M Shehata1, Amany M Fekry1,2, Alain Walcarius2. 1. Chemistry Department, Faculty of Science, Cairo University, Giza 12613, Egypt. 2. LCPME, Université de Lorraine, CNRS, F-54000 Nancy, France.
Abstract
It appeared that either the carbon paste or the screen-printed carbon electrodes that were modified with gold nanoparticles (AuNPs) gave rise to the largest current responses after a rapid screening of various nanomaterials as modifiers of carbon composite electrodes in view of designing an electrochemical sensor for Moxifloxacin Hydrochloride (Moxi). The screen-printed electrode (SPE) support was preferred over the carbon paste one for its ability to be used as disposable single-use sensor enabling the circumvention of the problems of surface fouling encountered in the determination of Moxi. The response of AuNPs modified SPE to Moxi was investigated by cyclic voltammetry (CV) (including the effect of the potential scan rate and the pH of the medium), chronoamperometry, and differential pulse voltammetry (DPV) after morphological and physico-chemical characterization. DPV was finally applied to Moxi detection in phosphate buffer at pH 7, giving rise to an accessible concentration window ranging between 8 µM and 0.48 mM, and the detection and quantification limits were established to be 11.6 µM and 38.6 µM, correspondingly. In order to estimate the applicability of Moxi identification scheme in actual trials, it was practiced in a human baby urine sample with excellent recoveries between 99.8 % and 101.6 % and RSDs of 1.1-3.4%, without noticeable interference.
It appeared that either the carbon paste or the screen-printedcarbon electrodes that were modified with gold nanoparticles (AuNPs) gave rise to the largest current responses after a rapid screening of various nanomaterials as modifiers of carbon composite electrodes in view of designing an electrochemical sensor for Moxifloxacin Hydrochloride (Moxi). The screen-printed electrode (SPE) support was preferred over the carbon paste one for its ability to be used as disposable single-use sensor enabling the circumvention of the problems of surface fouling encountered in the determination of Moxi. The response of AuNPs modifiedSPE to Moxi was investigated by cyclic voltammetry (CV) (including the effect of the potential scan rate and the pH of the medium), chronoamperometry, and differential pulse voltammetry (DPV) after morphological and physico-chemical characterization. DPV was finally applied to Moxi detection in phosphate buffer at pH 7, giving rise to an accessible concentration window ranging between 8 µM and 0.48 mM, and the detection and quantification limits were established to be 11.6 µM and 38.6 µM, correspondingly. In order to estimate the applicability of Moxi identification scheme in actual trials, it was practiced in a human baby urine sample with excellent recoveries between 99.8 % and 101.6 % and RSDs of 1.1-3.4%, without noticeable interference.
Moxifloxacin Hydrochloride (Moxi) (1-cyclopropyl-6-fluoro-1, 4-dihydro-8-methoxy-7-[(4aS, 7aS)-octahydro-6H-pyrrolo [3, 4-b] pyridin-6-yl] 4-oxo-3 quinoline carboxylic acid) is an antimicrobial fluoroquinolone (antibiotics family) of low toxicity and antimicrobial [1,2] versus bacteria, both gram (+ve) and gram (−ve) [3]. It is mainly mentioned in medical treatments [4,5]. Examining of pharmaceutical molecules, like Moxi, is an important request [6], so that several conventional protocols were operated for this purpose, such as HPLC [7,8], HPLC-UV [9], capillary electrophoresis [10], fluorimetric [11], and chemiluminescence [12]. These techniques were sensitive, but they are usually expensive and time consuming, and they may require quite sophisticated protocols [13]. It is the reason why electrochemical approaches for Moxi detection were proposed as possible alternatives for overcoming these disadvantages, thanks to their simplicity, rapidity, and good analytical performance. Previous examples in the literature are based on various electrodes, like glassy carbon electrode [14], carbon paste electrodes (CPEs) [15,16], and/or molecularly imprintedpolymers on electrodes [17].Heterogeneous carbon-based electrodes (CPEs or carbon screen-printed electrodes, SPEs) and their modified forms are suitable for the determination of different substances due to their fairly inert electrochemistry, negligible background