene were tested when mixed with the other five SARS‐CoV‐2–negative plasmas in minipools of six. The positive detection rate was 90%. Probit regression analysis in SPSS was used to determine the 95% cutoff value. According to the probit regression analysis (as shown in Figure S1), the N has a 95% cutoff value of 11.47 copies/mL (95% confidence interval [CI]: from 8.49 to 21.14 copies/mL). The ORF 1ab assay detects 16.36 copies/mL with 95% probability (95% CI, 12.78‐25.06 copies/mL).
TABLE 1

Determination of limits of detection based on artificial SARS‐CoV‐2 virus

Concentration of target gene(copies/mL)Number of detections /number of replicates (%)Mean Ct
NORF 1abNORF 1abNORF 1ab
800882.4820/20 (100)20/20 (100)29.2431.80
400441.2420/20 (100)20/20 (100)29.8431.90
200220.6220/20 (100)20/20 (100)30.9433.40
100110.3120/20 (100)20/20 (100)31.7634.23
5055.1520/20 (100)20/20 (100)32.9435.55
2527.58 a 20/20 (100)20/20 (100)34.1036.09
12.5 a 13.7920/20 (100)17/20 (85)35.0436.62
6.256.8916/20 (80)13/20 (65)35.1636.75
3.1253.4513/20 (65)8/20 (40)35.8537.74
1.56251.728/20 (40)3/20 (15)36.6238.28

Abbreviations: Ct, threshold cycle; ORF 1ab, open reading frame 1ab; SARS‐CoV‐2, severe acute respiratory syndrome coronavirus 2.

The highest dilution at which 100% of replicates were positive.

Determination of limits of detection based on artificial SARS‐CoV‐2 virus Abbreviations: Ct, threshold cycle; ORF 1ab, open reading frame 1ab; SARS‐CoV‐2, severe acute respiratory syndrome coronavirus 2. The highest dilution at which 100% of replicates were positive.

Reproducibility studies

Variability was evaluated using artificial SARS‐CoV‐2 virus of high, moderate, or low concentrations (N gene 20, 10 and 5 copies/mL; ORF 1ab gene 27.58, 68.94 and 137.89 copies/mL). The mean, SD, and coefficient of variation (CV) of the Ct value were determined. As shown in Table 2, The CV values of intra‐ and interassay for the N gene were 0.90‐2.34 and 0.56‐0.85. The CV values of intra‐ and interassay for ORF 1ab were 0.73‐2.24 and 0.80‐1.58. Overall, these results show that the assay is reliable, with different Ct values.
TABLE 2

Intra‐assay and interassay of SARS‐CoV‐2 RT‐PCR assay based on Ct

TargetConcentration (copies/mL)ReplicationsMean CtSDCV(%)
InterassayN25.002034.070.190.56
62.502033.030.280.85
125.002031.900.230.72
ORF 1ab27.582036.740.581.58
68.942035.150.521.48
137.892034.090.270.80
Intra‐assayN25.002034.120.802.34
62.502033.120.511.55
125.002031.850.290.90
ORF 1ab27.582036.430.822.24
68.942035.190.491.40
137.892034.050.250.73

Abbreviations: Ct, cycle threshold; CV, coefficient of variation; ORF 1ab, open reading frame 1ab; RT‐PCR, reverse transcription polymerase chain reaction; SARS‐CoV‐2, severe acute respiratory syndrome coronavirus 2.

Intra‐assay and interassay of SARS‐CoV‐2 RT‐PCR assay based on Ct Abbreviations: Ct, cycle threshold; CV, coefficient of variation; ORF 1ab, open reading frame 1ab; RT‐PCR, reverse transcription polymerase chain reaction; SARS‐CoV‐2, severe acute respiratory syndrome coronavirus 2.

Analytical specificity

The specificity of SARS‐CoV‐2 RT‐PCR assay was evaluated by using 61 other respiratory pathogens that were tested in SARS‐CoV‐2 negative plasma. These pathogens and their concentrations were listed in Table S1. Melting curve analysis of each sample showed that there was no evidence of amplification. No false‐positive test results were obtained with any sample, demonstrating high specificity for the detection of SARS‐CoV‐2. As shown in Table 3, high concentration of hemoglobin (≥20 000 mg/L) could affect the detection rate of SARS‐CoV‐2. The lipids interference study demonstrated that there was no detectable interference at the concentration of 8000, 4000, or 2000 mg/L (Table 4).
TABLE 3

