Pulsatile insulin from pancreatic islets is crucial for glucose homeostasis, but the mechanism behind coordinated pulsatility is still under investigation. One hypothesis suggests that cholinergic stimulation of islets by pancreatic ganglia resets these endocrine units, producing synchronization. Previously, it was shown that intracellular Ca2+ oscillations within islets can be entrained by pulses of a cholinergic agonist, carbachol (CCh). Although these proxy measurements of Ca2+ provided insight into the synchronization mechanism, measurement of insulin output would be more direct evidence. To this end, a fluorescence anisotropy competitive immunoassay for online insulin detection from single and grouped islets in a microfluidic system was developed using a piezoelectric pressure-driven fluid delivery system and a squaraine rotaxane fluorophore, SeTau-647, as the fluorescent label for insulin. Due to SeTau-647 having a longer lifetime and higher brightness compared to the previously used Cy5 fluorophore, a 45% increase in the anisotropy range was observed with enhanced signal-to-noise ratio (S/N) of the measurements. This new system was tested by measuring glucose-stimulated insulin secretion from single and groups of murine and human islets. Distinct islet entrainment of groups of murine islets by pulses of CCh was also observed, providing further evidence for the hypothesis that pulsatile output from the ganglia can synchronize islet behavior. We expect that this relatively straightforward, homogeneous assay can be widely used for examining not only insulin secretion but other secreted factors from different tissues.
Pulsatile insulin from pancreatic islets is crucial for glucose homeostasis, but the mechanism behind coordinated pulsatility is still under investigation. One hypothesis suggests that cholinergic stimulation of islets by pancreatic ganglia resets these endocrine units, producing synchronization. Previously, it was shown that intracellular Ca2+ oscillations within islets can be entrained by pulses of a cholinergic agonist, carbachol (CCh). Although these proxy measurements of Ca2+ provided insight into the synchronization mechanism, measurement of insulin output would be more direct evidence. To this end, a fluorescence anisotropy competitive immunoassay for online insulin detection from single and grouped islets in a microfluidic system was developed using a piezoelectric pressure-driven fluid delivery system and a squarainerotaxane fluorophore, SeTau-647, as the fluorescent label for insulin. Due to SeTau-647 having a longer lifetime and higher brightness compared to the previously used Cy5 fluorophore, a 45% increase in the anisotropy range was observed with enhanced signal-to-noise ratio (S/N) of the measurements. This new system was tested by measuring glucose-stimulated insulin secretion from single and groups of murine and human islets. Distinct islet entrainment of groups of murine islets by pulses of CCh was also observed, providing further evidence for the hypothesis that pulsatile output from the ganglia can synchronize islet behavior. We expect that this relatively straightforward, homogeneous assay can be widely used for examining not only insulin secretion but other secreted factors from different tissues.
Recent years have brought new
insights on the microstructure and innervation of the pancreas and
its endocrine fraction, the islets of Langerhans.[1−3] Insulin release
from single islets is pulsatile with a period of 3–7 min,[4−6] and similar periods have been reported in vivo when insulin levels
are measured in the portal vein.[7,8] For an oscillatory insulin
profile to be observed in vivo, the release of insulin from a large
majority of islets within the pancreas must be synchronized, yet how
this synchronization occurs is still unclear. It is clear, however,
that these in vivo oscillations are important, as disordered oscillations
are observed in individuals prior to, and during, Type 2 diabetes,[9] and these dynamic insulin profiles act more efficiently
than static levels on regulating glucose.[10,11]One potential mechanism of how islets may be synchronized
is by
continuous resetting of islet oscillations due to repeated pulses
of acetylcholine (ACh) from parasympathetic ganglia, clusters of autonomic
nerve cell bodies that innervate the pancreas. ACh potentiates glucose-stimulated
insulin secretion (GSIS) by activating M3-muscarinic receptors
(M3R) in islets. This activation leads to mobilization of Ca2+ from inositol-1,4,5-trisphosphate (IP3)-sensitive Ca2+ stores in the endoplasmic reticulum and the acceleration
of protein kinase C-mediated insulin exocytosis.[12,13]This hypothesis to explain islet synchronicity has been supported
by the observation of spontaneous bursts of electrical activity in
ganglia of a feline pancreas every 6–8 min,[14] correlating with the insulin oscillation periods found
in plasma. In addition, insulin oscillations at a similar period have
been observed during ex vivo perfusion of the pancreas,[7,15−17] implying the synchronizing agent must be local to
the pancreas. In vitro studies have also shown that a single pulse
of ACh or the M3R agonist carbachol (CCh) can promote transient synchronization
in groups of murine islets in a glucose-rich environment.[18,19] Building on these findings, we found that groups of murine islets
exposed to periodic or aperiodic pulses of CCh can be synchronized.[20] However, the only evidence reported has been
the synchronization of intracellular Ca2+ oscillations
as a marker of insulin release with no report demonstrating synchronized,
oscillatory hormone release.To shed light on how M3R activation
may synchronize insulin secretion,
the development and use of high-sensitivity and high temporal resolution
analytical methods are essential. The most established methods for
insulin measurement are heterogeneous antibody-based assays, chief
among which are enzyme-linked immunosorbent assays (ELISAs). These
heterogeneous assays have high sensitivity, yet typically require
multiple wash steps, forcing the assays to be offline due to their
long processing times. Other types of immunoassays have been developed
to limit these drawbacks, for example, by incorporating electrophoretic
immunoassays into microfluidic systems.[5,6,21−23] While this type of analysis has
many advantages over ELISA, including the ability to perform online
measurements, there are several drawbacks to the method that limit
its incorporation into other laboratories. Perhaps the biggest obstacle
is working with shallow microfluidic channels with depths less than
10 μm. These small channel dimensions are required to reduce
the electric current and subsequent Joule heating to acceptable levels,
but also lead to more arduous channel fabrication processes and higher
risks of clogging.In contrast, homogeneous immunoassays do
not require a wash or
separation step, easing the requirements for the channel dimensions.
