BACKGROUND: The global pandemic of Severe Acute Respiratory Syndrome-Related Coronavirus 2 (SARS-CoV2) has resulted in unprecedented challenges for healthcare systems. One barrier to widespread testing has been a paucity of traditional respiratory viral swab collection kits relative to the demand. Whether other sample collection kits, such as widely available MRSA nasal swabs can be used to detect SARS-CoV-2 is unknown. METHODS: We compared simultaneous nasal MRSA swabs (COPAN ESwabs ® 480C flocked nasal swab in 1mL of liquid Amies medium) and virals wabs (BD H192(07) flexible mini-tip flocked nasopharyngeal swabs in 3mL Universal Transport Medium) for SARS-CoV-2 PCR testing using Simplexa COVID-19 Direct assay on patients over a 4-day period. When the results were discordant, the viral swab sample was run again on the Cepheid Xpert Xpress ® SARS-CoV-2 assay. RESULTS: Of the 81 included samples, there were 19 positives and 62 negatives in viral media and 18 positives and 63 negative in the MRSA swabs. Amongst all included samples, there was concordance between the COPAN ESwabs ® 480C and the viral swabs in 78 (96.3%). CONCLUSION: We found a high rate of concordance in test results between COPAN ESwabs ® 480C in Amies solution and BD H192(07) nasopharyngeal swabs in in 3 mL of Universal Viral Transport medium viral media. Clinicians and laboratories should feel better informed and assured using COPAN ESwabs ® 480C to help in the diagnosis of COVID-19.
BACKGROUND: The global pandemic of Severe Acute Respiratory Syndrome-Related Coronavirus 2 (SARS-CoV2) has resulted in unprecedented challenges for healthcare systems. One barrier to widespread testing has been a paucity of traditional respiratory viral swab collection kits relative to the demand. Whether other sample collection kits, such as widely available MRSA nasal swabs can be used to detect SARS-CoV-2 is unknown. METHODS: We compared simultaneous nasal MRSA swabs (COPAN ESwabs ® 480C flocked nasal swab in 1mL of liquid Amies medium) and virals wabs (BD H192(07) flexible mini-tip flocked nasopharyngeal swabs in 3mL Universal Transport Medium) for SARS-CoV-2 PCR testing using Simplexa COVID-19 Direct assay on patients over a 4-day period. When the results were discordant, the viral swab sample was run again on the Cepheid Xpert Xpress ® SARS-CoV-2 assay. RESULTS: Of the 81 included samples, there were 19 positives and 62 negatives in viral media and 18 positives and 63 negative in the MRSA swabs. Amongst all included samples, there was concordance between the COPAN ESwabs ® 480C and the viral swabs in 78 (96.3%). CONCLUSION: We found a high rate of concordance in test results between COPAN ESwabs ® 480C in Amies solution and BD H192(07) nasopharyngeal swabs in in 3 mL of Universal Viral Transport medium viral media. Clinicians and laboratories should feel better informed and assured using COPAN ESwabs ® 480C to help in the diagnosis of COVID-19.
