Kieu The Loan Trinh1, Duc Anh Thai2, Woo Ri Chae3, Nae Yoon Lee2. 1. Department of Industrial Environmental Engineering, Gachon University, 1342 Seongnam-daero, Sujeong-gu, Seongnam-si, Gyeonggi-do 13120, Korea. 2. Department of BioNano Technology, Gachon University, 1342 Seongnam-daero, Sujeong-gu, Seongnam-si, Gyeonggi-do 13120, Korea. 3. BioNano Research Institute, Gachon University, 1342 Seongnam-daero, Sujeong-gu, Seongnam-si, Gyeonggi-do 13120, Korea.
Abstract
In the present study, we introduce a new approach for rapid bonding of poly(methyl methacrylate) (PMMA)-based microdevices using an acetic acid solvent with the assistance of UV irradiation. For the anticipated mechanism, acetic acid and UV irradiation induced free radicals on the PMMA surfaces, and acrylate monomers subsequently formed cross-links to create a permanent bonding between the PMMA substrates. PMMA devices effectively bonded within 30 s at a low pressure using clamps, and a clogging-free microchannel was achieved with the optimized 50% acetic acid. For surface characterizations, contact angle measurements and bonding performance analyses were conducted using predetermined acetic acid concentrations to optimize bonding conditions. In addition, the highest bond strength of bonded PMMA was approximately 11.75 MPa, which has not been reported before in the bonding of PMMA. A leak test was performed over 180 h to assess the robustness of the proposed method. Moreover, to promote the applicability of this bonding method, we tested two kinds of microfluidic device applications, including a cell culture-based device and a metal microelectrode-integrated device. The results showed that the cell culture-based application was highly biocompatible with the PMMA microdevices fabricated using an acetic acid solvent. Moreover, the low pressure required during the bonding process supported the integration of metal microelectrodes with the PMMA microdevice without any damage to the metal films. This novel bonding method holds great potential in the ecofriendly and rapid fabrication of microfluidic devices using PMMA.
In the present study, we introduce a new approach for rapid bonding of poly(methyl methacrylate) (PMMA)-based microdevices using an acetic acid solvent with the assistance of UV irradiation. For the anticipated mechanism, acetic acid and UV irradiation induced free radicals on the PMMA surfaces, and acrylate monomers subsequently formed cross-links to create a permanent bonding between the PMMA substrates. PMMA devices effectively bonded within 30 s at a low pressure using clamps, and a clogging-free microchannel was achieved with the optimized 50% acetic acid. For surface characterizations, contact angle measurements and bonding performance analyses were conducted using predetermined acetic acid concentrations to optimize bonding conditions. In addition, the highest bond strength of bonded PMMA was approximately 11.75 MPa, which has not been reported before in the bonding of PMMA. A leak test was performed over 180 h to assess the robustness of the proposed method. Moreover, to promote the applicability of this bonding method, we tested two kinds of microfluidic device applications, including a cell culture-based device and a metal microelectrode-integrated device. The results showed that the cell culture-based application was highly biocompatible with the PMMA microdevices fabricated using an acetic acid solvent. Moreover, the low pressure required during the bonding process supported the integration of metal microelectrodes with the PMMA microdevice without any damage to the metal films. This novel bonding method holds great potential in the ecofriendly and rapid fabrication of microfluidic devices using PMMA.
