Stephanie Choi1, Firas Jassim2, Edem Tsikata2, Ziad Khoueir2,3, Linda Y Poon2,4, Boy Braaf5, Benjamin J Vakoc5, Brett E Bouma1,5,5, Johannes F de Boer6,5, Teresa C Chen2.
Abstract
Purpose: To compare artifact rates in two-dimensional (2D) versus three-dimensional (3D) retinal nerve fiber layer (RNFL) scans using Spectralis optical coherence tomography (OCT).Entities:
Keywords: RNFL artifacts; glaucoma; optical coherence tomography
Mesh:
Year: 2020 PMID: 32714638 PMCID: PMC7351591 DOI: 10.1167/tvst.9.3.12
Source DB: PubMed Journal: Transl Vis Sci Technol ISSN: 2164-2591 Impact factor: 3.283
Figure 1.Examples of artifact types found in 2D and 3D images. (A) Incorrect segmentation of the anterior RNFL (red line). (B) Misidentification of the anterior RNFL due to a posterior vitreous detachment (red line). (C) Incorrect segmentation of the posterior RNFL (blue line). (D) Incorrect segmentation of the internal limiting membrane (red line), posterior RNFL (yellow line), and RPE (dark blue line). (E) Complete segmentation failure, despite clearly identifiable anatomic layers. (F) Degraded image quality where the signal is indistinguishable from the background noise. (G) Out of register artifact resulting in truncation of retinal layers. (H) Cut edge on left-hand side of the B-scan. (I) Decentration error was defined as when the optic nerve head was greater than 10% off center. (J) Area of PPA crosses the outer border of the smallest annulus (2.5–3.5 mm).
Demographic Characteristics of the Normal and Glaucoma Study Populations
| Characteristic | Glaucoma | Normal |
|
|---|---|---|---|
| Total number of eyes | 106 | 95 | — |
| Right eyes/left eyes ( | 58/48 | 57/38 | 0.45 |
| Male/female ( | 51/55 | 37/58 | 0.19 |
| Age (y), mean ± SD (range) | 67.6 ± 12.1 (21–87) | 53.6 ± 16.7 (18–87) | <0.001 |
| Refractive error | |||
| Spherical equivalent (diopters), mean ± SD | –0.6 ± 1.8 | –0.4 ± 1.7 | 0.45 |
| Visual field | |||
| Mean deviation (dB), mean ± SD | –12.5 ± 7.9 | –1.7 ± 2.3 | <0.001 |
| Pattern standard deviation (dB), mean ± SD | 8.4 ± 3.4 | 1.7 ± 0.5 | <0.001 |
Results are statistically significant.
Percentage of 2D Retinal Nerve Fiber Layer Scans with Artifacts
| Glaucoma ( | Normal ( | ||
|---|---|---|---|
| Artifact Type | Percentage of B-scans (%) | Percentage of B-scans (%) |
|
| Posterior RNFL misidentification | 46.2 | 5.3 | <0.001 |
| Decentration | 10.4 | 8.4 | 0.64 |
| Anterior RNFL misidentification | 9.4 | 3.2 | 0.07 |
| PVD/ERM | 4.7 | 2.1 | 0.32 |
| Degraded image | 2.8 | 0 | 0.10 |
| Peripapillary atrophy | 1.9 | 0 | 0.18 |
| Complete absence of segmentation | 0 | 0 | N/A |
| Out of register | 0 | 0 | N/A |
| Cut edge | 0 | 0 | N/A |
| Mirroring | 0 | 0 | N/A |
| Missing parts | 0 | 0 | N/A |
| Motion artifact | 0 | 0 | N/A |
| Any artifact (total percentage) | 58.5 | 14.7 | <0.001 |
Proportion of 2D retinal nerve fiber layer scans with the specified artifact type divided by 106 for the glaucoma group and 95 for the normal group.
Total percentage represents the total percentage of 2D retinal nerve fiber layer scans with at least one artifact type. Scans with multiple artifacts were only counted once.
Results are statistically significant.
Percentage of B-Scans with Artifacts in 3D Volume Scans (Smallest Annulus Size of 2.5–3.5 mm)
| Artifact Type | Glaucoma (%) | Normal (%) |
|
|---|---|---|---|
| Posterior RNFL misidentification | 21.9 | 15.5 | 0.002 |
| RPE misidentification | 5.8 | 4.9 | 0.38 |
| Anterior RNFL misidentification | 3.8 | 2.4 | 0.23 |
| PVD/ERM | 1.9 | 1.2 | 0.33 |
| Out of register | 2.9 | 2.7 | 0.89 |
| Peripapillary atrophy | 2.0 | 0.4 | 0.10 |
| Degraded | 1.3 | 0.7 | 0.20 |
| Cut edge | 0.3 | 0.6 | 0.39 |
| Complete absence of segmentation | 0.1 | 0.1 | 0.65 |
| Mirroring | 0.1 | 0.2 | 0.77 |
| Motion artifact | 0 | 0 | N/A |
| Any artifact (total percentage) | 35.4 | 21.8 | <0.001 |
Percentage of artifacts per eye was calculated as the number of B-scans with artifact type divided by 97, which was the total number of B-scans per volume scan that contributed to calculations for the smallest sized annulus (2.5–3.5 mm).
Total percentage represents the total percentage of B-scans within a volume scan with at least one artifact type. Scans with multiple artifacts were only counted once.
Results are statistically significant.
Global Retinal Nerve Fiber Layer Thickness Derived Using 3D Volume Scans, Pre- and Post-Interpolation (Annulus Size 2.5–3.5 mm)
| Mean Global RNFL Thickness Pre- Interpolation Thickness (µm) | Mean Global RNFL Thickness Post-Interpolation Thickness (µm) |
| Mean Absolute Percentage Difference (Pre- vs. Post- Interpolation) | Standard Deviation of Percentage Difference (Pre- vs. Post-Interpolation) |
| |
|---|---|---|---|---|---|---|
| Glaucoma ( | 77.0 ± 11.6 | 75.1 ± 11.2 | 0.23 | 0.03% | 4.7% | 0.94 |
| Normal ( | 104.0 ± 10.6 | 104.0 ± 10.0 | 0.98 | 0.1% | 3.3% | 0.68 |
Results are presented as mean ± standard deviation.
MATLAB software reported blank values for 8 patients after interpolation. Average global retinal nerve fiber layer thickness was recalculated using the remaining 98 of 106 glaucoma patients.