As early as December 2019, a cluster of patients with pneumonia of unknown etiology was identified in Wuhan, China [1] and shortly thereafter, the causative agent was identified as a novel betacoronavirus, the severe acute respiratory syndrome-related coronavirus-2 (SARS-CoV-2) [2]. The spectrum of disease severity ranges from asymptomatic disease to severe pulmonary disease including acute respiratory distress syndrome, multisystem organ failure, and death [1]. Genetic sequencing performed early in the course of the outbreak led to the development of diagnostic testing [3].With testing, SARS-CoV-2 was noted to be responsible for a global pandemic, and as of April 27, 2020, a total of nearly 3 million cases worldwide have been diagnosed, and at least 207,000 have died from COVID-19 (coronavirus disease 2019) [4]. Many experts have advocated for widespread testing. Due to the rapidity of and extensive spread of the infection however, many countries, including the United States, were ill-prepared to employ widespread testing as an effective public health tool [5]. Limitations related to the availability of testing were complex, and occurred on many levels: first, various assays had to be developed and deployed, second, physical testing sites (including drive-through) had to be operationalized, and lastly supplies of ancillary equipment had to be obtained. Key to ancillary equipment was adequate and sustained supply of viral sample collection kits. Many healthcare systems that have overcome the first two of these barriers have been faced with unexpected shortages of swabs and transport media, compounded by slow or non-existent replacements and unreliable supply chains.As with creative interventions to re-purpose personal protective equipment (PPE), healthcare systems have looked for ways to overcome barriers to testing. Interim guidelines issued by the CDC on April 14, 2020 clarified the allowance of other swab types with guidance on specimen collection and transport. In these guidelines, both FDA and CDC allowed for expansion of the specimen types and swab/transport media to accommodate demand for more testing. While nasopharyngeal specimens obtained with a “mini-tip” viral swab and transported in viral transport medium is still the preferred choice for initial testing, acceptable alternatives include nasopharyngeal aspirates and nasal washes as well as swabs of oropharynx, anterior nares, and nasal mid-turbinates. The latter two are only appropriate in symptomatic patients and both nares must be sampled. The anterior nares specimen should be collected using a spun polyester or flocked COPAN ESwabs® used for MRSA detection. These interim guidelines allow for anterior nares and mid-turbinate specimens to be transported in viral transport medium, Amies transport medium, or sterile saline. There is no data, however on the effect on diagnostic accuracy of SARS-CoV-2 tests based on these variable swabs and transport media used for specimen collection from sites other than the nasopharynx.Given the wide availability of specimen collection kits used for MRSA testing in our facility, we sought to determine whether nasal MRSA swabs (flocked polyester COPAN ESwab® 480C) and their transport medium (1mL of liquid Amies medium) can be used to detect SARS-CoV-2 when viral swabs are not available. We therefore sought to assess the concordance between the test results using these two different collection kits. This of paramount importance at this critical time.
Methods
Nasal/nasopharyngeal swabs samples were obtained at the VA Connecticut Healthcare System from either symptomatic patients or patients with a high-risk exposure to COVID-19 presenting to a pre-scheduled “drive-through” outpatient testing center, its emergency department, as well as the inpatient and medical intensive care COVID-19 units over a four-day period. Hospitalized inpatients on the inpatient unit and medical intensive care unit had been initially diagnosed with SARS-CoV-2 from one day to 24 days prior to re-testing.Each patient underwent concomitant testing with both a “viral swab” in Becton, Dickinson and Company H192(07) flexible fine-tip flocked nasopharyngeal swab, with 3 mL of Universal Viral Transport medium (Becton Dickinson H192(07) and an “MRSA swab” COPAN ESwab ® 480C with a flocked polyester nasal swab in 1mL of liquid Amies solution. Trained nurses or physicians collected the specimens. The viral swab specimens were nasopharyngeal; each MRSA swab specimen was collected by accessing a single nare, and extending the swab as far into the mid-turbinate area as the thickness of the swab would allow. After collection, specimens were stored at room temperature until received by the lab and subsequently at -20 degrees Celsius. All specimens were tested within 24 hours of collection. We found that the COPAN ESwab® was not compatible with the Roche Cobas 6800 machine due to the high concentration of mucus and particulate matter that is incompatible with that instrument’s sampling apparatus. We therefore tested both samples with the Simplexa ® COVID-19 Direct Kit (DiaSorin). This real-time RT-PCR assay used manufacturer-provided fluorescent probes with corresponding forward and reverse primers to target two regions of the SARS-CoV-2 genome, the S gene and ORF1ab [6]. When the MRSA and viral swabs were discordant, the viral swab was run on the Xpert® Xpress SARS-CoV-2 (Cepheid). The protocol was considered a quality improvement project and granted exemption by the VA Connecticut Research Office and deemed that review by the VA Connecticut Investigational Review Board (IRB)or R&D Committee was not needed since the objective was optimization of testing supplies. The VA Connecticut IRB provided written confirmation that IRB approval was not necessary. A waiver of informed consent was obtained from its Human Studies Subcommittee.