current, easily renewable surface (CPEs), or possible mass production as single-use sensors (SPEs), and simple preparation [18]. The modifiedCPE or SPE electrodes result in low detection limits, high sensitivities, and selectivity via accumulation or electrocatalysis or both of them [19,20].In particular, metalnanoparticles and other nanomaterials offer attractive features that can be exploited for electrochemical sensing of biologically-active compounds [21], notably based on modifiedcarbon electrodes. For instance, gold nanoparticles have been utilized in designing various types of sensors, owing to their unique structure and exceptional features, like good biocompatibility, high conductivity, high surface to volume ratio, and adsorption properties [22]. Graphene oxide (GO) has a large surface area, strong mechanical strength, good chemical stability, high conductivity, and excellent catalytic properties due to bearing two-dimensional plane rich with oxygen containing functional groups [23]. GO will be used here with Glutathione, which contains eight active binding sites to act as strong electron donors and antioxidant materials [24,25]. Ag nanoparticles have been utilized in catalysis and electrochemistry due to their great electrical conductivity and stability [26,27]. Ag nanoparticles can be mixed with cellulose acetate phthalate (also utilized in several drug delivery procedures as being nontoxic in human body usages [28]) or chitosan (promising for various biomedical applications [29]) to improve biocompatibility. Nano-Co(II, III) oxides have been significantly applied for electrochemical energy storage devices, electrocatalysis, biosensors, and electrochemically sensing devices [30,31,32,33,34,35,36], because of their high chemical stability, biocompatibility, and huge specific surface areas [37,38,39,40], which permit easy access to a large amount of active sites. Zeolites are promising materials with ion-exchange properties that can be exploited as modifying agents for electrochemical sensors [41,42]. Finally, CarbonNanotubes (CNTs) exhibit fascinating properties [43,44], such as excellent electronic characteristics, extremely large and active surface areas, and intrinsic electrocatalytic features, which make them unique candidates for electrode modification and sensors development [45,46].All of these modifiers are possibly interesting for Moxi detection, owing to their preconcentration ability and/or electrocatalytic properties and the first objective of this work was to evaluate them in order to select the most promising one. After a rapid screening by cyclic voltammetry (CV) utilizing carbon composite electrodes modified with the above components, the best system (i.e., based on gold nanoparticles, AuNPs, thanks to their unique structure and high conductivity, high surface to volume ratio, and adsorption properties) was selected and investigated for the electrochemical sensing of Moxi at AuNPs modified screen-printedcarbon electrodes. The modified electrodes were examined by various physico-chemical techniques and their electrochemical behavior was characterized by several electrochemical methods prior to Moxi detection by differential pulse voltammetry (DPV). The figures of merit were determined and the best sensors enabled the identification of Moxi in urine samples.
2. Materials and Methods
2.1. Chemicals and Solutions
The analyte (Moxi) was obtained from the National Organization for Drug Control and Research (Giza, Egypt). The graphite powder and paraffin oil that were utilized to build on the carbon paste (CP) were provided by Merck (Darmstadt, Germany), and the carbon ink used to fabricate the screen-printed electrodes was purchased from Alfa Aesar. Gold nanoparticles were prepared from a gold chloride salt (KAuCl4). Other electrode modifiers were Graphene Oxide (GO) (15–20 sheets, 4–10% edge oxidized), Ag nanoparticles, Nano-Co (II, III) oxide (NCO), CNTs (multi-walled, 3–20 nm OD, 1–3 nm ID, 0.1–10 micron long), l-reducedGlutathione, Cellulose acetate phthalate and chitosan powder (Aldrich, USA), and Zeolite Y (from Alfa Aesar). Britton Robinson (B-R) buffer solutions were prepared from CH3COOH, H3BO3, and H3PO4 (total concentration: 4.0 × 10−2 M), at pH values ranging from 2 to 9 (as adjusted by using 0.2 M NaOH) [47,48]. All of the solutions were prepared utilizing triply distilledwater.
