Manoja Kumar Das1, Narendra Kumar Arora1, Thomas Mathew2, Bhadresh Vyas3, Salam Kabita Devi1, Abhishek Yadav1.
Abstract
OBJECTIVE: To document the temperature integrity at the vaccine storage devices at various levels in three states of India.Entities:
Keywords: Cold chain; Datalogger; Freezing; Temperature excursion; Temperature maintenance; Vaccines
Year: 2020 PMID: 32518666 PMCID: PMC7268609 DOI: 10.1186/s40794-020-00109-z
Source DB: PubMed Journal: Trop Dis Travel Med Vaccines ISSN: 2055-0936
Fig. 1Process of selection of facilities and cold chain devices in the states and study districts
Description of vaccine stores and cold chain devices included for temperature monitoring
| Level/Devices | States | Total | |||
|---|---|---|---|---|---|
| Bihar | Gujarat | Kerala | |||
| Level of stores | State VS | 1 | 0 | 0 | 1 |
| Division VS | 7 | 6 | 3 | 16 | |
| District VS | 10 | 9 | 5 | 24 | |
| CHC/PHC | 76 | 49 | 30 | 155 | |
| Total | 94 | 64 | 36 | 196 | |
| Type of cold chain devices | WIC | 10 | 6 | 4 | 20 |
| ILR | 95a | 61 | 38 | 189 | |
| Refrigerator | 1 | 0 | 0 | 1 | |
| Cold box | 3 | 0 | 0 | 3 | |
| Total | 109 | 67 | 42 | 218 | |
VS Vaccine store, CHC Community health centre, PHC Primary health centre, ILR Ice lined refrigerator, WIC Walk-in cooler, Refrigerator Domestic refrigerator
aRecordings from 5 ILRs (from 3 PHCs and 2 District vaccine stores) were less than 7 complete days and excluded from analysis
Temperature integrity in cold chain devices according to type of vaccine stores
| Levels/ Devices | State | Durationa (hours) | % of time spent | Average temp (°C) | Temperature range (°C) | ||||
|---|---|---|---|---|---|---|---|---|---|
| <− 3 °C | -3 °C to | 0 °C to 2 °C | 2 °C to 8 °C | > 8 °C | |||||
| According to levels of stores | |||||||||
| State and Divisions | Bihar | 4107.5 | 4.5 | 8.6 | 12.8 | 66.6 | 7.4 | 2.7 | −16.1, 26.1 |
| Gujarat | 1395 | 0.1 | 0.3 | 11.2 | 86.2 | 1.9 | 4.8 | −7.2, 19.7 | |
| Kerala | 860.5 | 0 | 0 | 0 | 99.6 | 0.4 | 4.6 | 2.1, 33.2 | |
| Total | 6363 | 2.9 | 5.6 | 10.7 | 75.4 | 5.2 | 3.8 | −16.1, 33.2 | |
| District | Bihar | 5163.7 | 6.4 | 6.5 | 16.8 | 44.7 | 25.5 | 2.6 | −22.9, 23.9 |
| Gujarat | 3228.7 | 0 | 0.1 | 2 | 97.4 | 0.4 | 4.8 | −1.2, 25.3 | |
| Kerala | 1914.5 | 0 | 1 | 1.1 | 97.9 | 0 | 3.7 | −1.5, 12.8 | |
| Total | 10,307 | 3.2 | 3.5 | 9.3 | 71.1 | 12.9 | 3.6 | −22.9, 25.3 | |
| Sub-district | Bihar | 23,728.2 | 9.6 | 10.1 | 14.5 | 58 | 7.8 | 2.4 | −33.0, 34.9 |
| Gujarat | 12,115.2 | 6.3 | 13.7 | 25.1 | 53.4 | 1.5 | 2.0 | −14.1, 33.3 | |
| Kerala | 6526.2 | 0 | 6.5 | 36.8 | 53.9 | 2.8 | 2.8 | −2.6, 34.6 | |
| Total | 42,369.7 | 7.2 | 10.5 | 21 | 56 | 5.2 | 2.4 | − 33.0, 34.9 | |
| According to type of devices | |||||||||