Results
We sampled 81 unique patients, of which the viral swabs were positive for SARS-CoV-2 PCR on the DiaSorin test in 19. Of these samples, there were 19 positives and 62 negatives in viral media and 18 positives and 63 negatives in the MRSA collection kits (Fig 1). Amongst all included samples, there was concordance between the COPAN ESwabs ® 480C and the viral swabs in 78 (96.3%). In one patient the viral swab was negative and the COPAN ESwabs ® positive and in two patients the viral swab was positive and the COPAN ESwabs ® negative. When the results were discordant, the viral swab was re-tested with the Cepheid assay. When this was done, in two cases, the COPAN ESwab ® 480C result was concordant with the Cepheid platform. Concordance was achieved in one case because a negative result with the viral swab on the DiaSorin assay was presumptive positive on the Cepheid, and in the other case because a negative result on the DiaSorin with the COPAN ESwab® was positive with Cepheid.
Fig 1
Test esults on the Simplexa ® COVID-19 Direct Kit (DiaSorin) comparing viral swabs and MRSA swabs.
Discussion
SARS-CoV-2 infection has overwhelmed healthcare systems internationally and caused unparalleled morbidity and mortality. Efforts to mitigate this infection focus on social distancing, widespread testing, home isolation, and timely implementation of precautions for hospitalized patients. In order to meet the need for diagnosis of COVID-19, several testing platforms have been approved by FDA, and several more may become available in the near future. Currently available reverse-transcriptase PCR tests that have been granted emergency use authorization from the FDA target specific genes of SARS-CoV2, and they are thought to provide relatively comparable and accurate results. However, like most other healthcare systems across the US, our healthcare system is experiencing a shortage of viral respiratory swabs and viral transport medium. While CDC and FDA have modified their sample collection guidelines to include specimens from nares and mid-turbinate using polyester flocked swabs transported in a variety of media, the effect of these on the diagnostic accuracy of the RT-PCR tests is not known. The concept of assessment of validity of SARS-CoV2 testing using COPAN ESwabs ® 480C was therefore born out of necessity.Our primary concerns were the anticipated mismatch in the site of sampling, and the concern about RNA-survival in the Amies solution. The COPAN Eswab® is thicker than the viral respiratory swab because the targeted site of sampling for MRSA is the anterior nares. This is in contrast to the respiratory viral swab, which has a flexible mini-tip and is designed to sample beyond the turbinates, to the nasopharyngeal wall, where the yield of respiratory viruses is typically higher since the local ciliated epithelial cells support viral replication, in contrast to the squamous epithelium of the anterior nares. The yield of SARS-CoV2 from anterior nares sampling has been expected to be inadequate analogous to influenza testing, but data are limited and contradictory [7].Similarly, the Amies solution that accompanies the swabs for bacterial preservation does not contain any antimicrobial agent in order to prevent any compromise in the yield of bacterial genetic material. Instead, it contains salts to provide essential ions, inorganic phosphate to provide buffering, and sodium thioglycolate to provide reduced environment, meant to maintain viability of bacteria during transport. This, however, can be an undesirable medium for viral transport for that very reason. Abundant bacteria, when present, may commence a destructive effect on viral RNA almost immediately after sample collection. Time from specimen collection to receipt laboratory is therefore of utmost importance for such substitutions. To minimize such concerns we ensured that our samples were transported to the laboratory expeditiously, then frozen at -20 degrees Celsius to abort any bacterial activity. We find the high concordance rate between the swabs reassuring for reproducibility of this particular substitution.Concerns regarding SARS-CoV-2 RNA stability in liquid Amies, however, were recently addressed by Rogers et al [8]. They found that SARS-CoV-2 RNA, when spiked into liquid Amies specimen remnants from COVID-19 negative patients, was stable for at least seven days at room temperature and at least 14 days when refrigerated. They did note an increase in cycle threshold values of about 2–3 after two weeks of frozen storage, but those changes did not alter result interpretations. Taken together, their results with ours support the suitability of ESwabs with liquid Amies for COVID-19 testing.It is interesting to note that in two of the three cases when the results of the viral swab and COPAN ESwab ® were discordant on the DiaSorin platform, the Cepheid platform gave concordant results. In both cases the concordance was achieved because the