Several homogeneous-based approaches have been described,[24−31] with only a few being utilized for online measurement of insulin
release in microfluidic systems.[28−30] Fluorescence anisotropy
(or fluorescence polarization) immunoassays have been used for detection
of antibodies or antigens in biological samples for decades.[32−34] Fluorescence anisotropy is a measure of the depolarization of fluorescence
emission that occurs after a fluorophore has been excited by linearly
polarized light.[34] Immunoassays using this
detection system are typically competitive, whereby the target and
a tagged analogue of the target compete for binding sites to a limited
amount of antibody. In the case of insulin measurements, insulin (Ins)
released from islets competes with fluorescently tagged insulin (Ins*)
for a limiting amount of anti-insulin antibody (AbIns).
Quantification necessitates measuring the amount of antibody-bound
(B) and free (F) tagged species (Ins*), with the B/F ratio inversely
related to the concentration of target (Ins) in the mixture. This
ratio is determined using fluorescence anisotropy because each of
these species has a unique anisotropy as seen by the Perrin equation:where r0 is the
fundamental anisotropy of the fluorophore, τ is its fluorescence
lifetime, and θ is its rotational correlation time, defined
aswhere η is viscosity, V̅ is the molar volume, R is the
universal gas constant,
and T is temperature. In the case of the competitive
insulin immunoassay, the increased V̅ of the
bound Ins*–AbIns complex compared to free Ins* leads
to a higher anisotropy for the bound species (rB) than for the free Ins* (rF).
The immunoassay can be performed homogeneously because the total anisotropy
(⟨r⟩) of the mixture is a sum of rB and rF weighted
by their fractional amounts:where fB and fF are the fractional amount of B and F Ins*,
respectively. Every insulin concentration results in a unique combination
of fB and fF, and therefore, a distinct ⟨r⟩. The
competitive assay is ideal for this measurement technique due to the
large differences in V̅ between the B and F
Ins*. The ⟨r⟩ is determined experimentally
by exciting the solution with linearly polarized light and measuring
the degree of depolarization in the emission normalized to the total
fluorescence emission:with I∥ and I⊥ representing the fluorescence
emission intensities parallel and perpendicular, respectively, to
the polarization of the excitation light. Equation assumes equal detection sensitivity for the
two polarization states.We have recently applied this methodology
to monitor GSIS from
single and multiple islets of Langerhans using Cy5-labeled insulin
as Ins*.[28,29,35] In this report,
we describe the improvement of this previous assay using a brighter
fluorophore with a longer lifetime than Cy5. To demonstrate this new
system, similar GSIS dynamics were observed from murine and human
islets to those found using other analytical techniques. We then used
the method to provide the first reported evidence of coordinated insulin
pulses from batches of islets in response to periodic stimulation
by CCh. These applications not only show the benefit of this microfluidic-based
approach for measuring insulin release from either single or groups
of islets but also provide further evidence that cholinergic release
from parasympathetic nerves may help to coordinate islet behavior
resulting in periodic pulses of insulin.
Experimental Section
Chemicals
and Reagents
All reagents for microfluidic
assays and isolation and culture of islets were obtained from Sigma-Aldrich
(St. Louis, MO, U.S.A.) unless otherwise stated. NaOH, KCl, Tween-20,
KCl, and HF were from EMD Chemicals (San Diego, CA, U.S.A.). Glucose
(dextrose), RPMI 1640, and gentamicin sulfate were from Thermo Fisher
Scientific (Waltham, MA, U.S.A.). Collagenase P (from Clostrdium histolyticum) was acquired from Roche
Diagnostics (Indianapolis, IN, U.S.A.). Cosmic Calf Serum was acquired
from GE Healthcare Bio-Sciences (Pittsburgh, PA, U.S.A.). Monoclonal
AbIns was purchased from Meridian Life Science, Inc. (Saco,
ME, U.S.A.). All solutions were made with Milli-Q (Millipore, Bedford,
MA, U.S.A.) 18 MΩ·cm ultrapure water and filtered using
0.2 μm nylon syringe filters (Pall Corporation, Port Washington,
NY, U.S.A.). Immunoassay reagents (Ins* and AbIns) were
prepared in TEAT-40 composed of 25 mM Tricine, 40 mM NaCl, 1 mM EDTA
at pH 7.4 with an additional 0.1% Tween-20 (w/v) and 1 mg mL–1 BSA. A balanced salt solution (BSS) made to pH 7.4 was used for
preparing Ins standards and for islet perfusion. BSS was composed
of 125 mM NaCl, 2.4 mM CaCl2, 1.2 mM MgCl2,
5.9 mM KCl, 25 mM HEPES, 1 mg mL–1 BSA, and 3, 11,
or 12 mM glucose. Reagents for isolation and culture of murine islets
were prepared as previously described.[36]
Insulin Labeling and Purification of SeTau-647-Labeled Insulin
Labeling Ins with SeTau-647-NHS (SETA BioMedicals, Urbana, IL,
U.S.A.) and purifying Ins* involved chromatographic purification followed
by confirmation of antibody binding by capillary electrophoresis (CE),
adapted from what has been described previously for Cy5-labeled Ins.[37] Briefly, 1 mg of bovineinsulin in 0.1 M NaHCO3 (pH 9.3) was added to a vial containing 1 mg of SeTau-647.