In recent years, lab-on-a-chip
technology has attracted considerable
interest in various fields such as biomedical research, DNA analysis,
and healthcare. Because of the advantages such as low-cost and rapid
bonding methods at the industrial level, thermoplastics have attracted
much attention for fabricating microdevices. Among several thermoplastics,
poly(methyl methacrylate) (PMMA) is one of the essential thermoplastics
for microdevice fabrication because of its cost-effectiveness, optical
transparency, and biocompatibility.[1,2] In addition,
PMMA microfluidics have been widely used in applications in the various
biorelated fields.[3] There are different
types of bonding used for sealing PMMA–PMMA, such as thermal
bonding, solvent bonding, adhesive bonding, and chemical bonding.[4−7] Although thermal bonding is commonly used for fabricating PMMA devices
allowing for simple and uniform surface bonding, the use of bulky
press machines and heaters can lead to the deformation of the microchannel.[4] Adhesive and chemical bonding require low pressure
and low temperature; however, adhesives can clog the microchannel,
while chemical surface modifications are time-consuming, involving
multistep processes.[6,7] Among these methods, solvent bonding
generally possesses a robust and straightforward bonding process as
it achieves a high bond strength with a relatively simple bonding
strategy.[8] However, as a strong solvent
could dissolve PMMA, microchannel clogging is a universal drawback
of solvent bonding.[9,10] As an alternative, a solvent
with a moderate dissolution rate was applied for bonding PMMA with
the help of heat, and the overall process took more than 10 min.[8] In our previous studies, PMMAs were successfully
bonded using 90% ethanol as a solvent and UV light, and the bond strength
was approximately 6.17 MPa.[11] Although
it did not require any bulky press machine or heater, it needed to
control the solvent concentration to avoid clogging of the microchannel.
Moreover, we recently demonstrated the room-temperature bonding of
PMMA using acetic acid as a solvent.[12] However,
it took around 20 min to seal PMMAs using a press machine. Considering
these limitations, there is an urgent need for a new solvent bonding
approach for the rapid fabrication of microdevices.Over the
last decade, microsensor-based microfluidic devices have
attracted wide attention for the detection of various markers in the
environment, healthcare, temperature, biomedical, and chemical parameters.[13] These devices have a significantly increased
sensitivity and accuracy than commercial sensors. In particular, biosensor-based
devices containing a thin metal film as a microelectrode have been
used for temperature measurement, separation, and detection.[14,15] Mainly, thermal bonding and glue-assisted bonding were used to create
microdevices integrated with electrodes.[16−19] Although the thermal bonding
method is widely applicable, the risk of fractures on the integrated
metal electrode still remains because of the high bonding temperature
and pressure,[20] whereas glue-assisted bonding
faces the possibility of microchannel clogging. As an alternative,
an elastomeric substrate such as poly(dimethylsiloxane) (PDMS) was
used as a part of microdevices integrated with metal electrodes. However,
the pressure applied during the bonding process had to be finely controlled
to prevent the deformation or collapse of the PDMS microchannel.[21,22] Therefore, there is an urgent demand for a new method to integrate
thermoplastic devices with microelectrodes that can improve the biosensor
performance and further advance the biomedical fields.To overcome
these problems, in this study, we introduce a simple
and robust bonding method for fabricating a PMMA microdevice using
acetic acid as a solvent with the assistance of UV irradiation at
low pressure (see Figure ). In a previous study, we introduced the use of acetic acid
as a solvent for bonding PMMA, which was performed at room temperature
for 20 min under mild pressure conditions.[12] In the present study, acetic acid was still used as a solvent for
bonding PMMA; however, the overall bonding process was superior in
terms of time and pressure, requiring only 30 s of the UV irradiation
with the clamp-assisted low pressure. Based on the proof of concept,
we suggest that the highly energetic UV irradiation has reinforced
the activation of the PMMA surfaces in the presence of acetic acid,
and then the activated monomers of the acrylate functionalities on
two surfaces of PMMA were possibly re-cross-linked to realize a permanent
bonding. To demonstrate the bonding performance, various strength
measurement tests, such as bond strength analysis and leak test, were
performed. The water contact angle and Fourier transform infrared
spectroscopy (FTIR) were also analyzed. The tightly assembled microdevice
was used for two major applications—cell culture and integration
of a platinum-microelectrode into the device. The results showed that
the cells added into the PMMA microchannels showed a high viability
and decent adhesion to the surface even without postbonding surface
treatments. In addition, the integrated microelectrodes device was
successfully fabricated without the leak and fractures of the microelectrodes.
Figure 1
(a) Overall
procedure for bonding two PMMA substrates using acetic
acid treatment, followed by UV irradiation. (b,c) Chemical reaction
expected to occur on the surfaces of PMMA substrates after treating
with acetic acid and UV irradiation.
