| Literature DB >> 32614149 |
Kun Wei1, Zhiguo Xu1, Ruishi Mao1, Zulong Zhao1, Tiecheng Zhao1, Qianshun She1, Xincai Kang1, Jianli Wang1, Shengpeng Li1, Min Li1, Kai Song1, Herun Yang1, Limin Duan1.
Abstract
PURPOSE: The heavy-ion medical machine (HIMM), which is the first commercial medical accelerator designed and built independently by the institute of modern physics (IMP) in Wuwei, Gansu Province, China, had officially completed clinical trials at the time of this article's writing. Three types of detector systems were developed based on the ionization-chamber principle to monitor the beam parameters during treatment in real time, quickly verify the beam performance during a routine checkup, and ensure patient safety. METHODS AND MATERIALS: The above-mentioned detector systems were used for beam monitoring and quality assurance in the treatment system. The beam-monitoring system is composed of three integral ionization chambers (ICs) and two multistrip ionization chambers (MSICs) as a redundant design. The irradiation dose, beam position, and homogeneity of a lateral profile are monitored online by the beam-monitoring system, and safety interlocks are established to keep the test results under the predefined tolerance limitation. The quality-assurance equipment was composed of one MSIC and one IC stack. The IC stack was used for energy verification.Entities:
Keywords: CFDA; HIMM; beam monitoring; carbon-ion therapy; quality assurance
Mesh:
Substances:
Year: 2020 PMID: 32614149 PMCID: PMC7484831 DOI: 10.1002/acm2.12916
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Relevant parameters of HIMM facility for carbon‐ion therapy.
| Specification | Value |
|---|---|
| Treatment room | Room A, horizontal beam |
| Room B, both vertical and horizontal beams | |
| Energy range | 120–400 MeV/u |
| Beam flux | 2 × 106–4 × 108 pps |
| Radiation field | ≤200 × 200 mm2 |
| Dose rate | 0.001–1 Gy/s |
| Maximum range | 27 cm |
| Scanning technology | Room A, modulated scanning mode |
| Room B, uniform scanning mode |
FIG. 1Simplified layout of BMS and QAE. The BMS is composed of three integral ionization chambers (IC‐01, IC‐02, and IC‐03) and two multistrip ionization chambers (MSIC‐01 and MSIC‐02). The QAE is composed of a multistrip ionization chamber (MSIC‐03) and an IC stack.
FIG. 2Schematic layout of BMS; the blue part represents the main system, while the green part represents the redundant system. The signal obtained from the monitors is transmitted to the accelerator system or treatment system through front‐end electronics (FEC) and a data‐acquisition system (DAQ) for beam commissioning or online monitoring during the treatment process.
FIG. 3(a) Nonlinearity and (b) linearity correction of QFC output.
FIG. 4Top view of IC stack after assembly
Main properties of front‐end electronics.
| Specification | Value |
|---|---|
| Channels | 100 positive input channels |
| Integral capacitance | 10 pF |
| Dynamic charge range | 0 − 50 pC |
| Noise | <2 mV |
| Integral time range | minimum of 50
|
| Maximum linear error | Less than 1% |
FIG. 5(a) Spot distribution for off‐axis test. (b) Relative dose response in treatment room A.
FIG. 6(a) Noise of MSIC with baseline subtraction. (b) Minimum distinguishable signal.
FIG. 7Position resolution of MSIC in X and the Y directions.
FIG. 8Field profiles along the (a) X and (b) Y axis measured by MSIC and a medical film.
FIG. 9Relationship between response time of QFC (0.5 pC/pulse) and input signal from IC.
FIG. 10(a) Schematic of test. (b) Beam‐angle distribution with change of spot number (beam position) in X and Y directions.
FIG. 11(a) Depth dose distribution measured at different energies. (b) Relationship between beam energy and peak position of Bragg curves tested using an IC stack and water tank.
FIG. 12Relationship between relative position of Bragg peak and increased PMMA thickness: (a) 260 and (b) 330 MeV.