The mixture was incubated in the dark at room temperature for 30 min
with gentle stirring every 5 min. The mixture was then separated using
a PD-10 desalting column (GE Healthcare Bio-Sciences). The first visibly
eluting band was collected and purified.Further purification
of this band was performed by liquid chromatography (LC) using a Beckman
127S solvent module pump and a Beckman 166 UV detector (Beckman Coulter,
Brea, CA, U.S.A.). The column was a 25 cm × 4.6 mm i.d. Symmetry300
C4 (Waters Corp., Milford, MA, U.S.A.) with a particle diameter of
5 μm. Separation was performed using 70% water containing 0.1%
trifluoracetic acid (TFA)/30% acetonitrile with 0.1% TFA with UV–vis
detection at 280 nm. Injections of 200 μL were made onto the
column, and all detected peaks were collected.The same peaks
from multiple LC runs were pooled and concentrated
to dryness (Savant Speedvac, Thermo Fisher Scientific). Each pooled
fraction was checked for binding to AbIns by adding enough
AbIns in TEAT-40 to produce a 24 nM final concentration.
The resulting mixture was separated by CE (Beckman Coulter) with a
25 μm i.d. × 60 cm fused-silica capillary (Polymicro Technologies,
Phoenix, AZ, U.S.A.) using a 20 kV separation potential. Detection
was accomplished using laser-induced fluorescence with a 635 nm laser
(AixiZ, Houston, TX, U.S.A.) and a 663 nm long-pass filter. Separation
buffer contained 150 mM Tricine and 20 mM NaCl made to pH 7.4. After
determining which fraction bound to Ab, the Ins* was quantified by
UV–vis spectroscopy using the molar extinction coefficient
of SeTau-647 (ϵ = 200 000 M–1 cm–1 at 649 nm). Labeled insulin was aliquoted and stored
in the dark at −80 °C. Each aliquot contained 2.5 μL
of 20 μM Ins* which was then diluted to 45 nM in TEAT-40 for
microfluidic experiments.
Microfluidic Device and System
The
microfluidic device
was fabricated in borosilicate glass using photolithography and wet
etching techniques as previously detailed.[37,38] Microfluidic channels were 90 μm × 195 μm (depth
× width at middle) as measured by an SJ-410 surface profiler
(Mitutoyo Corp., Aurora, IL, U.S.A.). Fluidic inputs and the 120 nL
islet chamber were drilled using 0.02 in. and 0.012 in. diamond-tipped
drill bits, respectively (Wolfco Inc., Bozrah, CT, U.S.A.). The device
had four fluidic inlets: the two upstream of the islet chamber delivered
BSS with different concentrations of secretagogues, while the two
inlets downstream of the islet chamber delivered 45 nM Ins* and 90
nM AbIns, both in TEAT-40. Each inlet was connected to
a fluid reservoir pressurized by a piezoelectric flow controller (Elvesys,
Paris, France) with their flow rates regulated by in-line flow sensors
(Elvesys). The error associated with these sensors was specified at
2% RSD by the manufacturer and was periodically verified. The final
Ins* and AbIns concentrations were determined by the ratio
of flow rates from each of these inlets to the total flow rate. Throughout
the remainder of the text, the concentrations of the immunoassay reagent
are given which correspond to the fully mixed species.
Temperature
Control
To ensure physiological temperature
within the islet chamber, the entire microfluidic device was secured
in the center of a custom-built copper plate with 2 mm wide slits
cut in the shape of the channels to allow for optical detection within
the device.[28] On each corner of the plate,
a thermoelectric (Peltier) heater (TEC1-12706, Hebeit I.T., Shanghai,
China) was attached using a thermal adhesive paste (Arctic Silver
Inc., Visalia, CA, U.S.A.). On the opposite side of each Peltier was
attached a 40 mm × 40 mm × 10 mm (L × W × D) aluminum
heat sink. Each Peltier was powered by an 18 V, 3 A power supply (Extech
Instruments, Nashua, NY, U.S.A.). For temperature measurement, a calibrated
T-240C microthermocouple (Physitemp Instrument, Inc. Clifton, NJ,
U.S.A.) connected to a TAC80B-T thermocouple-to-analog converter (Omega
Engineering, Inc., Stamford, CT, U.S.A.) was sandwiched between the
copper plate and microfluidic device. The temperature was input into
a LabView program (National Instruments, Austin, TX, U.S.A.) written
in house via an NI-PCIe 6321 data acquisition (DAQ) card (National
Instruments). The program, through the DAQ card and a Crydom CMX60D10
relay, interrupted the power supply to the Peltier circuit to maintain
the desired temperature. Islet chamber temperature throughout a typical
60 min experiment was measured to be 37.0 ± 0.3 °C.
