Comparison of 2 portable human glucometers for the measurement of blood glucose concentration in White New Zealand rabbits.

Kassy G Silva1, Isabella Rotta1, Leandro B Costa1, Cristina S Sotomaior1.   

Abstract

We compared measurements of blood glucose concentrations in 30 healthy adult White New Zealand rabbits using 2 commercial portable glucometers (PGM1 and PGM2) and a laboratory chemical analyzer. Results were analyzed with Pearson correlation, Passing-Bablok regression analysis, Bland-Altman analysis, and a modified error grid. Measurements with PGM1 were significantly correlated (r = 0.37) with those obtained from the laboratory reference method (RM); Bland-Altman and Passing-Bablok analyses indicated no significant systematic or proportional differences (mean difference of -0.26, 95% CI of mean difference of -0.54 to 0.01, and LOA of -1.70 to 1.17); and error grid resulted in 100% of measurements in zone A. No significant correlation (r = -0.05) was detected between PGM2 and RM; Bland-Altman and Passing-Bablok analyses results indicated a mean difference of 2.14, 95% CI of mean difference of 1.67-2.60, and limit of agreement of -0.32 to 4.59, which overestimated blood glucose concentration, with 53% of glucose measurements in error grid zone A and 47% in zone B. PGM1 was considered accurate in normoglycemic rabbits, whereas the use of PGM2 could result in overestimations of glycemia.

Entities:  

Keywords:  blood glucose; glucometer; health; rabbits; spectrophotometry

Mesh:

Substances:

Year:  2020        PMID: 32723169      PMCID: PMC7488964          DOI: 10.1177/1040638720946702

Source DB:  PubMed          Journal:  J Vet Diagn Invest        ISSN: 1040-6387            Impact factor:   1.279


  16 in total

1.  Evaluation of six portable blood glucose meters for measuring blood glucose concentration in dogs.

Authors:  Todd A Cohen; Richard W Nelson; Philip H Kass; Mary M Christopher; Edward C Feldman
Journal:  J Am Vet Med Assoc       Date:  2009-08-01       Impact factor: 1.936

2.  Blood glucose: its measurement and clinical importance.

Authors:  V Marks
Journal:  Clin Chim Acta       Date:  1996-07-15       Impact factor: 3.786

3.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

4.  A new biometrical procedure for testing the equality of measurements from two different analytical methods. Application of linear regression procedures for method comparison studies in clinical chemistry, Part I.

Authors:  H Passing
Journal:  J Clin Chem Clin Biochem       Date:  1983-11

Review 5.  Hypoglycemia in the Newborn.

Authors:  Paul J Rozance; Joseph I Wolfsdorf
Journal:  Pediatr Clin North Am       Date:  2019-02-01       Impact factor: 3.278

Review 6.  13. Management of Diabetes in Pregnancy: Standards of Medical Care in Diabetes-2018.

Authors: 
Journal:  Diabetes Care       Date:  2018-01       Impact factor: 19.112

Review 7.  Practical use of home blood glucose monitoring in feline diabetics.

Authors:  Sara L Ford; Heather Lynch
Journal:  Vet Clin North Am Small Anim Pract       Date:  2013-03       Impact factor: 2.093

Review 8.  Point-of-Care Glucose and Ketone Monitoring.

Authors:  Siew Kim Chong; Erica L Reineke
Journal:  Top Companion Anim Med       Date:  2016-06-14

Review 9.  Common Emergencies in Rabbits, Guinea Pigs, and Chinchillas.

Authors:  Julie DeCubellis
Journal:  Vet Clin North Am Exot Anim Pract       Date:  2016-03-02

10.  Performance of two portable meters and a benchtop analyzer for blood glucose concentration measurement in rabbits.

Authors:  Paolo Selleri; Nicola Di Girolamo; Gianluca Novari
Journal:  J Am Vet Med Assoc       Date:  2014-07-01       Impact factor: 1.936

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.