Literature DB >> 25948592

The Trevo XP 3×20 mm retriever ('Baby Trevo') for the treatment of distal intracranial occlusions.

Diogo C Haussen1, Andrey Lima1, Raul G Nogueira1.   

Abstract

OBJECTIVE: To report our single-center initial experience using the Trevo XP ProVue Retriever 3×20 mm ('Baby Trevo') for distal intracranial occlusions.
METHODS: We performed a retrospective review of our interventional database for consecutive patients who underwent treatment for acute ischemic stroke with the Baby Trevo device between February and December 2014.
RESULTS: Of 134 patients treated during the study period, 8 underwent treatment with the Baby Trevo for distal occlusions. Their mean age was 51±20 years, 5 (62.5%) were male, mean baseline National Institutes of Health Stroke Scale was 19±5. The mean interval between the time the patient was last-known normal to groin puncture was 527±285 min, and the overall procedural length was 110±26 min. Intra-arterial tissue plasminogen activator was used in 5 (62.5%) cases. The device was used for a total of 10 branches: five middle cerebral artery (four superior M3 and one inferior M3), three anterior cerebral arteries (two pericallosal and one callosomarginal), and two posterior cerebral arteries (one P2 and one P3) occlusions. All patients achieved complete recanalization of the artery targeted by the Baby Trevo (arterial occlusive lesion 3). Good capillary reperfusion (TICI 2b-3) was noted in 6 (75%) cases. One pass was performed in 7 vessels and 2 passes in three branches. Vasospasm was noted in 5 (62.5%) of the vessels and fully responded to intra-arterial vasodilator infusion. Follow-up MRI revealed no infarct within the territory vascularized by the artery targeted by the Baby Trevo in 4 cases, partial infarct in 5, and complete infarct in 1. Two patients had parenchymal hematomas (one PH1 and one PH2). No vessel perforations, dissections, or subarachnoid hemorrhage were noted.
CONCLUSIONS: Our initial data suggest that treatment of distal cerebrovascular occlusions with the Trevo XP 3×20 mm Retriever is feasible. Although this device emerges as a promising technology for small and tortuous distal intracranial vessels, larger studies are still necessary to establish its safety and clinical benefit. Published by the BMJ Publishing Group Limited. For permission to use (where not already granted under a licence) please go to http://www.bmj.com/company/products-services/rights-and-licensing/

Entities:  

Keywords:  Intervention; Stroke; Thrombectomy

Mesh:

Year:  2015        PMID: 25948592     DOI: 10.1136/neurintsurg-2014-011613

Source DB:  PubMed          Journal:  J Neurointerv Surg        ISSN: 1759-8478            Impact factor:   5.836


  17 in total

1.  Acute ischemic stroke in a child due to basilar artery occlusion treated successfully with a stent retriever.

Authors:  Luis Savastano; Joseph J Gemmete; Aditya S Pandey; Christopher Roark; Neeraj Chaudhary
Journal:  BMJ Case Rep       Date:  2015-07-06

2.  Use of microcatheters for suction thrombectomy of acute distal occlusions.

Authors:  Hyon-Jo Kwon; Jong Wook Shin; Bum Soo Park; Jeong-Wook Lim; Hye Seon Jeong; Hee-Jung Song; Jei Kim; Hyeon-Song Koh
Journal:  Neuroradiology       Date:  2018-10-18       Impact factor: 2.804

3.  3MAX catheter for thromboaspiration of downstream and new territory emboli after mechanical thrombectomy of large vessel occlusions: initial experience.

Authors:  Fabio Settecase
Journal:  Interv Neuroradiol       Date:  2018-11-21       Impact factor: 1.610

Review 4.  Current Endovascular Approach to the Management of Acute Ischemic Stroke.

Authors:  Rakesh Khatri; Anantha R Vellipuram; Alberto Maud; Salvador Cruz-Flores; Gustavo J Rodriguez
Journal:  Curr Cardiol Rep       Date:  2018-05-07       Impact factor: 2.931

5.  Risks of Undersizing Stent Retriever Length Relative to Thrombus Length in Patients with Acute Ischemic Stroke.

Authors:  N F Belachew; T Dobrocky; T R Meinel; A Hakim; J Vynckier; M Arnold; D J Seiffge; R Wiest; E I Piechowiak; U Fischer; J Gralla; P Mordasini; J Kaesmacher
Journal:  AJNR Am J Neuroradiol       Date:  2021-10-14       Impact factor: 3.825

Review 6.  Mechanical thrombectomy in pediatric acute ischemic stroke: Clinical outcomes and literature review.

Authors:  Thomas P Madaelil; Akash P Kansagra; DeWitte T Cross; Christopher J Moran; Colin P Derdeyn
Journal:  Interv Neuroradiol       Date:  2016-03-04       Impact factor: 1.610

7.  Mechanical Thrombectomy of Distal Occlusions in the Anterior Cerebral Artery: Recanalization Rates, Periprocedural Complications, and Clinical Outcome.

Authors:  J Pfaff; C Herweh; M Pham; S Schieber; P A Ringleb; M Bendszus; M Möhlenbruch
Journal:  AJNR Am J Neuroradiol       Date:  2015-11-05       Impact factor: 3.825

8.  Endovascular administration of magnetized nanocarriers targeting brain delivery after stroke.

Authors:  Alba Grayston; Yajie Zhang; Miguel Garcia-Gabilondo; Mercedes Arrúe; Abraham Martin; Peter Kopcansky; Milan Timko; Jozef Kovac; Oliver Strbak; Laura Castellote; Sara Belloli; Rosa M Moresco; Maria Picchio; Anna Roig; Anna Rosell
Journal:  J Cereb Blood Flow Metab       Date:  2021-07-06       Impact factor: 6.960

9.  Blind Exchange With Mini-Pinning Technique Using the Tron Stent Retriever for Middle Cerebral Artery M2 Occlusion Thrombectomy in Acute Ischemic Stroke.

Authors:  Takeshi Yoshimoto; Kanta Tanaka; Junpei Koge; Masayuki Shiozawa; Hiroshi Yamagami; Manabu Inoue; Naruhiko Kamogawa; Tetsu Satow; Hiroharu Kataoka; Kazunori Toyoda; Masafumi Ihara; Masatoshi Koga
Journal:  Front Neurol       Date:  2021-05-19       Impact factor: 4.003

Review 10.  Reperfusion therapy in acute ischemic stroke: dawn of a new era?

Authors:  Sonu Bhaskar; Peter Stanwell; Dennis Cordato; John Attia; Christopher Levi
Journal:  BMC Neurol       Date:  2018-01-16       Impact factor: 2.474