Literature DB >> 32485473

Evaluation of nCoV-QS (MiCo BioMed) for RT-qPCR detection of SARS-CoV-2 from nasopharyngeal samples using CDC FDA EUA qPCR kit as a gold standard: An example of the need of validation studies.

Byron Freire-Paspuel1, Patricio Vega-Mariño2, Alberto Velez2, Paulina Castillo2, Marilyn Cruz2, Miguel Angel Garcia-Bereguiain3.   

Abstract

BACKGROUND: Several qPCR kits are available for SARS-CoV-2 diagnosis, mostly lacking of evaluation due to covid19 emergency.
OBJECTIVE: We evaluated nCoV-QS (MiCo BioMed) kit using CDC kit as gold standard.
RESULTS: We found limitations for nCoV-QS: 1) lower sensitivity 2) lack of RNA quality control probe.
CONCLUSIONS: Validation studies should be implemented for any SARS-CoV-2 RT-qPCR commercial kit to prevent unreliable diagnosis.
Copyright © 2020 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  CDC; RT-qPCR; SARS-CoV-2; Validation

Mesh:

Substances:

Year:  2020        PMID: 32485473      PMCID: PMC7243755          DOI: 10.1016/j.jcv.2020.104454

Source DB:  PubMed          Journal:  J Clin Virol        ISSN: 1386-6532            Impact factor:   3.168


Background

Multiple in vitro RT-qPCR diagnosis kits are available on the market for the detection of SARS-CoV-2. Some of them have received emergency use authorization (EUA) from the U.S. Food & Drug Administration (FDA) while others only report validations made by manufacturers, and in general little is known about their performances using clinical specimens. The CDC designed 2019-nCoV CDC EUA kit (IDT, USA) is based on N1 and N2 probes to detect SARS-CoV-2 that have received positive evaluation on recent reports [[1], [2], [3]], and and RNase P as an RNA extraction quality control. Other kit avalaible in the market is nCoV-QS (MiCo BioMed; South Corea) that include probes "ORF3a" and "N" probes for SARS-CoV-2 detection but no probe for RNA extraction quality control, with no EUA approval neither from FDA (USA) nor from Korean CDC [[4], [5], [6]].

Objective

This study compared the performance in terms of positive percent agreement (PPA) of nCoV-QS (MiCo BioMed; South Corea) and 2019-nCoV CDC EUA kit (IDT, USA) primers and probes for SARS-CoV-2 qPCR diagnosis from nasopharyngeal samples.

Study design

Fifty-four (54) clinical specimens (nasopharyngeal swabs collected on 0.5 mL TE pH 8 buffer) from patients selected as suspicious for SARS-CoV-2 infection were included on this study during the surveillance in Galapagos Islands started on April 8th 2020. Also, six negative controls (TE pH 8 buffer) were included as control for carryover contamination. Both CoV-QS and 2019-nCoV CDC EUA kits were used at SARS-CoV-2 diagnosis laboratory "LabGal" at "Agencia de Regulación y Control de la Bioseguridad y Cuarentena para Galápagos" at Puerto Ayora in Galapagos Islands (Ecuador), where we considered this validation necessary to guarantee the sensibility of SARS-CoV-2 during the surveillance.

Results

Twenty-five (25) samples were tested following an adapted version of the CDC protocol [1] using CFX96 BioRad instrument and PureLink Viral RNA/DNA Mini Kit (Invitrogen, USA) as an alternate RNA extraction method, and also interpreting as positive 3 samples where a probe was positive with Ct<40 and the second one with Ct values up to 41.15 (See Table 1 and Table 2 a). We performed this protocol for both nCoV-QS and 2019-nCoV CDC EUA primers and probes kits. Nine samples were negative for both kits; Sixteen samples were positive for 2019-nCoV CDC EUA (range of Ct values: 23.02–41.15 for N1; 24.08–40.12 for N2), but only ten (PPA 62.5 %; p < 0.001) of those ones were positive for nCoV-QS (range of Ct values: 28.71–39.98 for ORF3a; 24.48–35.44 for N). Results are detailed on Table 1a. The assay was validated to detect less than 10 viral RNA copies/uL by using 2019-nCoV N positive control (IDT, USA).
Table 1

Performance of nCoV-QS compared to 2019-nCoV CDC EUA for RT-qPCR SARS-CoV-2 diagnosis (% values: PPA).

CDC Probes SARS CoV-2 PositiveCDC Probes SARS CoV-2 Negative
Veri-Q ProbesSARS CoV-2 Positive22 (66.7 %)0
Veri-Q ProbesSARS CoV-2 Negative1121
Table 2

Ct values for nCoV-QS and 2019-nCoV CDC EUA RT-qPCR using CDC adapted protocol for 25 samples (a) and MiCoBioMed protocol for 29 samples (b).

