Abstract
OBJECTIVES: Prolonged survival of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) on environmental surfaces and personal protective equipment may lead to these surfaces transmitting this pathogen to others. We sought to determine the effectiveness of a pulsed-xenon ultraviolet (PX-UV) disinfection system in reducing the load of SARS-CoV-2 on hard surfaces and N95 respirators.Entities:
Year: 2020 PMID: 32741425 PMCID: PMC7443558 DOI: 10.1017/ice.2020.399
Source DB: PubMed Journal: Infect Control Hosp Epidemiol ISSN: 0899-823X Impact factor: 3.254
Experimental Design
| Test Microorganism | Test Device | Treatment Distance | Test Surface | Contact Time | Number |
|---|---|---|---|---|---|
| SARS-CoV-2 | PX-UV device | 1 m | Chamber slide | 0 min | 3 |
| Chamber slide | 1 min | 3 | |||
| Chamber slide | 2 min | 3 | |||
| Chamber slide | 5 min | 3 | |||
| N95 respirator | 0 min | 3 | |||
| N95 respirator | 5 min | 3 |
Note. SARS-CoV-2, severe acute respiratory syndrome coronavirus 2; PX-UV, pulsed-xenon ultraviolet light.
Impact of Pulsed-Xenon Ultraviolet Light on SARS-CoV-2 Inoculated Onto Hard Surfaces
| Test Surface | UV-C Exposure Time, min | Viral Titer [PFU/mL (Log10)] per Carrier | Average | % Reduction | Log10 Reduction Relative to Respective Controls | ||
|---|---|---|---|---|---|---|---|
| 1 | 2 | 3 | |||||
| Slide | 0 | 6.04 | 6.28 | 6.28 | 6.20 | N/A | N/A |
| Slide | 1 | 3 | 2.45 | 2.56 | 2.67 | 99.97 | 3.53 |
| Slide | 2 | 2.38 | <1.3 | <1.3 | <1.66 | >99.997 | >4.54 |
| Slide | 5 | <1.3 | 2.15 | 2.8 | <2.08 | >99.992 | >4.12 |
Note. UV-C, ultraviolet C light; PFU, plaque-forming units; N/A, not applicable.
Impact of Pulsed-Xenon Ultraviolet Light on SARS-CoV-2 Inoculated on N95 Respirators
| Test Surface | UV-C Exposure Time, min | Viral Titer [PFU/mL (Log10)] per Carrier | Average | % Reduction | Log10 Reduction Relative to Respective Controls | ||
|---|---|---|---|---|---|---|---|
| 1 | 2 | 3 | |||||
| N95 Respirator | 0 | 6.84 | 6.04 | 6.18 | 6.35 | N/A | N/A |
| N95 Respirator | 5 | <1.3 | <1.3 | 2.08 | <1.56 | >99.998 | >4.79 |
Note. UV-C, ultraviolet C light; PFU, plaque-forming units; N/A, not applicable.