Integration of the sensitivity-relevant electronics of nuclear magnetic resonance (NMR) and electron spin resonance (ESR) spectrometers on a single chip is a promising approach to improve the limit of detection, especially for samples in the nanoliter and subnanoliter range. Here, we demonstrate the cointegration on a single silicon chip of the front-end electronics of NMR and ESR detectors. The excitation/detection planar spiral microcoils of the NMR and ESR detectors are concentric and interrogate the same sample volume. This combination of sensors allows one to perform dynamic nuclear polarization (DNP) experiments using a single-chip-integrated microsystem having an area of about 2 mm2. In particular, we report 1H DNP-enhanced NMR experiments on liquid samples having a volume of about 1 nL performed at 10.7 GHz(ESR)/16 MHz(NMR). NMR enhancements as large as 50 are achieved on TEMPOL/H2O solutions at room temperature. The use of state-of-the-art submicrometer integrated circuit technologies should allow the future extension of the single-chip DNP microsystem approach proposed here up the THz(ESR)/GHz(NMR) region, corresponding to the strongest static magnetic fields currently available. Particularly interesting is the possibility to create arrays of such sensors for parallel DNP-enhanced NMR spectroscopy of nanoliter and subnanoliter samples.
Integration of the sensitivity-relevant electronics of nuclear magnetic resonance (NMR) and electron spin resonance (ESR) spectrometers on a single chip is a promising approach to improve the limit of detection, especially for samples in the nanoliter and subnanoliter range. Here, we demonstrate the cointegration on a single silicon chip of the front-end electronics of NMR and ESR detectors. The excitation/detection planar spiral microcoils of the NMR and ESR detectors are concentric and interrogate the same sample volume. This combination of sensors allows one to perform dynamic nuclear polarization (DNP) experiments using a single-chip-integrated microsystem having an area of about 2 mm2. In particular, we report 1H DNP-enhanced NMR experiments on liquid samples having a volume of about 1 nL performed at 10.7 GHz(ESR)/16 MHz(NMR). NMR enhancements as large as 50 are achieved on TEMPOL/H2O solutions at room temperature. The use of state-of-the-art submicrometer integrated circuit technologies should allow the future extension of the single-chip DNP microsystem approach proposed here up the THz(ESR)/GHz(NMR) region, corresponding to the strongest static magnetic fields currently available. Particularly interesting is the possibility to create arrays of such sensors for parallel DNP-enhanced NMR spectroscopy of nanoliter and subnanoliter samples.
Nuclear magnetic resonance (NMR)
spectroscopy is a powerful tool employed in research, industry, and
medicine. The use of NMR methodologies in an even wider range of applications
is often hindered by the relatively large minimum number of resonating
spins needed to achieve a sufficiently large signal-to-noise ratio
(SNR) in the available experimental time. In most of the situations
studied by NMR spectroscopy the samples are concentration limited.
In these conditions, the largest SNR is obtained with the largest
possible sample volume compatible with the high-homogeneity region
of the magnet at the strongest possible magnetic field. However, there
are also situations where the sample is volume limited. For volume-limited
samples, the use of an inductive detector having a sensitive volume
matched to the volume of the sample under investigation results in
a significant improvement of the SNR.[1−12] For samples in the nanoliter and subnanoliter range, noninductive
detection methods have been also proposed, such those based on nitrogen
vacancies in diamond[13−16] and magnetic resonance force microscopy.[17−21] Another approach to increase the SNR, applicable
to samples of any volume, is to increase the nuclear spin polarization,
e.g., by microwave-, optical-, and chemistry-based methodologies.[22−34] In the microwave DNP approach, the sample under investigation contains
unpaired electron spins which are excited into electron spin resonance
(ESR). The electron spin excitation allows one to enhance the nuclear
magnetization of several orders of magnitude above its thermal value,
improving the SNR in the NMR experiment by the same factor and reducing
the required experimental time as the square of this factor.During the last two decades, separate integration on a single chip
of the front-end electronics of inductive NMR spectrometers[35−49] as well as ESR spectrometers[50−55] has been demonstrated. These approaches are suitable, e.g., for
the miniaturization of the probe, for the reduction of the losses
and complexity of the connections, and for the realization of dense
arrays of detectors. In this work, we demonstrate the cointegration
on a single silicon chip of the front-end electronics of NMR and ESR
detectors. The excitation/detection planar spiral microcoils of the
NMR and ESR detectors are concentric and interrogate the same sample
volume. This combination of sensors allows one to perform dynamic
nuclear polarization (DNP) experiments using a single-chip-integrated
DNP microsystem having an area of about 2 mm2. In particular,
we report 1H DNP experiments on liquid samples having a
volume of about 1 nL performed at 10.7 GHz(ESR)/16 MHz(NMR). NMR enhancements
as large as 50 are achieved on TEMPOL/H20 solutions at
room temperature.
