Establishment of the upper reference limit for thyroid peroxidase autoantibodies according to the guidelines proposed by the National Academy of Clinical Biochemistry: comparison of five different automated methods.

Federica D'Aurizio1, Paolo Metus2, Annalisa Polizzi Anselmo2, Danilo Villalta3, Anna Ferrari4, Roberto Castello5, Graziella Giani6, Elio Tonutti7, Nicola Bizzaro8, Renato Tozzoli2.   

Abstract

AIM OF THE STUDY: The estimation of the upper reference limit (URL) for autoantibodies against thyroid peroxidase (TPOAbs) is a controversial issue, because of an uncertainty associated with the criteria used to correctly define the reference population. In addition, the URL of TPOAbs is method-dependent and often arbitrarily established in current laboratory practice. The aim of this study was to determine the reference limits of TPOAbs in a male sample according to the National Academy of Clinical Biochemistry (NACB) guidelines, and to compare them with those obtained in a female group, for five third-generation commercial-automated immunoassay (IMA) platforms.
METHODS: 120 healthy males and 120 healthy females with NACB-required characteristics (younger than 30 years, TSH between 0.5 and 2.0 mIU/L, normal thyroid ultrasound, absence of thyroid disease and absence of other autoimmune diseases) were studied. Sera were analyzed for TPOAbs concentration using five IMA methods applied in automated analyzers: Immulite 2000 XPi (IMM); Maglumi 2000 Plus (MAG); Kryptor Compact Plus (KRY); Phadia 250 (PHA) and Liaison XL (LIA).
RESULTS: A statistically significant difference (p < 0.05) between medians in male and female groups was observed for PHA (2.6 and 3.1 IU/mL, respectively) but not for the other four methods. Scatter plots of TPOAbs values revealed a wide dispersion with very different coefficients of variation between the five methods, varying from 48.6 % for KRY in females to 126.3 % for MAG in females. The URLs differed in males and females according to the method: 28.7 and 29.0 IU/mL for IMM, 24.6 and 25.4 IU/mL for MAG, 6.4 and 6.9 IU/mL for KRY, 8.3 and 10.0 IU/mL for PHA and 14.2 and 17.9 IU/mL for LIA, respectively. Such URLs were lower than those stated by the manufacturers except for LIA in females. The difference between URLs ranged from a minimum of 11.3 % (LIA in males) to a maximum of 66.8 % (PHA in males).
CONCLUSIONS: Differences in URLs could result from the different coating preparations of the TPO antigen (purified native or recombinant) on solid phase, which affect the proper exposure of the immunodominant epitopes recognized by the polyclonal antibodies present in serum of patients with autoimmune thyroid disease (AITD). Based on these findings, we suggest to overcome the proposal of the NACB guidelines which recommend to involve a single group of young male subjects, and propose, instead, to utilize two distinct groups: one of males and one of females. This new proposal removes the apparent contrast of an all-male reference group for a disease (such as AITD) that affects mainly females. However, in spite of the harmonization among methods provided by the use of an international standard preparation, the wide dispersion of quantitative results still observed in this study suggests the need for further efforts to better understand the cause of these discrepancies, focusing on TPO antigen preparations as the possible source of variability among different assays.

Entities:  

Keywords:  Autoimmune thyroid disease; Automation; Harmonization; Immunoassay; Reference limits; Thyroid peroxidase autoantibodies

Year:  2015        PMID: 26275915      PMCID: PMC4633414          DOI: 10.1007/s13317-015-0070-x

Source DB:  PubMed          Journal:  Auto Immun Highlights        ISSN: 2038-0305


