| Literature DB >> 32742757 |
Katsuhide Yamadera1, Hiroto Ishikawa1, Ayame Imai1, Mana Okamoto1, Akiko Kimura1, Osamu Mimura1, Fumi Gomi1.
Abstract
Purpose: The Glaucoma Module Premium Edition (GMPE) software for the SPECTRALIS optical coherence tomography (OCT) is able to automatically track the anatomic centers of both the fovea and the optic disc. We investigated the efficacy of the OCT in measuring the ocular torsion angle before and after strabismus surgery.Entities:
Keywords: Bland–Altman plots; cyclodeviation; cyclotropia; ocular torsion angle; optical coherence tomography
Mesh:
Year: 2020 PMID: 32742757 PMCID: PMC7354857 DOI: 10.1167/tvst.9.3.27
Source DB: PubMed Journal: Transl Vis Sci Technol ISSN: 2164-2591 Impact factor: 3.283
Figure 1.(a) OCT scans the retina, and GMPE automatically detects the centers of the optic disc and fovea, and then the software calculates the ocular torsion angle. (b) Magnification of the ocular torsion angle. The picture shows an ocular torsion angle of –16.5°.
Comparison of Measurements of Ocular Torsion Angle and Changes in Ocular Torsion Angle During Surgery
| Preoperative | Postoperative | |
|---|---|---|
| (Lower confidence limit 95%; upper confidence limit 95%) | (Lower confidence limit 95%; upper confidence limit 95%) | |
| Synoptophore | –8.8 ± 4.1 (–10.13 to –7.52) | –3.1 ± 2.8 (–3.95 to –2.15) |
| Fundus camera | –27.2 ± 6.4 (–29.21 to –25.11) | –18.6 ± 5.9 (–20.51 to –16.72) |
| OCT | –25.9 ± 6.8 (–28.05 to –23.70) | –18.2 ± 5.8 (–20.09 to –16.38) |
Figure 2.Preoperative correlation of ocular torsion angle (a). OCT is moderately correlated with synoptophore. (b) OCT is strongly correlated with fundus photography.
Comparisons Made Using the Bland–Altman plots
| Bland–Altman Plots | ||
|---|---|---|
| Preoperative | Postoperative | |
| (Lower confidence limit 95%; upper confidence limit 95%), | (Lower confidence limit 95%; upper confidence limit 95%), | |
| OCT and synoptophore | –18.88 to –15.22, | –16.94 to –13.43, |
| OCT and fundus camera | 0.34 to 2.23, | –0.34 to 1.11, |
Bold iatric indicates statistical significance (P < 0.05).
Figure 3.Bland–Altman plots. (a) Comparisons between OCT and the synoptophore did not show good agreement. (b) Comparisons between OCT and the fundus camera showed good agreement.