| Literature DB >> 32447879 |
Jee Hyeon Lee1, Sang Min Lee1, Jeong Hee Yoon2,3, Min-Jeong Kim1, Hong Il Ha1, Sung-Joon Park4, Eun Soo Kim1, Kwanseop Lee1, Jeong Min Lee2,3,5.
Abstract
PURPOSE: This study quantified the impact of respiratory motion on liver stiffness measurements according to different shear wave elastography (SWE) techniques and region of interest (ROI) methods, using liver fibrosis phantoms.Entities:
Keywords: Breath holding; Elasticity imaging techniques; Liver Cirrhosis; Motion; Phantom, Imaging; Region of interest
Year: 2020 PMID: 32447879 PMCID: PMC7758106 DOI: 10.14366/usg.19079
Source DB: PubMed Journal: Ultrasonography ISSN: 2288-5919
Fig. 1.Illustration of the phantom on the moving platform.
To reflect various degrees of respiratory motion, each liver fibrosis phantom was placed on a moving platform with three moving modes (static mode and moving mode at low speed and high speed).
Study design
| Operator[ | Phantoms[ | Moving platform[ | SWE technique | ROI method |
|---|---|---|---|---|
| 1 | 3.6 kPa | Static mode | pSWE (ElastPQ) | Fixed ROI |
| 2 | 16.1 kPa | Low-speed moving mode | 2D-SWE (ElastQ Imaging) | Circular ROI |
| 3 | 16.7 kPa | High-speed moving mode | Point ROI | |
| 18.4 kPa | Box ROI | |||
| Multiple ROIs |
SWE, shear wave elastography; pSWE, point shear wave elastography; 2D-SWE, 2-dimensional shear wave elastography; ROI, region of interest.
Three operators measured the stiffness values of the phantoms on the moving platforms with three movement conditions using two techniques of SWE. Operator 1 performed two measurement sessions at a 24-hour interval to assess intraobserver agreement.
Stiffness values of the phantoms measured using transient elastography.
Automatic electric cradle used to reflect various breathing conditions.
Fig. 2.Stiffness value measurements using point shear wave elastography (pSWE) (A) and 2-dimensional shear wave elastography (2D-SWE) (B-E).
The figures show pSWE (A) and the four methods of obtaining a region of interest (ROI) in 2D-SWE (B-E) performed on phantom 2. The four ROI methods in 2D-SWE included a circular ROI (B), a point ROI (C), a box ROI (D),and multiple ROIs (E).
Impact of motion on technical failure, reliability, stiffness values, and measurement time of pSWE and 2D-SWE
| Static mode | Moving mode | P-value[ | Low-speed moving mode | High-speed moving mode | P-value[ | ||
|---|---|---|---|---|---|---|---|
| Technical failure[ | pSWE | 0/12 (0) | 1/24 (4.2) | >0.99 | 0/12 (0) | 1/12 (8.3) | >0.99 |
| 2D-SWE | 0/12 (0) | 1/24 (4.2) | >0.99 | 0/12 (0) | 1/12 (8.3) | >0.99 | |
| P-value[ | >0.99 | >0.99 | >0.99 | >0.99 | |||
| Unreliable measurements[ | pSWE | 0/12 (0) | 12/24 (50.0) | 0.003 | 7/12 (58.3) | 5/12 (41.7) | 0.684 |
| 2D-SWE | 0/12 (0) | 1/24 (4.2) | >0.99 | 0/12 (0) | 0/12 (0) | >0.99 | |
| P-value[ | >0.99 | <0.001 | <0.001 | <0.001 | |||
| CVs (%)[ | pSWE | 7.6 (3.8) | 26.9 (21.0) | 0.013 | 30.9 (21.6) | 19.8 (21.9) | 0.429 |
| 2D-SWE | 2.9 (2.1) | 4.7 (6.5) | 0.022 | 6.1 (7.3) | 4.0 (2.6) | 0.288 | |
| P-value[ | 0.002 | <0.001 | 0.007 | <0.001 | |||
| Stiffness values (kPa)[ | pSWE | 20.8 (8.2) | 19.5 (11.5) | 0.190 | 19.5 (5.1) | 19.1 (15.0) | 0.931 |
| 2D-SWE | 21.9 (9.7) | 21.7 (2.7) | 0.566 | 21.6 (9.8) | 21.8 (1.7) | 0.316 | |
| P-value[ | 0.052 | 0.031 | 0.107 | 0.130 | |||
| Measurement time (s)[ | pSWE | 93.0 (75.5) | 102.0 (18.8) | 0.666 | 106.0 (21.8) | 92.5 (20.0) | 0.126 |
| 2D-SWE | 155.5 (59.0) | 172.0 (66.0) | 0.532 | 155.5 (74.0) | 180.0 (66.3) | 0.667 | |
| P-value[ | 0.053 | <0.001 | 0.218 | 0.015 |
Values are presented as number (%) unless otherwise indicated.