Introduction

Autoantibodies against thyroid peroxidase (TPOAbs) are both diagnostic hallmarks and early indicators of autoimmune thyroid disease (AITD), playing an important predictive role in healthy subjects, in pregnant women and in high-risk patients [1-3]. In recent years, refinements in autoantigen preparation and better selection of polyclonal and monoclonal antibodies have led to a new (third) generation of automated quantitative immunoassays (IMAs) with improved sensitivity and specificity for the measurement of TPOAbs [4, 5]. Despite their recently expanded use in clinical laboratories [6-8], it is yet unclear whether these more-sensitive automated third-generation IMAs have made improvements in terms of diagnostic accuracy and harmonization compared with previous methods [9, 10]. Hence, further efforts in defining the threshold values of positivity need to be made in order to avoid misclassification of patients with AITD [9]. The estimation of the upper reference limit (URL) for TPOAbs is a controversial issue, because of an uncertainty as to the criteria for correctly defining the reference population. The currently proposed direct methods are described in the National Academy of Clinical Biochemistry (NACB) guidelines, which recommend the use of a reference group composed of 120 men with the following features: younger than 30 years, biochemically euthyroid (serum TSH between 0.5 and 2.0 mIU/L), without goiter as proven by ultrasound investigation, no personal/family history of either thyroid disease or of non-thyroid autoimmune diseases [11]. These guidelines are in themselves difficult to apply or standardize. Furthermore, it is perplexing, in that it requires a male reference group for diseases (AITD) that predominantly affect females [12]. Hence, in developing commercial methods for TPOAbs quantification, manufacturers are faced with difficulties in following the NACB recommendations. Consequently, results obtained by different researchers using different analytical methods show a wide variation in URLs [8–10, 13, 14]. In addition, the definition of reference intervals for TPOAbs has limitations due to the natural history of AITD, because a large number of individuals may have specific autoantibodies in serum 10–15 years before the onset of these diseases. To overcome this problem and avoid the risk of enlisting apparently healthy subjects who may bear TPOAbs, an indirect method for the definition of reference limits was proposed by some authors [12, 15]. Using this approach, the URL values were lower than those obtained by direct methods. Also, they displayed sex-dependent differences [15]. Nevertheless, the indirect method also has limitations, being applicable only to the single method used in the population sample considered in that particular study. Taking into account that TPOAbs reference limits are method dependent and often arbitrarily established in the current laboratory practice [9, 10, 16], the aim of this study was to determine the URL of TPOAbs in a male sample according to the NACB guidelines, and to compare it with that obtained in a female group, applying and comparing five third-generation commercial automated IMA platforms.

Methods

120 males and 120 females with NACB-required characteristics (younger than 30 years, TSH between 0.5 and 2.0 mIU/L; normal thyroid ultrasound; absence of autoimmune and non-autoimmune thyroid disease and absence of other autoimmune diseases) were enrolled in the study. They came from a large health survey called ‘Thyroid takes to the square,’ carried out in the province of Verona (Italy) from 2008 to 2013, in which 7970 subjects with no previous or current thyroid disease symptoms were screened by clinical history, thyroid function test, and thyroid ultrasound. All participants gave their informed consent. The sera of the 120 males and 120 females were analyzed for TPOAbs concentration using five IMA methods applied in automated analyzers: Immulite 2000 XPi (IMM) (Siemens Healthcare Diagnostics, Flanders, NJ, USA); Maglumi 2000 Plus (MAG) (Shenzen New Industries Biomedical Engineering-SNIBE, Shenzen, China); Kryptor Compact Plus (KRY) (BRAHMS Thermo Scientific, Henningsdorf, Germany); Phadia 250 (PHA) (Phadia AB, Uppsala, Sweden) and Liaison XL (LIA) (DiaSorin, Saluggia, Italy). All assays were performed according to the manufacturers’ instructions at four different laboratories in Friuli Venezia Giulia region (Italy): Pordenone (IMM and MAG), S. Vito al Tagliamento (KRY), Tolmezzo (PHA), and Udine (LIA). The main features of the five methods are shown in Table 1.
Table 1

The main features of the five automated immunoassay methods, as declared by the manufacturers

