Test Agreement between Roche Cobas 6800 and Cepheid GeneXpert Xpress SARS-CoV-2 Assays at High Cycle Threshold Ranges.

Kari Broder1, Ahmed Babiker1, Charles Myers1, Terri White1, Heather Jones1, John Cardella1, Eileen M Burd1, Charles E Hill1, Colleen S Kraft2,3.   

Abstract

Entities:  

Keywords:  COVID-19; RT-PCR; SARS-CoV-2; diagnostics; virology

Mesh:

Year:  2020        PMID: 32444354      PMCID: PMC7383541          DOI: 10.1128/JCM.01187-20

Source DB:  PubMed          Journal:  J Clin Microbiol        ISSN: 0095-1137            Impact factor:   5.948


× No keyword cloud information.

LETTER

The SARS-CoV-2 pandemic has changed the face of the globe and upended the daily lives of billions. In an effort to bring mass testing to as many as possible, multiple diagnostic tests, including molecular, antigen detection, and serological assays, have been rapidly developed. Under U.S. Food and Drug Administration (FDA) Emergency Use Authorization (EUA), several reverse transcription-PCR (RT-PCR) assays have reached U.S. laboratories, each with its own testing capacity and proprietary methods (1, 2). Multiple logistical challenges have required laboratories to validate and implement multiple platforms for testing, but data on positive percent agreement across platforms are limited. Our institution utilizes the Roche Cobas 6800 SARS-CoV-2 assay, the Cepheid GeneXpert Xpress SARS-CoV-2 assay, and a laboratory-developed test (LDT) based on a modified CDC protocol, but there is no gold standard for the diagnostic accuracy of these assays. Thus, our objective was to determine the degree of agreement between two tests by running the same samples across both platforms and comparing the cycle threshold (C) values (E target to E target) among samples with high (>30) C values, corresponding to lower viral loads. We collected 35 positive (positivity determined per assay instructions) nasopharyngeal samples with an E target C value of ≥30 on the Roche Cobas 6800 assay; those samples then underwent secondary testing on the Cepheid GeneXpert assay within 3 days of initial testing. Discrepancies were resolved using the LDT. All specimens were tested using each manufacturer’s protocol. E target C values were compared by Bland-Altman analysis. Of 35 samples, 34 tested positive on both instruments. One sample tested positive on the Cobas 6800 assay (C = 37.9) and negative by the GeneXpert assay and was confirmed to be negative on the LDT. Among the positive samples, C values were similar between the two assays (P = 0.06). The values ranged from 30.1 to 37.9 (mean, 36.7 ± 1.88) on the Roche Cobas 6800 assay and 24.6 to 42.4 (mean, 32.8 ± 4.07) on the GeneXpert assay. Bland-Altman analysis revealed a bias of 0.33 ± 3.21 (mean difference of −1.59, 95% limits of agreement of −5.97 and 6.63), signifying close agreement between the 2 instruments with a high standard deviation (Fig. 1). Our findings corroborate recently published results which show close agreement between the Cobas 6800 and GeneXpert instruments, in particular, among samples with lower viral loads (3). Our in-house laboratory test determined that the limit of detection (LoD) of the Cobas 6800 assay is 100 copies/ml, and the LoD specified in the GeneXpert package insert is 250 copies/ml. This difference in LoD becomes significant at higher C values, where the negative bias becomes more pronounced (4, 5). Previous studies have shown that substantial viral loads can be detected around day 5 of infection and drop at differing rates over the course of the illness (6). At our institution, the GeneXpert is utilized as a rapid test (turnaround time, 45 min) to triage patients for admission among other criteria; therefore, it is imperative that our testing modality can capture patients who present at either end of their illness course and who may have lower viral counts/higher C values. Overall, the Cepheid GeneXpert Xpress SARS-CoV-2 assay and the Roche Cobas 6800 SARS-CoV-2 assay showed a high level of agreement among patients with high C values. This allows the laboratories to utilize both assays concurrently as fits with local testing algorithms. Further studies are required to confirm percent agreement across different platforms and specimen types to expand on our findings.
FIG 1

Bland Altman plot of agreement. The Bland Altman plot shows C values with bias of 0.33 ± 3.21 (95% limits of agreement, −5.97 and 6.63).

Bland Altman plot of agreement. The Bland Altman plot shows C values with bias of 0.33 ± 3.21 (95% limits of agreement, −5.97 and 6.63).
  6 in total

1.  Virological assessment of hospitalized patients with COVID-2019.

Authors:  Roman Wölfel; Victor M Corman; Wolfgang Guggemos; Michael Seilmaier; Sabine Zange; Marcel A Müller; Daniela Niemeyer; Terry C Jones; Patrick Vollmar; Camilla Rothe; Michael Hoelscher; Tobias Bleicker; Sebastian Brünink; Julia Schneider; Rosina Ehmann; Katrin Zwirglmaier; Christian Drosten; Clemens Wendtner
Journal:  Nature       Date:  2020-04-01       Impact factor: 49.962

Review 2.  Understanding Bland Altman analysis.