SampleConc. [copies/μL]CT (2019-nCoV CDC EUA)
Result (2019-nCoV CDC EUA)CT (nCoV-QS)
Result (nCoV-QS)
N1N2RPORF3aN
1193453.323.0224.825.04Positive28.7224.48Positive
2151418.723.3425.1622.69Positive32.1924.63Positive
328204.025.5328.0226.11Positive30.7827.3Positive
47157.327.3129.8225.43Positive34.7630.31Positive
5699.330.3433.2325.32Positive37.2932.25Positive
6684.130.3733.325.29PositiveN/A33.58Positive
7657.430.4232.3729.23Positive38.2632.16Positive
8261.631.6233.7627.32Positive39.8833.67Positive
9236.531.7534.421.73Positive39.9834.24Positive
10162.432.2434.8723.17Positive30.0435.44Positive
1115.235.3337.4726.37PositiveN/AN/ANegative
129.036.0140.0924.87PositiveN/AN/ANegative
132.837.5339.6225.27PositiveN/AN/ANegative
141.738.240.1226.97PositiveN/AN/ANegative
151.238.5839.9927.74PositiveN/AN/ANegative
160.241.1539.5127.98PositiveN/AN/ANegative
17N/AN/AN/A23.11NegativeN/AN/ANegative
18N/AN/AN/A26.6NegativeN/AN/ANegative
19N/AN/AN/A25.56NegativeN/AN/ANegative
20N/AN/AN/A25.12NegativeN/AN/ANegative
21N/AN/AN/A25.26NegativeN/AN/ANegative
22N/AN/AN/A27.63NegativeN/AN/ANegative
23N/AN/AN/A26.35NegativeN/AN/ANegative
24N/AN/AN/A25.6NegativeN/AN/ANegative
25N/AN/AN/A27.77NegativeN/AN/ANegative
Performance of nCoV-QS compared to 2019-nCoV CDC EUA for RT-qPCR SARS-CoV-2 diagnosis (% values: PPA). Ct values for nCoV-QS and 2019-nCoV CDC EUA RT-qPCR using CDC adapted protocol for 25 samples (a) and MiCoBioMed protocol for 29 samples (b). Twenty-nine (29) samples were tested following instructions manual from MiCo BioMed for nCoV-QS kit [6], using MiCo BioMed One RT-qPCR kit. PureLink Viral RNA/DNA Mini Kit (Invitrogen, USA) was used for RNA extraction. We performed this protocol for both nCoV-QS and 2019-nCoV CDC EUA primers and probes kits. Twelve samples were negative for both kits; Seventeen samples were positive for 2019-nCoV CDC EUA (range of Ct values: 23.1–39.05 for N1; 22.96–38.8 for N2), but only 12 (PPA 70.5 %; p < 0.001) of those ones were positive for nCoV-QS (range of Ct values: 26.45–39.43 for ORF3a; 24.03–39.89 for N). Results are detailed on Table 1 and Table 2b. We used CFX96 BioRad to run qPCR but also results were confirmed using Veri-Q PCR316 instrument from MiCo BioMed [4]. The assay sensitivity indicated on manufacturers manual (1.8 copies/uL for OFR3a and 4.24 copies/uL for N) could not be validated because positive control concentration was not provided. In summary, overall PPA for nCoV-QS was 66.7 % (22 out of 33 positives samples for 2019-nCoV CDC EUA; p < 0.001), and 70.5 % and 62.5 % for MiCo BioMed and adapted CDC protocols, respectively. Additionally, considering the viral loads calculated following adapted CDC protocol with 2019-nCoV N positive control (IDT, USA), the limit of detection (viral copies/uL) for nCoV-QS kit is much higher than the one indicated at manufacturer's manual [6].

Discussion

Although the main limitation of our study is the sample size (54 specimens), our results support that nCoV-QS kit had a significant lower performance in terms on PPA and sensitivity compared to 2019-nCoV CDC EUA. Also, the lack of any probe for RNA extraction quality control like RNase P and the unreported concentration of positive controls provided for the kit that does not allow viral load calculations, are limitations to be considered when using nCoV-QS kit. Considering the worldwide high demand of reagents for SARS-CoV RT-qPCR diagnosis, supplies shortage is a fact, actually affecting harder to developing countries like Ecuador. Under this scenario, validation studies are helpful to guarantee the quality of the supplies in the market for every country in the world.

Ethical considerations

All samples have been submitted for routine patient care and diagnostics. Ethical approval for this study was not required since all activities are according to legal provisions defined by the "Comité de Operaciones Especiales Regional de Galápagos" that is leading the Covid19 surveillance in Galapagos Islands. No extra specimens were specifically collected for this validation study. All data used in the current study was anonymized prior to being obtained by the authors.

Authorship contribution statement

All authors contributed to study conceptualization, experimental procedures and revision and approval of final version of the manuscript. Byron Freire-Paspuel and Miguel Angel García Bereguiain analyzed the data and wrote the manuscript.

Funding

None.

