A Comparison of the Visual Field Parameters of SITA Faster and SITA Standard Strategies in Glaucoma.

Raghavan Lavanya1, Mohammed Riyazuddin1, Srilakshmi Dasari1, Narendra K Puttaiah1, Jayasree P Venugopal2, Zia S Pradhan2, Sathi Devi2, Shruthi Sreenivasaiah2, Shonraj B Ganeshrao3, Harsha L Rao1.   

Abstract

PURPOSE: To compare the visual field (VF) parameters of the new Swedish Interactive Thresholding Algorithm (SITA), SITA Faster (SFR) with that of SITA Standard (SS) on the Humphrey Field Analyzer.
METHODS: Ninety-seven eyes of 97 subjects (63 glaucoma, 26 glaucoma suspects, and 8 normal eyes) underwent VF examination with SFR and SS strategies on the same day in random order. Agreement in VF parameters between SFR and SS strategies was assessed by Bland and Altman plots. In addition, some subjects underwent a second VF examination with SFR strategy to evaluate its test-retest variability.
RESULTS: The median test duration of SS strategy was 6 minutes 14 seconds, whereas SFR was 2 minutes 49 seconds (55% shorter, P<0.001). Median mean deviation (-7.3 vs. -7.6 dB, P=0.73) and VF index (88 vs. 88%, P=0.32) were similar between the 2 strategies, whereas pattern standard deviation was significantly higher (4.8 vs. 4.7 dB, P=0.01) with SS strategy. Overall average threshold sensitivity and Garway-Heath sector-wise threshold sensitivities were similar between the 2 strategies except for the nasal sector where SFR strategy had higher sensitivity (26 vs. 25 dB, P=0.02). Bland-Altman plots showed the mean difference in all VF parameters between the SS and SFR strategies were small (ranging from -1.0 dB for the nasal sector to -0.01 dB for superotemporal sector sensitivity). The test-retest variability of VF parameters with SFR strategy was low.
CONCLUSIONS: VF parameters with SFR showed good agreement with that of SS strategy. This, combined with low test-retest variability, suggests that SFR can be considered for diagnosis and monitoring of glaucoma.

Entities:  

Mesh:

Year:  2020        PMID: 32459685     DOI: 10.1097/IJG.0000000000001551

Source DB:  PubMed          Journal:  J Glaucoma        ISSN: 1057-0829            Impact factor:   2.503


  4 in total

1.  The Effect of Transitioning from SITA Standard to SITA Faster on Visual Field Performance.

Authors:  Alex T Pham; Pradeep Y Ramulu; Michael V Boland; Jithin Yohannan
Journal:  Ophthalmology       Date:  2021-03-30       Impact factor: 14.277

2.  The Frontloading Fields Study (FFS): Detecting Changes in Mean Deviation in Glaucoma Using Multiple Visual Field Tests Per Clinical Visit.

Authors:  Jack Phu; Michael Kalloniatis
Journal:  Transl Vis Sci Technol       Date:  2021-11-01       Impact factor: 3.283

3.  A Comparative Study between Swedish Interactive Thresholding Algorithm Faster and Swedish Interactive Thresholding Algorithm Standard in Glaucoma Patients.

Authors:  Núria Mendieta; Joel Suárez; Cristina Blasco; Romina Muñiz; Carmen Pueyo
Journal:  J Curr Ophthalmol       Date:  2021-10-22

4.  Differences in visual field loss pattern when transitioning from SITA standard to SITA faster.

Authors:  Christopher T Le; Jacob Fiksel; Pradeep Ramulu; Jithin Yohannan
Journal:  Sci Rep       Date:  2022-04-29       Impact factor: 4.996