| Literature DB >> 32439980 |
Sarah Hood1, Marshall K Ishmael2, Andrew Gunnell2, K B Foreman2,3, Tommaso Lenzi2.
Abstract
Motion capture is necessary to quantify gait deviations in individuals with lower-limb amputations. However, access to the patient population and the necessary equipment is limited. Here we present the first open biomechanics dataset for 18 individuals with unilateral above-knee amputations walking at different speeds. Based on their ability to comfortably walk at 0.8 m/s, subjects were divided into two groups, namely K2 and K3. The K2 group walked at [0.4, 0.5, 0.6, 0.7, 0.8] m/s; the K3 group walked at [0.6, 0.8, 1.0, 1.2, 1.4] m/s. Full-body biomechanics was collected using a 10-camera motion capture system and a fully instrumented treadmill. The presented open dataset will enable (i) clinicians to understand the biomechanical demand required to walk with a knee and ankle prosthesis at various speeds, (ii) researchers in biomechanics to gain new insights into the gait deviations of individuals with above-knee amputations, and (iii) engineers to improve prosthesis design and function.Entities:
Mesh:
Year: 2020 PMID: 32439980 PMCID: PMC7242470 DOI: 10.1038/s41597-020-0494-7
Source DB: PubMed Journal: Sci Data ISSN: 2052-4463 Impact factor: 6.444
Relevant Subject Information.
| Subject Code | Age(yrs) | Gender | Mass(kg) | Height(m) | Amputation side | Etiology | Age of Amputation (yrs) | K-Level | Prescribed Prosthesis | Socket Suspension | Training? (#) | Hand-rails? | |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Knee | Ankle | ||||||||||||
| TF01* | 26 | Male | 64.9 | 1.78 | Right | Traumatic | 5 | K3 | Plie FI | AllPro FI | Suction | No | No |
| TF02** | 79 | Male | 126.1 | 1.75 | Right | Infection | 1 | K2 | C-Leg Obk | Triton Obk | Lanyard | Yes (2) | Yes, All |
| TF05 | 72 | Male | 79.4 | 1.65 | Left | Traumatic | 4 | K2 | C-Leg Obk | Trition Low Profile Obk | Suction | No | No |
| TF06 | 60 | Male | 86.6 | 1.70 | Left | Dysvascular | 2 | K2 | C-Leg Obk | Kinterra FI | Lanyard | Yes (3) | Yes, All |
| TF07*** | 49 | Male | 102.1 | 1.91 | Left | Traumatic | 10 | K3 | C-Leg Obk | Triton Obk | Pin Lock | No | No |
| TF08 | 42 | Male | 95.3 | 1.85 | Right | Traumatic | 6 | K3 | Rheo Os | AllPro FI | Suction | No | No |
| TF09 | 65 | Male | 69.4 | 1.70 | Left | Traumatic | 2 | K2 | C-Leg Obk | Trias Obk | Suction | No | No |
| TF10 | 57 | Female | 58.5 | 1.65 | Left | Traumatic | 11 | K2 | C-Leg Obk | Trias Obk | Suction | No | Yes, All |
| TF11 | 51 | Male | 70.3 | 1.68 | Right | Traumatic | 33 | K3 | C-Leg Obk | Trias Obk | Suction | No | No |
| TF12 | 59 | Male | 99.8 | 1.83 | Left | Traumatic | 16 | K2 | C-Leg Obk | Trias Obk | Lanyard | Yes (1) | No |
| TF13 | 61 | Male | 88.5 | 1.88 | Left | Traumatic | 3 | K3 | Rheo Os | Proflex XC Os | Vacuum | No | Yes, LS |
| TF14 | 51 | Male | 108.9 | 1.73 | Right | Traumatic | 3 | K2 | Genium X3 Obk | Triton Obk | Lanyard | No | Yes, All |
| TF15 | 23 | Female | 68.0 | 1.75 | Right | Traumatic | 5 | K3 | Plie FI | Proflex XC Os | Suction | No | Yes, LS |
| TF16 | 36 | Male | 100.2 | 1.80 | Left | Traumatic | 8 | K3 | C-Leg Obk | AllPro FI | Suction | No | No |
| TF17 | 38 | Male | 104.3 | 1.91 | Left | Traumatic | 33 | K3 | Plie FI | Soleus ClgPk | Suction | No | No |
| TF18 | 69 | Male | 129.3 | 1.73 | Right | Traumatic | 50 | K2 | 3R46 Obk | Renegade FI | Suction | No | Yes |
| TF19 | 30 | Female | 59.0 | 1.60 | Left | Traumatic | 10 | K3 | 3R80 Obk | AllPro FI | Lanyard | No | No |
| TF20 | 59 | Male | 120.2 | 1.78 | Left | Traumatic | 42 | K2 | C-Leg Obk | Action Obk | Suction | No | Yes |
Obk – Ottobock, FI – Freedom Innovations LLC, Os – Ossur, ClgPk – College Park Industries, LS – subject only used handrails on the last speed.