The hemoglobin interference study of SARS‐CoV‐2 RT‐PCR assay

Hemoglobin concentration (mg/L)Number of positive tests/number of replicates (%)Mean Ct
ORF 1abNORF 1abN
20 0009/10 (90)10/10 (100)37.0134.48
10 00010/10 (100)10/10 (100)36.0833.13
500010/10 (100)10/10 (100)36.6833.22
250010/10 (100)10/10 (100)34.2832.98
125010/10 (100)10/10 (100)35.3232.96

Abbreviations: Ct, cycle threshold; ORF 1ab, open reading frame 1ab; RT‐PCR, reverse transcription polymerase chain reaction; SARS‐CoV‐2, severe acute respiratory syndrome coronavirus 2.

TABLE 4

The lipids interference study of SARS‐CoV‐2 RT‐PCR assay

Triglyceride concentration (mg/L)Number of positive tests/number of replicates (%)Mean Ct
ORF 1abNORF 1abN
800010/10 (100)10/10 (100)35.3333.28
400010/10 (100)10/10 (100)35.5733.09
200010/10 (100)10/10 (100)35.6333.23

Abbreviations: Ct, cycle threshold; ORF 1ab, open reading frame 1ab; RT‐PCR, reverse transcription polymerase chain reaction; SARS‐CoV‐2, severe acute respiratory syndrome coronavirus 2.

The hemoglobin interference study of SARS‐CoV‐2 RT‐PCR assay Abbreviations: Ct, cycle threshold; ORF 1ab, open reading frame 1ab; RT‐PCR, reverse transcription polymerase chain reaction; SARS‐CoV‐2, severe acute respiratory syndrome coronavirus 2. The lipids interference study of SARS‐CoV‐2 RT‐PCR assay Abbreviations: Ct, cycle threshold; ORF 1ab, open reading frame 1ab; RT‐PCR, reverse transcription polymerase chain reaction; SARS‐CoV‐2, severe acute respiratory syndrome coronavirus 2.

Clinical studies

Performance of the RT‐PCR assay with human clinical specimens from patients with confirmed/suspected COVID‐19. Two sputum and nine oropharyngeal swab specimens collected from 10 patients with confirmed/suspected COVID‐19 were evaluated for diagnostic performance. The clinical specimens were collected following the Guidance of Laboratory Testing for 2019 Novel Coronavirus issued by National Health Commission of the People's Republic of China. Oropharyngeal samples were collected as follows: Two oropharyngeal swabs were used simultaneously to rub the posterior wall of the oropharynx under direct visualization. Two oropharyngeal swabs were immediately placed into sterile tubes containing 2 mL of standard virus transport media, breaking the rod with one swift and controlled movement. The sputum sampling was easier. Patients forced out a deep cough and spit the sputum into sputum containers containing 2 mL of sampling solution. The clinical specimens were tested by our assay in a blinded manner. As shown in Table 5, all five specimens from four patients previously confirmed positive for COVID‐19 infection by RT‐PCR assay at the Public Health and Clinical Center of Chengdu were positive and the other specimens were negative, consistent with clinical diagnosis. The kappa coefficient between the two qualitative results was 100% (Table S2). However, the testing results of ORF 1ab and N gene for oropharyngeal swab sample for one patient were different (Tables S3 and S4).
TABLE 5

Diagnostic performance comparison with previous clinical NAT testing

NumberSample typeClinical diagnosisSARS‐CoV‐2 RT‐PCR assayClinical NAT testing
ORF 1ab CtN CtResultORF 1ab CtN CtResult
1SputumConfirmed38.2533.23Pos.39.0639.60Pos.
OP swabConfirmed33.55Pos.38.17Pos.
2SputumConfirmed35.8731.46Pos.38.2238.99Pos.
3OP swabConfirmed29.4128.51Pos.32.8333.61Pos.
4OP swabConfirmed33.53Pos.39Pos.
5OP swabSuspectedNeg.Neg.
6OP swabSuspectedNeg.Neg.
7OP swabSuspectedNeg.Neg.
8OP swabSuspectedNeg.Neg.
9OP swabSuspectedNeg.Neg.
10OP swabSuspectedNeg.Neg.

Abbreviations: Ct, cycle threshold; OP, oropharyngeal; ORF 1ab, open reading frame 1ab; N, nucleocapsid protein; NAT, nucleic acid test; Neg., negative; Pos., positive; RT‐PCR, reverse transcription polymerase chain reaction; SARS‐CoV‐2, severe acute respiratory syndrome coronavirus 2.

