Biomechanical analysis of the treatment of intertrochanteric hip fracture with different lengths of dynamic hip screw side plates.

Cheng-Chi Wang1,1, Cheng-Hung Lee1,2,1, Ning-Chien Chin1,1, Kun-Hui Chen1,3, Chien-Chou Pan1, Kuo-Chih Su4,3.   

Abstract

BACKGROUND: Dynamic hip screw (DHS) is a common implant used to treat stable-type intertrochanteric hip fractures. There are many factors that can affect the success rate of the surgery, including the length of side plates. It is therefore important to investigate the biomechanical effect of different DHS side plates on bones.
OBJECTIVE: In order to reduce the likelihood of an implant failure, the aim of this study was to use finite element analysis (FEA) to investigate and understand the effect of side plates with different lengths in DHS.
METHODS: In this FEA study, a 3D model with cortical bone, cancellous bone, side plate, lag screw, and cortical screws to simulate the implantation of DHS with different lengths of side plate (2-hole, 4-hole, and 6-hole) for intertrochanteric hip fractures was constructed. The loading condition was used to simulate the force (400 N) on the femoral head and the stress distribution on the lag screw, side plate, cortical screws, and femur was measured.
RESULTS: The highest stress points occured around the region of contact between the screw and the cortical bones. The stress on the femur at the most distal cortical screw was the greatest. The shorter the length of the side plate, the greater the stress on the cortical screws, resulting in an increased stress on the femur surrounding the cortical screws.
CONCLUSIONS: The use of DHS with 2-hole side plate may increase the risk of side plate pull-out. The results of this study provide a biomechanical analysis for selection of DHS implant lengths that can be useful for orthopaedic surgeons.

Entities:  

Keywords:  Biomechanics; dynamic hip screw; finite element analysis; intertrochanteric hip fracture

Year:  2020        PMID: 32716339     DOI: 10.3233/THC-202248

Source DB:  PubMed          Journal:  Technol Health Care        ISSN: 0928-7329            Impact factor:   1.285


  2 in total

1.  Novel Design of Minimal Incision Double Oblique Device for Osteosynthesis (DODO) of Hip: Results of an In-silico Study Based on the Femur Morphometrics of the Northeast (NE) Indian population.

Authors:  Pratik Nag; Bhaskar Borgohain; Souptick Chanda
Journal:  Indian J Orthop       Date:  2021-05-05       Impact factor: 1.251

Review 2.  Dynamic Hip Screw Plate Length in Stable Intertrochanteric Fracture Neck of Femur: A Systematic Review.

Authors:  Aditya Soni; Sandeep Munshi; Niranj G Radhamony; Rajiv Nair; Sachith Sreenivasan
Journal:  Cureus       Date:  2022-03-14
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.