Shashi B Singh1, Dewesh Kumar1, Vivek Kashyap1, Surendra Singh1.
Abstract
INTRODUCTION: Automated blood pressure (BP) monitor is widely used to assess the blood pressure (BP) of the study subjects in community-based researches. This study aims at the detection of hypertension by automated BP device and examines the concordance and validity between automated and mercury-free LED BP devices.Entities:
Keywords: Automated BP device; concordance correlation coefficient; kappa statistic; mercury-free LED BP instrument; validity
Year: 2020 PMID: 32509634 PMCID: PMC7266193 DOI: 10.4103/jfmpc.jfmpc_796_19
Source DB: PubMed Journal: J Family Med Prim Care ISSN: 2249-4863
Comparison of systolic and diastolic blood pressure measurements by two instruments
| Method | Mean | SD | Min | Max | R Squared | Correlation | CV | CCC | 95% CI | |
|---|---|---|---|---|---|---|---|---|---|---|
| Systolic blood pressure | Automated blood pressure monitor | 125.9 | 23.47 | 86 | 214 | |||||
| Mercury-free LED BP device | 118.3 | 22.9 | 77 | 189 | 0.88 | 0.94 | 0.07 | 0.89 | (0.86-0.91) | |
| Diastolic blood pressure | Automated blood pressure monitor | 74.45 | 12.2 | 48 | 117 | |||||
| Mercury-free LED BP device | 72.62 | 11.13 | 51 | 111 | 0.75 | 0.87 | 0.08 | 0.85 | (0.82-0.88) |
Regression Coefficient-systolic and diastolic blood pressure
| Parameter | Systolic BP | Diastolic BP | ||
|---|---|---|---|---|
| Intercept | Slope | Intercept | Slope | |
| Regression Coefficients | 12.31 | 0.96 | 5.40 | 0.95 |
| Lower 95% C.L. | 7.39 | 0.92 | 0.83 | 0.89 |
| Upper 95% C.L. | 17.23 | 1.0 | 9.97 | 1.01 |
| Standard Error | 2.50 | 0.02 | 2.32 | 0.03 |
| T | 4.92 | 46.26 | 2.32 | 30.06 |
| Probability level ( | 0.00 | 0.00 | 0.02 | 0.00 |
Figure 1Linear regression curve of systolic blood pressure
Figure 2Linear regression analysis of diastolic blood pressure
Comparison of patients based on the BP categorization by two different instruments
| Automated BP device | Mercury-free LED BP Device | Validity measures | |||
|---|---|---|---|---|---|
| Hypertensive | Non Hypertensive | Total | |||
| Hypertensive | 57 | 19 | 76 | Sensitivity=96.61%. | |
| Nonhypertensive | 2 | 222 | 224 | ||
| Total | 59 | 241 | 300 | ||
Characteristics of the ROC curve
| Systolic blood pressure | Sensitivity (%) | Specificity (%) | ROC characteristics |
|---|---|---|---|
| JNC VII (≥140 mmHg) | 98.1 | 92 | Area under ROC=0.984 |
| ROC curve cut of value at 139 mmHg | 100 | 91 | SE=0.005 |
| 95% CI = (0.973, 0.994) | |||
| JNC VII (≥90 mmHg) | 82.6 | 93 | Area under ROC=0.97 |
| ROC curve cut of value at 86 mmHg | 95.7 | 90 | SE=0.01 |
| 95% CI = (0.950, 0.989) | |||
Figure 3ROC curve of systolic blood pressure
Figure 4ROC curve of diastolic blood pressure