The CNAO dose delivery system for modulated scanning ion beam radiotherapy.

S Giordanengo1, M A Garella2, F Marchetto1, F Bourhaleb3, M Ciocca4, A Mirandola4, V Monaco3, M A Hosseini3, C Peroni3, R Sacchi3, R Cirio3, M Donetti2.   

Abstract

PURPOSE: This paper describes the system for the dose delivery currently used at the Centro Nazionale di Adroterapia Oncologica (CNAO) for ion beam modulated scanning radiotherapy.
METHODS: CNAO Foundation, Istituto Nazionale di Fisica Nucleare and University of Torino have designed, built, and commissioned a dose delivery system (DDS) to monitor and guide ion beams accelerated by a dedicated synchrotron and to distribute the dose with a full 3D scanning technique. Protons and carbon ions are provided for a wide range of energies in order to cover a sizable span of treatment depths. The target volume, segmented in several layers orthogonally to the beam direction, is irradiated by thousands of pencil beams which must be steered and held to the prescribed positions until the prescribed number of particles has been delivered. For the CNAO beam lines, these operations are performed by the DDS. The main components of this system are two independent beam monitoring detectors, called BOX1 and BOX2, interfaced with two control systems performing the tasks of real-time fast and slow control, and connected to the scanning magnets and the beam chopper. As a reaction to any condition leading to a potential hazard, a DDS interlock signal is sent to the patient interlock system which immediately stops the irradiation. The essential tasks and operations performed by the DDS are described following the data flow from the treatment planning system through the end of the treatment delivery.
RESULTS: The ability of the DDS to guarantee a safe and accurate treatment was validated during the commissioning phase by means of checks of the charge collection efficiency, gain uniformity of the chambers, and 2D dose distribution homogeneity and stability. A high level of reliability and robustness has been proven by three years of system activity needing rarely more than regular maintenance and working with 100% uptime. Four identical and independent DDS devices have been tested showing comparable performances and are presently in use on the CNAO beam lines for clinical activity.
CONCLUSIONS: The dose delivery system described in this paper is one among the few worldwide existing systems to operate ion beam for modulated scanning radiotherapy. At the time of writing, it has been used to treat more than 350 patients and it has proven to guide and control the therapeutic pencil beams reaching performances well above clinical requirements. In particular, in terms of dose accuracy and stability, daily quality assurance measurements have shown dose deviations always lower than the acceptance threshold of 5% and 2.5%, respectively.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 25563266     DOI: 10.1118/1.4903276

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  10 in total

1.  Re-irradiation With Carbon Ion Radiotherapy for Pelvic Rectal Cancer Recurrences in Patients Previously Irradiated to the Pelvis.

Authors:  Amelia Barcellini; Viviana Vitolo; Lorenzo Cobianchi; Andrea Peloso; Alessandro Vanoli; Alfredo Mirandola; Angelica Facoetti; Maria Rosaria Fiore; Alberto Iannalfi; Barbara Vischioni; Francesco Cuccia; Sara Ronchi; Maria Bonora; Giulia Riva; Rachele Petrucci; Emma D'Ippolito; Francesca Dal Mas; Lorenzo Preda; Francesca Valvo
Journal:  In Vivo       Date:  2020 May-Jun       Impact factor: 2.155

2.  INSIDE in-beam positron emission tomography system for particle range monitoring in hadrontherapy.

Authors:  Maria Giuseppina Bisogni; Andrea Attili; Giuseppe Battistoni; Nicola Belcari; Niccolo' Camarlinghi; Piergiorgio Cerello; Silvia Coli; Alberto Del Guerra; Alfredo Ferrari; Veronica Ferrero; Elisa Fiorina; Giuseppe Giraudo; Eleftheria Kostara; Matteo Morrocchi; Francesco Pennazio; Cristiana Peroni; Maria Antonietta Piliero; Giovanni Pirrone; Angelo Rivetti; Manuel D Rolo; Valeria Rosso; Paola Sala; Giancarlo Sportelli; Richard Wheadon
Journal:  J Med Imaging (Bellingham)       Date:  2016-12-02

Review 3.  National Effort to Re-Establish Heavy Ion Cancer Therapy in the United States.

Authors:  Arnold Pompos; Robert L Foote; Albert C Koong; Quynh Thu Le; Radhe Mohan; Harald Paganetti; Hak Choy
Journal:  Front Oncol       Date:  2022-06-14       Impact factor: 5.738

4.  Online proton therapy monitoring: clinical test of a Silicon-photodetector-based in-beam PET.

Authors:  Veronica Ferrero; Elisa Fiorina; Matteo Morrocchi; Francesco Pennazio; Guido Baroni; Giuseppe Battistoni; Nicola Belcari; Niccolo' Camarlinghi; Mario Ciocca; Alberto Del Guerra; Marco Donetti; Simona Giordanengo; Giuseppe Giraudo; Vincenzo Patera; Cristiana Peroni; Angelo Rivetti; Manuel Dionisio da Rocha Rolo; Sandro Rossi; Valeria Rosso; Giancarlo Sportelli; Sara Tampellini; Francesca Valvo; Richard Wheadon; Piergiorgio Cerello; Maria Giuseppina Bisogni
Journal:  Sci Rep       Date:  2018-03-06       Impact factor: 4.379

5.  Performances of the beam monitoring system and quality assurance equipment for the HIMM of carbon-ion therapy.

Authors:  Kun Wei; Zhiguo Xu; Ruishi Mao; Zulong Zhao; Tiecheng Zhao; Qianshun She; Xincai Kang; Jianli Wang; Shengpeng Li; Min Li; Kai Song; Herun Yang; Limin Duan
Journal:  J Appl Clin Med Phys       Date:  2020-07-02       Impact factor: 2.102

6.  Dosimetric Validation of a System to Treat Moving Tumors Using Scanned Ion Beams That Are Synchronized With Anatomical Motion.