*Subject only has 4 walking trials for speeds 0.6, 0.8, 1.0 m/s
**Subject was unable to walk at the last speed of 0.8 m/s.
***Subject only has 4 walking trials for speed 0.6 m/s.
Fig. 1Locations and names of the markers in the modified Plug In Gait model. Markers used in both the original and modified Plug In Gait model are shown in blue. Markers used only in the modified Plug In Gait model are shown in orange for tracking and green for static calibration.
Fig. 2Matlab structure in which the kinetic and kinematic data is saved.
Detailed Marker Information: Description, Type, Placement.
| Name | Description | Marker Type | Placement |
|---|---|---|---|
| LFHD | Left forehead | Tracking | Left temple |
| RFHD | Right Forehead | Tracking | Right temple |
| LBHD | Left Back Head | Tracking | Left posterior of the head |
| RBHD | Right Back Head | Tracking | Right posterior of the head |
| LEAR | Left Ear | Tracking | Left ear lobe |
| REAR | Right Ear | Tracking | Right ear lobe |
| C7 | C7 | Tracking | 7th cervical vertebra |
| T10 | T10 | Tracking | 10th thoracic vertebra |
| CLAV | Clavicle | Tracking | Jugular notch |
| STRN | Sternum | Tracking | Xiphoid process |
| RBAK | Right Back | Tracking | Anywhere over right scapula |
| LSHO | Left Shoulder | Tracking | Acromio-clavicular joint |
| LUPA* | Left Upper Arm | Tracking | Upper lateral 1/3 surface of left upper arm |
| LELB | Left Elbow | Tracking | Left Lateral Epicondyle |
| LELBM | Left Elbow Medial | Static Calibration | Left medial epicondyle |
| LFRM* | Left Forearm | Tracking | Lower Lateral 1/3 surface of left forearm |
| LWRA | Left Wrist A | Tracking | Lateral of the left wrist on thumb side |
| LWRB | Left Wrist B | Tracking | Medial of the left wrist on 5th digit side |
| LFIN | Left Finger | Tracking | Proximal knuckle of 3rd digit of left hand |
| L5HD | Left 5th digit | Tracking | Proximal knuckle of 5th digit (MCP joint) of left hand |
| RSHO | Right Shoulder | Tracking | Acromio-clavicular joint |
| RUPA* | Right Upper Arm | Tracking | Upper lateral 1/3 surface of right upper arm |
| RELB | Right Elbow | Tracking | Right Lateral Epicondyle |
| RELBM | Right Elbow Medial | Static Calibration | Right medial epicondyle |
| RFRM* | Right Forearm | Tracking | Lower Lateral 1/3 surface of right forearm |
| RWRA | Right Wrist A | Tracking | Lateral of the right wrist on thumb side |
| RWRB | Right Wrist B | Tracking | Medial of the right wrist on 5th digit side |
| RFIN | Right Finger | Tracking | Proximal knuckle of 3rd digit of right hand |
| R5HD | Right 5th digit | Tracking | Proximal knuckle of 5th digit (MCP joint) of right hand |
| LASI | Left ASIS | Tracking | Left anterior superior iliac spine |
| RASI | Right ASIS | Tracking | Right anterior superior iliac spine |
| LPSI | Left PSIS | Tracking | Left posterior superior iliac spine |
| RPSI | Right PSIS | Tracking | Right posterior superior iliac spine |