Description of the Single-Chip DNP Microsystem
The single-chip DNP microsystem is composed of two parts: the ESR
and NMR detectors (Figure a). The ESR- and NMR-integrated electronic circuits are very
similar to those previously reported.[42,54] The chip has
a size of 1.1 × 1.6 mm2, and it is fabricated using
a standard silicon CMOS technology (TSMC 180 nm, MS/RF). The NMR detector
consists of a broadband (10 MHz to 1 GHz) transmit/receive electronic
circuit directly connected (i.e., without tuning and matching capacitors)
to an excitation/detection microcoil (Figure a). Since the NMR detector does not contain
an rf source, it requires an external source to produce the 16 MHz
rf signals for the local oscillator (VLO) and the transmitter (VTX). The rf signal VTX is amplified by the integrated transmitter
rf electronics. When the integrated switch is in the transmit mode,
a peak-to-peak rf voltage of about 2.5 V and a peak-to-peak rf current
of about 90 mA is delivered to the NMR microcoil and the dissipated
power in the NMR microcoil is 56 mW. When the switch is in the receive
mode, the electromotive force induced in the NMR microcoil by the
spin precession is amplified by an rf receiver amplifier (30 dB, from
10 MHz to 1 GHz), downconverted to the kHz range using an integrated
mixer driven by the local oscillator signal, and further amplified
by an additional integrated low-frequency amplifier (20 dB, dc to
1 MHz). The NMR microcoil is concentrically placed inside the ESR
microcoil (Figure c). The two coupled concentric microcoils are simulated using a full-wave
electromagnetic simulator (Advanced Design System ADS, Keysight Technologies).
The obtained S-parameter files are used in an integrated circuits
simulator (Cadence, Cadence Design Systems Inc.) to simulate both
the ESR- and the NMR-integrated detectors. In these simulations, it
is found that an NMR coil with more than 15 turns significantly reduces
the quality factor of the ESR coil and quenches the oscillation of
the ESR circuitry. This is qualitatively explained by the losses caused
by the induced currents in the NMR coil. To keep a safety margin,
the number of turns is reduced from 22 to 10 and the outer diameter
is shrunk with respect to the NMR microcoil of the previous design.[42] The reduction of the number of turns and the
outer diameter reduces the inductance from 150 to 17 nH and the series
resistance from 110 to 30 Ω. The NMR microcoil has an external
diameter of 191 μm, a wire width of 3 μm, a wire thickness
2.3 μm, and a spacing of 2 μm between turns (Figure c). The dc power
consumption of the NMR system is 170 mW in the transmit mode and 40
mW in the receive mode.
Figure 1
Set-up for characterization of the single-chip-integrated
DNP microsystem
operating at 10.7 GHz(ESR)/16 MHz(NMR). (a) Schematic of the single-chip
DNP microsystem. Dashed red lines indicate the NMR/ESR circuits and
the concentric NMR/ESR microcoils. Black squares next to the port
names represent the bonding pads. VDDTR, VDDRX, and VSW ports are connected to several nets, and their connections are not
shown in the schematic for simplicity. VDDTR and VDDRX are the supply voltages of
all of the blocks in the transmitter and receiver of the NMR circuit,
respectively. VSW is the control signal
that is connected to all of the switches in both the transmitter and
the receiver of the NMR circuit. (b) Photograph of the single-chip
DNP microsystem. Blue lines indicate the position of the capillary
containing the sample under investigation. Dashed red lines indicate
the NMR/ESR circuits and the concentric NMR/ESR microcoils as in a.
