Data on characterization of glass biochips and validation of the label-free biosensor for detection of autoantibodies in human serum.

A V Pushkarev1,2, A V Orlov1,2, S L Znoyko1, D O Novichikhin1, V A Bragina1, A A Sizikov2, E Alipour3, H Ghourchian3, A I Nikitin4, G M Sorokin5, B G Gorshkov1, P I Nikitin1.   

Abstract

The data represent in-depth characterization of a novel method for highly sensitive simultaneous measuring in human serum of both critical parameters of autoantibodies: concentration and native kinetics. The latter refers to autoantibody interaction with free, not immobilized, antigen. The method and related biosensors are based on the spectral-correlation and spectral-phase interferometry. The data cover: multi-factor optimization and quantitative characterization of the developed affordable single-used biochips, including X-ray photoelectron spectroscopy (XPS) control of chemical modifications of the surface during fabrication; antibody screening; optimization and verification of protocols for label-free biosensing in human serum; mathematical model for fitting experimental data and calculation of kinetic constants of interaction of autoantibodies with free antigen; comprehensive verification of the method specificity; correlation between the data obtained with the developed biosensor and with enzyme linked immunosorbent assay (ELISA); comparison of analytical characteristics of the developed biosensor with the most advanced label-based methods. The data importance is confirmed by a companion paper (DOI 10.1016/j.bios.2020.112187), which shows that the combination of mentioned autoantibody parameters is promising for more accurate criteria for early diagnostics and efficient therapy of autoimmune disorders. The obtained data can be used in development of a wide range of biosensors, both label-free and based on various labels.
© 2020 The Authors.

Entities:  

Keywords:  autoantibodies; detection in human serum; diagnostics of autoimmune disorders; label-free detection; multiplex sensing; native kinetics

Year:  2020        PMID: 32426427      PMCID: PMC7225373          DOI: 10.1016/j.dib.2020.105648

Source DB:  PubMed          Journal:  Data Brief        ISSN: 2352-3409


Specifications Table

Value of the Data

The data provide a thorough and comprehensive optimization and verification of all stages of a biosensor development and functioning starting from characterization of biochip surface to validation using clinical samples. The data can be useful for those who deal with measuring the kinetic characteristics of biochemical reactions in complex mediums containing non-target components, as well as for those who use chemical modifications of glass surfaces in biosensing. The data represent a solid basis for further developments of high-precision biosensing systems for early diagnostics of autoimmune diseases.

Data Description

Calculation of kinetic constants of interaction of autoantibodies on the biochip with free antigen

Kinetic constants of interaction of autoantibodies on the biochip with free antigen [1] were calculated using the commonly applied model for determination of kinetic constants [2]. For this purpose, we used the sensograms recorded by the label-free biosensors based on the spectral-correlation interferometry (SCI) [3,4] and spectral-phase interferometry [5]. The sensogram fragments, which corresponded to the stage of binding of free antigen with the autoantibody on the biochip, were fit with the equationwhere R(t) – temporal dependence of biolayer thickness, R – maximum increment of the biolayer thickness at the stage of antigen binding with autoantibody, k – observed kinetic constant of association. The temporal dependence of biolayer thickness describes a bimolecular reaction between native antigen Ag in the sample with autoantibodies Ab on the biochip surface: The antigen concentration [Ag] in the solution was maintained constant, and the kinetic constants of association k and dissociation k were calculated from the equation:

Specificity of the biosensor

Specificity to potentially interfering molecules added to serum samples

We checked the following potentially interfering molecules: i) high-molecular-weight proteins: thyroid stimulating hormone (TSH) and prostate specific antigen; ii) low-molecular-weight substances: biotin (BIO) and chloramphenicol (CAP); iii) mixture of human immunoglobulins (IgG); iv) hepatitis B surface antigen (HBsAg) as a self-assembling protein; v) DNA and RNA molecules. The effect of each individual substance was tested separately by estimation of biomolecular binding [6,7]. The substance (10 IU/mL for TSH; 10 µg/mL for every other substance) was added to the serum samples containing both analyzed autoantibodies: anti-thyroglobulin (anti-TG), concentration of 971 IU/mL and anti-thyroid peroxidase (anti-TPO), concentration of 75 IU/mL. Then we measured the concentrations of anti-TG and anti-TPO. The obtained data can be seen in Fig. 1. In this and other bar plots in this paper, Y-axis shows relative Δd, which is the ratio of measured signal Δd to the specific signal obtained at the standard (initial) conditions for neat positive serum.
Fig. 1

Signals of the developed biosensor upon addition of various potentially interfering molecules to the serum samples (one interfering substance in each experiment): IgG – mixture of human immunoglobulins, CAP – chloramphenicol, PSA - prostate specific antigen, TSH - thyroid stimulating hormone, HBsAg - hepatitis B surface antigen, DNA - deoxyribonucleic acid, RNA - ribonucleic acid.

Signals of the developed biosensor upon addition of various potentially interfering molecules to the serum samples (one interfering substance in each experiment): IgG – mixture of human immunoglobulins, CAP – chloramphenicol, PSA - prostate specific antigen, TSH - thyroid stimulating hormone, HBsAg - hepatitis B surface antigen, DNA - deoxyribonucleic acid, RNA - ribonucleic acid.

