Jacinta E Browne1, Christin A Tiegs-Heiden1, Vance T Lehman1, Zaiyang Long1, Nicholas J Hangiandreou1, Robert E Watson1, Gina K Hesley1, Krzysztof R Gorny1.
Abstract
OBJECTIVE: To provide an initial report that patients with magnetic resonance imaging (MRI) non-conditional cardiac implanted electronic device (CIED) can undergo state-of-the-art magnetic resonance imaging-guided focused (MRgFUS) ablation procedures with careful planning and integration of the procedure into an established CIED MRI practice. PATIENT AND METHODS: We describe an MRgFUS ablation treatment of lumbar facet joints in a patient with an MRI non-conditional CIED (pacemaker), completed in accordance with our institutional CIED/MRI practice guidelines.Entities:
Keywords: CIED, cardiac implanted electronic device; MRI, magnetic resonance imaging; MRgFUS, magnetic resonance imaging-guided focused ultrasound; QA, quality assurance; SAR, specific absorption rate; T2, spin-spin relaxation time
Year: 2020 PMID: 32793874 PMCID: PMC7411156 DOI: 10.1016/j.mayocpiqo.2020.04.010
Source DB: PubMed Journal: Mayo Clin Proc Innov Qual Outcomes ISSN: 2542-4548
FigurePhotograph of the magnetic resonance imaging-guided focused ultrasound (MRgFUS) system with the pre-treatment quality assurance (QA) setup. MRgFUS table docked into the magnetic resonance imaging (MRI) scanner (black arrows) and connected to Exablate control electronics via “umbilical” cables (dashed orange oval). The pelvic coil with the pre-treatment QA phantom inside is indicated with the yellow dotted oval.
Details of MR Sequences and Image Acquisition That Were Used During the MRgFUS Ablation Procedure and the Corresponding Whole-Body SAR Values for Each Image Sequencea
| Sequence | Purpose | Whole-body SAR (W/kg) |
|---|---|---|
| Single-shot fast spin echo pulse sequence | Calibration –transducer localization in MRI coordinates performed using QA phantom as SAR exceeded pre-determined threshold of 1.5 W/kg | 1.73 |
| (TE/TR, 126/2108 ms; ETL, 16; FA, 90°; NEX, 1; slice, 6 mm; skip, 1.5 mm; FOV, 440 mm; BW, 31 KHz; matrix, 256 × 224 mm2) | ||
| 3D-localization T2 (TE/TR, 1.42/4.98 ms; ETL, 1; FA, 30°; NEX, 1; slice, 7 mm; skip, 12 mm; FOV, 440 mm; BW, 244 Hz/px; matrix, 256 × 224 mm2) | Anatomical reference images of treatment areas for MRgFUS treatment planning | 0.26 |
| Fast spin echo sequence – axial T2 (TE/TR, 81/3400 ms; ETL, 10; FA, 90°; NEX, 1; slice, 3 mm; skip, 3 mm; FOV, 160 mm; BW, 122 Hz/px; matrix, 256 × 192 mm2) | Pre-ablation planning images | 1.13 |
| Fast spin echo sequence – sagittal T2 (TE/TR, 81/4200 ms; ETL, 12; FA, 90°; NEX, 1; slice, 3 mm; skip, 4 mm; FOV, 200 mm; BW, 81.4 Hz/px; matrix, 256 × 224 mm2) | Pre-ablation planning images | 0.95 |
| Fast spin echo sequence – coronal T2 (TE/TR, 1.42/4.98 ms; ETL, 1; FA, 30°; NEX, 1; slice, 3 mm; skip, 3 mm; FOV, 440 mm; BW, 244 Hz/px; matrix, 256 × 224 mm2) | Pre-ablation planning images | 0.94 |
| Echo planar imaging – gradient echo (phase-sensitive gradient-recalled echo sequences) – axial (TE/TR, 18/250 ms; ETL, 1; FA, 35°; NEX, 2; slice, 3 mm; skip, 1 mm; 5 slices; BW, 62 kHz) | Thermal mapping of the treatment zone | 0.01 |
| Fast spin echo sequence with fat suppression – axial T2 (TE/TR, 40/4701 ms; ETL, 15; FA, 90°; NEX, 1; slice, 3 mm; skip, 3 mm; FOV, 160 mm; BW, 122 Hz/px; matrix, 320 × 192 mm2) | Post-procedure evaluation | 1.12 |
3D = three-dimensional; BW = bandwidth; ETL = echo train length; FA = flip angle; FOV = field-of-view; MR = magnetic resonance; MRgFUS = magnetic resonance imaging-guided focused ultrasound; MRI = magnetic resonance imaging; NEX = number of excitations; QA = quality assurance; SAR = specific absorption rate; T2 = spin-spin relaxation time; TE = echo time; TR = time to repetition.