Literature DB >> 32759627

Effectiveness and Technical Considerations of Solitaire Platinum 4×40 mm Stent Retriever in Mechanical Thrombectomy with Solumbra Technique.

Ho Jun Yi1,2, Jae Hoon Sung1, Dong Hoon Lee1, Seung Yoon Song1.   

Abstract

OBJECTIVE: The Solitaire Platinum 4×40 mm stent retriever contains radiopaque markers with a long length. We evaluated the effect of Solitaire Platinum 4×40 mm stent retriever in Solumbra technique thrombectomy, and compared it with shorter Solitaire stent retrievers.
METHODS: A total of 70 patients who underwent Solumbra technique thrombectomy with equal diameter (4 mm) and different length (40 vs. 20 mm) Solitaire stent retrievers were divided into two groups : the Solitaire Platinum 4×40 mm stent (4×40) group and the Solitaire FR 4×20 mm stent or Solitaire Platinum 4×20 mm stent (4×20) group. The clinical outcomes, Thrombolysis in Cerebral Infarction score, the first pass reperfusion status, and complications were evaluated and compared between the two groups. Multivariate analysis was performed to evaluate the predictive factors for reperfusion and complete reperfusion from the first pass.
RESULTS: Higher first-pass reperfusion and complete reperfusion were achieved in the 4×40 group (68.0% and 48.0%) than in the 4×20 group (46.7% and 33.3%; p=0.004 and 0.007, respectively). In multivariate analysis, radiopaque device and longer stent retriever were correlated with first-pass reperfusion (p=0.014 and 0.008, respectively) and first-pass complete reperfusion (p=0.022 and 0.012, respectively).
CONCLUSION: Our study demonstrated the usefulness of the Solitaire Platinum 4×40 mm stent retriever, which led to higher firstpass reperfusion and complete reperfusion rates than the Solitaire FR 4×20 mm stent or the Solitaire Platinum 4×20 mm stent, especially in Solumbra technique thrombectomy.

Entities:  

Keywords:  Angiography; Reperfusion; Stents; Stroke; Thrombectomy

Year:  2020        PMID: 32759627      PMCID: PMC7819789          DOI: 10.3340/jkns.2020.0046

Source DB:  PubMed          Journal:  J Korean Neurosurg Soc        ISSN: 1225-8245


INTRODUCTION

Mechanical thrombectomy (MT) is the primary modality for the treatment of acute ischemic stroke (AIS) caused by large artery occlusion (LAO). In addition, the MT technique continues to evolve in terms of its effectiveness and diversity. Initially, the effectiveness of stent retriever techniques was demonstrated via several randomized controlled trials and meta-analyses, and the usefulness of contact aspiration technique was shown in several studies [2,4,5,9,10,17,20,29,30]. In addition, a number of studies demonstrated that the chance of first-pass reperfusion was increased in Solumbra technique thrombectomy (stent retriever with simultaneous aspiration) [8,16,21,22]. Along with the evolution of MT techniques, devices for thrombectomy have also developed. The Solitaire AB stent (ev3; Covidien, Irvine, CA, USA) was the first stent-retriever device used in MT, and was developed further into the Solitaire flow restoration (FR) device and subsequently to the Solitaire 2 revascularization device. Solitaire is a nitinol stent retriever with laser-cut, attached to a push-wire with a closedcell scaffolding design [27,28]. However, the Solitaire FR stent retriever had a major disadvantage of poor visualization. The next generation Solitaire stent retriever, the Solitaire Platinum (Medtronic, Dublin, Ireland) has radiopaque platinum markers, which are spaced 10 mm apart. The use of a radiopaque device is known to be associated with first-pass reperfusion [12,18]. In addition to the characteristics of the stent retriever, the size of the stent may affect the outcomes of MT. The appropriate stent diameter is disputed, and several reports suggested conflictingly, that a stent with a larger or smaller diameter was better [13,31,33]. However, in terms of stent length rather than diameter, Haussen et al. [12] reported that longer stent retrievers enhance thrombectomy performance, probably due to the larger device-thrombus interaction area and smaller chances of missing the target site. For these reasons, the authors used Solitaire Platinum 4×40 mm stent retriever with a Solumbra technique. Therefore, the purpose of our study was to assess the potential benefits of SolitairePlatinum 4×40 mm stent retriever with its radiopacity and long length, and compared it with other Solitaire stent retrievers, especially in patients who underwent MT with Solumbra technique. In addition, the technical considerations involved in Solumbra technique thrombectomy with Solitaire Platinum 4×40 mm stent retriever are also discussed.