Authors:  Davide Giavarina
Journal:  Biochem Med (Zagreb)       Date:  2015-06-05       Impact factor: 2.313

3.  Detection of SARS-CoV-2 by Use of the Cepheid Xpert Xpress SARS-CoV-2 and Roche cobas SARS-CoV-2 Assays.

Authors:  Angelica Moran; Kathleen G Beavis; Scott M Matushek; Carol Ciaglia; Nina Francois; Vera Tesic; Nedra Love
Journal:  J Clin Microbiol       Date:  2020-07-23       Impact factor: 5.948

4.  Common pitfalls in statistical analysis: Intention-to-treat versus per-protocol analysis.

Authors:  Priya Ranganathan; C S Pramesh; Rakesh Aggarwal
Journal:  Perspect Clin Res       Date:  2016 Jul-Sep

5.  Triplex Real-Time RT-PCR for Severe Acute Respiratory Syndrome Coronavirus 2.

Authors:  Jesse J Waggoner; Victoria Stittleburg; Renee Pond; Youssef Saklawi; Malaya K Sahoo; Ahmed Babiker; Laila Hussaini; Colleen S Kraft; Benjamin A Pinsky; Evan J Anderson; Nadine Rouphael
Journal:  Emerg Infect Dis       Date:  2020-06-21       Impact factor: 6.883

6.  SARS-CoV-2 Testing.

Authors:  Ahmed Babiker; Charlie W Myers; Charles E Hill; Jeannette Guarner
Journal:  Am J Clin Pathol       Date:  2020-05-05       Impact factor: 2.493

  6 in total
  15 in total

1.  Interpretation of single target positivity among SARS-CoV-2 RT-PCR result tests.

Authors:  Honorine Fenaux; Théo Ghelfenstein-Ferreira; Maud Salmona; Nadia Mahjoub; Linda Feghoul; Sarah Maylin; Marie-Laure Chaix; Marine Minier; Audrey Gabassi; Jérôme Le Goff; Constance Delaugerre
Journal:  J Clin Virol Plus       Date:  2021-05-16

Review 2.  Tools and Techniques for Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2)/COVID-19 Detection.

Authors:  Seyed Hamid Safiabadi Tali; Jason J LeBlanc; Zubi Sadiq; Oyejide Damilola Oyewunmi; Carolina Camargo; Bahareh Nikpour; Narges Armanfard; Selena M Sagan; Sana Jahanshahi-Anbuhi
Journal:  Clin Microbiol Rev       Date:  2021-05-12       Impact factor: 26.132

Review 3.  Laboratory tests for the detection of SARS-CoV-2 infection: basic principles and examples.

Authors:  Khaled R Alkharsah
Journal:  Ger Med Sci       Date:  2021-05-27

4.  Rapid, point-of-care antigen and molecular-based tests for diagnosis of SARS-CoV-2 infection.

Authors:  Jacqueline Dinnes; Jonathan J Deeks; Sarah Berhane; Melissa Taylor; Ada Adriano; Clare Davenport; Sabine Dittrich; Devy Emperador; Yemisi Takwoingi; Jane Cunningham; Sophie Beese; Julie Domen; Janine Dretzke; Lavinia Ferrante di Ruffano; Isobel M Harris; Malcolm J Price; Sian Taylor-Phillips; Lotty Hooft; Mariska Mg Leeflang; Matthew Df McInnes; René Spijker; Ann Van den Bruel
Journal:  Cochrane Database Syst Rev       Date:  2021-03-24

5.  Diagnosing SARS-CoV-2 with Antigen Testing, Transcription-Mediated Amplification and Real-Time PCR.

Authors:  Sascha Dierks; Oliver Bader; Julian Schwanbeck; Uwe Groß; Michael S Weig; Kemal Mese; Raimond Lugert; Wolfgang Bohne; Andreas Hahn; Nicolas Feltgen; Setare Torkieh; Fenja R Denker; Peer Lauermann; Marcus W Storch; Hagen Frickmann; Andreas Erich Zautner
Journal:  J Clin Med       Date:  2021-05-29       Impact factor: 4.241

6.  Validation of High-Sensitivity Severe Acute Respiratory Syndrome Coronavirus 2 Testing for Stool-Toward the New Normal for Fecal Microbiota Transplantation.

Authors:  Ahmed Babiker; Jessica M Ingersoll; Max W Adelman; Andrew S Webster; Kari J Broder; Victoria Stittleburg; Jesse J Waggoner; Colleen S Kraft; Michael H Woodworth
Journal:  Clin Transl Gastroenterol       Date:  2021-06-09       Impact factor: 4.488

7.  Detection of SARS-CoV-2 at the point of care.

Authors:  Michael J Loeffelholz; Yi-Wei Tang
Journal:  Bioanalysis       Date:  2021-07-22       Impact factor: 2.681

8.  Rapid, point-of-care antigen and molecular-based tests for diagnosis of SARS-CoV-2 infection.

Authors:  Jacqueline Dinnes; Jonathan J Deeks; Ada Adriano; Sarah Berhane; Clare Davenport; Sabine Dittrich; Devy Emperador; Yemisi Takwoingi; Jane Cunningham; Sophie Beese; Janine Dretzke; Lavinia Ferrante di Ruffano; Isobel M Harris; Malcolm J Price; Sian Taylor-Phillips; Lotty Hooft; Mariska Mg Leeflang; René Spijker; Ann Van den Bruel
Journal:  Cochrane Database Syst Rev       Date:  2020-08-26

9.  Turnaround times - the Achilles' heel of community screening and testing in Cape Town, South Africa: A short report.

Authors:  James D Porter; Robert Mash; Wolfgang Preiser
Journal:  Afr J Prim Health Care Fam Med       Date:  2020-10-02

Review 10.  Advances in Molecular Diagnosis of Tuberculosis.

Authors:  Emily MacLean; Mikashmi Kohli; Stefan F Weber; Anita Suresh; Samuel G Schumacher; Claudia M Denkinger; Madhukar Pai
Journal:  J Clin Microbiol       Date:  2020-09-22       Impact factor: 5.948

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.