2.2. Preparation and Modification of Electrodes
2.2.1. Carbon Paste Electrodes (CPEs)
The bare carbon paste electrode (BCPE) was fabricated from mixing 5.0 g of graphite powder with 2.0 ml of paraffin oil into mortar for 10 min. to obtain a homogenous paste. The modifiedCPEs were prepared through several trials by adding selected amounts of different modifiers to the paste and hand blending the resulting mixtures for 10 min. more and checking their sensitivity toward Moxi in order to achieve the optimum sensor composition in each case. On that basis, 0.05 g of zeolite powder and 0.01 g of CNTs were used to obtain zeolite/carbonnanotubes modifiedCPE (ZCNTMCPE), 0.05 g of NCO for the nano-Co (II, III) oxide modifiedCPE (NCOMCPE), 0.05 g of cellulose acetate phthalate powder +0.05 g of chitosan and 0.05 g of silvernanoparticles for the silvernanoparticles/chitosan/cellulose acetate phthalate modifiedCPE (SCCMCPE), 0.03 g of GO, and 0.05 g reducedglutathione for the graphene oxide/reducedglutathione modifiedCPE (GORGMCPE). Each CP mixture was then packed into the end of Teflon tube (with diameter of 3 mm, corresponding to 7 mm2 electrode surface area) to prepare the corresponding sensors. The excess of composite material was removed and the electrode surface was smoothed while using ultrafine emery paper. Stainless steel ensured the electrical contact. Finally, the gold nanoparticles modifiedCPE (GNMCPE) was prepared by immersing a BCPE in 0.01 M of gold chloride solution and applying a deposition potential of −0.4 V for 300 s in order to deposit a nano-Au film, which was then air-dried for 5 min.
2.2.2. Screen-Printed Carbon Electrodes (SPEs)
Screen-printedcarbon electrodes (SPEs) were fabricated through ink screen-printing over the ceramic supports, with a silver reference electrode and carbon counter electrode. The PVC insulator was handled to cover the electrodes after being dried for half an hour at 60 °C [49]. BSPE (bare screen-printed electrode) was modified with gold films by means of an electrodeposition method in order to prepare gold nanoparticles modifiedSPE (GNMSPE). At potential of −0.4 V for 300 s [50], gold film was deposited over SP (screen-printed) carbon surface from 6 mM gold chloride solution (pH = 2.0) and left in air to dry for five minutes in order to be subsequently utilized at the electrochemical measurements. The geometrical surface area of the working electrode was 5 mm2.
2.3. Methods and Instrumentation
The electrochemical experiments were carried out using a 25 mL three-electrode cell comprising a platinumauxiliary electrode (CE), the modifiedCPE or SPE sensors as working electrode (WE), and a saturated calomel electrode (SCE) serving as reference electrode (RE) [51]. Differential Pulse Voltammetry (DPV), Cyclic Voltammetry (CV), Chronoamperometry (CA), and Electrochemical Impedance Spectroscopic (EIS) measurements were performed with the Bio-logic SAS model SP-150 potentiostat monitored by a computer-controlled EC-Lab® electrochemical software. The EIS tests were made at 10 mV ac amplitude in the frequency range of 1.0 mHz to 100 kHz [52] and the equivalent circuit models were built from the EC-Lab® software [50,51,52,53]. Adwa 1030 digitalized pH meter (Romania) that was linked to a glass electrode was used to adjust the solution pH.The materials and electrodes were characterized by various physico-chemical techniques. Surface morphology was observed by scanning electron microscopy (SEM) while using the Model Quanta 250 Field Emission Gun apparatus coupled with an energy dispersive analysis of X-rays (EDX) unit (FEI Company, Netherlands). Transmission electron microscopy (TEM) analysis was made using a JEM-1400 Electron Microscope (JEOL, Japan). The PAN-analytical X-Ray Diffraction equipment model X′Pert PRO with secondary monochromator, operating at 45 kV and 35 mA with Cu-radiation (λ = 1.542 Å) and scanning speed of 0.04° s-1, was used for structural investigations. The diffraction lines located at 2θ values between 2° and 60°, the corresponding spacing (d, Å), and relative intensities (I/Io), were obtained. The diffraction charts and relative intensities are obtained and compared with ICDD files.