Specificity of secondary antibody binding

In these experiments, we used serum samples that contained neither anti-thyroid peroxidase nor anti-thyroglobulin autoantibodies. The biolayer thickness increased during pumping such samples along the biochip with immobilized antigens (see characteristic sensograms in Fig. 2). However, at the next stage, when we pumped anti-human antibody that specifically recognized autoantibody-antigen complexes, the biolayer was practically unchanged. The slight decrease in the biolayer thickness at that stage was due to washing out the components that non-specifically immobilized at the previous stage. The obtained data do not exhibit non-specific binding of secondary antibodies.
Fig. 2

Sensograms of measuring serum samples that contained neither anti-thyroid peroxidase nor anti-thyroglobulin autoantibodies (verification of specific binding of secondary antibodies).

Sensograms of measuring serum samples that contained neither anti-thyroid peroxidase nor anti-thyroglobulin autoantibodies (verification of specific binding of secondary antibodies). The absence of immunoglobulins among the non-specific reactants bound to the surface was verified in a modified setup. The analyzed serum was replaced with immunoglobulin fraction of serum. The immunoglobulin concentration of 10 mg/mL was close to that in human blood serum. Free thyroglobulin (20 µg/mL) was added to serum immunoglobulin to block the autoantibodies that may be present. In these experiments, no biolayer increment was observed when pumping the immunoglobulin fraction followed by passing secondary anti-human antibodies (Fig. 3a). The data show no increment in the biolayer thickness due to non-specific binding of secondary antibody with antigen on the surface and no effect of potential interferents on the efficiency of recognition of target immunoglobulins by secondary antibody (Figs. 3b and 3c, respectively).
Fig. 3

Verification of specific binding of secondary antibodies: a – binding of serum immunoglobulins with antigen on the surface and related binding of secondary antibody; b – binding of secondary antibody with antigen on the surface; с – change in the efficiency of recognizing target immunoglobulins by secondary antibody upon addition of potential interferents.

Verification of specific binding of secondary antibodies: a – binding of serum immunoglobulins with antigen on the surface and related binding of secondary antibody; b – binding of secondary antibody with antigen on the surface; с – change in the efficiency of recognizing target immunoglobulins by secondary antibody upon addition of potential interferents.

Specific binding of target antibodies with antibody-antigen complexes

In these experiments, which were implemented in the single-channel mode of the SPI biosensor [8], various non-target antibodies in concentration 50 μg/mL were pumped instead of anti-human antibody. As non-target antibodies, we tested antibodies to: i) thyroid-stimulating hormone; ii) chloramphenicol; iii) biotin; iv) hepatitis B surface antigen. The data obtained under pumping the non-specific antibodies did not exceed the noise level (Fig. 4).
Fig. 4

Signals of the developed biosensor in the experiments, in which various non-target antibodies (concentration - 50 μg/mL) were pumped at the stage of passing anti-human antibody. The antibodies tested as non-target: anti-CAP, anti-BIO, anti-TSH, anti-HBsAg.

Signals of the developed biosensor in the experiments, in which various non-target antibodies (concentration - 50 μg/mL) were pumped at the stage of passing anti-human antibody. The antibodies tested as non-target: anti-CAP, anti-BIO, anti-TSH, anti-HBsAg.

Verification of absence of interference between immobilized proteins

This experimental series was implemented in the single-channel mode of the biosensor. The serum samples to be tested for anti-TPO were divided into two groups: the first one was measured as usual, while to the other samples, thyroglobulin was added before the measurements. The serum samples to be tested for anti-TG were prepared in the same way by addition of TPO and were measured in the similar setup. The obtained data are exhibited in Fig. 5. The statistical insignificance of differences in the signals was confirmed by p-values of 0.45 and 0.18 calculated for anti-TPO and anti-TG, respectively, both exceeding 0.1.
Fig. 5

Signals of the developed biosensor while assessment of interference between immobilized proteins.

Signals of the developed biosensor while assessment of interference between immobilized proteins.

Specificity of determination of native kinetics

We compared the data obtained in three series of experiments, which differed in the solutions pumped after passing the serum samples containing anti-TPO and anti-TG autoantibodies. The solutions were as follows: 1) the same serum containing anti-TPO and anti-TG autoantibodies with spiked TG (20 µg/mL) and TPO (20 µg/mL); 2) control - the same serum samples, no additions; 3) the same serum samples with addition of potentially interfering molecules (BIO, mixture of human immunoglobulins, HBsAg, DNA and RNA - each of these in concentration of 10 µg/mL; and 10 IU/mL TSH). According to the sensograms shown in Fig. 6, the biolayer increased only in the first series of experiments, in which the serum containing anti-TPO and anti-TG autoantibodies with spiked TG and TPO was pumped at the final step.
Fig. 6

Assessment of specificity of determination of native kinetics.

Assessment of specificity of determination of native kinetics.

Insignificance of steric effect

The kinetic constants of autoantibodies were compared for the same serum in three variants: undiluted, and diluted 3-fold, and 9-fold. This way, the serum samples had different concentrations of autoantibodies. The concentration of undiluted serum was close to the upper limit of clinically relevant range – 864 IU/mL of anti-TPO. The determined kinetic constant was the same within the experimental error (Table 1). The statistical insignificance of differences in the kon values was confirmed by p-values of 0.48 and 0.61 calculated for 3-fold and 9-fold dilutions, respectively, both exceeding 0.1. The data in Table 1 indicate that the higher the dilution factor (column 1), the lower the biolayer thickness of the native non-immobilized antigen binding Δd at the stage of determination of native kinetics (column 3). Thus, the superficial density of autoantibodies bound to the surface at the previous stage should also decrease with increasing of dilution factor.
Table 1

Kinetic constants for the same serum in various dilutions.