Performance of the RT‐PCR assay with plasma specimens from blood donors during COVID‐19 surveillance. A total of 16 287 blood specimens collected from donors between 27 January 2020 and 29 February 2020 in Chengdu were tested in minipools of six, and all results were negative. We also performed retrospective individual donation retesting of 15 stored samples from donors who notified us after blood donation that they had contacted patients with suspected COVID‐19 or had fever, cough, or other suspected SARS‐CoV‐2 symptoms. All tested negative, and these donors have not become patients with COVID‐19. Diagnostic performance comparison with previous clinical NAT testing Abbreviations: Ct, cycle threshold; OP, oropharyngeal; ORF 1ab, open reading frame 1ab; N, nucleocapsid protein; NAT, nucleic acid test; Neg., negative; Pos., positive; RT‐PCR, reverse transcription polymerase chain reaction; SARS‐CoV‐2, severe acute respiratory syndrome coronavirus 2.

DISCUSSION

SARS‐CoV‐2 RT‐PCR is widely used for clinical diagnostic purpose. Many reports have described RT‐PCR assays for detection of SARS‐CoV‐2 RNA targeting regions of ORF 1ab and N gene. , , No RT‐PCR assay for blood donor screening had been reported so far. Although there has not been definitive evidence demonstrating that SARS‐CoV‐2 can be transmitted through blood, SARS‐CoV‐2 RNA has been detected in asymptomatic blood donors. Some asymptomatic infected individuals may not be deferred by body temperature check, health history, travel/residence history, or contact history during predonation screening. NAT technology may be helpful to detect SARS‐CoV‐2 virus in these potential donors. , , , We urgently developed a NAT system to screen a blood donor's sample for SARS‐CoV‐2 with Sansure Biotechnology Company at the early stage of SARS‐CoV‐2 epidemic in Chengdu. To address a potential risk of SARS‐CoV‐2 transmission through blood, our NAT donor screening system was used for screening blood donors in Chengdu during the epidemic period. A total of 16 287 consecutive donor blood specimens were tested. All results were negative. In contrast, a donor screening study in Wuhan found SARS‐CoV‐2 RNA in blood samples from four asymptomatic blood donors. The difference between the results of these two studies most likely reflects the significant different level of infectious burden of the two populations. Comparing to Chengdu, Wuhan had a significantly higher population prevalence of COVID‐19 during this time period (27 January to 29 February 2020). The reported population prevalence in Wuhan was 42.4 in 10000 (47 532 infections in a population of 11.2 million), comparing to 0.06 in 10000 (106 infections in a population of 16.6 million) in Chengdu. The additional precautionary predonation nontesting screening measures implemented at our blood center prior to the initiation of NAT testing may have also contributed to our result of no positive samples. The nontesting measures included (a) body temperature check; (b) new questions in donor screening questionnaire about history (within past 28 days) of having fever, cough, or other suspected SARS‐CoV‐2 symptoms; contact with patients with suspected/confirmed COVID‐19 or traveling/residence to Hubei Province. Any “yes” answer would constitute a deferral for at least 28 days; and (c) blood donors were required to immediately notify the blood center if they were suspected or confirmed for COVID‐19 within 1 month after their donation. Protecting the safety of blood supply presents particular challenge during the early stage of an emerging infection when sufficient data were not available to address all safety concerns. Public health concerns demand the timely implementation of additional security measures in protecting blood safety. Both nontesting and testing strategies based on available information should be considered under the principle of abundance of caution while additional scientific information is being collected. The additional donor deferral criteria may have helped to screen out SARS‐CoV‐2–infected potential donors in regions with mild and moderate epidemic such as Chengdu. On the other hand, for regions with high infection prevalence, additional measures such as donor screening with NAT may be necessity if future additional data indicate a transfusion transmission risk for SARS‐CoV‐2. Implementation of a sensitive and specific donor screening test can also offer the benefit of relaxing nontesting deferral criteria therefore reduce the loss of otherwise eligible donors. Donor availability can be a critical challenge for blood services during emerging epidemics when both the safety and the availability of blood are of significant concern. Moreover, timely screening of blood donors during an emerging infectious epidemic is a useful tool to obtain valuable epidemiological data in a convenient representative sample of the health population. Such data are helpful for public health officials in understanding how an infection spreads in the general population, in addition to assessing local blood safety risk. This study has several limitations. First, the assay was validated with respiratory specimens from patients with confirmed COVID‐19. Up to now, data were limited on the viral load in plasma of asymptomatic carriers or individuals during the incubation period. Second, only 11 clinical specimens were used to compare assay performance across different NAT PCR platforms. Due to a relatively mild local epidemic of SARS‐CoV‐2 and strict population prevention measures, it was very difficult to get more clinical specimens during this study. Sansure Biotechnology Company provided the data about clinical evaluation between the Sansure Biotech Novel Coronavirus Nucleic Acid Diagnostic Kit (nucleic acid extraction and purification method was different with that used in this study but amplification method was the same) and the BGI Genomics comparator. Of 246 clinical samples, 52 specimens were tested positive. The results demonstrated a positive percent agreement of 94.34% and negative percent agreement of 98.96%. The data further demonstrate this assay's amplification ability of SARS‐CoV‐2. In conclusion, we developed a NAT screening system with demonstrated feasibility of screening large number of blood donors for SARS‐CoV‐2 on a routine basis if necessary.