Authors:  Michelle Lis; Wayne Newhauser; Marco Donetti; Moritz Wolf; Timo Steinsberger; Athena Paz; Christian Graeff
Journal:  Front Oncol       Date:  2021-09-08       Impact factor: 6.244

7.  Localization of anatomical changes in patients during proton therapy with in-beam PET monitoring: A voxel-based morphometry approach exploiting Monte Carlo simulations.

Authors:  Aafke Christine Kraan; Andrea Berti; Alessandra Retico; Guido Baroni; Giuseppe Battistoni; Nicola Belcari; Piergiorgio Cerello; Mario Ciocca; Micol De Simoni; Damiano Del Sarto; Marco Donetti; Yunsheng Dong; Alessia Embriaco; Veronica Ferrero; Elisa Fiorina; Marta Fischetti; Gaia Franciosini; Giuseppe Giraudo; Francesco Laruina; Davide Maestri; Marco Magi; Giuseppe Magro; Carlo Mancini Terracciano; Michela Marafini; Ilaria Mattei; Enrico Mazzoni; Paolo Mereu; Riccardo Mirabelli; Alfredo Mirandola; Matteo Morrocchi; Silvia Muraro; Alessandra Patera; Vincenzo Patera; Francesco Pennazio; Angelo Rivetti; Manuel Dionisio Da Rocha Rolo; Valeria Rosso; Alessio Sarti; Angelo Schiavi; Adalberto Sciubba; Elena Solfaroli Camillocci; Giancarlo Sportelli; Sara Tampellini; Marco Toppi; Giacomo Traini; Serena Marta Valle; Francesca Valvo; Barbara Vischioni; Viviana Vitolo; Richard Wheadon; Maria Giuseppina Bisogni
Journal:  Med Phys       Date:  2021-12-22       Impact factor: 4.506

8.  In-vivo range verification analysis with in-beam PET data for patients treated with proton therapy at CNAO.

Authors:  Martina Moglioni; Aafke Christine Kraan; Guido Baroni; Giuseppe Battistoni; Nicola Belcari; Andrea Berti; Pietro Carra; Piergiorgio Cerello; Mario Ciocca; Angelica De Gregorio; Micol De Simoni; Damiano Del Sarto; Marco Donetti; Yunsheng Dong; Alessia Embriaco; Maria Evelina Fantacci; Veronica Ferrero; Elisa Fiorina; Marta Fischetti; Gaia Franciosini; Giuseppe Giraudo; Francesco Laruina; Davide Maestri; Marco Magi; Giuseppe Magro; Etesam Malekzadeh; Michela Marafini; Ilaria Mattei; Enrico Mazzoni; Paolo Mereu; Alfredo Mirandola; Matteo Morrocchi; Silvia Muraro; Ester Orlandi; Vincenzo Patera; Francesco Pennazio; Marco Pullia; Alessandra Retico; Angelo Rivetti; Manuel Dionisio Da Rocha Rolo; Valeria Rosso; Alessio Sarti; Angelo Schiavi; Adalberto Sciubba; Giancarlo Sportelli; Sara Tampellini; Marco Toppi; Giacomo Traini; Antonio Trigilio; Serena Marta Valle; Francesca Valvo; Barbara Vischioni; Viviana Vitolo; Richard Wheadon; Maria Giuseppina Bisogni
Journal:  Front Oncol       Date:  2022-09-26       Impact factor: 5.738

9.  Monitoring Carbon Ion Beams Transverse Position Detecting Charged Secondary Fragments: Results From Patient Treatment Performed at CNAO.

Authors:  Marco Toppi; Guido Baroni; Giuseppe Battistoni; Maria Giuseppina Bisogni; Piergiorgio Cerello; Mario Ciocca; Patrizia De Maria; Micol De Simoni; Marco Donetti; Yunsheng Dong; Alessia Embriaco; Veronica Ferrero; Elisa Fiorina; Marta Fischetti; Gaia Franciosini; Aafke Christine Kraan; Carmela Luongo; Etesam Malekzadeh; Marco Magi; Carlo Mancini-Terracciano; Michela Marafini; Ilaria Mattei; Enrico Mazzoni; Riccardo Mirabelli; Alfredo Mirandola; Matteo Morrocchi; Silvia Muraro; Vincenzo Patera; Francesco Pennazio; Angelo Schiavi; Adalberto Sciubba; Elena Solfaroli-Camillocci; Giancarlo Sportelli; Sara Tampellini; Giacomo Traini; Serena Marta Valle; Barbara Vischioni; Viviana Vitolo; Alessio Sarti
Journal:  Front Oncol       Date:  2021-06-10       Impact factor: 6.244

10.  Innovations and the Use of Collimators in the Delivery of Pencil Beam Scanning Proton Therapy.

Authors:  Daniel E Hyer; Laura C Bennett; Theodore J Geoghegan; Martin Bues; Blake R Smith
Journal:  Int J Part Ther       Date:  2021-06-25
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.