3. Results and Discussion
3.1. Preliminary Observations and Surface Characterization of the AuNPs Modified Screen-Printed Electrodes
Carbon paste electrodes modified with various nanomaterials were first prepared and characterized (see Figure A1 and Figure A2 in Appendix A) and screened by cyclic voltammetry as a test to evidence the most appropriate additive to the carbon composite electrode for the electrochemical detection of Moxi (see Figure A3 in Appendix B). When investigated in 1 mM Moxi solution (buffered at pH 7.0), from CV, EIS and calibration data (see respectively parts A, B, and C in Figure A3 in Appendix B), and the best performance (most sensitive detection) was obtained with the composite carbon electrodes covered with gold nanoparticles. The screen-printed electrode modified with gold nanoparticles (GNMSPE) was selected here for the detection of Moxi and its surface characterization was achieved while using various techniques (Figure 1).
Figure A1
EDX spectra and SEM micrographs (insets) of (A) BCPE, (B) NCOMCPE, (C) GORGMCPE, (D) ZCNTMCPE, (E) SCCMCPE and (F) GNMCPE with its histogram, respectively.
Figure A2
EDX mapping pictures of (A) NCOMCPE, (B) GORGMCPE, (C) ZCNTMCPE, (D) SCCMCPE and (E) GNMCPE, respectively.
Figure A3
(A) CVs of 1.0 mM Moxi in 0.01 M B-R (pH = 7.0) and scan rate 50 mV s−1 measured on the various investigated electrodes as presented in the experimental section. (B) The corresponding Nyquist plots for the studied sensors (bottom right inset: the best fitted equivalent circuit). (C) The analytical curves obtained from performing DPV measurements at scan rate 10 mV s−1 with sensing platforms ZCNTMCPE, NCOMCPE, SCCMCPE, GORGMCPE, GNMCPE and GNMSPE.
Figure 1
Scanning electron microscopy (SEM) micrographs of the screen-printed electrode surfaces (A: screen-printed electrode (SPE) and B: gold nanoparticles modified SPE (GNMSPE) accompanied with its histogram); (C) transmission electron microscopy (TEM) micrograph of Au nanoparticles; and, (D) XRD pattern of GNMSPE.
SEM, TEM, and XRD characterizedGNMSPE (Figure 1). Comparing the SPE surface observed before (Figure 1A) and after gold electrodeposition (Figure 1B) indicates the successful formation of gold nanoparticles well dispersed onto the electrode surface, with spherical shape (and sizes of 13–58 nm, as evaluated from TEM examination, see Figure 1C). As compared to the GNMCPE (Figure A1F), it seems that the density of gold NPs is slightly larger on GNMSPE (Figure 1B). The XRD pattern of the GNMSPE shows sharp diffraction lines at 26.6°, 38.1°, and 42.1° (Figure 1D). The two sharp diffraction lines located at 2θ = 38.1° (with d-spacing of 2.6 Å) and 42.1° (with d-spacing of 2.0 Å) correspond to the Aunanoparticles, while the third one that is located at 2θ = 26.6° (with d-spacing of 3.36 Å) is attributed to carbon, the other main constituent of the modified electrode. No additional lines originating from any other crystalline elements can be noticed, which indicates the high purity of the depositedAu-nanoparticles and that their good crystallinity is supported by the sharpness of diffraction lines. The definite line broadening of the peaks suggests that Au-particles are in the nanometer size range.