CONFLICT OF INTEREST

The authors declare no conflicts of interest. Fig S1 Probit analysis of LOD for N and ORF 1ab gene. A, N gene assay, evaluated with artificial SARS‐CoV‐2 virus. B, ORF 1ab gene assay, evaluated with artificial SARS‐CoV‐2 virus. Click here for additional data file. Table S1 Analytical specificity of SARS‐CoV‐2 RT‐PCR assay by the test of known microorganism. Click here for additional data file. Table S2 The kappa coefficient between the assay and previous clinical NAT testing. Click here for additional data file. Table S3 The kappa coefficient of ORF 1ab detection between the assay and previous clinical NAT testing. Click here for additional data file. Table S4 The kappa coefficient of N gene detection between the assay and previous clinical NAT testing. Click here for additional data file.
  36 in total

1.  Identification of a novel coronavirus in patients with severe acute respiratory syndrome.

Authors:  Christian Drosten; Stephan Günther; Wolfgang Preiser; Sylvie van der Werf; Hans-Reinhard Brodt; Stephan Becker; Holger Rabenau; Marcus Panning; Larissa Kolesnikova; Ron A M Fouchier; Annemarie Berger; Ana-Maria Burguière; Jindrich Cinatl; Markus Eickmann; Nicolas Escriou; Klaus Grywna; Stefanie Kramme; Jean-Claude Manuguerra; Stefanie Müller; Volker Rickerts; Martin Stürmer; Simon Vieth; Hans-Dieter Klenk; Albert D M E Osterhaus; Herbert Schmitz; Hans Wilhelm Doerr
Journal:  N Engl J Med       Date:  2003-04-10       Impact factor: 91.245

2.  Genome Detective Coronavirus Typing Tool for rapid identification and characterization of novel coronavirus genomes.

Authors:  Sara Cleemput; Wim Dumon; Vagner Fonseca; Wasim Abdool Karim; Marta Giovanetti; Luiz Carlos Alcantara; Koen Deforche; Tulio de Oliveira
Journal:  Bioinformatics       Date:  2020-06-01       Impact factor: 6.937

3.  Early Transmission Dynamics in Wuhan, China, of Novel Coronavirus-Infected Pneumonia.

Authors:  Qun Li; Xuhua Guan; Peng Wu; Xiaoye Wang; Lei Zhou; Yeqing Tong; Ruiqi Ren; Kathy S M Leung; Eric H Y Lau; Jessica Y Wong; Xuesen Xing; Nijuan Xiang; Yang Wu; Chao Li; Qi Chen; Dan Li; Tian Liu; Jing Zhao; Man Liu; Wenxiao Tu; Chuding Chen; Lianmei Jin; Rui Yang; Qi Wang; Suhua Zhou; Rui Wang; Hui Liu; Yinbo Luo; Yuan Liu; Ge Shao; Huan Li; Zhongfa Tao; Yang Yang; Zhiqiang Deng; Boxi Liu; Zhitao Ma; Yanping Zhang; Guoqing Shi; Tommy T Y Lam; Joseph T Wu; George F Gao; Benjamin J Cowling; Bo Yang; Gabriel M Leung; Zijian Feng
Journal:  N Engl J Med       Date:  2020-01-29       Impact factor: 176.079

4.  A familial cluster of pneumonia associated with the 2019 novel coronavirus indicating person-to-person transmission: a study of a family cluster.