SystemMethodTracerAntigenImprecision (%): intra-assay; between-assayImprecision (%): totalLoD (IU/mL)LoQ (IU/mL)
IMMCLIALuminescence: adamantyl dioxetane phosphateNative TPO4.9–6.3; nr6.1–8.15.0nr
MAGCLIALuminescence: N-(aminobutyl)-N-(ethyl)-isoluminolNative TPO6.8–7.3; 7.4–7.6nr0.38nr
KRYFIAFluorescence: europium cryptate/XL 665Native TPO2.1–7.6; 5.8–16.0nr1.89.0
PHAFEIAFluorescence: β-D-galactosidase; 4-methyl-umbelliferyl-β-D-galactosideRecombinant TPO4.1–4.7; 2.5–4.8nrnrnr
LIACLIALuminescence: isoluminol derivativeRecombinant TPO3.6–6.2; 4.7–6.9nr0.61.0

CLIA chemiluminescence immunoassay, FEIA fluorescence enzyme immunoassay, FIA fluoroimmunoassay, IMM Immulite 2000 XPi, KRY Kryptor Compact Plus, LIA Liaison XL, LoD limit of detection, LoQ limit of quantitation, MAG Maglumi 2000 Plus, nr not reported, PHA Phadia 250, TPO thyroid peroxidase

The main features of the five automated immunoassay methods, as declared by the manufacturers CLIA chemiluminescence immunoassay, FEIA fluorescence enzyme immunoassay, FIA fluoroimmunoassay, IMM Immulite 2000 XPi, KRY Kryptor Compact Plus, LIA Liaison XL, LoD limit of detection, LoQ limit of quantitation, MAG Maglumi 2000 Plus, nr not reported, PHA Phadia 250, TPO thyroid peroxidase The Immulite 2000 XPi was a continuous random-access system designed around a proprietary assay tube that allowed for efficient washing of an integral antibody- or antigen-coated polystyrene beads as solid phase. This assay used an alkaline phosphatase enzyme label that was quantified with a sensitive chemiluminescent substrate (adamantyl-dioxetane phosphate) by creation of an unstable adamantyl-dioxetane anion. The breakdown of this unstable anion generated a prolonged glow of light [17]. The TPOAbs test was based on a sequential chemiluminescent IMA (CLIA), which used native purified TPO antigen coated on the solid phase. As stated by the manufacturer, the intra-assay imprecision was 4.9–6.3 %; the total imprecision was 6.1–8.1 %; the limit of detection (LoD) of TPOAbs assay was 5.0 IU/mL and the limit of quantitation (LoQ) was not declared. The Maglumi 2000 Plus was a continuous random-access IMA system that used N-(aminobutyl)-N-(ethyl)-isoluminol (ABEI) as luminescence substrate and magnetic particles serving both the solid phase and the separator in a liquid phase, with two different monoclonal antibodies labeled with either ABEI or fluorescein-5-isothiocyanate [18]. The TPOAbs test was a sandwich CLIA with native TPO coated to nanomagnetic microbeads. According to the manufacturer, the intra-assay imprecision was 6.8–7.3 %; the between-assay imprecision was 7.4–7.6 %; the LoD of the method was 0.38 IU/mL and the LoQ was not declared. The Kryptor Compact Plus was an IMA system, based on time-resolved amplified cryptate emission technology (TRACE), which measured the signal emitted from an immunocomplex with a time delay. The basis of the TRACE technology was non-radioactive energy transfer from a donor to an acceptor. The donor was a cage-like structure with an europium ion in the center (cryptate); the acceptor was part of a chemically modified, light-collecting algal protein (XL 665). The proximity of the donor to the acceptor intensified the fluorescent signal of the cryptate and extended the life-span of the acceptor signal, permitting the measurement of temporarily delayed fluorescence [19]. The TPOAbs test was a competitive fluorescence IMA (FIA) that used enzymatically active native purified TPO. As stated by the manufacturer, the intra-assay imprecision was 2.1–7.6 %; the between-assay imprecision was 5.8–16.0 %; the LoD was 1.8 IU/mL and the LoQ was 9.0 IU/mL. The Phadia (previously known as UniCAP, then ImmunoCAP) 250 was a fully integrated analyzer designed for the testing of allergy and autoimmune diseases. The reagents for the autoimmune diagnostics were based on the well-established EliA, normal ELISA-type-coated wells, that used β-D-galactosidase-labeled antibodies as tracer and 4-methyl-umbelliferyl-β-D-galactoside as developing agent, forming a fluorescent product (4-methyl-umbelliferone) fluorescence enzyme immunoassay (FEIA) [20]. The EliA TPOAbs test was a sandwich FEIA method that used human recombinant TPO antigen. As stated by the manufacturer, the intra-assay imprecision was 4.1–4.7 %; the between-assay imprecision was 2.5–4.8 %; the LoD and the LoQ were not declared. The Liaison XL was a fully automated CLIA analyzer, which adopted a flash technology with paramagnetic microparticles as solid phase [15, 21]. The anti-TPO Liaison XL 2-step assay was a sandwich CLIA that used directly coated magnetic microparticles with recombinant TPO as solid phase and a conjugate based on an isoluminol derivate. As declared by the manufacturer, the intra-assay imprecision was 3.6–6.2 %; the between-assay imprecision was 4.7–6.9 %, the LoD was 0.6 IU/mL and the LoQ was 1.0 IU/mL.