(c) Photograph of the NMR and ESR microcoils. ESR microcoil has 2
turns. NMR microcoil has 10 turns. (d) Block diagram of the complete
setup for characterization of the single-chip DNP microsystem: (A)
electromagnet (Bruker, 0–2.2 T); (B) homemade modulation coil
(0.33 mT/A); (C) rf amplification stage composed of two rf amplifiers
(Analog Devices HMC-C001) and a 3 dB attenuator; (D) three dc power
supplies (Keithley 2400); (E) rf generator (Rohde&Schwartz SMR-20); (F) mixer (Mini-Circuits ZX05-153-S+); (G) 100
MHz high-pass filter (Crystek CHPFL-0100); (H) 300 MHz low-pass filter
(Crystek CLPFL-0300); (I) frequency divider (Valon Technology 3010);
(J) homemade delay-line discriminator (200 MHz central frequency,
1 MHz detection range, 5 MHz FM bandwidth); (K) amplifier (Stanford
Research Systems SR560); (L) lock-in amplifier (EG&G 7265); (M) NMR magnetometer probe (Metrolab Instruments SA 1062
probe 3); (N) NMR magnetometer main electronic unit (Metrolab Instruments
SA PT2025); (O) rf splitter (Mini-Circuits ZFSC-2-11); (P) rf generator
(Stanford Research Systems SG384); (Q) frequency counter (Fluke PM6681);
(R) amplifier (EG&G 5113); (S) power amplifier
(Rohrer PA508); (T) photograph of the single-chip DNP-NMR microsystem
with the capillary containing the sample placed on the microcoils
and the bonding wires protected by glob-top (black material covering
the bottom part of the chip); (U) magnet power supply (Bruker, 0–150
A).
Set-up for characterization of the single-chip-integrated
DNP microsystem
operating at 10.7 GHz(ESR)/16 MHz(NMR). (a) Schematic of the single-chip
DNP microsystem. Dashed red lines indicate the NMR/ESR circuits and
the concentric NMR/ESR microcoils. Black squares next to the port
names represent the bonding pads. VDDTR, VDDRX, and VSW ports are connected to several nets, and their connections are not
shown in the schematic for simplicity. VDDTR and VDDRX are the supply voltages of
all of the blocks in the transmitter and receiver of the NMR circuit,
respectively. VSW is the control signal
that is connected to all of the switches in both the transmitter and
the receiver of the NMR circuit. (b) Photograph of the single-chip
DNP microsystem. Blue lines indicate the position of the capillary
containing the sample under investigation. Dashed red lines indicate
the NMR/ESR circuits and the concentric NMR/ESR microcoils as in a.
(c) Photograph of the NMR and ESR microcoils. ESR microcoil has 2
turns. NMR microcoil has 10 turns. (d) Block diagram of the complete
setup for characterization of the single-chip DNP microsystem: (A)
electromagnet (Bruker, 0–2.2 T); (B) homemade modulation coil
(0.33 mT/A); (C) rf amplification stage composed of two rf amplifiers
(Analog Devices HMC-C001) and a 3 dB attenuator; (D) three dc power
supplies (Keithley 2400); (E) rf generator (Rohde&Schwartz SMR-20); (F) mixer (Mini-Circuits ZX05-153-S+); (G) 100
MHz high-pass filter (Crystek CHPFL-0100); (H) 300 MHz low-pass filter
(Crystek CLPFL-0300); (I) frequency divider (Valon Technology 3010);
(J) homemade delay-line discriminator (200 MHz central frequency,
1 MHz detection range, 5 MHz FM bandwidth); (K) amplifier (Stanford
Research Systems SR560); (L) lock-in amplifier (EG&G 7265); (M) NMR magnetometer probe (Metrolab Instruments SA 1062
probe 3); (N) NMR magnetometer main electronic unit (Metrolab Instruments
SA PT2025); (O) rf splitter (Mini-Circuits ZFSC-2-11); (P) rf generator
(Stanford Research Systems SG384); (Q) frequency counter (Fluke PM6681);
(R) amplifier (EG&G 5113); (S) power amplifier
(Rohrer PA508); (T) photograph of the single-chip DNP-NMR microsystem
with the capillary containing the sample placed on the microcoils
and the bonding wires protected by glob-top (black material covering
the bottom part of the chip); (U) magnet power supply (Bruker, 0–150
A).The ESR detector is a slightly
modified version of the one previously
reported,[54] where the bonding pads are
distanced to facilitate placement of a capillary containing the liquid
sample under investigation (Figure a–c). The ESR detector is a differential Colpitts
LC oscillator operating at 10.7 GHz. The excitation/detection ESR