The P-value of <0.05 represent statistical significance.
pSWE, point shear wave elastography; 2D-SWE, 2-dimensional shear wave elastography; CV, coefficient of variation; SWE, shear wave elastography.
Comparison between static mode and moving mode at low speed and high speed in both SWE techniques.
Comparison between low-speed moving mode and high-speed moving mode in both SWE techniques.
The presence of technical failure and unreliable measurements were compared using the Fisher exact test or chi-square test.
Comparison between pSWE and 2D-SWE using a single-circle region of interest in each movement condition.
The CVs, stiffness values, and measurement time were presented as median values with interquartile range and compared using the Mann-Whitney test for data points without technical failure or unreliable measurements in both SWE techniques.
Fig. 3.Impact of motion on point shear wave elastography (pSWE) and 2-dimensional shear wave elastography (2D-SWE) measurements of stiffness values.
Graphs of moving modes and coefficient of variation (CV) (A), stiffness values (B), and measurement time (C) for each shear wave elastography technique. *Statistical significance.
Impact of motion on technical failure, reliability, stiffness values, and measurement time of four phantoms with different stiffness values
| Static mode | Moving mode | P-value[ | Low-speed moving mode | High-speed moving mode | P-value[ | ||
|---|---|---|---|---|---|---|---|
| Technical failure[ | |||||||
| Phantom 1 | pSWE | 0/3 (0) | 0/6 (0) | >0.99 | 0/3 (0) | 0/3 (0) | >0.99 |
| 2D-SWE | 0/3 (0) | 0/6 (0) | >0.99 | 0/3 (0) | 0/3 (0) | >0.99 | |
| Phantom 2 | pSWE | 0/3 (0) | 1/6 (16.7) | 0.480 | 0/3 (0) | 1/3 (33.3) | 0.317 |
| 2D-SWE | 0/3 (0) | 0/6 (0) | >0.99 | 0/3 (0) | 0/3 (0) | >0.99 | |
| Phantom 3 | pSWE | 0/3 (0) | 0/6 (0) | >0.99 | 0/3 (0) | 0/3 (0) | >0.99 |
| 2D-SWE | 0/3 (0) | 0/6 (0) | >0.99 | 0/3 (0) | 0/3 (0) | >0.99 | |
| Phantom 4 | pSWE | 0/3 (0) | 0/6 (0) | >0.99 | 0/3 (0) | 0/3 (0) | >0.99 |
| 2D-SWE | 0/3 (0) | 0/6 (0) | >0.99 | 0/3 (0) | 0/3 (0) | >0.99 | |
| P-value[ | pSWE | >0.99 | 0.372 | >0.99 | 0.352 | ||
| 2D-SWE | >0.99 | >0.99 | >0.99 | >0.99 | |||
| Unreliable measurements[ | |||||||
| Phantom 1 | pSWE | 0/3 (0) | 3/6 (50.0) | 0.157 | 2/3 (66.7) | 1/3 (33.3) | 0.456 |
| 2D-SWE | 0/3 (0) | 1/6 (16.7) | 0.480 | 0/3 (0) | 1/3 (33.3) | 0.317 | |
| Phantom 2 | pSWE | 0/3 (0) | 3/6 (50.0) | 0.157 | 3/3 (100) | 0/3 (0) | NA |
| 2D-SWE | 0/3 (0) | 0/6 (0) | >0.99 | 0/3 (0) | 0/3 (0) | >0.99 | |
| Phantom 3 | pSWE | 0/3 (0) | 3/6 (50.0) | 0.157 | 1/3 (33.3) | 2/3 (66.7) | 0.456 |
| 2D-SWE | 0/3 (0) | 0/6 (0) | >0.99 | 0/3 (0) | 0/3 (0) | >0.99 | |