Statistical analysis

All methods were standardized with the reference preparation MRC 66/387 and used international units (IU), except for KRY whose results were initially expressed in arbitrary units and subsequently corrected in IU (conversion factor = 0.175). To compare the data obtained by the five systems, the results were expressed as a median and the URL was established at the 99.0th ‰. The non-parametric Mann–Whitney U test was used to compare TPOAbs levels in males and females within the same method. A two-sided value of p < 0.05 was considered statistically significant. The difference between the manufacturer’s URL and the experimental URL was expressed as the ratio between them in percentage (Δ). The correlation studies were performed on all samples across all the five methods. GraphPad Prism, version 4.0 (GraphPad Prism Software, San Diego, CA, USA) and MedCalc, version 13.3.1 (MedCalc Software, Ostend, Belgium) were used for statistical analysis.

Results

Value distributions were not Gaussian with a positive skew in both males and females for all the five methods. A statistically significant difference (p < 0.05) between medians in male and female groups was observed for PHA (2.6 and 3.1 IU/mL, respectively) but not for the other four methods. The scatter plots of TPOAbs values revealed a wide dispersion with very different coefficient of variation (CV) between methods, varying from 48.6 % for KRY in females to 126.3 % for MAG in females (Fig. 1).
Fig. 1

Values of thyroid peroxidase autoantibodies, in males and females, measured by the five automated methods. a Immulite 2000 XPi, b Maglumi 2000 Plus, c Kryptor Compact Plus, d Phadia 250, e Liaison XL. M males, F females

Values of thyroid peroxidase autoantibodies, in males and females, measured by the five automated methods. a Immulite 2000 XPi, b Maglumi 2000 Plus, c Kryptor Compact Plus, d Phadia 250, e Liaison XL. M males, F females URLs differed in males and females according to the method (Table 2): 28.7 and 29.0 IU/mL for IMM, 24.6 and 25.4 IU/mL for MAG, 6.4 and 6.9 IU/mL for KRY, 8.3 and 10.0 IU/mL for PHA, and 14.2 and 17.9 IU/mL for LIA, respectively. Such URLs were lower than those stated by the manufacturers except for LIA in females. The Δ between URLs ranged from a minimum of 11.3 % (LIA in males) to a maximum of 66.8 % (PHA in males).
Table 2

The main statistical parameters of thyroid peroxidase autoantibodies, in males and females, measured by the five automated immunoassay methods

SexNo.Mean (CI 95 %)SDCVMedian p eURLmURLΔ
IMM
 Males1208.2 (7.1–9.3)6.073.66.40.828.735.018.0
 Females1208.9 (7.6–10.1)7.079.66.729.017.1
MAG
 Males1205.5 (4.3–6.6)6.5119.82.70.824.630.018.0
 Females1204.9 (3.8–6.1)6.2126.32.225.415.3
KRY
 Males1202.7 (2.5–2.9)1.450.62.60.66.410.539.0
 Females1202.6 (2.3–2.8)1.248.62.46.934.3
PHA
 Males1202.8 (2.6–3.1)1.450.92.6<0.058.325.066.8
 Females1203.7 (3.1–3.8)1.851.83.110.060.0
LIA
 Males1203.2 (2.6–3.9)3.4106.01.90.814.216.011.3
 Females1203.2 (2.6–3.9)3.9111.31.717.911.9

