| Literature DB >> 32399647 |
Delphine Vallot1, Elena De Ponti2, Sabrina Morzenti2, Anna Gramek3, Anna Pieczonka4, Gabriel Reynés Llompart5, Jakub Siennicki6, Paul Deak6, Chiranjib Dutta6, Jorge Uribe6, Olivier Caselles7.
Abstract
INTRODUCTION: Quantitative imaging biomarkers are becoming usual in oncology for assessing therapy response. The harmonization of image quantitation reporting has become of utmost importance due to the multi-center trials increase. The NEMA image quality test is often considered for the evaluation of quantitation and is more accurate with a radioactive solid phantom that reduces variability. The goal of this project is to determine the level of variability among imaging centers if acquisition and imaging protocol parameters are left to the center's preference while all other parameters are fixed including the scanner type.Entities:
Keywords: 18F-FDG NEMA body phantom; 68Ge solid phantom; Multi-center harmonization; PET/CT; quantitation
Year: 2020 PMID: 32399647 PMCID: PMC7218035 DOI: 10.1186/s40658-020-00294-y
Source DB: PubMed Journal: EJNMMI Phys ISSN: 2197-7364
Fig. 1a Scheme of the IEC61675/NEMA NU2-2012 Image Quality Body Phantom (in black the cold spheres and in gray the hot spheres) and b a picture of the QUICK phantom
Acquisition and reconstruction parameters for the standard NEMA test with 18F-FDG phantom
| Auto mA modulation | |
Summary of the test format (overview) realized with the QUICK phantom in each QUICK site. The rows denote the acquisition parameters (NEMA protocol/clinical protocol for the particular site), and the columns represent the reconstruction parameters (NEMA protocol/clinical protocol for the particular site) used in the test
| Reconstruction parameters | |||||
|---|---|---|---|---|---|
| NEMA | Clinical | ||||
| OSEM | BSREM ( | OSEM | BSREM | ||
Acquisition and reconstruction parameters for each test
| 256 | 256 | 256 | |
| 40 | 40–70 | 40–70 | |
| OSEM | OSEM | OSEM | |
| 8 | 4–6 | 4–6 | |
| 12 | 12 | 12 | |
| 2 | 4.8–6.4 | 4.8–6.4 | |
| None | Standard (weight 1 4 1) | Standard (weight 1 4 1) | |
| CTAC, scatter, randoms | CTAC, scatter, randoms | CTAC, scatter, randoms | |
| No | Yes/no | Yes/no | |
| 256 | 256 | 256 | |
| 40 | 40–70 | 40–70 | |
| BSREM | BSREM | BSREM | |
| 25 | 25 | 25 | |
| 25 | 350 | 350 | |
| CTAC, scatter, randoms | CTAC, scatter, randoms | CTAC, scatter, randoms | |
| Yes | Yes | Yes |
Fig. 2CR (%) values for the standard NEMA test performed with 18F-FDG phantom
Fig. 3BV (%) values for the standard NEMA test performed with an 18F-FDG phantom
Fig. 4Lung error (LE in %) values for the standard NEMA test performed with 18F-FDG phantom
CR, BV, and LE for the six tests performed with QUICK phantom as average ± standard deviation of datasets from all QUICK sites combined [17]
| NEMA ACQ and RECON OSEM (test 1) | NEMA ACQ and RECON BSREM (test 2) | NEMA ACQ and clinical RECON OSEM (test 3) | NEMA ACQ and clinical RECON BSREM (test 4) | Clinical ACQ and RECON OSEM (test 5) | Clinical ACQ and RECON BSREM (test 6) | |
|---|---|---|---|---|---|---|
| Sphere | ||||||
| 10 mm | 29.4 ± 2.0 | 39.9 ± 2.5 | 17.9 ± 1.6 | 17.6 ± 2.4 | 18.0 ± 2.1 | 18.8 ± 1.1 |
| 13 mm | 42.9 ± 1.6 | 53.5 ± 1.1 | 30.8 ± 3.0 | 34.1 ± 3.4 | 30.7 ± 3.0 | 36.1 ± 2.2 |
| 17 mm | 52.9 ± 1.6 | 61.4 ± 1.5 | 45.0 ± 4.3 | 49.5 ± 3.1 | 43.8 ± 4.4 | 50.2 ± 1.6 |
| 22 mm | 60.5 ± 1.2 | 67.5 ± 1.4 | 54.8 ± 4.5 | 60.7 ± 2.6 | 54.9 ± 4.3 | 61.6 ± 0.5 |
| 28 mm | 67.9 ± 0.7 | 77.7 ± 1.0 | 59.3 ± 1.6 | 66.8 ± 2.4 | 59.3 ± 1.7 | 67.5 ± 1.5 |
| 37 mm | 74.6 ± 0.4 | 83.1 ± 0.6 | 66.6 ± 2.3 | 75.9 ± 1.8 | 66.9 ± 1.7 | 76.9 ± 1.1 |
| Sphere | ||||||
| 10 mm | 7.4 ± 2.1 | 8.9 ± 0.6 | 4.3 ± 1.8 | 4.0 ± 1.2 | 4.7 ± 1.8 | 4.8 ± 1.6 |
| 13 mm | 5.9 ± 1.6 | 7.1 ± 0.5 | 3.8 ± 1.6 | 3.5 ± 1.1 | 4.2 ± 1.6 | 4.2 ± 1.4 |
| 17 mm | 4.6 ± 1.2 | 5.4 ± 0.4 | 3.2 ± 1.3 | 3.0 ± 0.8 | 3.6 ± 1.4 | 3.5 ± 1.2 |
| 22 mm | 3.6 ± 0.9 | 4.1 ± 0.3 | 2.8 ± 1.1 | 2.4 ± 0.6 | 3.1 ± 1.2 | 2.9 ± 0.9 |
| 28 mm | 2.8 ± 0.7 | 3.1 ± 0.3 | 2.3 ± 0.9 | 2.0 ± 0.4 | 2.6 ± 0.9 | 2.3 ± 0.7 |
| 37 mm | 2.2 ± 0.5 | 2.4 ± 0.2 | 2.0 ± 0.7 | 1.7 ± 0.4 | 2.3 ± 0.8 | 1.9 ± 0.5 |
| 16.4 ± 0.5 | 12.2 ± 0.6 | 23.1 ± 1.4 | 14.3 ± 1.1 | 22.6 ± 2.5 | 13.1 ± 1.3 | |
Fig. 5Contrast recovery and background variability for test 1 for the QUICK sites
Fig. 6CR (left column) and BV (right column). Top row, tests 1 and 2. Middle row, tests 3 and 4. Bottom row, tests 5 and 6. All tests were performed with the QUICK phantom
Fig. 7LE values for the six tests performed with the QUICK phantom