The Effect of Various β Values on Image Quality and Semiquantitative Measurements in 68Ga-RM2 and 68Ga-PSMA-11 PET/MRI Images Reconstructed With a Block Sequential Regularized Expectation Maximization Algorithm.

Lucia Baratto1, Heying Duan1, Valentina Ferri1, Mehdi Khalighi2, Andrei Iagaru1.   

Abstract

PURPOSE: To compare the block sequential regularized expectation maximization (BSREM) algorithm with the ordered subsets expectation maximization (OSEM) algorithm and to evaluate how different penalty factors (b values) influence image quality and SUV measurements.
METHODS: We analyzed data from 78 prostate cancer patients who underwent Ga-RM2 (n = 42) or Ga-prostate-specific membrane antigen (PSMA)-11 (n = 36) PET/MRI. The raw PET data were retrospectively reconstructed using both time-of-flight (TOF)-BSREM with b values of 250, 350, 500, 750, and 1000 and TOF-OSEM. Each reconstruction was reviewed independently by 3 nuclear medicine physicians and scored qualitatively using a Likert scale (1 = poor, 5 = excellent quality). SUV measurements were analyzed as well.
RESULTS: Fifty-seven lesions were detected (21 on Ga-RM2 and 36 on Ga-PSMA-11 PET/MRI); SUVmax decreased with the increase of β values for both tracers. Background noise (SUVsd) decreased with increasing of β values for both tracers. The mean ± SD scores for Ga-RM2 PET images were 2.4 ± 0.5 for b = 250 reconstructions, 3.2 ± 0.6 for b = 350, 4 ± 0.6 for b = 500, 4.5 ± 0.5 for b = 750, 4.4 ± 0.7 for b = 1000, and 3.4 ± 0.6 for TOF-OSEM. The mean ± SD scores for Ga-PSMA-11 PET images were 3.2 ± 0.8 for b = 250 reconstructions, 4.1 ± 0.8 for b = 350, 4.7 ± 0.6 for b = 500, 4.8 ± 0.4 for b = 750, 4.7 ± 0.6 for b = 1000, and 3.8 ± 0.5 for TOF-OSEM.
CONCLUSIONS: Time-of-flight-BSREM algorithm improves image quality. Different b values should be used for different Ga-labeled radiopharmaceuticals such as those targeting GRPR and PSMA receptors. Once selected, the same b value should be consistently used because SUVmax measurements differ with different b values.

Entities:  

Mesh:

Substances:

Year:  2020        PMID: 32433170     DOI: 10.1097/RLU.0000000000003075

Source DB:  PubMed          Journal:  Clin Nucl Med        ISSN: 0363-9762            Impact factor:   7.794


  6 in total

1.  High quality imaging and dosimetry for yttrium-90 (90Y) liver radioembolization using a SiPM-based PET/CT scanner.

Authors:  Heying Duan; Mohamed H Khalaf; Valentina Ferri; Lucia Baratto; Shyam M Srinivas; Daniel Y Sze; Andrei Iagaru
Journal:  Eur J Nucl Med Mol Imaging       Date:  2021-01-14       Impact factor: 9.236

2.  Impact of total variation regularized expectation maximization reconstruction on the image quality of 68Ga-PSMA PET: a phantom and patient study.

Authors:  Feng-Jiao Yang; Shu-Yue Ai; Runze Wu; Yang Lv; Hui-Fang Xie; Yun Dong; Qing-Le Meng; Feng Wang
Journal:  Br J Radiol       Date:  2021-02-16       Impact factor: 3.039

3.  Improved image reconstruction of 89Zr-immunoPET studies using a Bayesian penalized likelihood reconstruction algorithm.

Authors:  Julian Kirchner; Joseph A O'Donoghue; Anton S Becker; Gary A Ulaner
Journal:  EJNMMI Phys       Date:  2021-01-19

4.  The effect of Q.Clear reconstruction on quantification and spatial resolution of 18F-FDG PET in simultaneous PET/MR.

Authors:  Defeng Tian; Hongwei Yang; Yan Li; Bixiao Cui; Jie Lu
Journal:  EJNMMI Phys       Date:  2022-01-10

5.  Results of a Prospective Trial to Compare 68Ga-DOTA-TATE with SiPM-Based PET/CT vs. Conventional PET/CT in Patients with Neuroendocrine Tumors.

Authors:  Lucia Baratto; Akira Toriihara; Negin Hatami; Carina M Aparici; Guido Davidzon; Craig S Levin; Andrei Iagaru
Journal:  Diagnostics (Basel)       Date:  2021-05-30

6.  New PET technologies - embracing progress and pushing the limits.

Authors:  Nicolas Aide; Charline Lasnon; Adam Kesner; Craig S Levin; Irene Buvat; Andrei Iagaru; Ken Hermann; Ramsey D Badawi; Simon R Cherry; Kevin M Bradley; Daniel R McGowan
Journal:  Eur J Nucl Med Mol Imaging       Date:  2021-06-03       Impact factor: 9.236

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.