CV and Δ are expressed in percentage; Mean, Median, SD, eURL and mURL are expressed in IU/mL. eURL = 99.0th ‰; Δ = ∣mURL − eURL∣/mURL × 100

CI confidence intervals, CV coefficient of variation, eURL experimental upper reference limit, IMM Immulite 2000 XPi, KRY Kryptor Compact Plus, LIA Liaison XL, MAG Maglumi 2000 Plus, mURL manufacturer’s upper reference limit, PHA Phadia 250, SD standard deviation

The main statistical parameters of thyroid peroxidase autoantibodies, in males and females, measured by the five automated immunoassay methods CV and Δ are expressed in percentage; Mean, Median, SD, eURL and mURL are expressed in IU/mL. eURL = 99.0th ‰; Δ = ∣mURL − eURL∣/mURL × 100 CI confidence intervals, CV coefficient of variation, eURL experimental upper reference limit, IMM Immulite 2000 XPi, KRY Kryptor Compact Plus, LIA Liaison XL, MAG Maglumi 2000 Plus, mURL manufacturer’s upper reference limit, PHA Phadia 250, SD standard deviation Correlation analysis showed that all the five methods did not compare well with each other (Table 3), suggesting discrepancies between them.
Table 3

Method comparison

Methods n r (95 % CI)
IMM
 MAG2400.0011 (−0.1269 to 0.1291)
 KRY240−0.0235 (−0.1513 to 0.1050)
 PHA2400.0008 (−0.1266 to 0.1282)
LIA2400.0707 (−0.0576 to 0.1966)
KRY
 MAG240−0.0845 (−0.2108 to 0.0445)
 LIA240−0.0988 (−0.2247 to 0.0304)
 PHA240−0.0078 (−0.1362 to 0.1208)
LIA
 MAG2400.2628 (0.1390 to 0.3786)
 PHA2400.3679 (0.2521 to 0.4734)
MAG
 PHA2400.3226 (0.2028 to 0.4330)

CI confidence intervals, IMM Immulite 2000 XPi, KRY Kryptor Compact Plus, LIA Liaison XL, MAG Maglumi 2000 Plus, PHA Phadia 250, r correlation coefficient

Method comparison CI confidence intervals, IMM Immulite 2000 XPi, KRY Kryptor Compact Plus, LIA Liaison XL, MAG Maglumi 2000 Plus, PHA Phadia 250, r correlation coefficient

