Technical Delivery Parameters of 2000 Proton Treatment Courses.

Amanda J Deisher1, Thomas J Whitaker1, Chris J Beltran1, Robert L Foote1, Michael G Haddock1, Anita Mahajan1.   

Abstract

PURPOSE: To summarize the technical delivery parameters of proton plans delivered at the Mayo Clinic in Rochester, Minnesota.
MATERIALS AND METHODS: The database of treated patient proton plans was queried to extract field parameters such as gantry angle, patient support angle, minimum and maximum water-equivalent depth (WED) treated, number of layers, field size, patient orientation, and monitor units. The plans were analyzed in aggregate, by disease site, and by fractionation.
RESULTS: There were 2963 proton plans for 2023 distinct treatment sites delivered between June 2015 and September 2018. The mean number of fields per plan was 2.8. The mean number of energy layers per field was 51.9. The mean monitor unit per field was 117.4. The median maximum field dimension was 12.4 cm; 95% of the fields had a maximum dimension < 28.7 cm, and the maximum field dimension was 39.8 cm. The median maximum field WED was 16.4 cm; 95% of the fields reached a maximum WED of ≤ 26.4 cm, and the maximum field WED was 32.4 cm.
CONCLUSION: A large variety of disease sites were treated using the maximum field size (40 cm) and WED (32.4 cm) capabilities of our half-gantry system. © Copyright 2020 The Author(s).

Entities:  

Keywords:  gantry; pencil-beam scanning; proton therapy

Year:  2020        PMID: 32582812      PMCID: PMC7038916          DOI: 10.14338/IJPT-19-00066.1

Source DB:  PubMed          Journal:  Int J Part Ther        ISSN: 2331-5180


  2 in total

1.  Optimizing mini-ridge filter thickness to reduce proton treatment times in a spot-scanning synchrotron system.

Authors:  Lorraine Courneyea; Chris Beltran; Hok Seum Wan Chan Tseung; Juan Yu; Michael G Herman
Journal:  Med Phys       Date:  2014-06       Impact factor: 4.071

2.  Reassessment of the Necessity of the Proton Gantry: Analysis of Beam Orientations From 4332 Treatments at the Massachusetts General Hospital Proton Center Over the Past 10 Years.

Authors:  Susu Yan; Hsiao-Ming Lu; Jay Flanz; Judith Adams; Alexei Trofimov; Thomas Bortfeld
Journal:  Int J Radiat Oncol Biol Phys       Date:  2015-09-30       Impact factor: 7.038

  2 in total
  2 in total

1.  Design of a 3D patient-specific collision avoidance virtual framework for half-gantry proton therapy system.

Authors:  Jingjing M Dougherty; Thomas J Whitaker; Daniel W Mundy; Erik J Tryggestad; Chris J Beltran
Journal:  J Appl Clin Med Phys       Date:  2021-12-10       Impact factor: 2.102

2.  Operational Performance of a Compact Proton Therapy System: A 5-Year Experience.

Authors:  Omar A Zeidan; Ethan Pepmiller; Twyla Willoughby; Zhiqiu Li; James Burkavage; Brian Harper; Michael Fraser; Katie Moffatt; Sanford L Meeks; Naren Ramakrishna
Journal:  Int J Part Ther       Date:  2022-07-01
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.