Dilutionkon x 106, IU-1 x s-1 x mLΔd, nm
Undiluted94±170.43±0.03
3-fold79±130.31±0.02
9-fold86±150.22±0.01
Kinetic constants for the same serum in various dilutions.

Optimization of the assay

Chemical modification of glass surface of biochips

The process of chemical modification of the glass biochip surface was optimized toward maximization of the biosensor signal at the final step of the assay [9,10]. Carboxylation and epoxylation were regarded as the candidate techniques. Amination was not considered because it required activation of carboxyl groups on proteins, which might cause conjugation of carboxyl groups with amino groups of the proteins and protein agglomeration. The carboxylated and epoxylated glass cover slips were prepared. Onto the obtained slips, thyroglobulin was covalently immobilized directly in the biosensor liquid handling system with additional pre-activation of the carboxylated glass slips by N-(3-Dimethylaminopropyl)-N-ethylcarbodiimide hydrochloride. After that, the surface was blocked with 10 mg/mL of bovine serum albumin (BSA) in phosphate-buffered saline (PBS) buffer followed by pumping of the analyzed serum sample, diluted 10-fold in the same buffer. Each reagent was pumped until adsorption-desorption equilibrium determined by the sensogram plateauing. After the serum, we pumped 10 mg/mL PBS-BSA for a short time and then secondary antibodies diluted in the same buffer. The maximal increase of the biolayer on the sensor chip was measured (Fig. 7).
Fig. 7

Comparison of increments in the thickness of the secondary antibody biolayer for the epoxylated and carboxylated sensor chips.

Comparison of increments in the thickness of the secondary antibody biolayer for the epoxylated and carboxylated sensor chips.

Antibody screening

Another important optimization task was selection of secondary antibody, which was done similarly to the previously reported SCI screening of antibodies for magnetic immunoassays [11,12] interrogated by the frequency mixing technique [13,14]. In the present work, six different clones were screened. We carried out measurements with each single clone, as well as their different combinations in the total concentration of 50 µg/mL. The setup and protocols of these experiments were the same as those given in section 3.1 except changing antibodies instead of surface modification type. Fig. 8 exhibits the sensograms for the optimal variants, while the data for other variants can be found in Table 2.
Fig. 8

Comparison of the assay sensograms obtained with top six optimal antibody variants.

Table 2

Raw data on antibody screening.

Antibody cloneConcentration, µg/mLThickness, nm
Polyclone P45500.41 ± 0.03
52G1500.09 ± 0.02
3D3500.32 ± 0.02
5A9C500.48 ± 0.04
5A9C + 3D325+251.03 ± 0.08
5A9C + 3D3 + 52G116.3+16.3+16.31.17 ± 0.11
All 6 clones8.3 for each clone1.19 ± 0.07
2C11500.05 ± 0.01
5C7500.06 ± 0.01
5G12500.08 ± 0.02
Comparison of the assay sensograms obtained with top six optimal antibody variants. Raw data on antibody screening.

Durations of pumping at each step

To minimize the total assay time, we have optimized the durations of pumping of each reagent along the sensor chip surface. The assay consists of two stages: pumping the analyzed serum sample and pumping the secondary antibody. As the optimal duration, we considered the time between the start of pumping until the sensogram plateauing plus two minutes. These two minutes were added to assure the sensogram plateauing and for more precise determining the biolayer increment value. The optimal duration of serum pumping was found to be 14-18 min (the limiting factor was duration of binding of serum components with the spot where thyroglobulin was immobilized). The time for pumping of secondary antibodies was about 6 min.

Flow rate

In this experimental series, we measured anti-TG and anti-TPO at: i) low (3.35 μL/min); ii) medium (6.7 μL/min); iii) high (13.4 μL/min) and iv) extremely high 35 μL/min flow rates. The data for anti-TG shown in Fig. 9 were the same within experimental error. The similar data were obtained for anti-TPO.
Fig. 9

Effect of flow rate on the biosensor signal (signals for anti-TG are shown).

Effect of flow rate on the biosensor signal (signals for anti-TG are shown). The possible effect of mass transfer on the procedure of determining the constants (kob, kon, koff, KA, KD) was also studied at the flow rates in the range of (3.35 – 35) μL/min. As an example, the values of the observed constant kob at different flow rates during determination of native kinetic characteristics of binding TG antigen with autoantibodies in the clinical sample of patient #3 are shown in Table 3. The observed constant is the same within experimental error, as well as the constants kon, koff, KA, KD calculated based on the obtained values of kob. That suggests no significant effect of mass transfer limitations. The flow rate of 6.7 μL/min used in the research was optimal in view of the reagent consumption and reliable absence of mass transfer limitations.
Table 3

Values of the observed constant kob at different flow rates

Flow rate3.35 μL/min6.7 μL/min13.4 μL/min35 μL/min
kob x 103, s-116 ± 419 ± 517 ± 322 ± 4
Values of the observed constant kob at different flow rates