Discussion

To our knowledge, this is the first study on the definition and comparison of TPOAbs URLs between five different commercially available automated methods. The study was designed to accurately define and harmonize TPOAbs URLs in order to avoid misclassification of patients in the process of AITD diagnosis. Thus, according to the NACB recommendations, we evaluated TPOAbs concentrations and URLs in a reference male group comparing them with a female one. The first relevant result of this study was the demonstration of a significant difference between the TPOAbs URLs obtained with the five different automated IMA methods (both in males and females) and the results expected according to the corresponding package insert. With the exception of the LIA method in females, URLs values obtained in the present study were lower than those predicted in the insert, with Δ varying from 11.3 % (LIA in males) to 66.8 % (PHA in males) (Table 2). This finding was in contrast to the conclusion of a recent study which showed that the established URLs were very similar to those recommended by seven different manufacturers [14]. The difference between that study and the present one resides in the population sample: in the study by Springer et al. TPOAbs were measured in women in the first trimester of pregnancy; in the present study, TPOAbs were determined in 120 healthy males and in 120 healthy females. The discrepancies between the experimental and the manufacturer’s URLs, as found in the present study, could be linked to any of several factors. First of all, there are racial differences in autoantibody concentrations measured in different populations over the world: in most cases, studies, that were sponsored by the test’s manufacturer and were conducted in the geographical area of the production line, may be not reproducible in other geographical settings. An additional aspect was the lack of strict criteria in the selection of subjects for the reference group. In fact, the possibility of enrolling apparently healthy individuals with subclinical AITD and high levels of TPOAbs, whose prevalence is estimated around 3.3–25.8 % in the general population, raises the 99th ‰ of the reference value distribution [12, 13, 15, 22, 23] and produces an increase in the proportion of subjects classified as negative for TPOAbs. On the other hand, the demonstration of a significant difference between the TPOAbs URLs obtained with the five different automated IMA methods both in males and females, confirms the data of a previous study on the same topic, using indirect methods [15]. The second relevant consideration emerging from this study was the dependence of the URLs on the method used. A decade ago, when third-generation methods were introduced the differences between the URL values were on the order of 4–5 times (from 20.0 to 100.0 IU/mL) [8, 9]. In this study, the differences of the experimental URLs were similar, but at concentration levels that are four times lower than those reported in the past, ranging from 6.4 to 28.7 IU/mL in males and from 6.9 to 29.0 IU/mL in females (Table 2). The discrepancy of URLs between methods was associated with the dispersion of the results obtained by each method in both the male and female reference groups. Indeed, the dispersion expressed as CV % showed widely divergent and fluctuating values, from 48.6 % for KRY to 126.3 % for MAG (Table 2; Fig. 1). There are no clear explanations for these discrepancies. In fact, over the years, there has been an improvement in the harmonization among methods, due to automation of the analytical procedures and the use of the same reference preparation (MRC 66/387) [10]. Moreover, it does not seem that the intra-method analytical imprecision contributes to these differences, since the precision performances declared by the individual manufacturer are essentially overlapping (Table 1). In our opinion, these differences could lie in the different coating preparations of the TPO antigen (purified native or recombinant) on solid phase, which affect the proper exposure of the immunodominant epitopes recognized by the polyclonal antibodies present in serum of AITD patients (epitopic fingerprint), with the consequent lack of recognition of some of them [24-26]. Another important result to be noted was the difference between genders: medians were not statistically significant different between males and females for IMM, KRY, MAG, and LIA, while they were different for PHA. Based on these results, we suggest to overcome the proposal of NACB guidelines [11], which recommend to involve a single group of young male subjects in favor of using two distinct groups, male and female ones. This new proposal removes the apparent incongruity of using a male reference group for a disease (such as AITD) affecting mainly females. Moreover, in spite of the harmonization among methods, the wide dispersion of quantitative results, still observed in this study, suggests the need for further studies to better understand the cause of these discrepancies, focusing on TPO antigen preparation as the possible source of variability among different assays.
  23 in total

1.  Laboratory medicine practice guidelines. Laboratory support for the diagnosis and monitoring of thyroid disease.

Authors:  Zubair Baloch; Pierre Carayon; Bernard Conte-Devolx; Laurence M Demers; Ulla Feldt-Rasmussen; Jean-François Henry; Virginia A LiVosli; Patricia Niccoli-Sire; Rhys John; Jean Ruf; Peter P A Smyth; Carole A Spencer; Jan R Stockigt
Journal:  Thyroid       Date:  2003-01       Impact factor: 6.568

2.  Performance characteristics of 5 automated thyroglobulin autoantibody and thyroid peroxidase autoantibody assays.

Authors:  Sonia L La'ulu; Patricia R Slev; William L Roberts
Journal:  Clin Chim Acta       Date:  2006-07-25       Impact factor: 3.786

Review 3.  Autoimmune diagnostics: the technology, the strategy and the clinical governance.

Authors:  Nicola Bizzaro; Renato Tozzoli; Danilo Villalta
Journal:  Immunol Res       Date:  2015-02       Impact factor: 2.829

Review 4.  Is vitamin D a player or not in the pathophysiology of autoimmune thyroid diseases?