MATERIALS AND METHODS

Study population

This retrospective study with prospectively collected data was approved by the Local Institutional Review Board (IRB No. VC17RESI0048) of each participating center. The data of patients with LAO who underwent Solumbra technique thrombectomies with a Solitaire stent retriever of equal diameter (4 mm) between January 2018 and July 2019 were obtained from each institution’s stroke database and evaluated. All of the procedures were performed by two experienced neuro-interventionalists at two centers. A total of 70 patients were identified for enrollment. These patients were dichotomized into : 1) a 4×40 group using 4×40 mm Solitaire stent retrievers (Solitaire Platinum 4×40 mm stent retriever) and 2) a 4×20 group using 4×20 mm Solitaire stent retrievers (Solitaire FR 4×20 mm or Solitaire Platinum 4×20 mm stent retrievers). Prior to thrombectomy, intravenous tissue plasminogen activator (IV t-PA, alteplase) was administered to patients within 4.5 hours after stroke onset at a maximum dose of 0.9 mg/kg in accordance with the European Cooperative Acute Stroke Study (ECASS) III trial [11]. The inclusion criteria were as follows : 1) occlusion of the distal intracranial carotid artery (ICA), middle cerebral artery (MCA, M1, or M2), or posterior circulation (vertebral artery, or basilar artery) established by computed tomography angiography (CTA); 2) neurologic deficits; and 3) Solumbra technique thrombectomy with 4 mm Solitairestent retriever. The exclusion criteria were : 1) the detection of hemorrhage on the initial CT scan; 2) presence of a large ischemic core with an Alberta Stroke Program Early CT Score (ASPECTS) ≤6; 3) MT with a simple stent retriever or the contact aspiration technique. and (4) Solumbra technique thrombectomy with a stent retriever other than a 4 mm Solitaire stent retriever, such as a Solitaire FR 6×30 mm, a Solitaire Platinum 6×40 mm, a Trevo XP Provue (Stryker Neurovascular, Fremont, CA, USA), Eric (MicroVention Terumo, Tustin, CA, USA), or Revive (Codman Neurovascular, San Jose, CA, USA) devices. All patients underwent CT immediately after the intervention to evaluate hemorrhage. In addition, vessel status and hemorrhage were evaluated by CTA 24 hours after MT. Hemorrhagic transformation of the infarct and vessel status was confirmed by magnetic resonance angiography (MRA) with susceptibility-weighted imaging on day 7 after MT.

Device selection and Solumbra technique

Solumbra technique thrombectomy with a 4 mm Solitaire stent retriever was performed in patients with AIS caused by LAO. All of the MT procedures were performed using the Solumbra technique, entailing thrombus retraction with a stent retriever, such as the Solitaire FR 4×20 mm (ev3; Covidien) or the Solitaire Platinum 4×20 mm or 4×40 mm (Medtronic) and simultaneous aspiration with a 5 Fr SOFIA (Sofia 5; MicroVention-Terumo, Tustin, CA, USA) or a 6 Fr AXS Catalyst 6 (CAT6; Stryker Neurovascular, Mountain View, CA, USA). In almost cases, an 8 Fr balloon-guiding catheter (BGC) (FlowGate2 [FG2]; Stryker Neurovascular, Fremont) was applied. An illustration of MT with the Solumbra technique is shown in (Fig. 1). Under consciousness sedation, an 8 Fr FG2 BGC was placed in the proximal ICA. A microcatheter (Excelsior XT-18; Stryker Neurovascular, Fremont) with a microwire was advanced as distally as possible from the occlusion site, and the aspiration catheter was approached close to the occlusion site, followed by deployment of the stent retriever in the distal portion of the occlusion site. After the 3 to 5 minutes of waiting after stent deployment to promote stent clot integration [32], the aspiration catheter with suction and the stent retriever were slowly removed as a single unit. At this time, a dual aspiration technique was used at the BGC with manual aspiration [15].
Fig. 1.

