Abstract
BACKGROUND: The clinical symptoms and adverse events caused by pacemaker battery depletion are not uncommon, but it is easy to miss or misdiagnose them clinically. To raise the level of awareness towards this clinical situation, we report two cases. CASEEntities:
Keywords: Atrioventricular block-induced torsade de pointes; Electrocardiogram; Pacemaker battery depletion; Pacemaker syndrome
Mesh:
Year: 2020 PMID: 32703237 PMCID: PMC7376836 DOI: 10.1186/s12872-020-01622-x
Source DB: PubMed Journal: BMC Cardiovasc Disord ISSN: 1471-2261 Impact factor: 2.298
Fig. 1ECG (a: at presentation b: after pacemaker exchange). a Non-synchronous ventricular pacing at a fixed of 65 bpm with retrograded atrial capture (black arrow). b Atrial pacing following by a spontaneous ventricular rhythm
Fig. 2ECG (a and b: at presentation). a: Completely dissociated P waves (black arrows) and wide QRS complexes. The rhythm strip showed complete atrioventricular(AV) block. The QRS complexes were wide because of the presence of right bundle branch block and left posterior fascicular block, representing the AV block is infranodal. QT intervals were prolonged (QTc 550 ms) and premature ventricular contraction(PVC) was present (black star). b: Complete AV block with wide QRS complexes, frequent PVCs were present. The first PVC occurred on the downslope of the T wave, which induced a polymorphic ventricular tachycardia with changing QRS complex amplitudes, which was known as torsade de pointes(TdP)
Manufacturer-Specific Pacemaker Programming at Replacement Notification
| Pacing characteristic | Boston Scientific | Abbott | Biotronik | Meditronic | Microport |
|---|---|---|---|---|---|
| Loss of Rare Response? | Yes | Yes | Yes | Yes | Yes |
| Loss of AV Synchrony? | No | No | No | Yes(VVI) | Yes(VVI) |
| Change in Lower Rate Limit? | No | Yes (−10%) | Yes (−11%) | Yes (65 bpm) | Yes (70 bpm) |
| Magnet rate following replacement Notification | ≤85 bpm | ≤86.3 bpm | ≤80 bpm | ≤65 bpm | ≤80 bpm |