(a) Overall
procedure for bonding two PMMA substrates using acetic
acid treatment, followed by UV irradiation. (b,c) Chemical reaction
expected to occur on the surfaces of PMMA substrates after treating
with acetic acid and UV irradiation.
Results
and Discussion
Contact Angle Measurement
The change
in surface wettability
was analyzed by measuring the water contact angle, as shown in Figure . Prior to the measurement,
PMMA substrates were treated with acetic acid and UV irradiation. Figure shows the changes
in water contact angles of PMMA before and after treatment with varying
acetic acid concentrations from 0 to 100%. When PMMA was treated with
0% acetic acid and irradiated with UV for 30 s, the contact angle
of PMMA was around 72.6 ± 1.1°, which did not significantly
differ from that of pristine PMMA. By increasing the concentration
of the acetic acid solution, the contact angle values slightly decreased
to approximately 67.3 ± 1.1° when the acetic acid concentration
reached 100%. The use of acetic acid barely altered the surface property
of PMMA, which was also observed in our previous study,[12] indicating that the addition of UV treatment
did not affect the surface chemistry of the PMMA surface. Based on
this result, we can conclude that acetic acid is a preferable solvent
because of its minimal disturbance to the microchannel’s surface,
which reduces the risk of clogging frequently encountered by using
solvents with a high dissolution rate. In addition, by preserving
the original PMMA surface, the attractive properties of PMMA can be
readily used.
Figure 2
Water contact angle change of PMMA with varying acetic
acid concentrations,
followed by UV irradiation.
Water contact angle change of PMMA with varying acetic
acid concentrations,
followed by UV irradiation.
Bonding Performance
Figure shows the bonding performance of PMMA substrates
using different concentrations of acetic acid. An increase in the
acetic acid concentration resulted in the expansion of the bonded
area. Particularly, with 10–40% acetic acid, PMMA substrates
were only partially sealed, whereas using 50–100% acetic acid,
complete sealing was achieved. Based on these findings, 50% of acetic
acid was found to be the optimum condition to achieve the successful
homogeneous bonding of PMMA after 30 s of the UV irradiation.
Figure 3
Results of
bonding with different acetic acid concentrations (10–100%)
at 30 s of UV irradiation.
Results of
bonding with different acetic acid concentrations (10–100%)
at 30 s of UV irradiation.
FTIR Measurement
Figure shows the FTIR spectra of three samples, including
pristine PMMA (Figure a), UV-treated PMMA (Figure b), and acetic acid- and UV-treated PMMA (Figure c). Figure d compares the peaks for pristine PMMA, and
acetic acid- and UV-treated PMMA. According to the FTIR spectrum in Figure a, pristine PMMA
shows spectra regions at around 2950, 1723, 1142, and 750 cm–1, which corresponded to the presence of CH3, CO stretching,
COC, and CH2 twisting modes, respectively. The FTIR spectrum
obtained from pristine PMMA is consistent with our previous results.[23−26] As shown in Figure b,c, IR spectra of PMMA after the UV treatment and PMMA after acetic
acid and UV treatment displayed the same absorption peaks exhibited
by pristine PMMA, indicating that UV irradiation and acetic acid did
not change the chemical property of PMMA. This result differs from
our previous study on PMMA bonding mediated by acetic acid at room
temperature, where we hypothesized that an acetic acid molecule directly
participated in the reaction to bridge two PMMA substrates in 20 min.[12] In the current study, however, an additional
UV irradiation was applied to reduce the bonding time down to 30 s.
It is well-known that UV irradiation causes chain scissions of PMMA,
creating unstable radicals. Here, we hypothesized that though acetic
acid works to partially dissolve the PMMA surface, it does not directly
participate in the chemical reaction because the reaction between
broken PMMA chains was greater than that between acetic acid and PMMA.
Therefore, we speculate that when acetic acid was removed immediately
after UV irradiation, there was no sufficient time provided for acetic
acid to modify the PMMA surface.
Figure 4
FTIR spectra of (a) pristine PMMA, (b)
UV-treated PMMA, and (c)
acetic acid- and UV-treated PMMA. (d) Comparative FTIR spectra of
pristine PMMA and acetic acid- and UV-treated PMMA.