Effect of pH

The effect of buffer pH was investigated using buffers of different pH values: 5.0, 6.0, 7.4, 8.0, 9.6. For the measurements, we added 40 µL of 10 mg/mL BSA diluted in the chosen buffer to 50 µL of human blood serum. The resulting solution was pumped for 10 min at room temperature along the glass biochip having respective antigens pre-immobilized in different sensing spots. At this stage, the IgG autoantibodies in blood serum bound with the antigen on the biochip. Then, 10 mg/mL BSA diluted in the chosen buffer was pumped along the sensor chip. After that, a 50 µg/mL solution of secondary recognition antibody (goat antibody to human immunoglobulins) in 10 mg/mL BSA diluted in the chosen buffer was pumped. The data presented in Fig. 10 show no changes in signal within experimental error in the pH range of 6.0 – 8.0. At pH < 6.0 and pH > 8.0, the signal decreases. In the pH range of 7.35 – 7.45, relevant to blood (serum), in which autoantibodies should be measured, the signal is uniform within experimental error.
Fig. 10

The effect of pH on the biosensor signal.

The effect of pH on the biosensor signal.

Correlation of the developed biosensor with ELISA

We compared the data obtained by measuring the same serum samples by both the proposed assay and ELISA (Table 4 and Fig. 11).
Table 4

Raw data on correlation of the developed biosensor with ELISA.

Patient #Anti-thyroid peroxidase autoantibodyAnti-thyroglobulin autoantibody
ELISA, IU/mLDeveloped assay, IU/mLELISA, IU/mLDeveloped assay, IU/mL
10.91.855.86.7
28659717385
31.12.33212189
40.80.96384324
52.43.7507578
Fig. 11

Correlation of the data obtained by the developed multiplex label-free biosensor and ELISA for detection of anti-TG in clinical samples.

Raw data on correlation of the developed biosensor with ELISA. Correlation of the data obtained by the developed multiplex label-free biosensor and ELISA for detection of anti-TG in clinical samples.

Comparison of the developed biosensor with traditional methods

Table 5 exhibits the side-by-side comparison of analytical characteristics of the developed biosensor with the most sensitive modern techniques.
Table 5

Comparison of analytical characteristics of the developed biosensor for measuring anti-TG and anti-TPO in serum with those of label-based methods

Name (manufacturer)LabelsDetection methodLOD for anti-TPO, IU/mLLOD for anti-TG, IU/mLSimultaneous detection of several autoantibodies in a single sampleDetermination of kinetic characteristicsDuration, minRefs
The presented biosensorLabel-freeMultiplex spectral-correlation interferometry1.76YesYes25[1]
RIA kits (Institute of isotopes Ltd.)125I (radioactive)Radioimmunometric213NoNo180[15,16]
RIA kits (Demeditec Diagnostics GmbH)125I (radioactive)Radioimmunometric2.48.6NoNo120-180[17,18]
DYNOtest (BRAHMS Diagnostica)125I (radioactive)Radioimmunometric5.55.5NoNo180[19], [20], [21]
ELISA kits (Demeditec Diagnostics GmbH)Horseradish peroxidaseColorimetric510NoNo65[22,23]
AccuBind® ELISA Kits (Monobind Inc.)Horseradish peroxidaseColorimetric1.55NoNo105[24,25]
Elecsys® (Roche Diagnostics)RutheniumElectrochemiluminescence510NoNo20[20,26,27]
ARCHITECT i2000 (Abbott Diagnostics)Acridinium estersBead-based chemiluminescence11NoNo29[28], [29], [30]
ADVIA Centaur®, (Siemens Healthineer)Acridinium estersBead-based chemiluminescence1510NoNo18[28], [29], [30], [31]
TRACE Kryptor, Brahms DiagnosticaEuropium cryptate/ XL 665Fluorescence1.810NoNo29[27,31,32]
Comparison of analytical characteristics of the developed biosensor for measuring anti-TG and anti-TPO in serum with those of label-based methods Fig. 12 illustrates the difference between our principle and traditional label-free techniques. Our principle based on autoantibody polyvalency has permitted, for the first time, registration of autoantibody interaction with free antigens in serum rather than those immobilized on a surface. Our approach allows avoiding most of the surface-induced issues (described in the related Biosens. Bioelectron. paper [1]), which are due to unpredictable conformation of surface-deposited antigens.
Fig. 12

Difference between our principle and traditional label-free approaches for determination of kinetic parameters of autoantibody in serum.

Difference between our principle and traditional label-free approaches for determination of kinetic parameters of autoantibody in serum.

Linearity of dilution and recovery test

Since all experiments with autoantibodies are carried out in clinical serum, “spike and recovery tests” are not appropriate. We carried out the “linearity of dilution” test, and the obtained concentration data are given in Table 6.
Table 6

Linearity of dilution test.

anti-TPO concentrationanti-TG concentration
Serum from patient #2Serum from patient #5
Dilution factorMeasured, IU/mLExpected, IU/mL% ExpectedDilution factorMeasured, IU/mLExpected, IU/mL% Expected
19711578
2507.3485.5104.5%2284.2289.098.3%
4237.9242.898.0%4136.4144.594.4%
8113.5121.493.5%876.972.3106.4%
1661.160.7100.8%1637.636.1104.1%
Linearity of dilution test.