| Phantom 4 | pSWE | 0/3 (0) | 3/6 (50.0) | NA | 1/3 (33.3) | 2/3 (66.7) | 0.456 |
| 2D-SWE | 0/3 (0) | 0/6 (0) | >0.99 | 0/3 (0) | 0/3 (0) | >0.99 | |
| P-value[ | pSWE | >0.99 | >0.99 | 0.287 | 0.287 | ||
| 2D-SWE | >0.99 | >0.99 | >0.99 | 0.161 | |||
| CVs (%)[ | |||||||
| Phantom 1 | pSWE | 7.9 (3.3) | 6.8 (12.4) | 0.702 | 5.4 (0) | 12.3 (11.0) | NA |
| 2D-SWE | 2.9 (7.0) | 7.5 (7.1) | 0.252 | 7.6 (5.9) | 7.5 (7.0) | 0.800 | |
| Phantom 2 | pSWE | 7.2 (94.5) | 34.0 (0.9) | 0.800 | NA | 34.0 (0.9) | NA |
| 2D-SWE | 2.7 (2.0) | 3.4 (8.6) | 0.172 | 10.4 (8.2) | 2.9 (0.7) | 0.197 | |
| Phantom 3 | pSWE | 8.0 (5.2) | 26.9 (20.5) | 0.103 | 34.6 (15.4) | 21.8 (0.0) | NA |
| 2D-SWE | 3.8 (2.7) | 5.0 (9.4) | 0.906 | 4.6 (9.0) | 5.4 (3.5) | >0.99 | |
| Phantom 4 | pSWE | 5.2 (4.7) | 30.9 (34.2) | 0.100 | 38.4 (14.9) | 11.6 (0.0) | NA |
| 2D-SWE | 2.7 (3.0) | 5.2 (13.2) | 0.378 | 4.7 (13.2) | 5.7 (5.9) | >0.99 | |
| P-value[ | pSWE | 0.787 | 0.148 | 0.301 | 0.232 | ||
| 2D-SWE | 0.557 | 0.702 | 0.920 | 0.191 | |||
| Stiffness values (kPa)[ | |||||||
| Phantom 1 | pSWE | 6.2 (0.4) | 5.9 (0.2) | 0.403 | 5.8 (0) | 5.9 (0.1) | NA |
| 2D-SWE | 5.2 (0.2) | 5.1(0.5) | 0.142 | 4.9 (0.4) | 5.1(0.1) | 0.401 | |
| Phantom 2 | pSWE | 20.5 (1.8) | 19.9 (3.8) | >0.99 | NA | 19.9 (3.8) | NA |
| 2D-SWE | 22.1 (0.3) | 21.8 (1.4) | 0.361 | 21.9 (1.1) | 21.8 (0.7) | >0.99 | |
| Phantom 3 | pSWE | 21.1 (1.3) | 20.5 (0.4) | 0.702 | 20.5 (0.1) | 20.1 (0.0) | NA |
| 2D-SWE | 21.6 (0.7) | 21.6 (1.0) | 0.901 | 21.7 (0.4) | 21.6 (1.0) | >0.99 | |
| Phantom 4 | pSWE | 21.4 (0.6) | 19.5 (2.4) | 0.103 | 19 (0.9) | 21 (0.0) | NA |
| 2D-SWE | 24.1 (1.0) | 23.7 (5.9) | 0.700 | 23.3 (5.8) | 24.1 (1.9) | 0.702 | |
| P-value[ | pSWE | 0.041 | 0.108 | 0.169 | 0.310 | ||
| 2D-SWE | 0.021 | 0.002 | 0.082 | 0.036 | |||
| Measurement time (s)[ | |||||||
| Phantom 1 | pSWE | 88.0 (10.0) | 96.5 (29.0) | 0.710 | 106.0 (0) | 84.0 (6.0) | NA |
| 2D-SWE | 154.0 (56) | 134.5 (104.0) | >0.99 | 129.0 (78.0) | 140 (89) | 0.702 | |
| Phantom 2 | pSWE | 154.0 (114.0) | 101.0 (20.0) | 0.401 | NA | 101.0 (20.0) | NA |
| 2D-SWE | 118.0 (90.0) | 158.5 (156) | 0.702 | 145.0 (136.0) | 172.0 (89.0) | >0.99 | |
| Phantom 3 | pSWE | 80.0 (157.0) | 100.0 (20.0) | 0.441 | 100.0 (14.0) | 98.0 (15.0) | 0.268 |
| 2D-SWE | 193.0 (75.0) | 197.5 (183.0) | 0.908 | 197.0 (89.0) | 212.0 (183.0) | 0.701 | |
| Phantom 4 | pSWE | 89.0 (135.0) | 101.0 (60.0) | 0.550 | 126.5 (49.0) | 109.0 (0.0) | NA |
| 2D-SWE | 162 (25.0) | 188.5 (80.0) | 0.550 | 197.0 (60.0) | 180.0 (78.0) | >0.99 | |
| P-value[ | pSWE | 0.332 | 0.329 | 0.818 | 0.310 | ||
| 2D-SWE | 0.512 | 0.360 | 0.542 | 0.702 | |||
Values are presented as number (%) unless otherwise indicated.