FTIR spectra of (a) pristine PMMA, (b)
UV-treated PMMA, and (c)
acetic acid- and UV-treated PMMA. (d) Comparative FTIR spectra of
pristine PMMA and acetic acid- and UV-treated PMMA.
Bond Strength Measurement
Figure shows the results of bond strength measurements
of the PMMA substrates (20 mm × 20 mm × 2 mm). All bonded
PMMAs only partially overlapped and were sealed using acetic acid
and immediately treated with UV irradiation, as described above. Figure a,b shows the real
images of bonded PMMAs in which the overlapped lengths were 4 and
2 mm, respectively; the bonded area after the detachment is shown
in Figure c. According
to the results, when PMMA substrates were bonded with the overlapped
length of 4 mm, the two substrates were unable to detach; thus, the
bond strengths could not be measured for PMMAs bonded with 40–100%
acetic acid. In our previous study, where the PMMAs were bonded without
UV irradiation, the bond strength measurement was still possible even
with an overlapped length of 5 mm.[12] Therefore,
the bond strength in the present study was significantly increased
as compared to the previous bonding strategies.[1,11,12] To measure the strength, the overlapped
length was decreased correspondingly with the reduction of the acetic
acid concentration (see Figure d). In detail, when 40% acetic acid was used and the overlapped
length was 2 mm, the bond strength was around 4.25 ± 0.02 MPa.
However, for 50 and 60% acetic acid, the overlapped length was 1.5
mm, and the bond strength systematically increased to 6.85 ±
0.11 and 7.63 ± 0.08 MPa, respectively. For 70 and 100% acetic
acid, the overlapped length was 1 mm. As a result, the bond strength
could not be measured because of the undetached PMMAs in the case
of 100% acetic acid. From these results, the highest possible measure
for bond strength was exhibited to be approximately 11.75 ± 0.02
MPa using 70% acetic acid, which is considered the strongest bonding
method for PMMAs (see Table ). Based on the results, shown in Figures and 5d, 50% acetic
acid, which is considered to be the optimized bonding condition, was
used in the further test.
Figure 5
Results of bond strength measurement. Bonded
PMMAs with the overlapped
length: (a) 4; (b) 2 mm. (c) Enlarged photo showing the bonded area
after the detachment. (d) Results of bond strength measured at different
acetic acid concentrations. All experiments were performed in triplicate.
Table 1
Comparison Showing Several Solvent
Bonding Methods for Bonding PMMAsa
bonding method
bonding conditions
bond strength
applications
reference
solvent-assisted thermal bonding
•acid-,
amine-, and plasma-modified
5.5 MPa
•not mentioned
(27)
•85 °C for 30 min
•pressure (0.005 MPa)
plasticizer-assisted thermal
bonding
•dibutyl phthalate
not
mentioned
•capillary electrophoresis
(28)
•90 °C for 10 min
•pressure (0.6 MPa)
solvent bonding
•1,2-dichloroethane and
ethanol (2:8)
3.8 MPa
•micromixer
(29)
•RT for 5 min
•capillary electrophoresis
•pressure (0.1 MPa)
solvent bonding
•acetonitrile
not mentioned
•electrophoresis
(30)
•solvent exposure for 8 min
•pressure (0.5 MPa)
solvent bonding
•chloroform
and ethanol
2.675 MPa
•micromixer
(31)
•40 °C for 10 min
•pressure (using
quartz glass fixture)
solvent bonding
•acetic
acid 50%
8.55 MPab
•polymerase chain reaction
(12)
•RT for 20 min
•human cell culture
•pressure (0.4 MPa)
microwave-assisted solvent bonding
•methanol, ethanol, and isopropanol
not mentioned
•electrophoresis
(32)
•microwave (90 s)
•pressure (using clamp)
ultrasonic-assisted solvent
bonding
•isopropanol 100%
2.25 MPa
•not mentioned
(33)
•ultrasonic (8 s)
•pressure (0.32 MPa)
UV-assisted solvent bonding
•ethanol 90%
6.17 MPa
•not mentioned
(11)
•UV irradiation (1 min)
•pressure (using clamp)
UV-assisted solvent bonding
•acetic acid 50%
11.75 MPab
•integrated microelectrode device
this work
•UV irradiation (30 s)
•human
cell culture
•pressure (using clamp)
RT: room temperature.