Biochip characterization

Mechanism of antigen attachment to the biochip surface

The mechanism of antigen attachment to the biochip glass surface is illustrated in Fig. 13 (see also the full protocol in section 2.3 of the related Biosens. Bioelectron. paper [1]). The glass surface pre-treated with piranha solution (Fig. 13a) is modified with aminosylane (APTES) to introduce an amino group (NH2) as a basis for the following functionalization (Fig. 13b). Further treatment with succinic anhydride (SA) forms a carboxyl group (COOH) on the surface (Fig. 13c) that is activated at the next step with carbodiimide (EDC) to produce o-acylisourea active ester (Fig. 13d). The latter reacts with an amino group (NH2) of antigen with the formation of a covalent peptide bond (Fig. 13e).
Fig. 13

Mechanism of antigen attachment onto the glass surface.

Mechanism of antigen attachment onto the glass surface.

Instrumental control of chemical modifications of biochip surface

The process of chemical modifications of the slip during fabrication of the biochip was controlled with X-ray photoelectron spectroscopy (XPS). For this purpose, we employed the photoelectron spectrometer “KRATOS AXIS ULTRA DLD” with spherical analyzer, ion gun, UV- and X-ray sources. Charge neutralization was used for all samples. The binding energy scale was charge referenced to the C1s at 284.7 eV. The characteristic peaks in the spectra to be analyzed are as follows: nitrogen N1s - at 400 eV (see Fig. 14a, b, c), carbon C1s – at 286 eV and 289 eV for С-О and С=О bonds, respectively (Fig. 14 d, e, f).
Fig. 14

XPS-spectra of the biochip surface: a, d – before modification; b, e – after modification with APTES; c, f – after modification with succinic anhydride.

XPS-spectra of the biochip surface: a, d – before modification; b, e – after modification with APTES; c, f – after modification with succinic anhydride. The spectrum of an original, non-modified glass is shown in Fig. 14a. The Figs. 14b, c show pronounced peaks at 400 eV, corresponding to amino groups appeared on the surface after incubation with APTES. The carboxylated glass produced by subsequent incubation with succinic anhydride (Fig. 14f), in contrast to the aminated and unmodified ones (Figs. 14d, e), exhibits a peak at 288 eV that corresponds to С=О bond, which is available in carboxyl group and amide bond. The peak at 286 eV appears in the spectra of both aminated and carboxylated surfaces. That peak corresponds to С-О bond, which presents in carboxyl group and is formed during silanization of the surface.

Biochip activation

The effect of biochip activation on stability of protein immobilization was assessed as follows. The immobilization process was real-time monitored with non-activated biochips and those activated with carbodiimide. The characteristic sensograms recorded in these experiments are shown in Fig. 15.
Fig. 15

Sensograms of stable immobilization of proteins on the fabricated surface after its activation with carbodiimide (top, red line) and reversible immobilization on the unactivated surface (bottom, green line) with desorption after pumping PBS-BSA buffer at 30 min.

Sensograms of stable immobilization of proteins on the fabricated surface after its activation with carbodiimide (top, red line) and reversible immobilization on the unactivated surface (bottom, green line) with desorption after pumping PBS-BSA buffer at 30 min.

Biochip shelf-life

Here, the immobilization stage was implemented inside the biosensor to demonstrate real-time registration of all reactions. This stage, though, is not required for the biosensor operation. The biochips can be prepared in advance. To estimate the shelf-life of the biochips, the immobilization was done by spot deposition rather than inside the biosensor, and was followed by drying at room temperature and further refrigerating at +4 oC. After one-month storage, the biochips were used in the experiments. Fig. 16 shows similar (within experimental error) signals obtained with "fresh" and stored biochips. The statistical insignificance of differences in the signals is confirmed by p-value of 0.44 exceeding 0.1.
Fig. 16

Comparison of the signals obtained with serums of low (5.8 IU/mL, blue bars) and high (384 IU/mL, red bars) anti-TG levels using "fresh" and one-month stored biochips.

Comparison of the signals obtained with serums of low (5.8 IU/mL, blue bars) and high (384 IU/mL, red bars) anti-TG levels using "fresh" and one-month stored biochips.

Raw data from the sensor

Table 7 presents the fragments of raw data of sensograms obtained for both native kinetics determination and concentration measurement of different autoantibodies in the same serum sample.
Table 7

Raw data recorded by the developed biosensor for simultaneous detection and kinetics characterization of different autoantibodies in the same serum sample.