Cepheid obtained a positive result where the Diasorin had obtained a negative. This is perhaps related to the fact that the Cepheid interprets results as positive up to a cycle threshold value of 44, whereas the Diasorin’s cut-off for positives is 40. A recent news report of unpublished data suggests that the DiaSorin test has a false-negative rate of nearly 11% and Cepheid a false negative result in 1.8% [9].An unexpected finding was that the Ct values for detection of the ORF1ab sequence were significantly lower overall with the COPAN ESwab® than for the viral swab. Also, if it weren’t for two outliers, the same would have been true for the Ct values for S gene detection. We speculate that this might be related to the fact that the tip of the COPAN ESwab® is larger than that of the viral swab, and might collect a larger sample with a larger viral load. This would be consistent with our anecdotal observation that specimens collected with the COPAN ESwab® contained more mucus. This finding could also be related to the fact that viral RNA concentration may be higher in the COPAN ESwab® specimens because the volume of liquid Amies transport medium is one third the volume of UTM used with the viral swab (1 mL versus 3 mL). At a minimum, we can conclude that the COPAN ESwab® with liquid Amies transport medium appears as sensitive for detecting SARS-CoV-2 RNA as the viral swab in UTM.There are several limitations to our study. Our sample size is both small and from one center, and needs to be confirmed in future larger studies including multiple centers with multiple technicians obtaining samples. Furthermore, we used two proprietary testing formulations. Whether this is generalizable to other formulations of viral and non-viral sampling systems is not known. Ideally, we would have liked to test each sample on multiple platforms looking for concordance between several testing modalities, as it is conceivable that inhibitory substances may be present in nasal materials or different transport media. However, due to the ongoing relative shortage of available reagents, we could not justify this for this quality improvement project.In addition to its tremendous physical and emotional toll, infection caused by SARS-CoV-2 has stressed supply chains. While some may have used typical MRSA swabs when there have been shortages of viral swabs, heretofore there has been little evidence of their testing characteristics. We found a high rate of concordance in test results between COPAN ESwabs ® 480C in Amies solution and swabs in viral media. Clinicians and laboratories should feel better informed and assured that using COPAN ESwabs ® 480C to help in the diagnosis of COVID-19 is acceptable in resource-limited settings.(XLSX)Click here for additional data file.10 Jun 2020PONE-D-20-14345SARS-CoV-2 detection in setting of viral swabs scarcity: are MRSA swabs and viral swabs equivalent?PLOS ONEDear Dr. Federman,Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.As mentioned by Reviewer#1, this study was done using limited number of samples from one location. This should be mentioned in the text as a limitation of this study. All other comments of all reviewers should be addressed on the revised manuscript in addition to the rebuttal letter.Please ensure that your decision is justified on PLOS ONE’s publication criteria and not, for example, on novelty or perceived impact.Please submit your revised manuscript by Jul 25 2020 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.Please include the following items when submitting your revised manuscript:A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocolsWe look forward to receiving your revised manuscript.Kind regards,Selvakumar Subbian, Ph.D.Academic EditorPLOS ONEJournal Requirements:When submitting your revision, we need you to address these additional requirements.1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found athttps://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf andhttps://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf2. Thank you for stating in your ethics statement"The protocol was deemed a quality improvement project and granted exemption by the VA Connecticut Research Office and deemed that review by the VA Connecticut Investigational Review Board and R and D Committee was not needed."Please clarify whether the VA Connecticut Research Office is your ethics committee.Please also clarify whether informed consent was specifically waived by your ethics committee.If not, please discuss whether all data were fully anonymized before you accessed them or whether written informed consent was obtained from patients.3. In your Data Availability statement, you have not specified where the minimal data set underlying the results described in your manuscript can be found. PLOS defines a study's minimal data set as the underlying data used to reach the conclusions drawn in the manuscript and any additional data required to replicate