3.2. Parameters Affecting the Detection, Optimization and Analytical Performance
3.2.1. Influence of Potential Scan Rate
Figure 2 illustrates the effect of potential scan rate on the voltammetric real">sponse of Moxi at GNMSPE. The anodic peak currents increased with rising the scan rate (ν = 10–200 mV/s, see Figure 2A), giving a linear variation with respect to the square root of scan rate (see inset in Figure 2A). The linear regression equations for both the bare and GNMSPE electrodes are respectively: Ip (μA) = −10.66 + 5.39 ν1/2 (mV s−1) (r2 = 0.956) and Ip (μA) = 14.40 + 9.75 ν1/2 (mV s−1) (r2 = 0.960). This indicates that the oxidation process of Moxi is under diffusion control. It is also noticeable that the Moxi signals are shifted to the positive direction while increasing the potential scan rates (Figure 2B), confirming the irreversibility of the electron transfer reaction [54].
Figure 2
(A) Cyclic voltammetry’s (CVs) of 1.0 mM Moxi on GNMSPE recorded at various potential scan rates (v). Inset: variations of anodic peak current (Ipa) with the square root of scan rate (v1/2) for BSPE and GNMSPE. (B) Variation of the anodic signal potential of GNMSPE with the logarithm of scan rate.
3.2.2. pH Effect and Sensor Stability
The effect of pH on Moxi oxidation was evaluated in B-R buffer (pH 2.0–9.0) at BSPE and GNMSPE, respectively (Figure 3A). The optimum pH corresponding to the largest currents was 7.0 for both electrodes, but the signal intensity was much larger on the modified electrode (in agreement with the results of Figure A3A in Appendix B). The best performance at neutral pH was also evidenced from impedance data (Figure 3B). Though not easily measurable due to a peak-to-wave shape, the voltammetric signals shifted to lower potential values at higher pH, with a slope change around pH 6, consistent with the Moxi pKa value of 6.25 [55]. Such behavior also agrees with the transfer of protons that are involved in the electrochemical oxidation of structurally related compounds [56].
Figure 3
(A) CVs of 1.0 mM Moxi on GNMSPE in 0.01 M B-R buffer with different pH values at scan rate 50 mV s−1. Inset: variation of anodic peak current (Ipa) for BSPE and GNMSPE as function of pH. (B) Nyquist plots of Moxi on GNMSPE at different pH values. (C) Variation of anodic peak current (Ipa) obtained for successive analyses at increasing times (1–30 min.) using the same GNMSPE in B–R buffer pH 7.0 containing 1.0 mM Moxi. Inset: the corresponding CVs.
The stability of GNMSPE was tested in B-R buffer solution comprising 1 mM of the analyte through CV technique (Figure 3C). It was demonstrated that the signal intensity decreased rapidly with lengthening the immersion time of sensor: starting around 100 µA, the signal intensity dropped by one half of its value after only one more CV cycle, to level down to about one-third of its initial value after some minutes of successive analyses. This is probably due to the irreversible adsorption of the Moxi oxidation products onto the electrode surface and/or the poisoning of the surface active sites (Aunanoparticles) by chelation with Moxi [15,57], justifying the need for a single-use sensor, as GNMSPE, thanks to the screen-printing technology to produce such electrodes.
3.2.3. Chronoamperometry
Figure 4A shows chronoamperograms that were recorded at a constant poten<span class="Chemical">tial of +1.03 V vs. SCE utilizing GNMSPE in BR buffer (pH = 7.0) containing various Moxi concentrations.
Figure 4
(A) Chronoamperograms at GNMSPE for different concentration of Moxi (0.0–1.0 mM). Insets: Variations of I vs. t−1/2 built from chronoamperograms (right) ant plot of the corresponding slopes against Moxi concentration (left). (B) Variation of Icat/If ratio versus t1/2.
Utilizing the Cottrell Equation (1), the diffusion coefficient (D) can be evaluated from plotting I versus t−1/2 [58] (see right inset in Figure 4A).