Time, minΔd, nmTime, minΔd, nm
Native kinetics determinationConcentration measurementNative kinetics determinationConcentration measurement
TPOTGTPOTGTPOTGTPOTG
25.024481.825281.344502.102711.0960828.924351.905881.463562.901771.98803
25.073311.827911.344722.104011.0973128.973181.905861.466652.902501.99066
25.122141.830631.345712.104921.1004929.021881.907131.469162.909281.99355
25.170961.832271.345992.103591.1008129.070711.909321.472172.913241.99403
25.219791.833741.345012.106221.1017329.119531.913291.476112.916991.99364
25.268491.834061.344432.105961.1024029.168231.916391.479382.916071.99817
25.317191.833951.344272.107331.1011129.217061.921951.482802.920241.99817
25.366141.834631.344432.108081.1021329.265761.927901.486232.922281.99835
25.414841.835411.345632.106351.1024929.314581.931731.489302.924731.99823
25.463541.837051.345752.104891.1022029.363421.937741.492002.926871.99738
25.512381.837941.344772.098041.1004729.412111.942601.494572.929561.99679
25.561081.838551.343582.096831.0996529.460941.949921.498532.930471.99580
25.610031.838071.343282.096231.1002129.509641.955021.501682.931731.99679
25.658591.835961.344542.096341.1007629.558461.962511.504672.933901.99489
25.707561.837351.343952.098441.0990629.607161.970661.505812.936241.99585
25.756251.834961.344672.100861.1012329.655991.979751.507752.935861.99553
25.804951.837831.344762.101091.1018929.704691.987761.511292.938521.99449
25.853781.840701.343872.101431.1028929.753521.996391.515372.940421.98936
25.902481.842631.344012.102081.1030429.802342.007371.518282.940641.98914
25.951311.845651.344462.102901.1052229.851042.015761.521182.942781.98808
26.000131.844421.344912.102321.1094329.899742.023171.522772.943521.98767
26.048831.843301.345062.099961.1138929.948702.029651.523112.943021.98718
26.097661.846191.345712.095941.1204929.997532.041521.524772.944621.98800
26.146361.844821.346562.090461.1249830.046232.052241.527422.946451.98628
26.195181.845381.348132.086301.1319430.094932.060941.530772.945581.98673
26.243881.844001.348022.084361.1414630.143752.069021.532762.944901.98513
26.292841.845161.348712.080451.1526430.192452.079661.533552.947501.98338
26.341411.843511.348562.078321.1643830.241282.089661.536022.946961.98302
26.390371.841651.348292.079271.1790230.290112.099401.536112.948011.98429
26.438931.841231.348882.080311.1946230.338812.109141.536772.949241.98406
26.487761.843311.350512.083701.2101130.387632.117851.538222.951511.98262
26.536591.844981.351872.089191.2285230.436332.124461.539282.951321.98400
26.585291.846871.352052.096481.2470430.485162.128741.540012.952401.98541
26.634121.847091.351272.107031.2670430.533862.135321.540822.952531.98675
26.682941.845741.351392.119321.2889630.582682.139591.542792.954101.98772
26.731641.845621.351352.135241.3144230.631512.145941.545402.953171.98980
26.780471.849081.351832.149481.3399130.680212.147471.548002.954271.99039
26.829291.851941.353392.167661.3652730.728912.151521.550452.953281.99091
26.877991.854691.353112.186771.3934630.777742.151661.551702.952251.99209
26.926831.856451.354422.208211.4213730.826562.153981.553312.951241.99115
26.975531.860961.354552.229701.4481330.875262.156031.553912.952941.99052
27.024351.861131.355622.255451.4757030.924232.158931.553662.951391.99262
27.073181.859671.356332.280981.5041230.972922.160211.553942.950741.99539
27.121881.860161.357502.305181.5322431.021622.164661.554512.948281.99396
27.170701.859121.358702.328301.5606231.070442.166991.554112.948981.99326
27.219531.860701.359092.354831.5884131.119142.167991.554632.949371.99299
27.268231.863331.360542.386151.6156631.167982.169301.555402.951621.99153
27.317061.865461.361212.418391.6409631.216792.167161.555992.951661.99306
27.376961.870171.363262.450411.6693231.265632.168581.555022.950531.99423
27.436731.871321.365372.482691.6949031.314332.168441.555742.951221.99349
27.485421.873591.368702.513671.7188231.363162.170461.556272.950621.99542
27.534381.872201.371262.548151.7435631.411852.173041.557372.950451.99666
27.583081.875011.372682.580371.7656331.460682.172871.558612.950101.99502
27.631781.879281.374042.609971.7849931.509382.173521.559422.948591.99584
27.680601.882181.377732.638361.8044231.558212.173871.559572.948341.99672
27.729561.882501.381072.664541.8227531.606912.173941.560482.948951.99444
27.778131.882841.383492.685471.8387431.655732.172851.561812.947301.99251
27.826961.883221.386502.707451.8553531.704562.174171.562912.945151.99096
27.875661.882971.388202.727081.8708531.753262.175391.563202.944161.98945
27.924481.884641.391382.746551.8825231.802082.176741.563782.942781.99028
27.985421.885461.395392.761051.8933131.850912.175341.563782.943261.98809
28.046361.884701.400102.776291.9042231.899612.176311.563442.943191.98876
28.095181.886191.403032.788581.9139531.948442.176741.563302.942691.98946
28.144011.888131.406072.799211.9218431.997142.178441.563762.943341.99113
28.192711.889281.409182.809751.9297832.045962.179421.565612.944351.99407
28.241411.889171.411542.819451.9376132.094662.177121.566872.943851.99467
28.290371.891091.413842.828921.9437332.154562.178751.567322.945841.99205
28.339061.895071.418032.836551.9502832.214462.181001.566872.946361.99180
28.387891.894441.420922.845261.9547932.263162.178081.568652.947611.99283
28.436591.896851.424722.854431.9602232.311982.178411.568632.946291.99589
28.485421.898821.428262.860921.9653232.360682.178221.570852.947991.99836
28.534121.899601.431882.865341.9727032.409642.180021.572792.948241.99910
28.582941.896421.436022.872321.9784832.458332.181961.574312.948641.99868
28.631781.896501.441302.878401.9804532.507032.180471.574322.949251.99877
28.680481.899281.444482.883291.9823932.555862.182661.574782.949321.99768
28.729171.902331.448422.888541.9867832.604692.180351.575062.949701.99639
28.777871.902951.451382.889741.9837132.653522.175641.576642.949461.99792
28.826831.902791.455692.893021.9837732.702212.173511.577692.949901.99693
28.875661.903161.459712.899071.9870932.750912.172521.579572.951201.99541
Raw data recorded by the developed biosensor for simultaneous detection and kinetics characterization of different autoantibodies in the same serum sample.