the reported study findings in their entirety. All PLOS journals require that the minimal data set be made fully available. For more information about our data policy, please see http://journals.plos.org/plosone/s/data-availability.Upon re-submitting your revised manuscript, please upload your study’s minimal underlying data set as either Supporting Information files or to a stable, public repository and include the relevant URLs, DOIs, or accession numbers within your revised cover letter. For a list of acceptable repositories, please see http://journals.plos.org/plosone/s/data-availability#loc-recommended-repositories. Any potentially identifying patient information must be fully anonymized.Important: If there are ethical or legal restrictions to sharing your data publicly, please explain these restrictions in detail. Please see our guidelines for more information on what we consider unacceptable restrictions to publicly sharing data: http://journals.plos.org/plosone/s/data-availability#loc-unacceptable-data-access-restrictions. Note that it is not acceptable for the authors to be the sole named individuals responsible for ensuring data access.We will update your Data Availability statement to reflect the information you provide in your cover letter.4. We noted in your submission details that a portion of your manuscript may have been presented or published elsewhere:'The entire manuscript was submitted simultaneous with MedRxiv and published there. There was supposed to be simultaneous submission to your journal but there was a problem with submission (linking of my account with the account of a Dr. David Feillin- your office knows about this)'Please clarify whether this publication was peer-reviewed and formally published.If this work was previously peer-reviewed and published, in the cover letter please provide the reason that this work does not constitute dual publication and should be included in the current manuscript.5. PLOS requires an ORCID iD for the corresponding author in Editorial Manager on papers submitted after December 6th, 2016. Please ensure that you have an ORCID iD and that it is validated in Editorial Manager. To do this, go to ‘Update my Information’ (in the upper left-hand corner of the main menu), and click on the Fetch/Validate link next to the ORCID field. This will take you to the ORCID site and allow you to create a new iD or authenticate a pre-existing iD in Editorial Manager. Please see the following video for instructions on linking an ORCID iD to your Editorial Manager account: https://www.youtube.com/watch?v=_xcclfuvtxQ6. Please ensure that you refer to Figure 1 in your text as, if accepted, production will need this reference to link the reader to the figure.[Note: HTML markup is below. Please do not edit.]Reviewers' comments:Reviewer's Responses to QuestionsComments to the Author1. Is the manuscript technically sound, and do the data support the conclusions?The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.Reviewer #1: PartlyReviewer #2: Yes**********2. Has the statistical analysis been performed appropriately and rigorously?Reviewer #1: NoReviewer #2: N/A**********3. Have the authors made all data underlying the findings in their manuscript fully available?The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.Reviewer #1: YesReviewer #2: Yes**********4. Is the manuscript presented in an intelligible fashion and written in standard English?PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.Reviewer #1: YesReviewer #2: Yes**********5. Review Comments to the AuthorPlease use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)Reviewer #1: The manuscript entitled SARS-CoV-2 detection in setting of viral swab scarcity: are MRSA swabs and viral swabs equivalent? describes the use of MRSA swabs with Amies media and Viral swab with universal transport media for sample collection from SARS-CoV-2 infectedpatients. Authors indicated that the MRSA swabs yields the comparable results as that of viral swabs. However, this manuscript failed to address the following;1. Sample size is less2. Sampling was done at only one center; sampling in different screening center by different technicians will require to exclude personal variation in sample collection.3. As the Amies media is not known as medium of choice for sample collection for viral diagnosis and it is expected that the sample will be stored for varying time period at room temperature prior to transport to lab to store at -20C; the duration of stability of RNA in the Amies media must be tested prior to recommending them for SARS-CoV-2 sample collection.4. References not included for several sentences in the introduction; For example- "The spectrum of disease severity ranges from asymptomatic disease to severe pulmonary disease including acute respiratory distress syndrome, multisystem organ failure, and death. Genetic sequencing performed early in the course of the outbreak led to thedevelopment