Solumbra technique thrombectomy with a Solitaire platinum 4×40 mm stent retriever. A : Initial angiography revealed left M1 occlusion. B : A non-subtracted image showed a deployed Solitaire platinum 4×40 mm stent retriever (black arrow : the proximal end of the working device; black dotted arrow : the distal end of the working device) and a 6 French Catalyst (CAT6) aspiration catheter (white arrow : distal tip of the aspiration catheter). C : Using a weak pull of the Solitaire platinum 4×40 mm stent retriever, the distal tip of the CAT6 aspiration catheter is moved to the proximal end of the working length of the stent retriever. To contact the thrombus, the distal tip of the CAT6 should be placed just proximal to the thrombus (black arrow : the proximal end of the working length of the stent; white arrow : distal tip of the aspiration catheter). D : After retrieval of the Solitaire platinum 4×40 mm and CAT6 with simultaneous aspiration, angiography revealed the complete reperfusion status of the left M1.

Multivariable factors and outcome and complications

Multivariable factors including patient sex, age, risk factors, prior stroke or transient ischemic attack (TIA), history of previous anti-platelet or anti-coagulant use, ASPECTS score, application of IV t-PA, stroke etiology by Trial of ORG 10172 in Acute Stroke Treatment (TOAST) classification [1], site of the arterial occlusion, and procedure time (time from groin puncture to reperfusion) were reviewed. All patients underwent clinical assessment using the National Institutes of Health Stroke Scale (NIHSS) (range, 0–42; with higher scores indicating more severe neurologic deficit) and modified Rankin Scale (mRS) scores at 3 months (a favorable clinical outcome was defined as mRS ≤2). The radiologic results were evaluated according to the thrombolysis in cerebral infarction (TICI) grading system, successful recanalization (TICI grade of 2b or 3), first-pass reperfusion (modified TICI, mTICI 2b or 3 with the first pass), and first-pass complete reperfusion (mTICI 3 with the first pass). Various complications, such as post-thrombectomy hemorrhage at 24 hours, symptom-related hemorrhage at 24 hours, vessel perforation, arterial dissection, distal emboli, hemorrhagic transformation of the infarct on MRA 7 days after MT, and 3-month mortality were reviewed. Post-thrombectomy hemorrhage at 24 hours was defined as the occurrence of intracerebral hemorrhage or subarachnoid hemorrhage on a CTA scan obtained 24 hours after the procedure. All multimodal factors and clinical data were analyzed by all authors.

Statistical analyses

All data were analyzed using Stata Statistical Software, release 15 (Stata, College Station, TX, USA). Between-group comparisons were calculated using Student’s t-test/Mann-Whitney U test or chi-squared test/Fisher’s exact test. Multivariate logistic regression analyses were performed for variables with <0.2 level of significance in univariate analysis. Two-tailed p-value of ≤0.05 were considered to indicate significant difference.

RESULTS

Baseline characteristics and procedure detail

Twenty-five patients (14 males, 56.0%) were treated via Solumbra technique thrombectomy with 4×40 mm Solitaire Platinum stent retriever (4×40 group), and 45 (27 males, 60.0%) underwent Solumbra technique thrombectomies with 4×20 mm Solitaire stent retrievers (4×20 group). The mean age of the patients in the 4×40 group was 72 years (standard deviation [SD], 12.7) and that of the 4×20 group was 69 years (SD, 13.1). There were no significant differences in multiple risk factors such as hypertension, diabetes mellitus, atrial fibrillation, coronary artery disease, dyslipidemia, smoking, history of prior stroke or TIA, history of anti-platelet use, previous anti-coagulation use, the proportion of left hemisphere strokes, and the rate of IV t-PA application, between 4×40 and 4×20 groups. The median initial NIHSS value, and ASPECTS scores were 10 and 9, respectively, in the 4×40 group and 8 and 9, respectively, in the 4×20 group, without significant difference. There was no significant difference in stroke etiology between the two groups. The site of arterial occlusion in each group was as follows (4×40 group : 4×20 group) : MCA, 13 : 27; distal ICA, 10 : 13; and posterior circulation, 2 : 5. In terms of procedure, there were no statistically significant differences in procedure time and the rate of BGC application between the two groups (p=0.247 and 0.776, respectively). The distribution ratio of the two-aspiration catheters (CAT6 : Sofia5) was 21 : 4 in the 4×40 group, and 33 : 12 in the 4×20 group (p=0.186). Radiopaque devices were significantly more common in the 4×40 group (p=0.001), suggesting that 25 non-radiopaque Solitaire FR 4×20 mm stents were included in the 4×20 group (Table 1).
Table 1.