| WIC | Bihar | 3483.2 | 0.5 | 8.2 | 10.6 | 72 | 8.6 | 4.4 | −4.6, 26.1 |
| Gujarat | 1036.7 | 0.1 | 0.4 | 15.1 | 81.5 | 2.5 | 4.6 | −7.2, 19.7 | |
| Kerala | 718.5 | 0 | 0. | 0 | 99.5 | 0.5 | 4.3 | 2.1, 33.2 | |
| Total | 5238.5 | 0.4 | 5.5 | 10.1 | 77.6 | 6.3 | 4.4 | −7.2, 33.2 | |
| ILR | Bihar | 28,452.2 | 9.7 | 9.8 | 14.9 | 56.4 | 9.1 | 1.9 | −33.0, 34.6 |
| Gujarat | 15,702.2 | 4.8 | 10.6 | 19.8 | 63.5 | 1.3 | 2.7 | −14.1, 33.3 | |
| Kerala | 8582.7 | 0 | 5.1 | 28.2 | 64.5 | 2.1 | 3.0 | −2.6, 33.2 | |
| Total | 52,737.2 | 6.7 | 9.3 | 18.5 | 59.9 | 5.6 | 2.4 | −33.0, 34.6 | |
| Other devicesb | Bihar | 1064 | 0 | 0.3 | 21.9 | 22.1 | 55.5 | 9 | −2.3, 24.6 |
| Total | 1064 | 0 | 0.3 | 21.9 | 22.1 | 55.5 | 9 | −2.3, 24.6 | |
| Pooled for all levels and all types of devices | |||||||||
| Pooled all levels | Bihar | 32,999.5 | 8.5 | 9.3 | 14.7 | 57 | 10.5 | 2.4 | −33.0, 34.9 |
| Gujarat | 16,739 | 4.5 | 9.9 | 19.5 | 64.6 | 1.4 | 2.8 | −14.1, 33.3 | |
| Kerala | 9301.2 | 0 | 4.7 | 26 | 67.2 | 2 | 3.2 | −2.61, 34.6 | |
| Total | 59,039.7 | 6 | 8.8 | 17.8 | 60.8 | 6.6 | 2.7 | −33.0, 34.9 | |
WIC Walk-in cooler, ILR Ice lined refrigerator
aDuration of temperature recording by data loggers in the cold chain devices in hours
b Other devices (domestic refrigerator and cold boxes) were used for vaccine storage in Bihar only at the time of the study
Continuous episodes of temperature excursion below 0 °C in cold chain devices
| Total observation perioda | Exposure to temperature < 0 °C | ||||||
|---|---|---|---|---|---|---|---|
| Total durationa (hours) | Total continuous episodes durationa (hours) | Duration of continuous episodesb | |||||
| ≤1 h | > 1–9 h | ≥10 h | |||||
| State and Division | Bihar | 4107.5 | 540 | 28.9 | 12.7 | 14 | 2.2 |
| Gujarat | 1395 | 5.75 | 0.7 | 0.2 | 0.5 | 0 | |
| Kerala | 860.5 | 0 | 0 | 0 | 0 | 0 | |
| Total | 6363 | 545.7 | 29.7 | 13 | 14.5 | 2.2 | |
| District | Bihar | 5163.7 | 666.7 | 5.9 | 0.7 | 2.7 | 2.5 |
| Gujarat | 3228.7 | 4.5 | 0.7 | 0 | 0.7 | 0 | |
| Kerala | 1914.5 | 19 | 1 | 0 | 1 | 0 | |
| Total | 10,307 | 690.2 | 7.7 | 0.7 | 4.5 | 2.5 | |
| Sub-district | Bihar | 23,728.2 | 4663.2 | 41.9 | 5.2 | 19 | 17.7 |
| Gujarat | 12,115.2 | 2415.7 | 62.9 | 29.2 | 26.2 | 7.5 | |
| Kerala | 6526.2 | 421.2 | 14.9 | 3 | 10.2 | 1.7 | |
| Total | 42,369.7 | 7500.2 | 120 | 37.5 | 55.5 | 27 | |
| WICs | Bihar | 3483.2 | 303.2 | 22.2 | 12.2 | 8 | 2 |
| Gujarat | 1036.7 | 5.75 | 0.7 | 0.2 | 0.5 | 0 | |
| Kerala | 718.5 | 0 | 0 | 0 | 0 | 0 | |
| Total | 5238.5 | 309 | 23 | 12.5 | 8.5 | 2 | |
| ILRs | Bihar | 28,452.2 | 5563.2 | 53.7 | 5.7 | 27.5 | 20.5 |
| Gujarat | 15,702.2 | 2420.2 | 63.7 | 29.2 | 27 | 7.5 | |
| Kerala | 8582.7 | 440.2 | 16.4 | 3 | 11.7 | 1.7 | |
| Total | 52,737.2 | 8423.7 | 133.4 | 38 | 65.7 | 29.7 | |
