Universal nonselective excitation and refocusing pulses with improved robustness to off-resonance for Magnetic Resonance Imaging at 7 Tesla with parallel transmission.

L Van Damme1,2, F Mauconduit2, T Chambrion1,3, N Boulant2, V Gras2.   

Abstract

PURPOSE: In MRI at ultra-high field, the k T -point and spiral nonselective (SPINS) pulse design techniques can be advantageously combined with the parallel transmission (pTX) and universal pulse techniques to create uniform excitation in a calibration-free manner. However, in these approaches, pulse duration is typically increased as compared to standard hard pulses, and excitation quality in regions exhibiting large resonance frequency offsets often suffer. This limitation is inherent to structure of k T -point or SPINS pulse, and likely can be mitigated using parameterization-free pulse design approaches.
METHODS: The Gradient Ascent Pulse Engineering (GRAPE) algorithm was used to design parameterization-free RF and magnetic field gradient (MFG) waveforms for creating 8 ∘ excitation, up to 105 ∘ scalable refocusing and inversion, nonselectively across the brain. Simulations were performed to provide flip angle normalized root-mean-squares error (FA-NRMSE) estimations for the 8 ∘ and the 180 ∘ k T -point, SPINS, and GRAPE pulses. GRAPE pulses were tested experimentally with anatomical head scans at 7T.
RESULTS: As compared to k T -points and SPINS, GRAPE provided substantial improvement of excitation, refocusing, and inversion quality at off-resonance while at least preserving the same global FA-NRMSE performance. As compared to k T -points, GRAPE allowed for a substantial reduction of the pulse duration for the 8 ∘ excitation and the 105 ∘ refocusing.
CONCLUSIONS: Parameterization-free universal nonselective pTX-pulses were successfully computed using GRAPE. Performance gains as compared to k T -points were validated numerically and experimentally for three imaging protocols. In its current implementation, the computational burden of GRAPE limits its use to applications where pulse computations are not subject to time constraints.
© 2020 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  GRAPE; RF pulse design; SPINS; kT-point; optimal control; parallel transmission; ultra high field

Mesh:

Year:  2020        PMID: 32755064     DOI: 10.1002/mrm.28441

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  1 in total

1.  Physical limits to human brain B0 shimming with spherical harmonics, engineering implications thereof.

Authors:  Bruno Pinho Meneses; Alexis Amadon
Journal:  MAGMA       Date:  2022-07-13       Impact factor: 2.533

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.