Baseline characteristics of patients in Solumbra technique thrombectomy, according to the length of Solitaire stent retriever

Variable4×40 (25 patients)4×20 (45 patients)p-value[*]
Dermographic
 Sex, male14 (56.0)27 (60.0)0.601
 Age (years)72.0±12.769.0±13.10.449
 HTN15 (60.0)25 (55.6)0.552
 DM5 (20.0)10 (22.2)0.680
 Af8 (32.0)12 (26.7)0.346
 CAD5 (20.0)9 (20.0)0.889
 Dyslipidemia7 (28.0)16 (35.6)0.465
 Smoking5 (20.0)13 (28.9)0.263
 Prior stroke or TIA4 (16.0)7 (15.6)0.793
 Previous anti-platelet use6 (24.0)9 (20.0)0.571
 Previous anti-coagulant use3 (12.0)4 (8.9)0.466
 Occlusions in the left hemisphere14 (56.0)24 (53.3)0.702
 IV t-PA8 (32.0)18 (40.0)0.109
 Initial NIHSS10 (2–18)8 (2–18)0.127
 ASPECTS score9 (6–10)9 (6–10)0.924
Etiology of stroke
 Cardio-embolic10 (40.0)22 (48.9)0.297
 Atherosclerosis7 (28.0)10 (22.2)0.382
 Dissection1 (4.0)1 (2.2)0.307
 Other or undetermined7 (28.0)12 (26.7)0.693
Site of arterial occlusion
 MCA13 (52.0)27 (60.0)0.419
 Distal ICA10 (40.0)13 (28.9)0.310
 Posterior circulation2 (8.0)5 (11.1)0.507
Procedure detail
 Procedure time (minutes)34 (14–82)46 (15–91)0.247
 Use of BGC24 (96.0)42 (93.3)0.776
 Aspirationcatheter,CAT6:Sofia521 : 433 : 120.186
 Radiopaque stent retriever25 (100.0)20 (44.4)0.001[]

Values are presented as mean±standard deviation, median (interquartile range), or number (%).

p-values are calculated by chi-square test or Fisher’s exact test and Mann-Whitney U test or Student’s t-test.

Statistically significant.

4×40 : Solitaire Platinum 4×40 mm stent retriever, 4×20 : Solitaire 2 FR 4×20 mm stent retriever + Solitaire Platinum 4×20 mm stent retriever, HTN : hypertension, DM : diabetes mellitus, Af : atrial fibrillation, CAD : coronary artery disease, TIA : transient ischemic attack, IV t-PA : intravenous tissue-plasminogen activator, NIHSS : National Institutes of Health Stroke Scale, ASPECT : Alberta Stroke Program Early CT, MCA : middle cerebral artery, ICA : internal carotid artery, BGC : balloon guide catheter, CAT6 : 6 Fr Catalyst, Sofia5 : 5 Fr Sofia

Outcomes and complications

No statistically significant difference was found in the rate of successful recanalization, between the two groups (4×40 group : 4×20 group, 96.0% : 95.6%; p=0.842). However, the proportion of first-pass reperfusions and first-pass complete reperfusions was significantly higher in the 4×40 group, compared to the 4×20 group (68.0% vs. 46.7% and 48.0% vs. 33.3%, p=0.004 and 0.007, respectively). There were no significant differences in complications, between the two groups (Table 2).
Table 2.