Experimental Design, Materials, and Methods

Multiplex biosensor based on the spectral-correlation interferometry that uses a microscope cover glass slip as a biochip, affordable to be single-used in medical applications, has been designed. The microarray imaging label-free biosensor employs radiation from a broadband superluminescent diode (SLD-381-MP, Superlum Diodes, Ltd., Russia) in an optical scheme that comprises two interferometers. The base (inter-mirror distance) of one Fabry-Perot interferometer is periodically changed with a piezoelectric driver. The mentioned glass biochip serves simultaneously as the second two-beam reflective interferometer. During biochemical reactions on the biochip, the optical thickness of the slip with the bound biological layer changes. To measure such changes, the interference between a reference beam reflected from the bottom surface of the slip and a beam reflected from the "biolayer – analyzed sample" interface is used. Variations in optical thickness of the biomolecule layer on the recognition spots on the biochip are calculated by phase changes of correlation signals from a 12-bit monochrome CCD-camera (Basler, Germany). For simultaneous registration of biochemical reactions in several recognition spots, their image signals are averaged over the area of each recognition spot. The following reagents were used in this work: microscope cover glasses, bovine serum albumin (BSA), (3-Aminopropyl)triethoxysilane (APTES), N-(3-Dimethylaminopropyl)-N-ethylcarbodiimide hydrochloride (EDC) (Sigma Aldrich, USA); sulphuric acid, methanol, dimethylformamide (Chimmed, Russia); thyreoperoxidase, thyroglobulin, goat monoclonal antibodies against human IgG (Xema medica, Russia; GeneTex, USA; Russian Research Center for Molecular Diagnostics and Therapy). The serum standards and samples from patients were provided by the Russian Cardiology Research and Production Complex of Russian Ministry of Health (Moscow, Russia), and the present experiments were approved by the ethics committee of this institution. Immunoglobulin fraction of serum for intravenous use was purchased in a local pharmacy. As the low-cost single-used biochips, microscope glass cover slips that do not require any additional metal or dielectric coatings were used. The biochip surface was carboxilated for efficient and rapid immobilization of biomolecules. For this purpose, the thoroughly cleaned microscope cover slips were incubated for 16 h under an exhaust hood in 3% solution of APTES in methanol and then for 2 h in 15 mM solution of succinic anhydride in dimethylformamide. Then the glasses were thermally processed in a dry heat oven at 105 ºC for 1 h with further washing at room temperature. The chips were activated by 15-min incubation in 20 mg/mL solution of EDC in MES. Then 50 µg/mL of antigens in PBS was deposited in different sensing spots to achieve covalent immobilization of antigen onto the biochip due to interaction of antigen amino groups with activated carboxyl groups on the surface. The surface was blocked with a mixture of 1% glycine and 0.5% of bovine serum albumin (BSA) in PBS buffer. The prepared biochips can be stored without deterioration for a long time until using. Quality control of immunoreagents was implemented as described in Ref. [33]. The process of chemical modifications of the slip during fabrication of the biochip was controlled with X-ray photoelectron spectroscopy (XPS). For this purpose, we employed the photoelectron spectrometer “KRATOS AXIS ULTRA DLD” with spherical analyzer, ion gun, UV- and X-ray sources described in section 7 above. All experiments in this research were carried out in human blood serum. Each of the samples contained various amounts of anti-TG and anti-TPO, naturally produced by the immune system of the patients (not spiked). 50 µL of the human blood serum was diluted with 40 µL of 10 mg/mL of PBS-BSA and then pumped for 10 min at room temperature along the glass biochip having respective antigens pre-immobilized in different sensing spots. At this stage, some of the IgG autoantibodies in blood serum, which had two binding fragments (Fab), bound with corresponding antigen on the surface with one Fab. The other free Fab was still capable to interact with free antigen. To provide the unique information on autoantibody binding with native antigen, which was not bound to the surface, the same serum samples with the spiked free antigen (both TPO and TG) were pumped. The kinetic constants of autoantibody interaction with native antigen were calculated using the sensograms obtained in these experiments and the model described in section 1 above. Other procedures and protocols are described in details in Ref. [1].

Declaration of Competing Interest

P.I.N. and B.G.G. are the named inventors on SCI- and SPI-related patents.
SubjectAnalytical chemistry
Specific subject areaSimultaneous measuring both critical parameters of several autoantibodies in human serum: kinetics of autoantibody interaction with non-immobilized antigens and concentration

Type of dataTables and figures

How data were acquiredSpectral-correlation interferometry (SCI); spectral-phase interferometry (SPI); SCI and SPI biosensors; enzyme linked immunosorbent assay (ELISA); X-ray photoelectron spectroscopy (XPS); proprietary software for SPI; data fitting with commonly used software and a mathematical model that describes adsorption kinetics of molecules on solid phase; statistical t-test.