microcoil has two turns, an external diameter of 270 μm, a wire
width of 12 μm, a wire thickness of 2.3 μm, and a spacing
of 2 μm between turns (Figure c). The NMR microcoil placed inside the ESR microcoil
reduces the effective inductance of the ESR microcoil from 2.8 to
2.3 nH and thus increases the operating frequency of the ESR detector
from 10.1 to 10.7 GHz with respect to the ESR detector previously
reported.[54] The maximum microwave power
dissipated into the ESR microcoil is about 17 mW with a maximum microwave
field B1 of about 0.1
mT at the center of the microcoil. Since the excitation microwave
field is produced by the integrated microwave LC oscillator, no external
microwave sources are required to perform the ESR excitation. The
ESR detection is also performed by the integrated microwave LC oscillator.
The external microwave electronics are used only to measure the frequency
variation of the integrated LC oscillator due to the ESR phenomenon
of the sample placed in close proximity with the inductor of the oscillator
(i.e., the ESR microcoil). The dc power consumption of the ESR circuit
is 90 mW at the maximum supply voltage of the oscillator of 2 V and
5 mW at the minimum supply voltage of the oscillator of 0.85 V.The complete setup for the characterization of the single-chip
DNP microsystem is shown in Figure d. The frequency-to-voltage conversion of the ESR detector
output is performed by a delay-line discriminator (DLD) whose central
frequency is 200 MHz. In order to match the DLD central frequency
and improve the spectral purity, the signal at the output of the detector
is amplified, mixed with an external reference, filtered, and shaped
through a divide by 1 frequency divider. The dc-coupled output of
the DLD is amplified and digitized by an analog to digital converter
(ADC). This signal is used to monitor the ESR oscillator frequency.
The ac-coupled output of the DLD is amplified and sent to a lock-in
amplifier for further amplification and synchronous demodulation.
The lock-in amplifier also generates the reference signal which is
used for magnetic field modulation. An NMR magnetometer is used to
track the variations of the magnetic field produced by the electromagnet
in which the DNP experiments are performed. The frequency lock with
the NMR magnetometer is necessary due to the relatively large drift
(about 1 ppm/h) of the electromagnet. The NMR frequency measured by
the magnetometer is used to set the frequency of the rf generator
connected to the integrated transmitter VTX and to the local oscillator VLO of the
integrated receiver. In particular, the rf generator frequency is
set 20 kHz below the frequency measured by the NMR magnetometer. This
allows one to obtain an NMR signal above the 1/f-noise
corner frequency of the integrated NMR receiver. A switch signal VSW determines the pulse width of the NMR excitation.
The output of the integrated NMR receiver VNMR is amplified, digitized, and digitally processed. The maximum NMR
signal amplitude is obtained with a rf pulse length τrf≅ 5 μs. The maximum microwave magnetic
field B1 at the center
of the ESR coil is about 70 μT, as estimated from measurements
performed on a BDPA sample (see below).The samples are contained
in borosilicate glass capillaries (BGCT
0.2, Capillary Tube Supply Ltd.) with a 0.2 mm outer diameter (o.d.)
and 0.18 mm inner diameter (i.d.). The capillaries are sealed with
a torch (Microtorch, Prodont Holliger). The nondegassed solutions
of pure water (H2O) (W3500, Sigma-Aldrich) and 4-hydroxy-2,2,6,6-tetramethylpiperidine
1-oxyl (TEMPOL) (176141, Sigma-Aldrich) are obtained by dilution at
room temperature in air starting from a 1 M solution. The degassed
solutions are prepared as follows. The degassing of water is performed
by bubbling with a nitrogen flow for about 1 h. A 200 mL glass vial,
with inlet/outlet pipes for the nitrogen flow, is filled with pure
water heated to 70 ◦C by a hot plate. A pipet is
used to transfer the degassed water into a 2 mL glass vial containing
the appropriate amount of TEMPOL molecules to produce a 100 mM solution.