Optical
Detection System for the Microfluidic Device
A 635 nm laser
(Coherent Inc. Santa Clara, CA, U.S.A.) was used as
the excitation source. The laser power was reduced from 25 to 7.8
mW with a neutral density filter (Thorlabs Inc., Newton, NJ, U.S.A.).
An achromatic fiberport collimator (PAF2S-7A, Thorlabs) coupled the
attenuated beam into a multimode fiber-optic bundle (Ceramoptec, Sunnyvale,
CA, U.S.A.) which then fed into a telescoping lens tube (SM1NR1, Thorlabs).
Contained in the lens tube was an achromatic doublet (AC254-080-A-ML,
Thorlabs) to collimate the excitation beam followed by a quartz-wedge
achromatic depolarizer (DPU-25-A, Thorlabs) to randomize the optical
polarization of the beam. The beam entered the back of an Eclipse
TS-100 microscope (Nikon Instruments Inc., Melville, NY, U.S.A.),
was polarized in the desired orientation using a linear polarizer
(WP25M-VIS, Thorlabs), and reflected by a dichroic mirror (XF2035,
Omega Optical Inc., Brattleboro, VT, U.S.A.). The reflected light
was focused by a 40×, 0.6 NA objective (Nikon) into the microfluidic
channel at the detection point. Fluorescence emission was collected
with the same objective, transmitted through the dichroic, and directed
into a two-channel microscope photometer (Horiba Scientific, Piscataway,
NJ, U.S.A.). Within the photometer, the light passed through a spatial
filter, a 665 nm long-pass filter (HQ665LP, Chroma Technology Corp.,
Bellows Falls, VT, U.S.A.), and a 635 nm notch filter (ZET635NF, Chroma
Technology Corp.). The emission beam was split by a polarizing beam
splitter cube (PBS101, Thorlabs) into its parallel and perpendicular
polarized components (with respect to the excitation polarization).
Each polarized component passed through a complementary linear polarizer
prior to impinging on separate photomultiplier tubes (PMTs) (R10699,
Hamamatsu Photonics, Middlesex, NJ, U.S.A.). Because most of the fluorescence
emission collected by the objective was parallel to the polarization
axis of the excitation beam, the gain of the PMT for detection of
the perpendicular component was set to 0.1 μA V–1, whereas the gain for the PMT used for detection of the parallel
component was 1.0 μA V–1. Data from both PMTs
were collected at 1000 Hz with a DAQ card (National Instruments, USB
6009) using a LabView program written in house.
Procurement
and Culture of Islets
The islet isolation
protocol was approved by the Florida State University Animal Care
and Use Committee (Protocol 1813) in a similar manner as previously
reported.[35−38] Islets from two mice were pooled and incubated in RPMI 1640 containing
11 mM glucose, l-glutamine, 10% Cosmic Calf serum, 100 U
mL–1 penicillin, 100 μg mL–1 streptomycin, and 10 μg mL–1 gentamycin
at 37 °C and 5% CO2. Islets were used no more than
4 days after isolation. Prior to each experiment, islets were rinsed
in BSS containing 3 mM glucose for at least 5 min before loading into
the islet chamber by sedimentation with a pipette. After loading,
the islet chamber was sealed with a piece of adhesive PCR film.Human islets, procured from Prodo Laboratories (Aliso Viejo, CA,
U.S.A.), were obtained from a deidentified cadaveric organ donor and,
therefore, were exempt from Institutional Review Board approval. The
donor was a 40 year old Hispanic female, 67 in., 203 lbs., 31.8 BMI,
4.8% HbA1c, and without a history of diabetes. Culture of human islets
was performed at 37 °C and 5% CO2 in PIM(S) islet-specific
media (Prodo Laboratories).
Data Analysis
Data collected from
both PMTs during
experiments were converted to anisotropy using eq . Anisotropy traces were smoothed by a 1000-point
moving boxcar average. To quantify insulin, online calibration curves
were performed using 24 nM Ins* and AbIns and 0–400
nM insulin for experiments with multiple islets. For single-islet
experiments, the insulin concentrations were halved and immunoassay
reagent concentrations were reduced to 18 nM. Calibration curves were
generated after islet experiments and were performed at 37 °C.
Each point on a curve was the average anisotropy after a 1 min collection
with error bars representing ±1 standard deviation (SD). Calibration
points were fitted to a four-parameter logistic curve. The equation
from the curve was used to convert anisotropy to an insulin concentration
that was subsequently normalized by the flow rate through the islet
chamber. Limit of detection (LOD) was taken as the concentration of
insulin required to decrease anisotropy to a value lower than 3 times
the SD of the blank. For single and grouped islet experiments, LOD
was 10 and 20 nM, respectively, due to the different immunoassay reagent
concentrations used for the experiments. All error bars are ±1
SD unless indicated otherwise. The amount of insulin released from
islets was quantified by measuring the area under the curve for specified
durations using Origin 9.0 (OriginLab, Northampton, MA, U.S.A.).
