Three-dimensional inhomogeneous magnetization transfer with rapid gradient-echo (3D ihMTRAGE) imaging.

Gopal Varma1, Fanny Munsch1, Brian Burns2, Guillaume Duhamel3, Olivier M Girard3, Arnaud Guidon2, R Marc Lebel2, David C Alsop1.   

Abstract

PURPOSE: To demonstrate the feasibility of integrating the magnetization transfer (MT) preparations required for inhomogeneous MT (ihMT) within an MPRAGE-style acquisition. Such a sequence allows for reduced power deposition and easy inclusion of other modules.
METHODS: An ihMT MPRAGE-style sequence (ihMTRAGE) was initially simulated to investigate acquisition of the 3D ihMT data sequentially, or in an interleaved manner. The ihMTRAGE sequence was implemented on a 3T clinical scanner to acquire ihMT data from the brain and spine.
RESULTS: Both simulations and in vivo data provided an ihMT signal that was significantly greater using a sequential ihMTRAGE acquisition, compared with an interleaved implementation. Comparison with a steady-state ihMT acquisition (defined as having one MT RF pulse between successive acquisition modules) demonstrated how ihMTRAGE allows for a reduction in average power deposition, or greater ihMT signal at equal average power deposition. Inclusion of a prospective motion-correction module did not significantly affect the ihMT signal obtained from regions of interest in the brain. The ihMTRAGE acquisition allowed combination with a spatial saturation module to reduce phase wrap artifacts in a cervical spinal cord acquisition.
CONCLUSIONS: Use of preparations necessary for ihMT experiments within an MPRAGE-style sequence provides a useful alternative for acquiring 3D ihMT data. Compared with our steady-state implementation, ihMTRAGE provided reduced power deposition, while allowing use of the maximum intensity from off-resonance RF pulses. The 3D ihMTRAGE acquisition allowed combination of other modules with the preparation necessary for ihMT experiments, specifically motion compensation and spatial saturation modules.
© 2020 International Society for Magnetic Resonance in Medicine.

Keywords:  MPRAGE; MT; brain; inhomogeneous magnetization transfer; myelin; spinal cord

Mesh:

Year:  2020        PMID: 32602958      PMCID: PMC7722070          DOI: 10.1002/mrm.28324

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  37 in total

1.  Mixed-effects and fMRI studies.

Authors:  K J Friston; K E Stephan; T E Lund; A Morcom; S Kiebel
Journal:  Neuroimage       Date:  2005-01-01       Impact factor: 6.556

2.  Measurement of signal-to-noise ratios in MR images: influence of multichannel coils, parallel imaging, and reconstruction filters.

Authors:  Olaf Dietrich; José G Raya; Scott B Reeder; Maximilian F Reiser; Stefan O Schoenberg
Journal:  J Magn Reson Imaging       Date:  2007-08       Impact factor: 4.813

3.  Three-dimensional magnetization-prepared rapid gradient-echo imaging (3D MP RAGE).

Authors:  J P Mugler; J R Brookeman
Journal:  Magn Reson Med       Date:  1990-07       Impact factor: 4.668

4.  Low duty-cycle pulsed irradiation reduces magnetization transfer and increases the inhomogeneous magnetization transfer effect.

Authors:  G Varma; O M Girard; S Mchinda; V H Prevost; A K Grant; G Duhamel; D C Alsop
Journal:  J Magn Reson       Date:  2018-08-31       Impact factor: 2.229

5.  Tract-specific and age-related variations of the spinal cord microstructure: a multi-parametric MRI study using diffusion tensor imaging (DTI) and inhomogeneous magnetization transfer (ihMT).

Authors:  Manuel Taso; Olivier M Girard; Guillaume Duhamel; Arnaud Le Troter; Thorsten Feiweier; Maxime Guye; Jean-Philippe Ranjeva; Virginie Callot
Journal:  NMR Biomed       Date:  2016-04-21       Impact factor: 4.044

6.  Magnetization transfer from inhomogeneously broadened lines: A potential marker for myelin.

Authors:  Gopal Varma; Guillaume Duhamel; Cedric de Bazelaire; David C Alsop
Journal:  Magn Reson Med       Date:  2014-03-06       Impact factor: 4.668

7.  PROMO: Real-time prospective motion correction in MRI using image-based tracking.

Authors:  Nathan White; Cooper Roddey; Ajit Shankaranarayanan; Eric Han; Dan Rettmann; Juan Santos; Josh Kuperman; Anders Dale
Journal:  Magn Reson Med       Date:  2010-01       Impact factor: 4.668

8.  Interpretation of magnetization transfer from inhomogeneously broadened lines (ihMT) in tissues as a dipolar order effect within motion restricted molecules.

Authors:  G Varma; O M Girard; V H Prevost; A K Grant; G Duhamel; D C Alsop
Journal:  J Magn Reson       Date:  2015-09-07       Impact factor: 2.229

9.  Optimization of white-matter-nulled magnetization prepared rapid gradient echo (MP-RAGE) imaging.

Authors:  Manojkumar Saranathan; Thomas Tourdias; Ersin Bayram; Pejman Ghanouni; Brian K Rutt
Journal:  Magn Reson Med       Date:  2014-05-29       Impact factor: 4.668

Review 10.  FSL.

Authors:  Mark Jenkinson; Christian F Beckmann; Timothy E J Behrens; Mark W Woolrich; Stephen M Smith
Journal:  Neuroimage       Date:  2011-09-16       Impact factor: 6.556

View more
  2 in total

1.  Spinal cord and brain tissue impairments as long-term effects of rugby practice? An exploratory study based on T1 and ihMTsat measures.

Authors:  Arash Forodighasemabadi; Guillaume Baucher; Lucas Soustelle; Thomas Troalen; Olivier M Girard; Maxime Guye; Jean-Baptiste Grisoli; Jean-Philippe Ranjeva; Guillaume Duhamel; Virginie Callot
Journal:  Neuroimage Clin       Date:  2022-07-23       Impact factor: 4.891

2.  Combining inhomogeneous magnetization transfer and multipoint Dixon acquisition: Potential utility and evaluation.

Authors:  Ece Ercan; Gopal Varma; Ivan E Dimitrov; Yin Xi; Marco C Pinho; Fang F Yu; Shu Zhang; Xinzeng Wang; Ananth J Madhuranthakam; Robert E Lenkinski; David C Alsop; Elena Vinogradov
Journal:  Magn Reson Med       Date:  2020-10-26       Impact factor: 4.668

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.