The 10 and 1 mM solutions are prepared in 2 mL glass vials by subsequent
dilution in degassed water. The preparation of the solutions and the
filling/sealing of the capillaries is performed in a few seconds to
minimize the absorption of O2 in contact with air (the
diffusion length of O2 in water in 1 s is about 60 μm).Electromagnetic and electrical simulations of the integrated microwave
oscillator, performed with ADS and Cadence, show that the maximum
achievable microwave current Imw in the
ESR microcoil is 74 mA, obtained with an oscillator supply voltage
of 2 V. Figure shows
the result of a COMSOL simulation performed with microwave current Imw = 74 mA in the ESR microcoil. The indicated
magnetic field is one-half of the magnitude of the component of the
microwave magnetic field perpendicular to the static magnetic field
(i.e., B1) in two orthogonal
cross sections of the capillary where the sample is confined. The
microwave magnetic field B1 in the center of the coil is about 100 μT. The obtained
microwave magnetic field is the result of the superposition of the
microwave magnetic field created by the ESR coil and the microwave
magnetic field caused by the induced microwave currents in the NMR
microcoil. Moving vertically from the chip (and coils) surface into
the sample region, the total microwave magnetic field first decreases,
cancels out in the dark blue region, then increases, and finally decreases
again further away from the chip surface. This is the expected magnetic
field produced by two concentric coils of different diameters carrying
currents flowing in opposite directions. Simulations of the microwave
magnetic field produced by the ESR microcoil with the NMR microcoil
terminated with a high impedance (or without the NMR microcoil) show
that the average microwave magnetic field in the sample region could
be increased by almost 1 order of magnitude (and, as expected, in
the center of the coil would be approximately given by B1 = μ0Imw/d ≅ 400 μT).
Figure 2
Simulations
and experiments for characterization of the microwave
magnetic field produced by the ESR microcoil. (a) Three-dimensional
representation of the ESR and NMR microcoils together with the map
of the microwave magnetic field B1, defined as one-half of the component perpendicular to the
static magnetic field B0 (indicated by
the red arrow) of the microwave magnetic field produced by the microwave
current into the ESR microcoil. In a, the two perpendicular cross
sections correspond to the region occupied by the water solution inside
the capillary. (b and c) Maps of B1 in two larger regions. Black dashed lines indicate the region
occupied by the water solution inside the capillary. Simulations are
performed using COMSOL Multiphysics (COMSOL Inc.). Amplitude of the
microwave current in the ESR microcoil is set to 74 mA, according
to the combined results of simulations of the ESR/NMR-integrated electronics
performed with an Advanced Design System (ADS, Keysight Technologies)
and Cadence (Cadence Design Systems Inc.). (d) ESR spectra of a sample
of BDPA placed in the center of the ESR microcoil for different ESR
oscillator supply voltages VDD. Increase
of the resonance static magnetic field is due to the increase of the
oscillator frequency with the oscillator supply voltage VDD. (e) Microwave magnetic field B1 in the center of the ESR coil extracted
from measurement of the line width of the ESR signals shown in d according
to the equation ΔB0, = (2/γT2)(1 + γ2B12T1T2)1/2,[56] where ΔB0, is the field difference
between the two zero crossings of the ESR signal and corresponds to
the peak-to-peak line width of the dispersion signal measurable without
magnetic field modulation. Experimental conditions: fmw≅ 10.7 GHz. Modulation frequency: fm = 16.7 kHz. Modulation magnetic field: Bm = 6 μT.