Results and Discussion
Fluorescence anisotropy is an all-optical,
quantitative method
for analysis of the degree of rotational depolarization of a fluorophore.
For decades, this technique (or a similar technique, fluorescence
polarization) has been used for quantitative measurements of homogeneous
immunoassays, typically in plate reader-based systems.[32−34] Using this method in a flow-based system allows dynamic concentration
changes to be identified without the need for separation of the B
and F Ins*. We have previously developed a fluorescence anisotropy
immunoassay for quantification of insulin release from islets using
Cy5–insulin as the labeled Ins* and gravity as the means to
drive fluid flow in the device.[28,39] In this report, we
describe the use of SeTau-647 as the fluorescent label for Ins*, which
provided a larger dynamic range to the insulin assay compared to Cy5,
and an actively controlled perfusion system that improved the robustness
of the device and resulted in an increased temporal resolution of
the assay.
Microfluidic Device and Perfusion System Characterization
In the previous iteration of the system,[28] gravity-driven flow was used to deliver reagents to the microfluidic
chip. Although there are several advantages to this passive pressure-driven
mechanism of fluid delivery, one major drawback is its inability to
automatically correct flow rate fluctuations when bubbles or clogs
develop in the fluidic lines or microfluidic device. To improve the
robustness of flow, a variable-pressure system was implemented. Presented
in Figure S1 is a schematic of the fluidic
control system. Fluid reservoirs containing assay reagents were pressurized
with filtered air by a piezoelectric pressure regulator. As liquid
flowed from the reservoirs into the microfluidic device, flow rates
were continuously monitored by an in-line flow sensor connected in
a feedback loop to the piezoelectric regulator which adjusted the
pressure output to maintain flow stability. In Figure A, the positions of fluidic ports, islet
chamber (green dot), and detection point (blue dot) on the chip are
annotated, together with the direction of flow. For all experiments,
the perfusion flow rate through the islet chamber was maintained at
0.3 μL min–1. The Ins* and AbIns flow rates were set at 0.4 and 0.8 μL min–1, respectively, for a total flow rate at the detection point of 1.5
μL min–1.
Figure 1
Microfluidic chip, flow profile, and pulse
train characterization.
(A) Perfusion solutions or insulin standards were delivered to inlets
1 and 2 for islet experiments or calibrations, respectively, and passed
through the closed islet chamber (green dot). The two remaining inlets
were used to deliver AbIns and Ins* as noted. The reagents
mixed and equilibrated in the 160 mm long mixing channel before fluorescence
measurement at the detection point (blue dot). Black arrows indicate
the direction of flow. (B) Flow of SeTau-647 was initiated from at
0 min, while signal was measured at the detection point. The x-axis break at 0.5 min is for ease in viewing. (C) The
blue trace (right axis) shows the programmed flow rates of SeTau-647
delivered from inlet 2. The black trace (left axis) is the fluorescence
signal at the islet chamber during a pulsing experiment. A 53 ±
3% pulse attenuation was observed at the islet chamber.
Microfluidic chip, flow profile, and pulse
train characterization.
(A) Perfusion solutions or insulin standards were delivered to inlets
1 and 2 for islet experiments or calibrations, respectively, and passed
through the closed islet chamber (green dot). The two remaining inlets
were used to deliver AbIns and Ins* as noted. The reagents
mixed and equilibrated in the 160 mm long mixing channel before fluorescence
measurement at the detection point (blue dot). Black arrows indicate
the direction of flow. (B) Flow of SeTau-647 was initiated from at
0 min, while signal was measured at the detection point. The x-axis break at 0.5 min is for ease in viewing. (C) The
blue trace (right axis) shows the programmed flow rates of SeTau-647
delivered from inlet 2. The black trace (left axis) is the fluorescence
signal at the islet chamber during a pulsing experiment. A 53 ±
3% pulse attenuation was observed at the islet chamber.The microfluidic system was held on the stage of a microscope,
and the optical system (Figure S2) was
used to evaluate the temporal resolution of the system. Reservoirs
connected to inlets 1 and 2 contained BSS and 20 nM SeTau-647, respectively.
BSS was delivered to the AbIns and Ins* inlets (Figure A) at 0.4 and 0.8
μL min–1, respectively, with the islet chamber
sealed. Initially, BSS flowed into the device through inlet 1 at 0.3
μL min–1, while the flow of SeTau-647 from
inlet 2 was set to 0.0 μL min–1. The flow
rates from inlets 1 and 2 were then reversed so that SeTau-647 was
delivered and fluorescence was measured at the detection point of
the device. As shown in Figure B, after a delay time (td), the
fluorescence increased until the signal plateaued. Because time 0
corresponded to the time at which SeTau-647 began to flow, td quantified the time for travel from point
1 in Figure A to the
detection point. The td was determined
as the intensity to reach 10% of the final signal at the detection
point and was 3.2 ± 0.2 min (n = 4 trials).