Authors:  Jasper Fuk-Woo Chan; Shuofeng Yuan; Kin-Hang Kok; Kelvin Kai-Wang To; Hin Chu; Jin Yang; Fanfan Xing; Jieling Liu; Cyril Chik-Yan Yip; Rosana Wing-Shan Poon; Hoi-Wah Tsoi; Simon Kam-Fai Lo; Kwok-Hung Chan; Vincent Kwok-Man Poon; Wan-Mui Chan; Jonathan Daniel Ip; Jian-Piao Cai; Vincent Chi-Chung Cheng; Honglin Chen; Christopher Kim-Ming Hui; Kwok-Yung Yuen
Journal:  Lancet       Date:  2020-01-24       Impact factor: 79.321

5.  Clinical characteristics of 24 asymptomatic infections with COVID-19 screened among close contacts in Nanjing, China.

Authors:  Zhiliang Hu; Ci Song; Chuanjun Xu; Guangfu Jin; Yaling Chen; Xin Xu; Hongxia Ma; Wei Chen; Yuan Lin; Yishan Zheng; Jianming Wang; Zhibin Hu; Yongxiang Yi; Hongbing Shen
Journal:  Sci China Life Sci       Date:  2020-03-04       Impact factor: 10.372

6.  The Incubation Period of Coronavirus Disease 2019 (COVID-19) From Publicly Reported Confirmed Cases: Estimation and Application.

Authors:  Stephen A Lauer; Kyra H Grantz; Qifang Bi; Forrest K Jones; Qulu Zheng; Hannah R Meredith; Andrew S Azman; Nicholas G Reich; Justin Lessler
Journal:  Ann Intern Med       Date:  2020-03-10       Impact factor: 25.391

Review 7.  Unique epidemiological and clinical features of the emerging 2019 novel coronavirus pneumonia (COVID-19) implicate special control measures.

Authors:  Yixuan Wang; Yuyi Wang; Yan Chen; Qingsong Qin
Journal:  J Med Virol       Date:  2020-03-29       Impact factor: 20.693

8.  Updated rapid risk assessment from ECDC on the novel coronavirus disease 2019 (COVID-19) pandemic: increased transmission in the EU/EEA and the UK.

Authors: 
Journal:  Euro Surveill       Date:  2020-03

9.  Characteristics of and Important Lessons From the Coronavirus Disease 2019 (COVID-19) Outbreak in China: Summary of a Report of 72 314 Cases From the Chinese Center for Disease Control and Prevention.

Authors:  Zunyou Wu; Jennifer M McGoogan
Journal:  JAMA       Date:  2020-04-07       Impact factor: 56.272

10.  Estimating the asymptomatic proportion of coronavirus disease 2019 (COVID-19) cases on board the Diamond Princess cruise ship, Yokohama, Japan, 2020.

Authors:  Kenji Mizumoto; Katsushi Kagaya; Alexander Zarebski; Gerardo Chowell
Journal:  Euro Surveill       Date:  2020-03
View more
  2 in total

1.  Rapid, point-of-care antigen and molecular-based tests for diagnosis of SARS-CoV-2 infection.

Authors:  Jacqueline Dinnes; Jonathan J Deeks; Sarah Berhane; Melissa Taylor; Ada Adriano; Clare Davenport; Sabine Dittrich; Devy Emperador; Yemisi Takwoingi; Jane Cunningham; Sophie Beese; Julie Domen; Janine Dretzke; Lavinia Ferrante di Ruffano; Isobel M Harris; Malcolm J Price; Sian Taylor-Phillips; Lotty Hooft; Mariska Mg Leeflang; Matthew Df McInnes; René Spijker; Ann Van den Bruel
Journal:  Cochrane Database Syst Rev       Date:  2021-03-24

2.  Development and evaluation of a Novel RT-PCR system for reliable and rapid SARS-CoV-2 screening of blood donations.

Authors:  Meng Li; Yuwei Zhao; Ying Li; Xue Chen; Dongxia Luo; Mei Luo; Jue Hou; Juan Liu; Humin Liu; Huan Wang; Yufang Dong; Lanjiang Zhang; Maosheng Ji; Xin Zhao; Caibing Wei; Wen Li; Jialiang Gao; Hua Shan; Xuemei Fu
Journal:  Transfusion       Date:  2020-09-10       Impact factor: 3.337

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.