Simulations
and experiments for characterization of the microwave
magnetic field produced by the ESR microcoil. (a) Three-dimensional
representation of the ESR and NMR microcoils together with the map
of the microwave magnetic field B1, defined as one-half of the component perpendicular to the
static magnetic field B0 (indicated by
the red arrow) of the microwave magnetic field produced by the microwave
current into the ESR microcoil. In a, the two perpendicular cross
sections correspond to the region occupied by the water solution inside
the capillary. (b and c) Maps of B1 in two larger regions. Black dashed lines indicate the region
occupied by the water solution inside the capillary. Simulations are
performed using COMSOL Multiphysics (COMSOL Inc.). Amplitude of the
microwave current in the ESR microcoil is set to 74 mA, according
to the combined results of simulations of the ESR/NMR-integrated electronics
performed with an Advanced Design System (ADS, Keysight Technologies)
and Cadence (Cadence Design Systems Inc.). (d) ESR spectra of a sample
of BDPA placed in the center of the ESR microcoil for different ESR
oscillator supply voltages VDD. Increase
of the resonance static magnetic field is due to the increase of the
oscillator frequency with the oscillator supply voltage VDD. (e) Microwave magnetic field B1 in the center of the ESR coil extracted
from measurement of the line width of the ESR signals shown in d according
to the equation ΔB0, = (2/γT2)(1 + γ2B12T1T2)1/2,[56] where ΔB0, is the field difference
between the two zero crossings of the ESR signal and corresponds to
the peak-to-peak line width of the dispersion signal measurable without
magnetic field modulation. Experimental conditions: fmw≅ 10.7 GHz. Modulation frequency: fm = 16.7 kHz. Modulation magnetic field: Bm = 6 μT.To cross-check these simulation results, we performed experiments
with a single crystal of 1:1 α,γ-bisdiphenylene-β-phenylallyl:benzene
(BDPA/benzene, 152560, Sigma-Aldrich) having a size of about 70 ×
70 × 5 μm3 placed in the center of the ESR microcoil
(Figure d). At room
temperature, BDPA has relaxation times T1≅ T2≅ 100 ns.[57] Since the line width depends
only on B1 and the relaxation
times,[56] knowledge of the relaxation times
and measurement of the line width allows one to estimate the value
of B1. From the measured
ESR signals reported in Figure d, the extracted value of B1 in the center of the ESR microcoil is about 70 μT
at the maximum supply voltage of the oscillator of 2 V (Figure e), corresponding to the simulated
maximum current in the ESR coil of 74 mA. As discussed above, the
full-wave COMSOL simulation gives a B1 in the center of the ESR microcoil of about 100
μT, in good agreement with the 70 μT estimated from these
BDPA measurements.
Experimental Results
The capillary-encapsulated
samples of TEMPOL/H2O solutions
are fixed on top of the single-chip DNP microsystem with a small drop
of vacuum grease (high-vacuum grease, Dow Corning). As shown on the
left side of Figure , at concentrations of 1 and 10 mM the ESR spectra consist of three
hyperfine lines due the 15N nucleus (I = 1). At concentrations of 100 mM and 1 M a single line is observed,
as also reported by Gafurov et al.[58] After
obtaining the ESR spectrum, the B0 magnetic
field is set to one of the three maxima for concentrations of 1 and
10 mM and to the single maximum for the 100 mM and 1 M concentrations.