The response time (tr) was the time required
for the fluorescence signal to change from 10% to 90% of the final
signal at the detection point. This value was 36 ± 1 s (n = 4), a 3-fold improvement on the previous gravity-driven
perfusion system,[28] and was taken as the
temporal resolution of the device. The values provided here are for
trials within a single microfluidic device; the interdevice numbers
varied slightly, but the intradevice variation was similar.To evaluate how pulses of CCh would attenuate when delivered to
the islet chamber, sequential pulses of SeTau-647 were produced and
measured at the islet chamber. Initially, BSS was delivered through
inlet 1 at 0.3 μL min–1. The flow was then
stopped, and SeTau-647 was delivered for 30 s through inlet 2. After
the 30 s had elapsed, BSS was again delivered. This pattern was repeated
every 4.5 min until four pulses of SeTau-647 were generated followed
by a constant flow of SeTau-647. The PMT voltage (Figure C) showed the arrival of all
four pulses as well as a plateau when constant dye was perfused. The
delay time for travel between point 1 and the islet chamber was measured
to be 0.71 ± 0.01 min (n = 4) using the pulse
profile shown in Figure C. Due to travel through the channels during that time, a 53 ±
3% attenuation in signal was observed as measured by the height of
the pulses as compared to the constant dye delivery.
Assay Improvement
with Squaraine Rotaxane Fluorophore
Previously, Cy5 was used
as the fluorescent label for Ins*.[28] Since
fluorescence anisotropy is a spectroscopic
measurement, the photophysical features of the fluorescent label play
an important role in the assay. Therefore, we sought to replace Cy5
with a commercially available fluorescent label with superior optical
properties. This new fluorophore, SeTau-647, is a squarainerotaxane
fluorescent probe which has similar excitation and emission maxima
to Cy5 but a larger Stokes shift (46 nm), a larger molar extinction
coefficient (200 000 M–1 cm–1), a larger quantum yield (0.65), and a longer fluorescence lifetime
(3.2 ns) in aqueous solutions.[40,41] Encapsulation of the
squaraine chromophore by the rotaxane macrocycle provides protection
of the dye, thereby increasing chemical stability and photobleaching
resistance.[42]To further justify
switching the fluorophore, predicted immunoassay calibration curves
were made by calculating the anisotropy (⟨r⟩calc) for Cy5-labeled Ins (Ins*Cy5)
and SeTau-647-labeled Ins (Ins*SeTau-647) at different
insulin concentrations. To perform these calculations, rB and rF for Ins*Cy5 and Ins*SeTau-647 were calculated using eqs and 2 with θ values for each species assuming 0 degrees of hydration
and at 37 °C. Values for τ and r0 were obtained from the literature.[43−46]Table S1 summarizes the properties of the B and F species for both labels
used to calculate rB and rF. Once these two anisotropies were calculated, fB and fF at different
unlabeled insulin concentrations were calculated using mass action
and equilibrium equations (see the Supporting Information for details). These values were then used to determine
⟨r⟩calc using eq .Figure A shows
the calculated immunoassay calibration curves for both fluorophores.
All data points were plotted relative to the anisotropy at 0 nM insulin,
and both curves show the sigmoidal decrease in anisotropy as a function
of insulin as expected for a competitive assay. To quantify assay
improvement, the changes in anisotropies (Δ⟨r⟩calc) for both curves from 0 to 600 nM insulin
were compared. A 57% larger change was found with SeTau-647 compared
to Cy5 due to the larger τ of the former. Conceptually, the
larger value permits a higher degree of depolarization resulting in
a greater change in anisotropy between rB and rF in eq .
Figure 2
Competitive immunoassay improvement with squaraine
rotaxane fluorophore.
(A) Calculated calibration curves using Ins*Cy5 and Ins*SeTau-647 were computed using their photophysical properties and mathematically
calculated B/F ratios as explained in the text and Supporting Information. (B) Experimental calibration curves
generated with 24 nM AbIns and 24 nM of either Ins*Cy5 or Ins*SeTau-647. Δ⟨r⟩ values in panels A and B are taken as relative
to the anisotropy at 0 nM. SeTau-647 is shown to produce improved
assay range from both calculated (panel A) and experimental (panel
B) results.
Competitive immunoassay improvement with squarainerotaxane fluorophore.
(A) Calculated calibration curves using Ins*Cy5 and Ins*SeTau-647 were computed using their photophysical properties and mathematically
calculated B/F ratios as explained in the text and Supporting Information. (B) Experimental calibration curves
generated with 24 nM AbIns and 24 nM of either Ins*Cy5 or Ins*SeTau-647. Δ⟨r⟩ values in panels A and B are taken as relative
to the anisotropy at 0 nM. SeTau-647 is shown to produce improved
assay range from both calculated (panel A) and experimental (panel
B) results.Following these calculations,
a comparison of experimentally obtained
calibration curves using Ins*Cy-5 and Ins*SeTau-647 was made (Figure B). Measurements were relative to the mean anisotropy at 0 nM insulin.