Maximum bond strength could not
be measured.
Results of bond strength measurement. Bonded
PMMAs with the overlapped
length: (a) 4; (b) 2 mm. (c) Enlarged photo showing the bonded area
after the detachment. (d) Results of bond strength measured at different
acetic acid concentrations. All experiments were performed in triplicate.RT: room temperature.Maximum bond strength could not
be measured.
Leak Test
Figure shows the
results of the leak test performed to confirm
the stability of the bonded PMMA microdevice. The PMMA microdevice
(40 × 40 mm2) with a serpentine microchannel (500
× 1000 μm2) was fabricated using the introduced
bonding method (see Figure a). To perform the test, the ink solution was introduced into
the microchannel, and the leak phenomena were continuously observed
from 0 to 180 h at room temperature (see Figure b). There was no leak during the test period
in which the ink was maintained inside the microchannel. However,
the red arrows indicated the evaporation phenomena observed at 12
h after the introduction of the ink solution into the microchannel.
This could be explained by the permeability of the silicone tubes
used for the introduction of the ink solution at the inlet and outlet
of the microdevice. From those results, we can conclude that using
the introduced bonding method, a highly stable bond between two PMMAs
was formed without a leak.
Figure 6
(a) Design of the microdevice used for the leak
test. (b) Results
of the leak test performed when observed over a 0–180 h period.
The red arrows indicate ink which was lost because of natural evaporation.
(a) Design of the microdevice used for the leak
test. (b) Results
of the leak test performed when observed over a 0–180 h period.
The red arrows indicate ink which was lost because of natural evaporation.
Channel Profile: Cross-Sectional View of
the Microchannel
Figure shows the
cross-sectional view of the bonded PMMAs fabricated by the method
introduced in this study. The engraved microchannels possessed curved
morphologies as the ball mills were used for computer numerical control
(CNC) milling. Microchannels with various dimensions were fabricated,
whose width and depth were identical—300, 500, and 1000 μm.
The results showed that the bonding process did not significantly
deform the microchannel morphology, verified by almost identical channel
morphologies. We can speculate that the acetic acid in the optimized
conditions influences only the surface of the PMMA substrate without
melting the bulk itself. Moreover, only a mild pressure exerted by
the use of clamps was applied for the bonding. Therefore, we can conclude
that channel deformations often caused by PMMAswelling during other
solvent bonding did not occur when treated with acetic acid followed
by UV irradiation. Acetic acid and UV irradiation can sufficiently
activate the PMMA surface to produce monomers to participate in a
permanent bonding, without any adverse effect on the morphology of
the microchannel. Therefore, the introduced bonding method can be
used for rapid bonding of PMMA, particularly having small microchannels.
Figure 7
Bright-field
photos of the cross-sectional view of the microchannels
engraved on the PMMA substrate before and after the PMMA–PMMA
bonding based on the introduced method.
Bright-field
photos of the cross-sectional view of the microchannels
engraved on the PMMA substrate before and after the PMMA–PMMA
bonding based on the introduced method.
Bonded PMMA Microdevice for Human Cell Culture
The
feasibility of the bonded PMMA microdevice for human cell culture
was assessed by directly seeding mesenchymal stem cells (MSCs) into
the channels and incubating for 3 days. Figure a illustrates the design of the microdevice.
The device consists of four straight semicircular microchannels with
inlets and outlets. The biocompatibility of the acetic acid- and the
UV-treated PMMA was confirmed by performing a cell viability test
using calcein-AM/ethidium homodimer-1 (EthD-1) on day 3 of the incubation
period (see Figure b). It was observed that the attached MSCs were highly viable and,
judging from the cell stretching, showed a decent adhesion to the
PMMA surface. This result indicates that, without further surface
treatment to induce cell viability and adhesion, the acetic acid-
and UV-treated PMMA microdevice can be directly used for cell-related
applications.