For all NMR spectra shown on the right side of Figure , the rf pulse length is τrf = 5 μs, the acquisition time is Tdaq = 400 ms, and the pulse repetition time is Tr = 500 ms. The nonenhanced NMR spectra are the average of Navg = 100 000 spectra obtained in about
14 h. The DNP-enhanced NMR spectra are the average of Navg = 1000 spectra obtained in about 9 min. In the DNP-enhanced
NMR measurements, the microwave excitation is present also during
the NMR detection. The frequency shift between the nonenhanced and
the enhanced NMR spectra is caused by the dc current flowing in the
two-turn ESR microcoil. Even though the coil is designed such that
the magnetic field created by the dc current running in each section
is almost entirely canceled by the magnetic field created by the dc
current running in the adjacent section, a small field is still created
in the direction of B0 due to the nonzero
distance (about 14 μm) between the two wires. The observed shift
(up to 4 ppm) is much larger than that attributable to temperature
effects. Since the temperature-induced frequency shift for the 1H nucleus in water is about 0.01 ppm/◦C,[59] the observed shift would correspond to a temperature
increase of 400 ◦C, which is obviously impossible
for a liquid water sample. In a future version of the integrated ESR
detector, this shift will be entirely suppressed by an ESR oscillator
design in which no dc current runs through the ESR microcoil, such
as the Colpitts oscillator.[54]
Figure 3
ESR (left column)
and NMR (right column) spectra of TEMPOL/H2O solutions
for different concentrations (1 mM, 10 mM, 100
mM, and 1 M). In the NMR spectra, the red curves are the nonenhanced
(B1 = 0) NMR spectra
enlarged 5 times whereas the blue curves are the DNP-enhanced NMR
spectra (B1≅ 60 μT). DNP enhancement ε values for
each spectra are given on the top-right corner. Enhancement is defined
as the ratio of the integrals of the enhanced and nonenhanced NMR
signals in the frequency domain. ESR measurements are performed in
the following conditions. Modulation frequency: fm = 16.7 kHz. Modulation magnetic field: Bm ≅ 6 μT. Microwave frequency: fmw ≅ 10.7 GHz. Microwave magnetic field: B1 ≅ 60 μT. NMR
measurements are performed in the following conditions. frf ≅ 16 MHz. Pulse length: τrf = 5 μs. Pulse repetition time: Tr = 500 ms. Time-domain match filter time constant: Tm = 100 ms. Acquisition time: Tdaq = 400 ms. Number of averaging: Navg =
100 000 (for the nonenhanced signal) and Navg = 1000 for the enhanced signal.
ESR (left column)
and NMR (right column) spectra of TEMPOL/H2O solutions
for different concentrations (1 mM, 10 mM, 100
mM, and 1 M). In the NMR spectra, the red curves are the nonenhanced
(B1 = 0) NMR spectra
enlarged 5 times whereas the blue curves are the DNP-enhanced NMR
spectra (B1≅ 60 μT). DNP enhancement ε values for
each spectra are given on the top-right corner. Enhancement is defined
as the ratio of the integrals of the enhanced and nonenhanced NMR
signals in the frequency domain. ESR measurements are performed in
the following conditions. Modulation frequency: fm = 16.7 kHz. Modulation magnetic field: Bm ≅ 6 μT. Microwave frequency: fmw ≅ 10.7 GHz. Microwave magnetic field: B1 ≅ 60 μT. NMR
measurements are performed in the following conditions. frf ≅ 16 MHz. Pulse length: τrf = 5 μs. Pulse repetition time: Tr = 500 ms. Time-domain match filter time constant: Tm = 100 ms. Acquisition time: Tdaq = 400 ms. Number of averaging: Navg =
100 000 (for the nonenhanced signal) and Navg = 1000 for the enhanced signal.Figure reports
the DNP enhancements (Figure a) and the NMR line widths (Figure b) obtained with TEMPOL/H2O solutions
having different concentrations. A maximum enhancement of about −50
is obtained with the largest B1 (i.e., about 70 μT) with a 10 mM degassed solution.
Enhancements on the order of −100 have been previously reported
at 0.3 T.[60−66] In particular, Höfer et al. reported an enhancement of −100
for a 10 mM solution of TEMPOL/H2O at room temperature
(i.e., with the same test sample and experimental conditions of this
work).[60] The lower enhancement measured
in our work is probably due to the lower average B1 which, as shown in Figure , is insufficient to reach
the saturation region. This is caused by the NMR coil inside the ESR
coil, which reduces significantly the microwave magnetic field B1 produced by the ESR coil
as explained in the description of the system. Figure shows also that the degassing of water increases
the enhancement, especially for low radical concentrations. The enhancement
of the NMR signal for the 1 mM degassed solution is almost three times
larger than that measured with the nondegassed solution. For the 10
mM sample, the enhancement increase caused by degassing is much lower,
probably due to the negligible additional relaxation induced by the
presence of the O2 molecules (0.2 mM at room temperature)
with respect to the relaxation due to 10 mM TEMPOL molecules. As shown
in Figure b, for 1
and 100 mM concentrations, the NMR line width is about 20 Hz (i.e.,
1 ppm), mainly limited by the lack of shim coils in the electromagnet
(a very similar chip operated in a 7 T magnet equipped with shimming
coils shows spectral resolutions down to about 2 Hz[46]). The NMR line width of the 1 M solution is about 90 Hz,
presumably due to the large concentration of TEMPOL, which significantly
reduces the T2 value.[60,67]
Figure 4
DNP
enhancement and NMR line width. (a) Enhancement ε and
(b) NMR line width of TEMPOL/H2O solutions with different
concentrations (1 mM, 10 mM, 100 mM, and 1 M) at different microwave
magnetic fields B1.