| Other devicesc | Bihar | 1064 | 3.5 | 0.9 | 0.7 | 0.2 | 0 |
| Total | 1064 | 3.5 | 0.9 | 0.7 | 0.2 | 0 | |
| Pooled all levels | Bihar | 32,999.5 | 5870 | 76.9 | 18.7 | 35.7 | 22.5 |
| Gujarat | 16,739 | 2426 | 64.5 | 29.5 | 27.5 | 7.5 | |
| Kerala | 9301.2 | 440.2 | 16.4 | 3 | 11.7 | 1.7 | |
| Total | 59,039.7 | 8736.2 | 157.7 | 51.5 | 74.5 | 31.7 | |
WIC Walk-in cooler, ILR Ice lined refrigerator
a Duration in hours
b A continuous episode below 0 °C was counted when readings were continuously below 0 °C for at least 45 min or more
c Other devices (domestic refrigerator and cold boxes) were used for vaccine storage in Bihar only at the time of the study
Continuous episodes of temperature excursion above 8 °C in cold chain devices
| Total observation perioda (hours) | Exposure to temperature > 8 °C | ||||||
|---|---|---|---|---|---|---|---|
| Total durationa (hours) | Total continuous episodes durationa (hours) | Duration of continuous episodesb | |||||
| ≤1 h | > 1–9 h | ≥10 h | |||||
| State and Divisions | Bihar | 4107.5 | 302.7 | 14.5 | 4.25 | 8 | 2.25 |
| Gujarat | 1395 | 26 | 7 | 7 | 0 | 0 | |
| Kerala | 860.5 | 3.5 | 0.2 | 0 | 0.25 | 0 | |
| Total | 6363 | 332.2 | 21.7 | 11.25 | 8.25 | 2.25 | |
| District | Bihar | 5163.7 | 1318.7 | 2.2 | 0 | 0.5 | 1.75 |
| Gujarat | 3228.7 | 14.5 | 0.5 | 0.25 | 0 | 0.25 | |
| Kerala | 1914.5 | 0.5 | 0.2 | 0.25 | 0 | 0 | |
| Total | 10,307 | 1333.7 | 3 | 0.5 | 0.5 | 2 | |
| Sub-district | Bihar | 23,728.2 | 1846.2 | 28.5 | 8.75 | 11.75 | 8 |
| Gujarat | 12,115.2 | 186 | 6.5 | 4 | 1.2 | 1.2 | |
| Kerala | 6526.2 | 182.7 | 2.2 | 0.5 | 0 | 1.7 | |
| Total | 42,369.7 | 2215 | 38.2 | 13.5 | 12.7 | 12 | |
| WICs | Bihar | 3483.2 | 299.2 | 13.7 | 3.7 | 8 | 2 |
| Gujarat | 1036.7 | 26 | 7 | 7 | 0 | 0 | |
| Kerala | 718.5 | 3.5 | 0.2 | 0 | 0.2 | 0 | |
| Total | 5238.5 | 328.7 | 21 | 10.7 | 8.2 | 2 | |
| ILRs | Bihar | 28,452.2 | 2578 | 27.2 | 9 | 11.2 | 7.5 |
| Gujarat | 15,702.2 | 200.5 | 7.5 | 4.5 | 1.2 | 1.5 | |
| Kerala | 8582.7 | 183.2 | 2.5 | 0.7 | 0 | 1.7 | |
| Total | 52,737.2 | 2961.7 | 37.2 | 14.2 | 12.5 | 10.7 | |
| Other devicesc | Bihar | 1064 | 590.5 | 4.75 | 0.7 | 0.7 | 3.2 |
| Total | 1064 | 590.5 | 4.7 | 0.7 | 0.7 | 3.2 | |
| Pooled all levels | Bihar | 32,999.5 | 3467.7 | 46.2 | 13.5 | 20 | 12.7 |
| Gujarat | 16,739 | 226.5 | 14.5 | 11.5 | 1.2 | 1.5 | |
| Kerala | 9301.2 | 186.7 | 2.7 | 0.7 | 0.2 | 1.7 | |
| Total | 59,039.7 | 3881 | 63.2 | 25.7 | 21.5 | 16 | |
WIC Walk-in cooler, ILR Ice lined refrigerator
a Duration in hours
b A continuous episode above 8 °C was counted when readings were continuously above 8 °C for at least 45 min or more
c Other devices (domestic refrigerator and cold boxes) were used for vaccine storage in Bihar only at the time of the study