A 45% increase in the range of anisotropy values was observed with
Ins*SeTau-647, similar to the calculated increase
in anisotropy values. Additionally, due to the superior quantum yield
of the squarainerotaxane compared to Cy5, the signal-to-noise ratios
(S/N) of the measurements were enhanced, resulting in a 44% decrease
in the average SD across the 6-point calibration curves. The LOD,
limited in competitive assays by the equilibrium constant of the antibody–antigen
reaction, was 20 nM for both curves since both used the same AbIns.
Initial islet experiments were performed with
murine islets to compare
insulin secretion profiles using this new system to what has been
reported previously.[5,6,22,23,37,38] For single-islet experiments, fluidic reservoirs
leading to inlets 1 and 2 contained BSS with 3 and 11 mM glucose,
respectively. After loading and sealing an islet in the microfluidic
chamber, 3 mM glucose was applied for 5 min to condition the islet
to flow. After this rinse, anisotropy measurements commenced with
3 mM glucose for 2 min, followed by 11 mM for 40 min, and then a return
to 3 mM. Figure A
shows a representative GSIS profile from a single murine islet. The
black trace, corresponding to the left y-axis, represents
the insulin profile; the blue trace, corresponding to the right y-axis, represents the glucose profile. During the initial
2 min perfusion with 3 mM glucose, 42 pg of insulin was released.
Glucose was then increased to 11 mM, and a burst in insulin release
followed. Over the initial 13 min of exposure to 11 mM glucose, 900
pg of insulin was released. After the initial burst, insulin was released
in a pulsatile manner, consistent with other reports.[5,6,22,23,36] Over the next 28 min, 900 pg of insulin
was secreted. The insulin release rate returned to basal levels when
the glucose challenge was removed. Figure S3 shows the insulin secretion profiles of three additional murine
islet GSIS experiments. Similar trends were observed with biphasic
insulin release from each islet, although phase 2 oscillations were
not as marked as that shown in Figure A.
Figure 3
Online GSIS measurement from single and grouped murine
islets.
In both experiments, the black trace (left y-axis)
is the insulin secretion rate per islet, whereas the blue profile
(right axis) is the glucose stimulus. (A) A solution of 11 mM glucose
was delivered for 40 min to a single murine islet during which biphasic,
pulsatile insulin release was observed. (B) A group of five murine
islets showed biphasic insulin secretion release during the 25 min
exposure to 12 mM glucose.
Online GSIS measurement from single and grouped murine
islets.
In both experiments, the black trace (left y-axis)
is the insulin secretion rate per islet, whereas the blue profile
(right axis) is the glucose stimulus. (A) A solution of 11 mM glucose
was delivered for 40 min to a single murine islet during which biphasic,
pulsatile insulin release was observed. (B) A group of five murine
islets showed biphasic insulin secretion release during the 25 min
exposure to 12 mM glucose.A representative example of GSIS monitoring from a group of five
murine islets in the microfluidic chamber is shown in Figure B. For this experiment, fluidic
reservoirs connected to inlets 1 and 2 contained BSS with 3 and 12
mM glucose, respectively. The islets were rinsed with 3 mM glucose
for 5 min resulting in 570 pg of insulin secreted. Glucose levels
then rose to 12 mM for a total of 25 min. During the initial 13 min
of high glucose exposure, 2900 pg of insulin was released followed
by a more decreased output at 2000 pg for the remaining 12 min. Secretion
rates returned to basal levels when glucose was lowered to 3 mM. Figure S4 shows three additional GSIS experiments.
In each experiment, four islets were loaded into the chamber and exposed
to 12 mM glucose. Biphasic insulin release was observed during the
glucose challenges. For both single and multiple islet experiments,
the insulin secretion rates per islet were comparable with previously
reported values.[5,6,22,23,25,26,28,37,38]Due to the increased significance
in understanding the biology
of human disease, we also tested the ability to measure GSIS from
human islets using this system. As a representative case (Figure S5), three similarly sized (∼150
μm) islets from a healthy female donor were placed in the chamber
and insulin secretion recorded for 2 min at 3 mM glucose. Insulin
output was 100 pg during the low glucose rinse. GSIS was initiated
by delivering 11 mM glucose for 15 min with a total of 1600 pg of
insulin measured. Insulin secretion did not readily return to baseline
levels with 3 mM glucose. This secretion profile for human islets
is consistent with what has been observed previously.[29,47−50]Thus far, we have shown the capabilities of the homogeneous,
online
fluorescence anisotropy competitive immunoassay to quantify biphasic
GSIS dynamics from both single and grouped murine and human islets.
