Zhipeng Zhao1, Ning Zhang1, Ying Liu1, Ning Wu1, Zhuang Mao1, Wei Yang1, Guanghui Cheng1.
Abstract
PURPOSE: To analyze the results of clinical utilization of ring applicator for combined intracavitary (IC)/interstitial (IS) image-guided brachytherapy (BT) technique in locally advanced cervical cancer (LACC), and to explore the application conditions of IC/IS BT approach with this applicator for Chinese patients.Entities:
Keywords: brachytherapy; cervical cancer; interstitial; intracavitary; ring applicator
Year: 2020 PMID: 32695197 PMCID: PMC7366014 DOI: 10.5114/jcb.2020.96866
Source DB: PubMed Journal: J Contemp Brachytherapy ISSN: 2081-2841
Fig. 1Distribution of 4 planes and 8 directions. A) The 4 planes in sagittal planes. Plane I at the level of ring tube surface; Plane II at the parallel layer of 1 cm above ring tube surface; Plane III at the parallel layer of 2 cm above ring tube surface (ICRU point A); Plane IV at the parallel layer of 1 cm above plane III (3 cm above ring tube surface); B) The 8 directions in each plane
Comparison of dose parameters of target and OARs with ring applicator in two plans
| IC/IS treatment | IC research | ||||
|---|---|---|---|---|---|
| HR-CTV | D90 | 7.94 ±0.65 | 6.98 ±0.75 | 8.48 | < 0.001 |
| D98 | 6.66 ±0.75 | 5.85 ±0.66 | 7.27 | < 0.001 | |
| IR-CTV | D90 | 4.86 ±0.53 | 4.37 ±0.63 | 7.07 | < 0.001 |
| D98 | 3.75 ±0.58 | 3.34 ±0.69 | 3.99 | < 0.001 | |
| Bladder | D2.0cm3 | 5.18 ±0.66 | 5.61 ±0.50 | –6.71 | < 0.001 |
| Rectum | D2.0cm3 | 3.83 ±0.60 | 4.05 ±0.49 | –3.05 | 0.005 |
| Sigmoid | D2.0cm3 | 3.56 ±0.90 | 3.48 ±0.93 | 1.76 | 0.091 |
| Small bowel | D2.0cm3 | 2.68 ±1.32 | 2.72 ±1.36 | –1.32 | 0.199 |
Dose-volume histogram parameters in 43 patients treated with combined IC/IS BT
| Parameters | EQD2 dose (EBRT + BT) |
|---|---|
| HR-CTV D90 | 90.33 ±9.20 |
| IR-CTV D90 | 67.89 ±3.59 |
| HR-CTV D98 | 80.17 ±7.45 |
| IR-CTV D98 | 62.15 ±4.11 |
| Bladder D2.0cm3 | 76.53 ±6.49 |
| Rectum D2.0cm3 | 64.25 ±6.76 |
| Sigmoid D2.0cm3 | 61.79 ±8.01 |
| Small bowel D2.0cm3 | 60.15 ±6.80 |
Fig. 2Distribution of holes in the ring applicator
Total number of implanted needles and average insertion depth in 65 cases treated with ring applicator
| Needle channel | Number of needles | Average depth (cm) |
|---|---|---|
| 4 | 24 | 4.0 |
| 5 | 58 | 4.2 |
| 6 | 55 | 4.3 |
| 7 | 17 | 2.9 |
| 8 | 55 | 4.2 |
| 9 | 51 | 4.3 |
| 10 | 17 | 4.0 |
Fig. 3Regression fitting of power function for target volume and HR-CTV D90
Blue curve: IC/ISBT fitting curve, red curve: ICBT fitting curve According to the EQD2value of the LQ model, when the EQD2of HR-CTV D90was 85 Gy in terms of four equivalent calculations, the physical dose of HR-CTV D90 was 7.13 Gy. When HR-CTV D90dose value was 7.13 Gy, the target volume was 36 cm3 in red curve
Fig. 4The relationship between HR-CTV and PD coverage on different planes in ICIS treatment and IC research plans
The blue : the average distance from HR-CTV to the tandem. The red : the average distance from prescribed dose isodose curve of ICIS treatment plans to the tandem. The green : the average distance from prescribed dose isodose curve of IC research plans to the tandem