The P-value <0.05 represent statistical significance.
Phantoms 1-4: phantoms with different stiffness values (3.6, 16.1, 16.7, and 18.4 kPa, respectively) as measured using transient elastography.
pSWE, point shear wave elastography; 2D-SWE, 2-dimensional shear wave elastography; CV, coefficient of variation; NA, not applicable; SWE, shear wave elastography.
Comparison between static mode and moving mode at low speed and high speed in both SWE techniques.
Comparison between low-speed moving mode and high-speed moving mode in both SWE techniques.
The presence of technical failure and unreliable measurements were compared using the Fisher exact test or the chi-square test.
Comparison between pSWE and 2D-SWE in each moving mode.
The CVs, stiffness values, and measurement time were presented as median values with interquartile range and compared using the Mann-Whitney test or Kruskal-Wallis test for data points without technical failure or unreliable measurements in both SWE techniques.
Impact of motion on technical failure, reliability, stiffness values, and measurement time of four ROI methods in 2D-SWE
| Static mode | Moving mode | P-value[ | Low-speed moving mode | High-speed moving mode | P-value[ | |
|---|---|---|---|---|---|---|
| Technical failure[ | ||||||
| Circular ROI | 0/12 (0) | 0/24 (0) | >0.99 | 0/12 (0) | 0/12 (0) | >0.99 |
| Point ROI | 0/12 (0) | 0/24 (0) | >0.99 | 0/12 (0) | 0/12 (0) | >0.99 |
| Box ROI | 0/12 (0) | 0/24 (0) | >0.99 | 0/12 (0) | 0/12 (0) | >0.99 |
| Multiple ROIs | 0/12 (0) | 0/24 (0) | >0.99 | 0/12 (0) | 0/12 (0) | >0.99 |
| P-value[ | >0.99 | >0.99 | >0.99 | >0.99 | ||
| Unreliable measurements[ | ||||||
| Circular ROI | 0/12 (0) | 1/24 (4.2) | >0.99 | 0/12 (0) | 1/12 (8.3) | >0.99 |
| Point ROI | 0/12 (0) | 0/24 (0) | >0.99 | 0/12 (0) | 0/12 (0) | >0.99 |
| Box ROI | 0/12 (0) | 0/24 (0) | >0.99 | 0/12 (0) | 0/12 (0) | >0.99 |
| Multiple ROIs | 0/12 (0) | 0/24 (0) | >0.99 | 0/12 (0) | 0/12 (0) | >0.99 |
| P-value[ | >0.99 | 0.387 | >0.99 | 0.382 | ||
| CVs (%)[ | ||||||
| Circular ROI | 2.9 (2.1) | 4.7 (6.5) | 0.022 | 6.1 (7.3) | 4.0 (2.6) | 0.268 |
| Point ROI | 7.1 (5.2) | 6.1 (3.0) | 0.840 | 7.1 (3.1) | 6.6 (3.5) | 0.356 |
| Box ROI | 2.7 (5.3) | 5.7 (4.0) | 0.023 | 6.4 (4.2) | 5.3 (3.8) | 0.863 |
| Multiple ROIs | 5.0 (2.1) | 6.9 (4.0) | 0.159 | 6.9 (3.0) | 6.0 (4.8) | 0.686 |
| P-value[ | 0.004 | 0.363 | 0.880 | 0.169 | ||
| Stiffness values (kPa)[ | ||||||
| Circular ROI | 21.9 (9.7) | 21.7 (2.7) | 0.566 | 21.6 (9.8) | 21.8 (1.7) | 0.295 |
| Point ROI | 22.8 (10.8) | 22.2 (10.4) | 0.801 | 22.1 (10.6) | 22.2 (10.2) | 0.795 |
| Box ROI | 22.2 (9.4) | 20.8 (9.8) | 0.174 | 20.8 (9.1) | 20.8 (9.9) | 0.931 |
| Multiple ROIs | 22.4 (9.9) | 21.3 (9.5) | 0.356 | 21.3 (9.5) | 21.7 (9.5) | 0.418 |
