The Symptoms and Clinical events associated with Automatic Reprogramming (SCARE) at replacement notification study.

Sunil K Sinha1, Daniel Carlson1, Jonathan Chrispin1, Andreas S Barth1, John Jack Rickard2, David D Spragg1, Ronald Berger1, Charles Love1, Hugh Calkins1, Gordon F Tomaselli3, Joseph E Marine1.   

Abstract

BACKGROUND: Pacemaker patients experience battery depletion that activates pacemaker's alert for replacement notification. Automatic reprogramming at replacement notification can result in loss of rate response and atrioventricular (AV) synchrony.
OBJECTIVE: To determine if relevant symptoms or clinical events may be associated with automatic reprogramming at replacement notification.
METHODS: Electronic medical record review was undertaken for 298 patients referred for pacemaker generator replacement. Primary endpoints were symptoms or clinical events during replacement notification period.
RESULTS: Following elimination of duplicate pacemaker replacements (n = 12), "near-replacement notification" or "recalled" (n = 15) and pacemakers at "end of life" (n = 5), 266 subjects were included. Three distinct reprogramming cohorts were identified; those with no change (control) in pacing mode (n = 46), those with loss of rate response (n = 154), and those with loss of AV synchrony ± rate response (n = 66). In total, 83 subjects (31.2%) had symptoms with significant differences seen between groups (control = 4.3%, loss of rate response = 26.0%, loss of AV synchrony ± rate response = 62.1%, P < 0.001). Overall, 28 subjects (10.5%) experienced clinical events with significant differences seen between groups (control = 0.0%, loss of rate response = 6.5%, loss of AV synchrony ± rate response = 27.3%, P < 0.001).
CONCLUSIONS: Automatic reprogramming at replacement notification was associated with significant symptoms in 26% of those who lost rate response and in 62% of those who lost AV synchrony ± rate response. Additionally, 27% of the latter cohort required nonelective clinical care.
© 2018 Wiley Periodicals, Inc.

Entities:  

Keywords:  elective replacement indication; elective replacement time; pacemaker; pacemaker mode; pacemaker programming; recommended replacement time

Mesh:

Year:  2018        PMID: 30375674     DOI: 10.1111/pace.13532

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

1.  Adverse clinical events caused by pacemaker battery depletion: two case reports.

Authors:  Junpeng Liu; Li Wen; Simin Yao; Peipei Zheng; Shuang Zhao; Jiefu Yang
Journal:  BMC Cardiovasc Disord       Date:  2020-07-23       Impact factor: 2.298

2.  Torsade de pointes caused by citalopram during the pacemaker battery-depletion phase: A case report.

Authors:  Junwen Wang; Ziyi Sun; Siming Tao
Journal:  Ann Noninvasive Electrocardiol       Date:  2022-03-24       Impact factor: 1.485

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.