Declaration of Competing Interest

All authors have no conflict of interest to declare.
  22 in total

Review 1.  The crucial contribution of the universities on the SARS-CoV-2 surveillance in Ecuador: Lessons for developing countries.

Authors:  Daniela Santander-Gordon; Gabriel A Iturralde; Byron Freire-Paspuel; Marlon S Zambrano-Mila; Diana C Morales; Paolo A Vallejo-Janeta; Barbara Coronel; Heberson Galvis; Tatiana Jaramillo; Christian D Bilvao; Maria B Rodriguez-Paredes; Sebastian Rodriguez-Pazmiño; Juan C Laglaguano; Henry Herrera; Ana Maria Tito; Esteban Ortiz-Prado; Ismar A Rivera-Olivero; Aquiles R Henriquez-Trujillo; Tannya Lozada; Miguel Angel Garcia-Bereguiain
Journal:  One Health       Date:  2021-05-25

2.  Rapid, point-of-care antigen and molecular-based tests for diagnosis of SARS-CoV-2 infection.

Authors:  Jacqueline Dinnes; Jonathan J Deeks; Sarah Berhane; Melissa Taylor; Ada Adriano; Clare Davenport; Sabine Dittrich; Devy Emperador; Yemisi Takwoingi; Jane Cunningham; Sophie Beese; Julie Domen; Janine Dretzke; Lavinia Ferrante di Ruffano; Isobel M Harris; Malcolm J Price; Sian Taylor-Phillips; Lotty Hooft; Mariska Mg Leeflang; Matthew Df McInnes; René Spijker; Ann Van den Bruel
Journal:  Cochrane Database Syst Rev       Date:  2021-03-24

3.  Analytical and clinical evaluation of a heat shock SARS-CoV-2 detection method without RNA extraction for N and E genes RT-qPCR.

Authors:  Alfredo Bruno; Domenica de Mora; Byron Freire-Paspuel; Angel S Rodriguez; Maria Belen Paredes-Espinosa; Maritza Olmedo; Martha Sanchez; Jennifer Romero; Michelle Paez; Manuel Gonzalez; Alberto Orlando; Miguel Angel Garcia-Bereguiain
Journal:  Int J Infect Dis       Date:  2021-06-21       Impact factor: 3.623

4.  Analytical and Clinical Evaluation of "AccuPower SARS-CoV-2 Multiplex RT-PCR kit (Bioneer, South Korea)" and "Allplex 2019-nCoV Assay (Seegene, South Korea)" for SARS-CoV-2 RT-PCR Diagnosis: Korean CDC EUA as a Quality Control Proxy for Developing Countries.

Authors:  Byron Freire-Paspuel; Miguel Angel Garcia-Bereguiain
Journal:  Front Cell Infect Microbiol       Date:  2021-06-10       Impact factor: 5.293

5.  COVID-19 outbreaks among isolated Amazonian indigenous people, Ecuador.

Authors:  Aquiles R Henriquez-Trujillo; Esteban Ortiz-Prado; Ismar A Rivera-Olivero; Nemonte Nenquimo; Andrés Tapia; Mitchell Anderson; Tannya Lozada; Miguel Angel Garcia-Bereguiain
Journal:  Bull World Health Organ       Date:  2021-07-01       Impact factor: 9.408

Review 6.  Diagnostic techniques for COVID-19 and new developments.

Authors:  Elham Sheikhzadeh; Shimaa Eissa; Aziah Ismail; Mohammed Zourob
Journal:  Talanta       Date:  2020-07-14       Impact factor: 6.057

7.  "One health" inspired SARS-CoV-2 surveillance: The Galapagos Islands experience.

Authors:  Byron Freire-Paspuel; Patricio Vega-Mariño; Alberto Velez; Paulina Castillo; Carlos Masaquiza; Ronald Cedeño-Vega; Tannya Lozada; Marilyn Cruz; Miguel Angel Garcia-Bereguiain
Journal:  One Health       Date:  2020-10-20

8.  "Sample pooling of RNA extracts to speed up SARS-CoV-2 diagnosis using CDC FDA EUA RT-qPCR kit".

Authors:  Byron Freire-Paspuel; Patricio Vega-Mariño; Alberto Velez; Marilyn Cruz; Miguel Angel Garcia-Bereguiain
Journal:  Virus Res       Date:  2020-09-24       Impact factor: 3.303

9.  Analytical sensitivity and clinical performance of a triplex RT-qPCR assay using CDC N1, N2, and RP targets for SARS-CoV-2 diagnosis.

Authors:  Byron Freire-Paspuel; Miguel Angel Garcia-Bereguiain
Journal:  Int J Infect Dis       Date:  2020-10-25       Impact factor: 3.623

Review 10.  Considerations for diagnostic COVID-19 tests.

Authors:  Olivier Vandenberg; Delphine Martiny; Olivier Rochas; Alex van Belkum; Zisis Kozlakidis
Journal:  Nat Rev Microbiol       Date:  2020-10-14       Impact factor: 78.297