Figure 8
PMMA microdevice for human cell culture (MSCs). (a) Illustration
of the bonded microdevice having semicircular straight microchannels.
(b) Fluorescent image of the MSCs cultured inside the semicircular
microchannel at day 3. Cell viability was characterized by the calcein
AM/EthD-1 staining kit distinguishing live cells (green) and dead
cells (red). The cell density was 5 × 105 cells mL–1.
PMMA microdevice for human cell culture (MSCs). (a) Illustration
of the bonded microdevice having semicircular straight microchannels.
(b) Fluorescent image of the MSCs cultured inside the semicircular
microchannel at day 3. Cell viability was characterized by the calcein
AM/EthD-1 staining kit distinguishing live cells (green) and dead
cells (red). The cell density was 5 × 105 cells mL–1.
PMMA Microdevice Integrated
with Platinum Microelectrodes
Figure shows the
results of fabricating a PMMA microdevice integrated with a platinum
microelectrode. Close-up images confirmed that microelectrode was
not fractured, ensuring the ability to bond the metal electrode layer
and the microchannel plate while maintaining the electrode intact. Figure a,b shows the illustration
and real image of the integrated microdevice using the introduced
bonding method, respectively. To validate the operation of the microelectrode,
the overall experiment was set up as shown in Figure c. The microchamber was first filled by the
electrolyte solution, and then the resistance was measured using a
Fluke 117 digital multimeter. The presented 4.568 MΩ value indicated
that the device retained an excellent electric conductivity between
platinum electrodes inside the microchamber; therefore, the proposed
bonding strategy led to the successful fabrication of the platinum-integrated
microdevice. This bonding strategy possesses potential applications
in microanalytical systems such as separation, detection, and electrical
lysis. Moreover, the fracture phenomenon of platinum microelectrodes
was observed. As shown in Figure d, no fracture or minimal fracture was observed at
the edge of the PMMA plate adjoined to the metal film (zone 1 and
3) as well as the center of the electrode (zone 2). In contrast, by
using thermal bonding, the fracturedplatinum electrode was indicated
at the edge of the cover layer, and extensive cracks appeared on the
metal electrode (see Figure e) induced by thermal stress (105 °C, 20 min, and 0.1
MPa). Thermal bonding was achieved by following the conditions specified
in a previous study.[34] The difference in
the thermal expansion coefficient between the microelectrode and PMMA
may lead to a larger thermal plastic deformation. Consequently, the
plastic deformation because of a high bonding temperature is mainly
responsible for the fracture of the microelectrode.[14] Based on the results, as compared with other bonding strategies,
the proposed bonding approach showed many improvements and advantages
for bonding the integrated microelectrode PMMA device in terms of
a simple and robust process at mild pressure (see Table ). Therefore, to prevent the
cracking of electrodes during the bonding process, the proposed method
holds high potential for the rapid and straightforward fabrication
of metal-electrode-integrated microdevices.
Figure 9
Results of the bonding
method for a rapid prototype PMMA microdevice
with integrated metal microelectrodes. (a) Illustration of the PMMA
microdevice containing the platinum microelectrodes. (b) Real image
of the microdevice using the proposed bonding method. (c) Setup and
performance of resistance measurement using the fabricated microdevice.
(d) Enlarged photos of three squares marked in (b) showing fracture-free
platinum electrodes at the edge of the microdevice. (e) Fractured
platinum electrodes in the same zones of (b) using thermal bonding
as a control experiment.
Table 2
Comparison
of Some Bonding Methods
for Fabricating Integrated Microelectrode PMMA Devices
bonding method
bonding conditions
metallic microelectrode
reference
thermal
bonding
•108 °C for 25 min
gold
(20)
•pressure (tightly clamped)
plasma-assisted thermal bonding
•oxygen plasma
copper
(14)
•85 °C for 15 min
•pressure (0.002 MPa)
UV-assisted thermal bonding
•UV irradiation (8 min)
gold
(21)
•80 °C for 3 min
•pressure (0.00135 MPa)
UV/O3-assisted thermal
bonding
•UV/O3 treatment (8 min)
gold
(35)
•80 °C for 5 min
•pressure (0.02 MPa)
solvent-assisted thermal
bonding
•isopropanol
gold
(36)
•80 °C for 10 min
•pressure (0.0025 MPa)
UV-assisted solvent
bonding
•acetic acid 50%
platinum
this work
•UV
irradiation (30 s)
•pressure (using
clamp)
Results of the bonding
method for a rapid prototype PMMA microdevice
with integrated metal microelectrodes. (a) Illustration of the PMMA
microdevice containing the platinum microelectrodes. (b) Real image
of the microdevice using the proposed bonding method. (c) Setup and
performance of resistance measurement using the fabricated microdevice.