where i = current, in unit A, F = Faraday constant (96485 C/mol), n = number of electrons, A = area of the electrode in cm2, c0 = bulk concentration of the analyte in mol/cm3; t = time in s, and D = diffusion coefficient for species in cm2/s.A calibration curve can be obtained as a linear relationship between different Moxi concentrations (0 to 1.0 mM) and the slopes obtained from the current relation versus t−1/2 (Figure 4A, left inset), leading to the calculation of the diffusion coefficient, D value for Moxi equal to 3.57 × 10−6 cm2 s−1. The catalytic rate constant kcat (mol–1 L–2 s−1) was obtained from the relation (2) [59]:
with Icat/If being the ratio between the currents sampled in the presence (Icat) or absence of Moxi (If), C0 the bulk concentration of Moxi (mol L−1), and t the elapsedtime (s).The mean catalytic rate constant was evaluated from the slope of Icat/If versus t1/2 plot (Figure 4B), and a value of kcat = 7.91 × 103 mol−1 L−2 s−1 was calculated for a Moxi concentration of 1.0 mM. The high Kcat value obtained indicates quite fast charge transfer processes between the analyte and the modified electrode surface and supports the interest of AuNPs for Moxi determination.
3.3. Calibration Plot and Detection Limit
Differential pulse voltammetry (DPV) was used as sensitive technique for building the calibration plot for Moxi (Figure 5). As shown, a linear variation of peak currents with Moxi concentration was obtained in the 8.0–480 µM range (in B-R buffer, pH 7.0): Ip (µA)= 2.38 + 0.026 C (µM), r2 = 0.984. The limit of detection (LOD = 3 σ/m) and limit of quantification (LOQ = 10 σ/m), where, σ is the standard deviation and m is the calibration plot slope were obtained as 11.6 µM and 38.6 µM, respectively. Comparing the obtaineddata for Moxi determination by GNMSPE against others analytical or methods [7,60,61] or electrochemical techniques [12,13,14] shows acceptable performance of the proposed novel sensor, offering opportunities for on-site analyses thanks to the possible mass production of low-cost screen-printed electrodes and the suggested electrochemical sensor for Moxi detection also compared with others [17,62,63,64,65,66,67] in terms of detection limits, as shown in Table 1.
Figure 5
Calibration plot for Moxi, Inset: corresponding differential pulse voltammetry’s (DPVs) for successive addition of Moxi in B-R buffer solution of pH 7.0 at scan rate = 10 mV s−1 using GNMSPE and illustrative picture of a typical SPE used in this work.
Table 1
Comparison of the analytical figures of merit for the present method with others approaches for the determination of Moxi.
Bifunctional monomer molecularly imprinted film at graphene modified glassy carbon electrode
Voltammetric
5.12 × 10−10
1.0 × 10−9–5.0 × 10−5
[63]
Carbon paste modified with silver nanoparticles
Electrochemical
2.9 × 10−9
7.0 × 10−7–1.8 × 10−4
[15]
Molecularly imprinted polymer
Potentiometric based PVC membranes
1.7 × 10−6
1.0 × 10–5–1.0 × 10−2
[64]
New coated platinum selective electrode
Potentiometric titration
9.2 × 10−6
1.0 × 10-5–1.0 × 10−3
[65]
A green HPLC assay method
HPLC
0.23 µg/mL
2.0–40.0 µg/mL
[66]
molecularly imprinted polymer
PLC-MS/MS method
0.03 µg/mL
0.2–1.2 µg/mL
[67]
GNMSPE
Electrochemical
11.6 × 10−6
8 × 10−6–0.48 × 10−3
This work
3.4. Real Sample Analysis, Interfering Species and Simultaneous Determination Test
A calibration plot was established utilizing the GNMSCPE in a diluted urine medium, giving a linear correlation over a concentration range of 20–450 µM, in order estimate the method applicability to real sample analysis (Figure 6). The linear regression equation that was obtained from the variation of DPV peak currents as a function of the Moxi concentration s was: Ipa (µA) = 0.028C (µM) + 3.84, r2 = 0.98). The LOD (3σ) was 10.7 µM, and the LOQ (10σ) was 35.6 µM. Four different concentrations were chosen on the calibration plot through standard addition method for the examination of Moxi in buffered solution of pH 7.0, and each measurement was repeated five times in order to evaluate the accuracy and precision of the approach, as represented in Table 2. The detected concentrations of Moxi were always very near to the expected values, giving acceptable recoveries for all samples between 99.8 % and 101.6 % with RSDs of 1.1–3.4 %.