Outcomes and complications of patients in Solumbra technique thrombectomy, according to the length of Solitaire stent retriever

4×40 (25 patients)4×20 (25 patients)p-value[*]
Outcome
 Successful recanalization24 (96.0)43 (95.6)0.842
 First-pass reperfusion17 (68.0)21 (46.7)0.004[]
 First-pass complete reperfusion12 (48.0)15 (33.3)0.007[]
 Favorable 3 months mRS14 (56.0)23 (51.1)0.592
Complication
 Post thrombectomy hemorrhage2 (8.0)4 (8.9)0.627
 Symptomatic hemorrhage1 (4.0)2 (4.4)0.793
 Vessel perforation0 (0.0)0 (0.0)1.000
 Arterial dissection0 (0.0)0 (0.0)1.000
 Distal emboli2 (8.0)4 (8.9)0.679
 Hemorrhagic transformation of infarct4 (16.0)9 (20.0)0.391
 3 months mortality2 (8.0)4 (8.9)0.497

Values are presented as number (%).

p-values are calculated by chi-square test or Fisher’s exact test and Mann-Whitney U test or Student’s t-test.

Statistically significant.

4×40 : Solitaire Platinum 4×40 mm stent retriever, 4×20 : Solitaire 2 FR 4×20 mm stent retriever + Solitaire Platinum 4×20 mm stent retriever, Successful recanalization : thrombolysis in cerebral infarction 2b or 3, mRS : modified Rankin Scale, Favorable mRS : 90 days mRS ≤2, First-pass reperfusion : modified thrombolysis in cerebral infarction 2b or 3 with the first pass, First-pass complete reperfusion : modified thrombolysis in cerebral infarction 3 with the first pass

Predictors of reperfusion and full reperfusion with first pass

In multivariate logistic regression analysis, the use of radiopaque Solitaire stent retriever (odds ratio [OR], 2.08; 95% confidence interval [CI], 1.22–3.84; p=0.014), and longer Solitaire Platinum 4×40 mm stent retrievers (OR, 2.81; 95% CI, 1.34–4.10; p=0.008) were correlated with first-pass reperfusion. In addition, multivariate logistic regression analysis for first-pass complete reperfusion indicated that the use of radiopaque devices (OR, 1.88; 95% CI, 1.20–3.12; p=0.022) and longer Solitaire Platinum 4×40 mm stents (OR, 2.12; 95% CI, 1.48–3.76; p=0.012) were independent predictors (Table 3).
Table 3.

Multivariable regression for first-pass reperfusion and first-pass TICI3 reperfusion

OR95% CIp-value[*]
First-pass reperfusion
 Initial NIHSS2.950.61–4.380.442
 IV t-PA0.970.58–2.260.367
 CAT6 Aspiration catheter1.860.78–3.140.172
 Radiopaque stent retriever2.081.22–3.840.014[]
 Long stent retriever (Platinum 4×40 mm)2.811.34–4.100.008[]
First-pass complete reperfusion
 Initial NIHSS2.720.58–4.460.407
 IV t-PA0.820.66–3.080.266
 CAT6 Aspiration catheter2.060.82–3.900.312
 Radiopaque stent retriever1.881.20–3.120.022[]
 Long stent retriever (Platinum 4×40 mm)2.121.48–3.760.012[]

p-values are calculated by multivariate logistic regression analysis.

Statistically significant.

TICI : thrombolysis in cerebral infarction, OR : odds ratio, CI : confidence interval, First-pass reperfusion : modified TICI 2b or 3 with the first pass, NIHSS : National Institutes of Health Stroke Scale, IV t-PA : intravenous tissue-plasminogen activator, CAT6 : 6 French Catalyst, First-pass complete reperfusion : modified TICI 3 with the first pass