(d) Enlarged photos of three squares marked in (b) showing fracture-free
platinum electrodes at the edge of the microdevice. (e) Fracturedplatinum electrodes in the same zones of (b) using thermal bonding
as a control experiment.
Conclusions
In the present study,
a novel bonding strategy for bonding PMMA
using acetic acid as a solvent followed by UV irradiation for 30 s
was introduced. This approach is based on the use of acetic acid as
a solvent for the partial dissolution of PMMA, which is followed by
UV irradiation for re-cross-linking monomers of PMMA during the bonding
process. Also, the introduced method is superior to other solvent
bonding methods in terms of simplicity in the bonding process, biocompatibility,
and integrity of the microchannel (see Table S1). Of note, the bonded PMMAs have achieved a bond strength as high
as 11.75 MPa without the use of external hot embossing instruments,
and with the mild conditions applied, clogging-free microchannels
were attained. When the device was applied to the cell culture, the
microchannels facilitated cells with high viability and decent adhesion
after 3 days of culture. Moreover, the bond of the microelectrode-integrated
PMMA device was strong, and the electrodes stayed fracture-free after
the bonding process, which helped to maintain the performance of the
microelectrodes. Therefore, the introduced bonding method is a powerful
approach for the rapid fabrication of PMMA microdevices as well as
to expand the applications for biosensing to advance the healthcare
system.
Experimental Section
Materials
Acetic acid (99.7%) was
purchased from Sigma-Aldrich
(MO, USA). PMMA substrates (2 mm thicknesses) were obtained from Goodfellow
(Coraopolis, USA). Acetic acid solutions diluted in deionized water
to the appropriate concentrations were used for all bonding experiments
and analytical measurements. MSC and MSC medium were acquired from
ScienCell (CA, USA). Calcein AM and EthD-1 for the cell viability/cytotoxicity
test were purchased from Thermo Fisher Scientific (MA, USA). Silicone
tubes (i.d. 1.0 mm, o.d. 2.0 mm) were obtained from Dow Corning (MI,
USA). Platinum 99.95% was acquired from Vacuum Thin Film Materials
(Incheon, Korea).
Bonding Mechanism
Figure a shows the overall scheme
for bonding two
PMMA substrates using acetic acid treatment, followed by UV irradiation.
First, acetic acid was placed on the surface of one PMMA substrate
and then covered by the other PMMA. With the help of two clamps, the
two PMMA substrates were tightly fastened. Then, the pair of PMMA
was irradiated with 234 nm UV light for 30 s. Finally, the two clamps
were removed from the bonded PMMAs after the UV treatment process. Figure b shows the PMMA
substrates producing free radicals on the surfaces when treated with
the acetic acid solution under UV irradiation. Subsequently, acrylate
monomers of PMMA substrates were formed (see Figure b), which possibly formed cross-links to
create a permanent bond between the two PMMA substrates (see Figure c).
Surface Characterization
(1): Water Contact Angle Measurement
The surface wettability
of the acetic acid-treated PMMA followed
by the UV irradiation was evaluated by the water contact angle measurement.
For the measurements, PMMA substrates (20 mm × 20 mm) were covered
with 0, 10, 20, 30, 40, 50, 60, 70, and 100% acetic acid solution
and immediately treated with a 234 nm UV lamp for 30 s using a UV
light curing system. After the UV treatment, the PMMA substrates were
carefully dried by compressed air before the measurement of water
contact angles. By using a Phoenix 300 contact angle analyzer, the
water contact angle was measured by depositing static water droplets
onto the substrates. The results were analyzed with the Image-Pro
300 software. Five measurements were made and averaged for further
assessment. For comparison, the water contact angle was also measured
on the surface of a pristine PMMA as a control.