Enhancement is defined as the ratio of the integrals of the enhanced
and nonenhanced NMR signals in the frequency domain. NMR line width
is defined as the full width at half-maximum of the NMR signal in
the frequency domain. At an operating frequency of 16 MHz, a line
width of 20 Hz corresponds to 1.25 ppm.
DNP
enhancement and NMR line width. (a) Enhancement ε and
(b) NMR line width of TEMPOL/H2O solutions with different
concentrations (1 mM, 10 mM, 100 mM, and 1 M) at different microwave
magnetic fields B1.
Enhancement is defined as the ratio of the integrals of the enhanced
and nonenhanced NMR signals in the frequency domain. NMR line width
is defined as the full width at half-maximum of the NMR signal in
the frequency domain. At an operating frequency of 16 MHz, a line
width of 20 Hz corresponds to 1.25 ppm.
Outlook
In this work we demonstrated, for the first time, the integration
of a DNP microsytem consisting of an NMR transceiver and an ESR oscillator
on a single silicon chip of less than 2 mm2. Measurements
on TEMPOL/H2O solutions, performed at 10.7 GHz(ESR)/16
MHz(NMR), show enhancements as large as −50 on samples having
an effective volume of about 1 nL. In the following, we discuss the
improvements and extensions of the approach demonstrated in this work.
A straightforward but rather modest increase of the SNR can be obtained
by a narrowband design of the integrated NMR receiver. Despite the
generally lower enhancement factors observed in liquid-state DNP at
high magnetic fields, a very significant improvement of the SNR (and
of the spectral separation of chemically shifted signals) can be obtained
by increasing the operating frequency of the NMR/ESR subsystems.[22] A moderate increase of the frequency to the
40 GHz(ESR)/60 MHz(NMR) region could allow for low-cost DNP-enhanced
NMR spectrometers in permanent magnets. The use of state-of-the-art
submicrometer-integrated circuit technologies should allow the extension
of the single-chip DNP microsystem approach proposed here up the THz(ESR)/GHz(NMR)
region,[68−70] corresponding to the strongest static magnetic fields
currently available. We previously reported single-chip ESR detectors
operating at up to 146 GHz[51] and single-chip
NMR detectors operating at up to 300 MHz.[42,44,47] However, the combination of single-chip
NMR/ESR detectors at these frequencies (and above) has been not demonstrated
yet. The main technical challenge for extension of the proposed single-chip
DNP microsystem approach to higher frequencies is the coupling between
the NMR and the ESR excitation/detection structures, which influence
the strength of the microwave field B1. In order to obtain a sufficiently large B1, the NMR structure and its
impedance termination should be carefully codesigned with the ESR
structure with possible drawbacks in terms of NMR sensitivity. Another
interesting opportunity offered by the single-chip approach is the
possibility to create dense arrays of such sensors for parallel DNP-enhanced
NMR spectroscopy of a large number of nanoliter and subnanoliter different
samples (or a bigger volume of the same sample). In addition, preliminary
measurements performed with the DNP microsystem proposed in this work
show that it can be operated also at temperatures down to 4 K, at
least. Hence, the single-chip DNP approach proposed here could be
well suited also for the study of DNP processes other than the Overhauser
effect in liquids at room temperature.
Authors: Tomas Orlando; Rıza Dervişoğlu; Marcel Levien; Igor Tkach; Thomas F Prisner; Loren B Andreas; Vasyl P Denysenkov; Marina Bennati Journal: Angew Chem Int Ed Engl Date: 2018-12-20 Impact factor: 15.336