DISCUSSION

Currently, the development of various thrombectomy techniques has enhanced the clinical and radiologic outcomes of patients with LAO [6,7,19]. Recent advances in MT pursuit the first pass effect (first-pass reperfusion or first-pass complete reperfusion), beyond simple successful recanalization [25]. The importance of the first-pass effect has been reported in various studies, where it was associated with a significantly higher chance of favorable clinical outcomes with the retrieval of the entire thrombus as a single mass on the first attempt [23,24,34]. Fewer stent passages reduced the risk of procedure-related complications, such as parenchymal hematoma and other vessel damage [3]. Therefore, the utility of the Solumbra technique contributes to reperfusion or complete reperfusion from the first pass [14,21,25]. Based on these studies, the authors used Solumbra technique for MT in most cases. In addition to thrombectomy techniques, the device may be considered a factor affecting the outcomes of MT. There are several controversies related to the selective advantages of stent retrievers with an appropriate diameter. However, recent studies have suggested that stent retrievers with longer lengths and radiopaque characteristics yield better outcomes [12,13,18,31]. In our study, longer and radiopaque stent retrievers were found to increase the first-pass effect. The reperfusion and complete reperfusion rates with first pass were significantly higher in the group with 4×40 mm long stent retrievers than in those with 4×20 mm short stent retrievers. Furthermore, the multivariate analysis in our study showed that stent retrievers with longer length and radiopacity improved the first-pass effect in MT with Solumbra technique. These findings revealed the effectiveness of Solitaire Platinum 4×40 mm stent retrievers compared to other 4 mm Solitaire stent retrievers, especially in MT with the Solumbra technique. Longer stent retrievers have theoretical advantages because of the larger interaction area between the device and the thrombus, with lower chances of missing the target site than their shorter counterparts. In addition, the radiopaque characteristics of the device allow the interventionalists to view the entire stent, and confirm its optimal location and adequate vessel-wall apposition of the stent. Based on these characteristics, the usefulness of the Solitaire Platinum 4×40 mm stent retriever and its technical considerations during the Solumbra technique thrombectomy was shown in (Fig. 2). Under roadmap guidance, the microcatheter was navigated distal to the occlusion site, and the Solitaire Platinum 4×40 mm stent retriever was deployed to cover the distal and proximal ends of the occlusion. The aspiration catheter was navigated close to the clot, and the microcatheter was removed to increase the aspiration flow (Fig. 2A) [26]. Next, the stent retriever was gently pulled with weak force and the distal tip of the aspiration catheter was moved to the distal lesion as close to the occlusion site as possible to contact the thrombus. At this time, the distal tip of the aspiration catheter should be placed just proximal to the site of the thrombus, between the two markers (mid-portion of the marker and the proximal end of the working length marker) of the Solitaire platinum 4×40 mm stent retriever (Fig. 2B). After waiting several minutes to enable stent-clot integration, retrieval of the whole Solumbra system was performed for clot retrieval, with simultaneous continuous suction at the aspiration catheter and dual aspiration through the BGC. When the distal tip of the aspiration catheter was adequately positioned between the mid-portion marker and the proximal end of the working length marker without crossing the mid-portion marker, the thrombus could be retrieved as a single mass (Fig. 2C). However, frequent failures associated with thrombus retrieval occurred when the distal tip of the aspiration catheter was located more distally, beyond the mid-portion of the stent, and re-captured more than half of the stent retriever (Fig. 2D).
Fig. 2.

Representative case of solumbra technique thrombectomy with a Solitaire platinum 4×40 mm stent retriever. A : Roadmap image shows the location of the devices, before their retrieval (black arrow : the proximal end of the working length of the stent; black dotted arrow : mid-portion marker of the stent; white arrow : distal tip of the aspiration catheter). B : By a weak pull of the Solitaire platinum 4×40 mm stent retriever, the distal tip of aspiration catheter (white arrow) should be placed just proximal to the thrombus and between the two markers (black and black dotted arrows) of the Solitaire platinum 4×40 mm stent. C : After retrieval of the Solumbra technique devices, the thrombus was captured by the aspiration catheter and stent retriever. D : Image of the solumbra technique system after retrieval, in which the tip of the aspiration catheter was located distally beyond the midportion marker of the stent retriever.

