Reducing metal artifacts between implants in cone-beam CT by adjusting angular position of the subject.

Chang-Ki Min1, Kyoung-A Kim2.   

Abstract

OBJECTIVES: To reduce inter-implant metal artifacts in cone-beam CT (CBCT) imaging by adjusting angular position of the subject relative to the source-detector plane.
MATERIALS AND METHODS: Two dental implants were placed in a block made of homogeneous dental impression material. Using a custom-made apparatus, the specimen was scanned with a CBCT machine at seven different angles (0°, 15°, 30°, 45°, 60°, 75°, and 90°) along three different spatial axes, yielding 21 experimental groups. Thirteen volumes of interest (VOI) including inter- and peri-implant areas were selected from each axial reconstruction perpendicular to the implants. Gray values (GVs) of each pixel within these VOIs were measured. Mean differences in GV (ΔGV) between the VOIs and control area were calculated and expressed as a percentage. These ΔGVs from different spatial angle were compared and analyzed by Welch's analysis of variance (ANOVA) and linear regression using SPSS 25.0 software.
RESULTS: As alpha angle increased, the ΔGV of the inter-implant area increased from - 62.02% to near-zero while the standard deviation decreased. Welch's ANOVA and linear regression analysis revealed ΔGV increased significantly with alpha angle (p < 0.001, R2 = 0.406). VOIs on the extension line of two implants showed similar results. After adjusting the beta and gamma angles, there was no significant change in ΔGV.
CONCLUSIONS: Increasing the alpha angle can reduce metal artifacts in the inter-implant area in CBCT images.

Entities:  

Keywords:  Artifacts; Cone-beam computed tomography; Dental implant; Diagnostic imaging

Mesh:

Year:  2020        PMID: 32638201     DOI: 10.1007/s11282-020-00458-7

Source DB:  PubMed          Journal:  Oral Radiol        ISSN: 0911-6028            Impact factor:   1.852


  2 in total

1.  Reduction of aneurysm clip artifacts on CT angiograms: a technical note.

Authors:  J H Brown; E S Lustrin; M H Lev; C S Ogilvy; J M Taveras
Journal:  AJNR Am J Neuroradiol       Date:  1999-04       Impact factor: 3.825

2.  Normal range of motion of the cervical spine.

Authors:  B Lind; H Sihlbom; A Nordwall; H Malchau
Journal:  Arch Phys Med Rehabil       Date:  1989-09       Impact factor: 3.966

  2 in total
  3 in total

1.  Does the size of an object containing dental implant affect the expression of artifacts in cone beam computed tomography imaging?

Authors:  Mahkameh Moshfeghi; Yaser Safi; Ingrid Różyło-Kalinowska; Shiva Gandomi
Journal:  Head Face Med       Date:  2022-06-29       Impact factor: 2.246

2.  Quantitative analysis of metal artefacts of dental implant in CBCT image by correlation analysis to micro-CT: A microstructural study.

Authors:  Chang-Ki Min; Kyoung-A Kim
Journal:  Dentomaxillofac Radiol       Date:  2020-10-08       Impact factor: 2.419

3.  Effect of the Field of View Size on CBCT Artifacts Caused by the Presence of Metal Objects in the Exomass.

Authors:  Yaser Safi; Mitra Ghazizadeh Ahsaie; Maede Jafarian Amiri
Journal:  Int J Dent       Date:  2022-09-09
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.