Surface Characterization
(2): FTIR Analysis
PMMA samples
(pristine and acetic acid-treated PMMAs, followed by UV light) were
analyzed by FTIR (JASCO 4700 FTIR spectrophotometer), which offers
quantitative and qualitative analyses to characterize the chemical
nature of the modified surfaces. The FTIR indicated He–Ne as
the reference at the resolution of 4 cm–1 and the
mode of diffuse reflectance with scanning from 4500 to 400 cm–1 of resolution.
Bond Strength Analysis
and Leak Tests
For the bond
strength analyses, partially bonded PMMAs were measured using a texture
analyzer (QTS 25, Brookfield, Middleborough, MA, USA). All samples
used for strength measurements were treated with acetic acid solution
(40, 50, 60, 70, and 100%), and the UV exposure time was fixed at
30 s. Before bond strength analysis, the samples with through-holes,
which were punctured on each PMMA substrate using a drilling machine
for the insertion of twines, were prepared. Then, the partially overlapped
PMMA substrates were fixed in the strength measurement analyzer using
the inserted twines and then pulled apart at the speed of 100 mm min–1. All experiments were triplicated to examine bonding
reproducibility. For the leak test analysis, the PMMA microdevice
(40 mm × 40 mm × 4 mm) was used to observe the ink solution
inside the microchannel in 0, 6, 12, 24, 48, 72, 96, and 180 h, successively.In order
to test the biocompatibility of the microdevice with human cell culture,
semicircular channels with a diameter of 1 mm were engraved on the
PMMA and bonded with another flat PMMA substrate using the method
mentioned above. The cell culture chips were sterilized by placing
them in a clean bench under UV light for 1 h. The trypsinized MSCs
(passage 4) were collected from a cell-culture plate and directly
inserted to the channel after dilution. After incubating in a 5% humidified
CO2 incubator at 37 °C for 3 days, cell viability
was examined by staining them with 2 μM calcein AM and 4 μM
EthD-1 following the manufacturer’s instruction to indicate
the live and dead cells in green and red fluorescence, respectively.
The captured microscopic images were analyzed using the ProgRes CapturePro
2.8 software (Jenoptik).
PMMA Microdevice Integrated with Platinum
Microelectrodes
The microelectrode-integrated microdevice
consisted of channel
and electrode layers. The overall schematic of the fabrication of
the PMMA microdevice integrated with microelectrodes is shown in Figure S1. Briefly, the patterns were engraved
on the PMMA substrate by using a CNC machine.[12] After ultrasonic cleaning, the engraved PMMA was dried by compressed
air. A paper mask with electrode patterns was cut by a laser cutter
and placed on the PMMA substrate. The platinum microelectrode on the
PMMA plate was created by sputter deposition, with a 100 nm thick
platinum layer coated on a 0.1 mm platinum target using a Cressington
sputter coater 108 auto machine. Subsequently, two halves of the device
were carefully aligned and bonded using acetic acid assisted with
UV irradiation. Particularly, 50% acetic acid was covered between
the channel and the electrode layer. The assembled device was clamped
and treated with a 234 nm UV for 30 s to realize bonding. After bonding,
the microdevice was washed to remove any excess acid inside the channel.
At each end of the channel, a 2 mm diameter hole was bored and connected
with a silicon tube to form the inlet/outlet. The fracture phenomenon,
as well as the electric conductivity of the microelectrode of the
complete microdevice, was later assessed.
Authors: Jayna J Shah; Jon Geist; Laurie E Locascio; Michael Gaitan; Mulpuri V Rao; Wyatt N Vreeland Journal: Anal Chem Date: 2006-05-15 Impact factor: 6.986
Authors: Sergio Leonardo Florez; Ana Lucia Campaña; M Juliana Noguera; Valentina Quezada; Olga P Fuentes; Juan C Cruz; Johann F Osma Journal: Micromachines (Basel) Date: 2022-06-19 Impact factor: 3.523