In the thrombectomy procedures using the Solumbra technique, the aspiration catheter should re-capture less than half or one-third of the stent retriever to ensure an adequate stent retriever capture area to fully integrate with the clot. In addition, during the Solumbra system retrieval, the aspiration catheter is moved further distally depending on its tension, and the stent retriever is captured further by the aspiration catheter. In addition to the advantage of a long stent, the radiopacity of the Solitaire Platinum 4×40 mm stent retriever with its markers 10 mm apart facilitate the correct positioning of the aspiration catheter and adjust the adequate capture range of the stent retrievers, especially during Solumbra technique thrombectomy. Therefore, a long and radiopaque stent retriever such as the Solitaire Platinum 4×40 mm stent retriever, is preferred for MT by Solumbra technique. The major limitations of this study were its relatively small sample size, and the retrospective non-blinded format. Since thrombectomy in the 4×20 mm Solitaire stent retriever group was performed earlier than in the 4×40 mm Solitaire Platinum stent retriever group, it may be assumed that the learning curve affected the outcome. However, all thrombectomy procedures in this study were performed by neuro-interventionalists with several years of experience. Therefore, the learning curve would not have a significant influence on the outcome. Furthermore, the baseline characteristics in the two groups were well balanced and there were no statistically significant differences. The disadvantages of longer stent retrievers include vessel damage or larger interface area with the vessel and consequently more friction. However, in our study, there were no significant differences in hemorrhage and dissection between the two groups. In addition, our study did not include the data associated with Solitaire FR 4×40 mm retriever, which is a non-radiopaque and long Solitaire stent retriever. Other potential unmeasured confounding variables were not controlled for, although every effort was made to adjust for the possibility of spurious results.

CONCLUSION

In this study, the use of the Solitaire Platinum 4×40 mm stent retriever for Solumbra technique thrombectomy in patients with LAO was highly effective, without increased periprocedural complications, compared with prior versions of the Solitaire FR 4×20 mm stent and the Solitaire Platinum 4×20 mm stent. The radiopacity and long length of the Solitaire Platinum 4×40 mm stent retriever may contribute to more achievement of first-pass reperfusion and first-pass complete reperfusion, especially with Solumbra technique thrombectomy. Future prospective multicenter and in vitro studies are needed to corroborate our findings.
  34 in total

1.  Modified Thrombolysis in Cerebral Infarction 2C/Thrombolysis in Cerebral Infarction 3 Reperfusion Should Be the Aim of Mechanical Thrombectomy: Insights From the ASTER Trial (Contact Aspiration Versus Stent Retriever for Successful Revascularization).

Authors:  Cyril Dargazanli; Robert Fahed; Raphael Blanc; Benjamin Gory; Julien Labreuche; Alain Duhamel; Gaultier Marnat; Suzana Saleme; Vincent Costalat; Serge Bracard; Hubert Desal; Mikael Mazighi; Arturo Consoli; Michel Piotin; Bertrand Lapergue
Journal:  Stroke       Date:  2018-04-06       Impact factor: 7.914

2.  PROTECT: PRoximal balloon Occlusion TogEther with direCt Thrombus aspiration during stent retriever thrombectomy - evaluation of a double embolic protection approach in endovascular stroke treatment.

Authors:  Christian Maegerlein; Sebastian Mönch; Tobias Boeckh-Behrens; Manuel Lehm; Dennis M Hedderich; Maria Teresa Berndt; Silke Wunderlich; Claus Zimmer; Johannes Kaesmacher; Benjamin Friedrich
Journal:  J Neurointerv Surg       Date:  2017-12-08       Impact factor: 5.836

3.  Effect of Retrievable Stent Size on Endovascular Treatment of Acute Ischemic Stroke: A Multicenter Study.

Authors:  D Yang; Y Hao; W Zi; H Wang; D Zheng; H Li; M Tu; Y Wan; P Jin; G Xiao; Y Xiong; G Xu; X Liu
Journal:  AJNR Am J Neuroradiol       Date:  2017-06-08       Impact factor: 3.825

4.  Mechanical Thrombectomy Using the new Solitaire™ Platinum Stent-retriever : Reperfusion Results, Complication Rates and Early Neurological Outcome.

Authors:  Johannes Pfaff; Stefan Rohde; Tobias Engelhorn; Arnd Doerfler; Martin Bendszus; Markus Alfred Möhlenbruch
Journal:  Clin Neuroradiol       Date:  2018-01-10       Impact factor: 3.649

5.  Endovascular therapy for ischemic stroke with perfusion-imaging selection.

Authors:  Bruce C V Campbell; Peter J Mitchell; Timothy J Kleinig; Helen M Dewey; Leonid Churilov; Nawaf Yassi; Bernard Yan; Richard J Dowling; Mark W Parsons; Thomas J Oxley; Teddy Y Wu; Mark Brooks; Marion A Simpson; Ferdinand Miteff; Christopher R Levi; Martin Krause; Timothy J Harrington; Kenneth C Faulder; Brendan S Steinfort; Miriam Priglinger; Timothy Ang; Rebecca Scroop; P Alan Barber; Ben McGuinness; Tissa Wijeratne; Thanh G Phan; Winston Chong; Ronil V Chandra; Christopher F Bladin; Monica Badve; Henry Rice; Laetitia de Villiers; Henry Ma; Patricia M Desmond; Geoffrey A Donnan; Stephen M Davis
Journal:  N Engl J Med       Date:  2015-02-11       Impact factor: 91.245