The improved assay range and higher S/N offered by SeTau-647, as well
as the subminute temporal resolution of this microfluidic system,
allowed observations of insulin secretion dynamics that were previously
unobservable in most homogeneous assays[24−29] except the most recently reported droplet-based molecular pincer
assay.[30]
Repetitive Activation of
M3 Receptors Synchronize Murine Islets
CCh, an M3R agonist,
synchronizes glucose-induced islet activity
as measured by synchronized oscillations of [Ca2+]i levels.[18−20] Because no recording of insulin release has been
reported after delivery of CCh pulses, synchronized hormone secretion
is yet to be verified. Using the newly developed anisotropy method,
we set out to test if GSIS from grouped islets can be entrained during
periodic application of CCh.For this experiment, reagent reservoirs
connected to inlets 1 and 2 contained 11 mM glucose and 10 μM
CCh in 11 mM glucose, respectively. A group of five murine islets
were loaded into the islet chamber that had been filled with 11 mM
glucose. After sealing the chamber, islets were continuously perfused
with constant 11 mM glucose from inlet 1. CCh pulses were generated
by switching the flow from inlet 1 to inlet 2. Pulse profiles consisted
of five pulses (denoted by “×” in Figure ) each delivered for 30 s every
5 min. Considering broadening of each pulse (Figure C), islets were exposed to an effective CCh
concentration of 4.7 μM.
Figure 4
Entraining effect of periodic CCh pulses
on GSIS in murine islets.
In both panels, five murine islets were perfused with a constant 11
mM glucose concentration. (A) A train of five CCh pulses were delivered
for 30 s each, every 5 min, as denoted by an “×”.
The numbered insulin peaks correlate to those mentioned in the text.
(B) A representative control experiment is shown with flow switches
denoted by “×”. Because no CCh was present, no
large pulses of insulin were observed.
Entraining effect of periodic CCh pulses
on GSIS in murine islets.
In both panels, five murine islets were perfused with a constant 11
mM glucose concentration. (A) A train of five CCh pulses were delivered
for 30 s each, every 5 min, as denoted by an “×”.
The numbered insulin peaks correlate to those mentioned in the text.
(B) A representative control experiment is shown with flow switches
denoted by “×”. Because no CCh was present, no
large pulses of insulin were observed.As shown in Figure A, in the time prior to the first CCh pulse, only small insulin pulses
were observed, which were likely due to the random overlap of insulin
oscillations from the individual islets. Once the CCh pulses began,
however, the measured insulin secretion profile was noticeably different.
Whereas the first insulin spike after CCh delivery (labeled as peak
1 in Figure A) was
slightly dampened, all four subsequent pulses induced significantly
higher spikes of insulin. The total insulin secreted over the final
three pulses (15–30 min experimental time) was 1600 pg. These
CCh-potentiated oscillations promptly declined after delivery of the
final pulse at 25 min.To ensure that synchronization was due
only to the periodic application
of CCh and not to any feature of the microfluidic system itself, control
experiments were performed by performing a similar pulse profile,
but without CCh present. This ensured that the islets remained perfused
with only 11 mM glucose even while the pulse profile was being delivered.
As shown in Figure B, no discernible synchronization of insulin release was observed,
only minor pulses similar to that observed prior to initiation of
CCh pulses in the earlier experiment. A total of four control experiments
were performed in this way, with the remaining three shown in Figure S6. These results complement our previous
work[20] that showed potentiation and 1:1
entrainment of [Ca2+]i by 5 min periodic CCh
pulses.
Conclusions
Significant improvements
to a previously described system were
essential for allowing this online fluorescence anisotropy competitive
insulin immunoassay to quantify biphasic GSIS dynamics from single
and grouped pancreatic islets and for continued exploration of the
hypothesis of ganglia-induced islet synchronization. The overall design
of the microfluidic system produces subminute temporal resolution
for insulin measurements, rapid enough to observe the fast oscillation
dynamics of single islets. Key to improving the system was the use
of SeTau-647 with its longer fluorescence lifetime than Cy5. This
resulted in increased fluorescence emission depolarization and a greater
anisotropy shift between low and high antigen concentrations. To demonstrate
the applicability of the system, online dynamic insulin release from
murine and human islets after glucose induction was measured. Synchronized
GSIS from a group of islets entrained by periodic pulses of CCh was
observed, providing new support for the hypothesis of ganglia-induced
islet synchronization. Although our ensemble insulin secretion measurements
do not currently allow for identifying potential heterogeneous effects
among the pooled islets, future work could incorporate islet equivalent
calculations to account for interislet heterogeneity. Implementation
of an automated protocol to load islets into the device would also
help to provide a more convenient platform for higher-throughput islet
analyses. These shortcomings notwithstanding, our online homogeneous
competitive immunoassay could be applicable as a user-friendly and
affordable approach for the quantification of biologically relevant
targets other than insulin with slight changes to the immunoassay
reagents.
Authors: Giovanni Lenguito; Deborah Chaimov; Jonathan R Weitz; Rayner Rodriguez-Diaz; Siddarth A K Rawal; Alejandro Tamayo-Garcia; Alejandro Caicedo; Cherie L Stabler; Peter Buchwald; Ashutosh Agarwal Journal: Lab Chip Date: 2017-02-28 Impact factor: 6.799
Authors: Adrian M Schrell; Nikita Mukhitov; Lian Yi; Joel E Adablah; Joshua Menezes; Michael G Roper Journal: Anal Methods Date: 2016-10-28 Impact factor: 2.896
Authors: Aleksey V Matveyenko; David Liuwantara; Tatyana Gurlo; David Kirakossian; Chiara Dalla Man; Claudio Cobelli; Morris F White; Kyle D Copps; Elena Volpi; Satoshi Fujita; Peter C Butler Journal: Diabetes Date: 2012-06-11 Impact factor: 9.461