Figure 6
Calibration plot of Moxi in a urine sample prepared in B-R buffer solution (pH 7.0) using GNMSPE. Inset A: separate detection of 200 µM Moxi with 200 µM paracetamol by DPV in the same solution. Inset B: I) calibration plot for Moxi in the presence of 200 µM paracetamol, and II) calibration for paracetamol in the presence of 200 µM Moxi. Inset C: detection of 200 µM Moxi alone or in the presence of 200 µM ascorbic acid, dopamine, uric acid or estradiol).
Table 2
Recovery and standard deviation data for Moxi detection in urine sample with GNMSPE.
Spiked Moxi (µM)
Detected Moxi 1 (µM)
Av. Recovery 1 (%)
RSD 2 (%)
25.0
25.4
101.6
1.1
75.0
74.9
99.8
3.4
150.0
149.8
99.9
1.9
300.0
300.6
100.2
2.6
1 Average values from 5 independent measurements (n = 5). 2 Relative Standard Deviation.
In order to investigate the interference effects for other compounds that can co-exist with Moxi in urine samples such as uric acid (UA), ascorbic acid (AA), dopamine (DA) and hormones like estradiol (ED), a fixed amount of Moxi (200 µM) spiked with the same concentration of these substrates and evaluated under the same experimental conditions through DPV. The results (see bottom inset C in Figure 6) indicate the negligible effect of interfering substrates (i.e., 0.8%, 1.8%, 3.7%, and 5.2% signal intensity decrease for ascorbic acid, dopamine, uric acid, and estradiol, respectively), which ensures good sensor selectivity. However, it can be anticipated that structurally relatedfluoroquinolones could interfere and the sensor would be not selective to Moxi in this case.Typically, a patient suffering from dry cough and high fever rate should receive 200 mg moxifloxacin orally each half a day and 500 mg paracetamol when necessary [68], justifying the need for simultaneous detection. Hence, the immediate identification of 200 µM Moxi with 200 µM paracetamol in pH 7.0 utilizing GNMSPE was achieved (see top left inset A in Figure 6). The curve demonstrates a sufficient separation between the two signals, i.e. paracetamol (at 0.49 V) and Moxi (at 0.98 V), with ΔE of 0.49 V, enabling the possible simultaneous detection of both Moxi and paracetamol in the same medium. To support these findings, a calibration for Moxi in the presence of a constant concentration of paracetamol and, the opposite, a calibration for paracetamol in the presence of a constant concentration of Moxi, were made (see right bottom inset B in Figure 6), confirming the interest of GNMSPE for the simultaneous detection of these two pharmaceuticals.
3.5. Reproducibility and Long-Term Stability
Five distinct electrodes were tested for the voltammetric determination of 50 µM Moxi without obvious dispersion of the data (RSD value of 2.3%), which ensures a preparation process highly trustworthy, in order to validate the reproducibility of GNMSPE in terms of RSD.The long-term stability of the proposed electrode was evaluated through storing GNMSPE in a refrigerator (4 ℃) for seven days. After that, a single voltammetric measurement for Moxi was carried out, revealing a current response at 95% of the value measured directly after fresh preparation, indicating good storage stability of the electrode.
4. Conclusions
This work brings together a comparison between several modifiers of heterogeneous carbon electrodes, pointing out the interest of gold nanoparticles on carbon paste or screen-printedcarbon electrodes for the electrochemical sensing of Moxi. SPE that was modified with simple electrochemically depositedgold nanoparticles, GNMSPE, showed a high sensitivity and applicability for detecting Moxi in aqueous solutions and urine samples under physiological conditions with acceptable LOD (11.6 µM), wide linear concentration range (8–480 µM), and excellent recoveries (ranging between 99.8% and 101.6%) and RSDs (1.1–3.4%, n = 5). Sensor optimization was achieved from data that were collected with the aid of numerous electrochemical techniques (CV, DPV, EIS, and CA). The GNMSPE electrochemical sensor can be also applied to Moxi determination in the presence of paracetamol.