6.  Randomized assessment of rapid endovascular treatment of ischemic stroke.

Authors:  Mayank Goyal; Andrew M Demchuk; Bijoy K Menon; Muneer Eesa; Jeremy L Rempel; John Thornton; Daniel Roy; Tudor G Jovin; Robert A Willinsky; Biggya L Sapkota; Dar Dowlatshahi; Donald F Frei; Noreen R Kamal; Walter J Montanera; Alexandre Y Poppe; Karla J Ryckborst; Frank L Silver; Ashfaq Shuaib; Donatella Tampieri; David Williams; Oh Young Bang; Blaise W Baxter; Paul A Burns; Hana Choe; Ji-Hoe Heo; Christine A Holmstedt; Brian Jankowitz; Michael Kelly; Guillermo Linares; Jennifer L Mandzia; Jai Shankar; Sung-Il Sohn; Richard H Swartz; Philip A Barber; Shelagh B Coutts; Eric E Smith; William F Morrish; Alain Weill; Suresh Subramaniam; Alim P Mitha; John H Wong; Mark W Lowerison; Tolulope T Sajobi; Michael D Hill
Journal:  N Engl J Med       Date:  2015-02-11       Impact factor: 91.245

7.  Mechanical thrombectomy after intravenous alteplase versus alteplase alone after stroke (THRACE): a randomised controlled trial.

Authors:  Serge Bracard; Xavier Ducrocq; Jean Louis Mas; Marc Soudant; Catherine Oppenheim; Thierry Moulin; Francis Guillemin
Journal:  Lancet Neurol       Date:  2016-08-23       Impact factor: 44.182

8.  First Pass Effect: A New Measure for Stroke Thrombectomy Devices.

Authors:  Osama O Zaidat; Alicia C Castonguay; Italo Linfante; Rishi Gupta; Coleman O Martin; William E Holloway; Nils Mueller-Kronast; Joey D English; Guilherme Dabus; Tim W Malisch; Franklin A Marden; Hormozd Bozorgchami; Andrew Xavier; Ansaar T Rai; Michael T Froehler; Aamir Badruddin; Thanh N Nguyen; M Asif Taqi; Michael G Abraham; Albert J Yoo; Vallabh Janardhan; Hashem Shaltoni; Roberta Novakovic; Alex Abou-Chebl; Peng R Chen; Gavin W Britz; Chung-Huan J Sun; Vibhav Bansal; Ritesh Kaushal; Ashish Nanda; Raul G Nogueira
Journal:  Stroke       Date:  2018-02-19       Impact factor: 7.914

9.  More than three passes of stent retriever is an independent predictor of parenchymal hematoma in acute ischemic stroke.

Authors:  Romain Bourcier; Suzana Saleme; Julien Labreuche; Mikael Mazighi; Robert Fahed; Raphael Blanc; Benjamin Gory; Maeva Kyheng; Gaultier Marnat; Serge Bracard; Hubert Desal; Arturo Consoli; Michel Piotin; Bertrand Lapergue
Journal:  J Neurointerv Surg       Date:  2018-11-02       Impact factor: 5.836

10.  Intracranial thrombectomy using the Solitaire stent: a historical vignette.

Authors:  Marta Aguilar Pérez; Elina Miloslavski; Sebastian Fischer; Hansjörg Bäzner; Hans Henkes
Journal:  J Neurointerv Surg       Date:  2011-12-14       Impact factor: 5.836

View more
  1 in total

1.  Effectiveness of Anchoring with Balloon Guide Catheter and Stent Retriever in Difficult Mechanical Thrombectomy for Large Vessel Occlusion.

Authors:  Ho Jun Yi; Bum-Tae Kim; Dong-Sung Shin
Journal:  J Korean Neurosurg Soc       Date:  2022-06-22
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.