Authors:  Federica D'Aurizio; Danilo Villalta; Paolo Metus; Paolo Doretto; Renato Tozzoli
Journal:  Autoimmun Rev       Date:  2014-10-12       Impact factor: 9.754

5.  Immunoassay of thyroid peroxidase autoantibodies: diagnostic performance in automated third generation methods. A multicentre evaluation.

Authors:  Federica D'Aurizio; Renato Tozzoli; Danilo Villalta; Giampaola Pesce; Marcello Bagnasco
Journal:  Clin Chem Lab Med       Date:  2015-02       Impact factor: 3.694

6.  Experience with the first fully automated chemiluminescence immunoassay for the quantification of 1α, 25-dihydroxy-vitamin D.

Authors:  Josef van Helden; Ralf Weiskirchen
Journal:  Clin Chem Lab Med       Date:  2015-04       Impact factor: 3.694

7.  [A new technique for measuring 17beta-estradiol using Kryptor (Cis Bio International): utilization to monitor ovulation stimulation].

Authors:  J Taïeb; C Sarnel; C Benattar; A Lindenbaum
Journal:  Ann Biol Clin (Paris)       Date:  2000 Jan-Feb       Impact factor: 0.459

8.  Autoantibodies to thyroperoxidase (TPOAb) in a large population of euthyroid subjects: implications for the definition of TPOAb reference intervals.

Authors:  Klaus Zöphel; Bernhard Saller; Gerd Wunderlich; Thomas Grüning; Rainer Koch; Jürgen Wilde; Klaus Mann; Wolf-Gunter Franke
Journal:  Clin Lab       Date:  2003       Impact factor: 1.138

9.  Thyroid autoimmunity in schoolchildren in an area with long-standing iodine sufficiency: correlation with gender, pubertal stage, and maternal thyroid autoimmunity.

Authors:  Irene Kaloumenou; George Mastorakos; Maria Alevizaki; Leonidas H Duntas; Emilia Mantzou; Charalambos Ladopoulos; Aristides Antoniou; Dimitrios Chiotis; Ioannis Papassotiriou; George P Chrousos; Catherine Dacou-Voutetakis
Journal:  Thyroid       Date:  2008-07       Impact factor: 6.568

10.  Reference intervals for thyroid markers in early pregnancy determined by 7 different analytical systems.

Authors:  Drahomira Springer; Vladimir Bartos; Tomas Zima
Journal:  Scand J Clin Lab Invest       Date:  2014-03       Impact factor: 1.713

View more
  4 in total

Review 1.  Challenges in the Standardization of Autoantibody Testing: a Comprehensive Review.

Authors:  Renato Tozzoli; Danilo Villalta; Nicola Bizzaro
Journal:  Clin Rev Allergy Immunol       Date:  2017-08       Impact factor: 8.667

2.  Definition of the upper reference limit for thyroglobulin antibodies according to the National Academy of Clinical Biochemistry guidelines: comparison of eleven different automated methods.

Authors:  F D'Aurizio; P Metus; A Ferrari; B Caruso; R Castello; D Villalta; A Steffan; K Gaspardo; F Pesente; N Bizzaro; E Tonutti; S Valverde; C Cosma; M Plebani; R Tozzoli
Journal:  Auto Immun Highlights       Date:  2017-06-19

3.  Data on characterization of glass biochips and validation of the label-free biosensor for detection of autoantibodies in human serum.

Authors:  A V Pushkarev; A V Orlov; S L Znoyko; D O Novichikhin; V A Bragina; A A Sizikov; E Alipour; H Ghourchian; A I Nikitin; G M Sorokin; B G Gorshkov; P I Nikitin
Journal:  Data Brief       Date:  2020-04-30

4.  No Effect of Thyroid Dysfunction and Autoimmunity on Health-Related Quality of Life and Mental Health in Children and Adolescents: Results From a Nationwide Cross-Sectional Study.

Authors:  Raphael Hirtz; Anne Keesen; Heike Hölling; Berthold P Hauffa; Anke Hinney; Corinna Grasemann
Journal:  Front Endocrinol (Lausanne)       Date:  2020-09-02       Impact factor: 5.555

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.