Data formatRaw and analyzed

Parameters for data collectionAffordable single-used microscope cover glass slips were used as the biochips. Each stage of the biochip preparation was controlled with X-ray photoelectron spectroscopy. Each stage of the biosensor operation was monitored in real time. Kinetic parameters were calculated by fitting the data with a mathematical model. The biosensor operation was validated by correlation with ELISA. Analytical characteristics of the developed biosensor were compared with those of other advanced methods.

Description of data collectionA multiplex microarray biosensor designed for this research recorded in real time sensograms of the biolayer thickness changes in separate recognition spots on a glass biochip. A 50-µL sample of human blood serum containing autoantibodies such as anti-thyroglobulin (anti-TG) and/or anti-thyroid peroxidase (anti-TPO) was diluted and pumped along the biochip for 10 min at room temperature. The respective antigens were pre-immobilized in different sensing spots of the biochip. Using the sensograms, concentration and kinetic constants of autoantibodies were determined at pumping anti-human antibodies or antigens, respectively.

Data source locationMoscow, Russia

Data accessibilityWith the article

Related research articleA.V. Orlov, A.V. Pushkarev, S.L. Znoyko, D.O. Novichikhin, V.A. Bragina, B.G. Gorshkov and P.I. Nikitin, Multiplex label-free biosensor for detection of autoantibodies in human serum: tool for new kinetics-based diagnostics of autoimmune diseases. Biosens. Bioelectron. 159 (2020) 112187, doi: 10.1016/j.bios.2020.112187

  16 in total

Review 1.  Clinical analyzers. Immunoassays.

Authors:  L J Sokoll; D W Chan
Journal:  Anal Chem       Date:  1999-06-15       Impact factor: 6.986

2.  Reversible conformational transitions of a polymer brush containing boronic acid and its interaction with mucin glycoprotein.

Authors:  Alexander E Ivanov; Nadezda Solodukhina; Marie Wahlgren; Lars Nilsson; Alexander A Vikhrov; Maxim P Nikitin; Alexey V Orlov; Petr I Nikitin; Marina V Kuzimenkova; Vitali P Zubov
Journal:  Macromol Biosci       Date:  2010-12-06       Impact factor: 4.979

3.  Performance characteristics of 5 automated thyroglobulin autoantibody and thyroid peroxidase autoantibody assays.

Authors:  Sonia L La'ulu; Patricia R Slev; William L Roberts
Journal:  Clin Chim Acta       Date:  2006-07-25       Impact factor: 3.786

4.  Analytical Platform with Selectable Assay Parameters Based on Three Functions of Magnetic Nanoparticles: Demonstration of Highly Sensitive Rapid Quantitation of Staphylococcal Enterotoxin B in Food.

Authors:  Vera A Bragina; Sergey L Znoyko; Alexey V Orlov; Averyan V Pushkarev; Maxim P Nikitin; Petr I Nikitin
Journal:  Anal Chem       Date:  2019-07-22       Impact factor: 6.986

5.  Multiplex label-free biosensor for detection of autoantibodies in human serum: Tool for new kinetics-based diagnostics of autoimmune diseases.

Authors:  A V Orlov; A V Pushkarev; S L Znoyko; D O Novichikhin; V A Bragina; B G Gorshkov; P I Nikitin
Journal:  Biosens Bioelectron       Date:  2020-04-11       Impact factor: 10.618

6.  Natural course of autoimmune thyroiditis in type 1 diabetes: association with gender, age, diabetes duration, and puberty.

Authors:  O Kordonouri; R Hartmann; D Deiss; M Wilms; A Grüters-Kieslich
Journal:  Arch Dis Child       Date:  2005-04       Impact factor: 3.791

7.  Lectin-carbohydrate affinity measured using a quartz crystal microbalance.

Authors:  Kateryna Lebed; Andrzej J Kulik; László Forró; Małgorzata Lekka
Journal:  J Colloid Interface Sci       Date:  2006-03-10       Impact factor: 8.128

8.  Relationship between antithyroid antibody and pregnancy outcome following in vitro fertilization and embryo transfer.

Authors:  Yi-ping Zhong; Ying Ying; Hai-tao Wu; Can-quan Zhou; Yan-wen Xu; Qiong Wang; Jie Li; Xiao-ting Shen; Jin Li
Journal:  Int J Med Sci       Date:  2012-01-05       Impact factor: 3.738

9.  Establishment of the upper reference limit for thyroid peroxidase autoantibodies according to the guidelines proposed by the National Academy of Clinical Biochemistry: comparison of five different automated methods.

Authors:  Federica D'Aurizio; Paolo Metus; Annalisa Polizzi Anselmo; Danilo Villalta; Anna Ferrari; Roberto Castello; Graziella Giani; Elio Tonutti; Nicola Bizzaro; Renato Tozzoli
Journal:  Auto Immun Highlights       Date:  2015-08-15

10.  Definition of the upper reference limit for thyroglobulin antibodies according to the National Academy of Clinical Biochemistry guidelines: comparison of eleven different automated methods.

Authors:  F D'Aurizio; P Metus; A Ferrari; B Caruso; R Castello; D Villalta; A Steffan; K Gaspardo; F Pesente; N Bizzaro; E Tonutti; S Valverde; C Cosma; M Plebani; R Tozzoli
Journal:  Auto Immun Highlights       Date:  2017-06-19
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.