| Literature DB >> 32472220 |
Felicity Stokes1, Cecile Acquaviva-Bourdain2, Bernd Hoppe3, John C Lieske4, Elisabeth Lindner5, Greg Toulson6, Frédéric M Vaz7, Gill Rumsby8.
Abstract
Measurement of oxalate in the blood is essential for monitoring primary hyperoxaluria patients with progressive renal impairment and on dialysis prior to transplantation. As no external quality assurance scheme is available for this analyte, we conducted a sample exchange scheme between six laboratories specifically involved with the investigation of primary hyperoxaluria to compare results. The methodologies compared were gas chromatography/mass spectrometry (GCMS), ion chromatography with mass spectrometry (ICMS), and enzymatic methods using oxalate oxidase and spectrophotometry. Although individual laboratories performed well in terms of reproducibility and linearity, there was poor agreement (absolute values) between centres as illustrated by a longer-term comparison of patient results from two of the participating laboratories. This situation was only partly related to differences in calibration and mainly reflected the lower recoveries seen with the ultrafiltration of samples. These findings lead us to conclude that longitudinal monitoring of primary hyperoxaluria patients with deteriorating kidney function should be performed by a single consistent laboratory and the methodology used should always be defined. In addition, plasma oxalate concentrations reported in registry studies and those associated with the risk of systemic oxalosis in published studies need to be interpreted in light of the methodology used. A reference method and external quality assurance scheme for plasma oxalate analysis would be beneficial.Entities:
Keywords: Method comparison; Plasma oxalate; Primary hyperoxaluria; Ultrafiltration
Mesh:
Substances:
Year: 2020 PMID: 32472220 PMCID: PMC7666277 DOI: 10.1007/s00240-020-01197-4
Source DB: PubMed Journal: Urolithiasis ISSN: 2194-7228 Impact factor: 3.436
Sample preparation steps for the individual laboratories
| Laboratory | Dilution/diluent | Filter (cut off) | Centrifugation | Extraction | Method (platform) | Stated linearity (µmol/L) |
|---|---|---|---|---|---|---|
| 1 | 1:1 with 0.24 M HCl prior to filtration | Amicon ultra (30 kDa) | 18,000 g 4 °C, 10 min | No | Oxalate oxidase (Ilab Aries) | 2–60 (up to 200 on dilution) |
| 2 | 1:1 with water prior to filtration | Amicon ultra (30 kDa) | 15,000 g 4 °C, 10 min | No | Oxalate oxidase (Indiko Plus) | 4–200 |
| 3 | 1:1 with NaCl + 200 µL 6 M HCl | None | Ethyl acetate | GCMS | 2–60 | |
| 4 | 200 μL + 30 μL 12 M HCl | None | Ethyl acetate | GCMS | 0.5–420 | |
| 5 | 500 μL + 20 μL 2 M HCl (added to filtrate after ultrafiltration) | Centrisart I ultrafiltration vial (10 kDa) | 1500 g 4 °C, 20 min | ICMS | 0.3-upper limit not defined | |
| 6 | 1 mL plasma + 10 μL 12 M HCl | Amicon ultra (30 kDa) | 2300 g, 20 °C, 30 min | Oxalate oxidase (microwell plate) | 1–300 |
Fig. 1Comparison of plasma oxalate results obtained from a PH patient over a period of months. Solid symbols (filled square) GCMS, open (unfilled square) ICMS results
Fig. 2Bias plot of results obtained by ICMS and GCMS. The solid line denotes mean % difference; dashed lines denote ± 2SD
Results from spiked samples
| Laboratory | Mean result (SD) | ||
|---|---|---|---|
| Neat plasma | + 11 μmol/L | + 35 μmol/L | |
| 1 | 7 (0.2) | 17 (0.5) | 39 (0.2) |
| 2 | 13 (0.3) | 23 (0.1) | 47 (0.6) |
| 3 | 6 (1.0) | 18 (0.7) | 47 (1.4) |
| 4 | 9 (0.2) | 21 (0.1) | 47 (0.6) |
| 5 | 15 (0.3) | 24 (0.5) | 45 (0.8) |
| 6 | 6 (0.2) | 14 (0.1) | 33 (0.3) |
| All lab mean (%CV) | 9 (41.6) | 19.3 (17.7) | 43.2 (12.7) |
Oxalate results are reported in whole numbers, standard deviation to 1 dp. SD were calculated for the blinded duplicate samples for each lab. Inter-laboratory CV was calculated for each sample
CV coefficient of variation
Recovery from spiked samples
| Laboratory | % Recovery | |
|---|---|---|
| + 11 μmol/L | + 35 μmol/L | |
| 1 | 90 | 91 |
| 2 | 92 | 98 |
| 3 | 105 | 117 |
| 4 | 105 | 108 |
| 5 | 77 | 84 |
| 6 | 77 | 77 |
Recovery = (spiked result-neat plasma)/amount spiked)
Analysis of a high oxalate blood sample diluted with plasma to minimise the change in matrix
| Laboratory | Plasma oxalate result (μmol/L) | |||
|---|---|---|---|---|
| PH patient plasma | 1 in 2 | 1 in 4 | Diluent plasma | |
| 1 | 28 | 21 | 15 | 9 |
| 2 | 45 | 34 | 25 | 15 |
| 3 | 33 | 29 | 28 | 12 |
| 4 | 50 | 37 | 26 | 17 |
| 5 | 29 | 21 | 17 | 14 |
| 6 | 27 | 21 | 15 | 9 |
| All lab mean (%CV) | 36 (25.1) | 28 (24.8) | 21 (26.4) | 13 (25.0) |
Fig. 3Results obtained following dilution of human plasma
Results from analysis of duplicate aqueous standards
| Laboratory | Added oxalate (µmol/L) | ||
|---|---|---|---|
| 150 | 50 | 25 | |
| Mean measured oxalate μmol/L (SD) | |||
| 1 | 164 (1.4) | 52 (0) | 26 (0) |
| 2 | 152 (2.0) | 50 (0.1) | 25 (0.6) |
| 3 | 161 (4.2) | 60 (0.8) | 28 (0) |
| 4 | 158 (0.6) | 52 (0.7) | 25 (0.1) |
| 5 | 143 (0) | 39 (0.3) | 19 (0) |
| 6 | 150 (0.6) | 53 (3.5) | 25 (0.2) |
| All lab mean (%CV) | 155 (4.9) | 51 (12.2) | 25 (12.1) |
Oxalate results are reported in whole numbers, standard deviation to 1 dp. SD was calculated for the blinded duplicate samples for each lab. Inter-laboratory CV was calculated for each sample
CV coefficient of variation
Fig. 4Comparison of aqueous standards by the six different laboratories. The % difference is plotted against the known oxalate concentration
Fig. 5Precision plot of inter-laboratory %CVs from each stage of the method comparison