of diagnostic testing".....include references.5. The following sentence is not clear and need to be re-written : We sampled 81 unique patients, of which the viral swabs were positive on the DiaSorin test in 19.6. It is needless to mention "Additionally, while our study was conducted only in Veterans, we do not believe thereshould be a difference in test results between Veterans and non-Veterans for the positive ornegative predictive value of testing for SARS-CoV-2 infection"7. Under methods, the sentence "virals wabs (BD H192(07) flexible mini-tip flocked " need to be written as "viral swabs (BD H192(07) flexible mini-tip flocked"8. The very short and incomplete sentences [ for example "The viral swab specimens were nasopharyngeal" needs to be merged with neighboring sentence or can be re-written appropriately.Reviewer #2: The tremendous need for swab test justifies the work by Daniel et.al. and this would for sure be helpful in terms of increasing the capacity to test. The manuscript is technically sound and the data support the conclusions. Although the sample number is small, the importance of work makes it acceptable.Percentage correlation of the positive and negative test of two different swabs together with results from Cepheid Xpert Xpress ® SARS-CoV-2 assay justify the conclusions with appropriate control.Minor revisions: The authors should describe the primers used for the amplification of CoV2 RNA with details of the amplification conditions in the method section.The manuscript is well written and acceptable with minor reversions.**********6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.If you choose “no”, your identity will remain anonymous but your review may still be made public.Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.Reviewer #1: NoReviewer #2: Yes: Alok Choudhary[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.30 Jun 2020see accompanying documentSubmitted filename: Response to Reviewers PLOS ONE.docxClick here for additional data file.22 Jul 2020SARS-CoV-2 detection in setting of viral swabs scarcity: are MRSA swabs and viral swabs equivalent?PONE-D-20-14345R1Dear Dr. Federman,We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.Kind regards,Selvakumar Subbian, Ph.D.Academic EditorPLOS ONEAdditional Editor Comments (optional):Reviewers' comments:Reviewer's Responses to QuestionsComments to the Author1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.Reviewer #1: (No Response)**********2. Is the manuscript technically sound, and do the data support the conclusions?The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.Reviewer #1: (No Response)**********3. Has the statistical analysis been performed appropriately and rigorously?Reviewer #1: (No Response)**********4. Have the authors made all data underlying the findings in their manuscript fully available?The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.Reviewer #1: (No Response)**********5. Is the manuscript presented in an intelligible fashion and written in standard English?PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.Reviewer #1: (No Response)**********6. Review Comments to the AuthorPlease use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)Reviewer #1: Authors have addressed all the comments.However, due to the small sample size tested for the evaluation of MRSA swabs, it will be more appropriate if the title of manuscript changed as "Pilot study on the applicability of MRSA swabs for SARS-CoV-2 sampling in setting of viral swabs scarcity".**********7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.If you choose “no”, your identity will remain anonymous but your review may still be made public.Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.Reviewer #1: Yes: Santhamani Ramasamy28 Jul 2020PONE-D-20-14345R1SARS-CoV-2 detection in setting of viral swabs scarcity: are MRSA swabs and viral swabs equivalent?Dear Dr. Federman:I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.If we can help with anything else, please email us at plosone@plos.org.Thank you for submitting your work to PLOS ONE and supporting open access.Kind regards,PLOS ONE Editorial Office Staffon behalf ofDr. Selvakumar SubbianAcademic EditorPLOS ONE
Authors: Amy A Rogers; Russell E Baumann; Gwynngelle A Borillo; Ron M Kagan; Hollis J Batterman; Marzena M Galdzicka; Elizabeth M Marlowe Journal: J Clin Microbiol Date: 2020-07-23 Impact factor: 5.948
Authors: Jo-Ann L Stanton; Rory O'Brien; Richard J Hall; Anastasia Chernyavtseva; Hye Jeong Ha; Lauren Jelley; Peter D Mace; Alexander Klenov; Jackson M Treece; John D Fraser; Fiona Clow; Lewis Clarke; Yongdong Su; Harikrishnan M Kurup; Vyacheslav V Filichev; William Rolleston; Lee Law; Phillip M Rendle; Lawrence D Harris; James M Wood; Thomas W Scully; James E Ussher; Jenny Grant; Timothy A Hore; Tim V Moser; Rhodri Harfoot; Blair Lawley; Miguel E Quiñones-Mateu; Patrick Collins; Richard Blaikie Journal: Front Public Health Date: 2022-01-24