| P-value[ | 0.974 | 0.458 | 0.697 | 0.667 | ||
| Measurement time (s)[ | ||||||
| Circular ROI | 155.5 (59.0) | 172.0 (66.0) | 0.532 | 155.5 (74.0) | 180.0 (66.3) | 0.667 |
| Point ROI | 151.5 (40.5) | 183.0 (67.0) | 0.041 | 168.0 (54.5) | 192.5 (68.5) | 0.106 |
| Box ROI | 164.0 (50.5) | 208.0 (61.5) | 0.004 | 184.0 (56.5) | 238.0 (64.5) | 0.002 |
| Multiple ROIs | 41.0 (8.0) | 50.5 (33.0) | <0.001 | 46.5 (15.5) | 62.5 (51.0) | 0.021 |
| P-value[ | <0.001 | <0.001 | <0.001 | <0.001 |
Values are presented as number (%) unless otherwise indicated.
The P-value of <0.05 represent statistical significance.
The post hoc analysis of CV and measurement times among the four ROI methods is presented in Table 5.
ROI, region of interest; 2D-SWE, 2-dimensional shear wave elastography; CV, coefficient of variation; SWE, shear wave elastography.
Comparison between static mode and moving mode at low speed and high speed across four ROI methods in 2D-SWE.
Comparison between low-speed moving mode and high-speed moving mode across four ROI methods in 2D-SWE.
The presence of technical failure and unreliable measurements were compared using the Fisher exact test or chi-square test.
Comparison between pSWE and 2D-SWE in each moving mode.
The CVs, stiffness values, and measurement time were presented as median values with interquartile range and compared using the Mann-Whitney test or Kruskal-Wallis test for data points without technical failure or unreliable measurements in both SWE techniques.
Post hoc analysis of CVs and measurement time among four ROI methods in 2D-SWE
| P-value | |||||
|---|---|---|---|---|---|
| CV | Measurement time | ||||
| Static mode | Static mode | Moving mode | Low-speed moving mode | High-speed moving mode | |
| Circular ROI vs. point ROI | 0.001 | 0.773 | 0.456 | 0.644 | 0.388 |
| Circular ROI vs. box ROI | 0.630 | 0.665 | 0.011 | 0.402 | 0.006 |
| Circular ROI vs. multiple ROIs | 0.028 | <0.001 | <0.001 | <0.001 | <0.001 |
| Point ROI vs. box ROI | 0.006 | 0.370 | 0.036 | 0.340 | 0.024 |
| Point ROI vs. multiple ROIs | 0.160 | <0.001 | <0.001 | <0.001 | <0.001 |
| Box ROI vs. multiple ROIs | 0.089 | <0.001 | <0.001 | <0.001 | <0.001 |
The P-values were calculated using the Mann-Whitney test and corrected using the Bonferroni method. The P-value <0.008 represent statistical significance.
CV, coefficient of variation; ROI, region of interest; 2D-SWE, 2-dimensional shear wave elastography.
Fig. 4.Impact of motion in various region of interest (ROI) methods for 2-dimensional shear wave elastography on stiffness value measurements.
Graphs of the moving modes and coefficient of variation (CV) (A), stiffness values (B), and measurement time (C) in each ROI method. *Statistical significance.