Chwee Ming Lim1,2,3, Gangodu Nagaraja Rao1, Hongliang Ren4.
Abstract
OBJECTIVE: To test the feasibility of a real time miniature endoscope system for imaging the nasopharynx. STUDYEntities:
Keywords: Imaging; In vivo miniature endoscope; Nasopharyngeal carcinoma
Year: 2020 PMID: 32426698 PMCID: PMC7221205 DOI: 10.1016/j.wjorl.2020.02.004
Source DB: PubMed Journal: World J Otorhinolaryngol Head Neck Surg ISSN: 2095-8811
Fig. 1A: Prototype 1 designed using medical grade stainless steel 316L; B: Prototype 2 designed using ObjetVeroclear™ (Strata Sys) 3D printing material; C: Prototype 3 with endoscope case, screws and backing for anchorage.15, 16, 17
Fig. 2A 3D printed material encasing the miniature camera device.
Fig. 3The quality of image captured and transmitted by the miniature camera in real time on skull model.
Fig. 4Cadaver with piercing of space near vomer bone and screw holes for fixation of screws from the 3 miniature camera prototypes.
Fig. 5The position of miniature camera on the posterior edge of the vomer bone on skull model. A: Anterior posterior view from the camera insertion site; B: Posterior view from the camera insertion site.
The qualities and characteristics of 3 prototypes.
| Testing standards | Prototype 1 | Prototype 2 | Prototype 3 |
|---|---|---|---|
| Dimensions (mm, L × W × H) | 70.0 × 5.5 × 6.0 | 90.0 × 6.2 × 1.5 | 35.0 × 7.5 × 5.0 |
| Weight (g) | 23.60 | 36.84 | 20.00–30.00 |
| Material | Titanium alloy (Ti–6Al–4V) | Medical grade stainless steel (316 L) | Titanium alloy (Ti–6Al–4V) |
| Anchorage | Vomer bone | N/A | Vomer bone |
| Angle (degree) | 180 (wide angle) | 90 (wide angle) | 180 (wide angle) |
The comparison of titanium alloy and stainless steel.
| Factors | Titanium alloy | Stainless steel |
|---|---|---|
| Type & grade | Ti6 Al4V (grade 5) | Stainless steel medical grade 316 L |
| Tensile strength (Mpa) | 950 | 485 |
| Hardness (HB) | 334 | 217 |
| Elastic modulus (Gpa) | 113.8 | 193 |
| Property | Bio-compatible, corrosive and heat resistant | Biocompatible, corrosive and heat resistant |
| Risk | Easily contaminated, poor shear strength and surface wear property | Components may induce immune reaction and also interfere with MRI due to presence of ferrite |
| Application | Medical devices and implants | Medical devices and implants |