| LILC | Left Iliac Crest | Tracking | Superior surface of the left iliac crest |
| RILC | Right Iliac Crest | Tracking | Superior surface of the right iliac crest |
| LGTR | Left Greater Trochanter | Tracking | Left greater trochanter |
| LTHI | Left Thigh | Tracking | Lateral surface of left thigh |
| LTHIA | Left Thigh Anterior | Tracking | On rigid cluster, anterior to LTHI |
| LTHII | Left Thigh Inferior | Tracking | On rigid cluster, inferior to LTHI |
| LTHIP | Left Thigh Posterior | Tracking | On rigid cluster, posterior to LTHI |
| LKNE | Left Knee | Tracking | Flexion-extension axis of left knee |
| LKNEM | Left Knee Medial | Static Calibration | Medial left knee joint |
| LTIB | Left Tibia | Tracking | Lateral surface of the left shank |
| LTIBA | Left Tibia Anterior | Tracking | On rigid cluster, anterior to LTIB |
| LTIBI | Left Tibia Inferior | Tracking | On rigid cluster, inferior to LTIB |
| LTIBP | Left Tibia Posterior | Tracking | On rigid cluster, posterior to LTIB |
| LANK | Left Ankle | Tracking | Lateral malleolus of left foot |
| LANKM | Left Ankle Medial | Static Calibration | Medial lateral malleolus of the left foot |
| LHEE | Left Heel | Tracking | Calcaneus at same height as LTOE |
| LTOE | Left Toe | Tracking | Second metatarsal head |
| L5FT | Left 5th digit toe | Tracking | Posterior calcaneus at level of LTOE |
| RGTR | Right Greater Trochanter | Tracking | Right greater trochanter |
| RTHI | Right Thigh | Tracking | Lateral surface of right thigh |
| RTHIA | Right Thigh Anterior | Tracking | On rigid cluster, anterior to RTHI |
| RTHII | Right Thigh Inferior | Tracking | On rigid cluster, inferior to RTHI |
| RTHIP | Right Thigh Posterior | Tracking | On rigid cluster, posterior to RTHI |
| RKNE | Right Knee | Tracking | Flexion-extension axis of right knee |
| RKNEM | Right Knee Medial | Static Calibration | Medial right knee joint |
| RTIB | Right Tibia | Tracking | Lateral surface of the right shank |
| RTIBA | Right Tibia Anterior | Tracking | On rigid cluster, anterior to RTIB |
| RTIBI | Right Tibia Inferior | Tracking | On rigid cluster, inferior to RTIB |
| RTIBP | Right Tibia Posterior | Tracking | On rigid cluster, posterior to RTIB |
| RANK | Right Ankle | Tracking | Lateral malleolus of right foot |
| RANKM | Right Ankle Medial | Static Calibration | Medial lateral malleolus of right foot |
| RHEE | Right Heel | Tracking | Calcaneus at same height as RTOE |
| RTOE | Right Toe | Tracking | Second metatarsal head |
| R5FT | Right 5th digit Toe | Tracking | Posterior calcaneus at level of RTOE |
*Markers are intentionally placed offset in height between the left and the right side.
| Measurement(s) | Biomechanics • Kinetics • kinematics • walking • Above-Knee Amputation • transfemoral amputation |
| Technology Type(s) | motion capture system • treadmill |
| Factor Type(s) | Medicare Functional Classification Level • walking speed |
| Sample Characteristic - Organism | Homo sapiens |