Validation of novel automatic ultra-wideband radar for sleep apnea detection.

Yong Zhou1,2, Degui Shu1, Hangdi Xu1, Yuanhua Qiu1, Pan Zhou1, Wenjing Ruan1, Guangyue Qin3, Joy Jin2, Hao Zhu4, Kejing Ying1, Wenxia Zhang1, Enguo Chen1.   

Abstract

BACKGROUND: To validate the accuracy of ultra-wideband (UWB) wireless radar for the screening diagnosis of sleep apnea.
METHODS: One hundred and seventy-six qualified participants were successfully recruited. Apnea-hypopnea index (AHI) results from polysomnography (PSG) were reviewed by physicians, while the radar device automatically calculated AHI values with an embedded chip. All results were statistically analyzed.
RESULTS: A UWB radar-based AHI algorithm was successfully developed according to respiratory movement and body motion signals. Of all 176 participants, 63 exhibited normal results (AHI <5/hr) and the remaining 113 were diagnosed with obstructive sleep apnea. Significant correlation was detected between radar AHI and PSG AHI (Intraclass correlation coefficient 0.98, P<0.001). Receiver operating characteristic curve (ROC) analysis revealed high sensitivity and specificity. High concordance in participants with varying gender, age, BMI, and PSG AHI was reached.
CONCLUSIONS: The UWB radar may be a portable, convenient, and reliable device for obstructive sleep apnea screening. 2020 Journal of Thoracic Disease. All rights reserved.

Entities:  

Keywords:  Non-contact; apnea-hypopnea index (AHI); screening; ultra-wideband radar

Year:  2020        PMID: 32395265      PMCID: PMC7212156          DOI: 10.21037/jtd.2020.02.59

Source DB:  PubMed          Journal:  J Thorac Dis        ISSN: 2072-1439            Impact factor:   2.895


Introduction

Sleep apnea is a disease condition that has serious health and financial implications. Obstructive sleep apnea contributes largely. The prevalence of obstructive sleep apnea (OSA), defined at an apnea-hypopnea index (AHI) ≥5, fell at a mean of 22% (range, 9–37%) in men and 17% (range, 4–50%) in women based on a published epidemiological study between 1993 and 2013 (1). It is believed that the prevalence of sleep apnea is still increasing around the world. At present, sleepiness scale evaluation followed by laboratory polysomnography (PSG) is a classic diagnostic procedure for sleep apnea. However, this process has several limitations, including relatively expensive costs, full night-time occupation of the sleep lab, discomfort due to attached sensors, and long waiting lists. It requires trained technician and physician attendance, and does not allow professionals to monitor patients continuously for long-term or distant follow-up. Furthermore, the most important problem is the fact that a considerable number of sleep apnea patients remain undiagnosed and untreated. This underestimated diagnosis is chiefly caused by the lack of public awareness and shortage of sleep centers. Untreated patients may thus experience poor quality of life, adverse cardiovascular events (2). Intermittent hypoxia, intrathoracic pressure changes, arousals, and sleep fragmentation, which occur in association with obstructive sleep apneas, are likely to contribute to increased risk of cardiovascular events through the synergistic combination of heterogeneous mechanisms, including enhanced sympathetic activity, upregulation of the atrial natriuretic peptide and renin-angiotensin-aldosterone systems, and recurrent intermittent hypoxia-reoxygenation, which resembles the ischemia-reperfusion cycle and favors the release of reactive oxygen species, cytokines, and vasoactive mediators (3,4). Over time, the autonomic and neurohumoral abnormalities perpetuate beyond the offending obstructive events and persist into the daytime, resulting in a disturbance of the overall circadian blood pressure rhythm and an increase in short-and long-term blood pressure variability, which actually promote damage to target organs, including hearth, brain and kidney (5). The impact on the neurocognitive functioning was associated with the blood pressure variability (6,7). Therefore, advances have been made to develop economical and more accessible diagnostic tools, especially those that can be used in patients’ homes. However, a recently published review on screening for sleep apnea in adults suggested that there is uncertainty about the accuracy or clinical utility of all potential screening tools (8). The lack of suitable observational methods for large-sample, long-running longitudinal investigations is likely the bottleneck (2). Recently, a non-contact radar sleep testing technology has emerged. Radar detection integrates radar and biomedical engineering technologies, and can wirelessly penetrate nonmetal substances within certain ranges of distance, detecting signals such as respiration, heartbeat, blood stream, bowel movements, and body motion. The ultra-wideband (UWB) wireless radar system in particular is based on emission and reception of sub-nanosecond pulses, and can provide a low-cost, small-sized, low power consumption and high data rate alternative (9). It is theoretically more sensitive than the traditional belts in detecting respiratory movements. Even the doppler radar can detect thoracic and abdominal movements with high accuracy, up to almost 100% of sensitivity and specificity. The device accurately detected central sleep apnea, obstructive sleep apnea and mixed sleep apnea (10). Usually obstructive sleep apnea presents respiratory effort throughout the event. Central sleep apnea shows absence of respiratory effort. And mixed sleep apnea presents absence of respiratory effort at the beginning of the event followed by respiratory effort during the second half. Studies have previously reported that satisfying AHI score values can be approximated from thoracic movements alone (11-15). In addition, body motion sensors such as hand-wearing Jawbone showed good agreement with PSG in measures of total sleep time and wake-after-sleep onset (16). Since radar can detect both respiratory movement and body motion, advantages in sleep apnea syndrome screening and diagnosis are obvious; the technology might be more accurate and can be applied outside sleep centers without trained technician attendance. An algorithm based on certain parameters can classify the severity of sleep apnea syndrome and assess sleep quality (17). However, no large-scale study validating such a promising device in sleep apnea screening has been published thus far. This study adopted simultaneous overnight sleep monitoring with a UWB wireless radar system and PSG, and results were compared to validate the AHI of the UWB wireless radar system.

Methods

Participants

The protocol of our study was approved by the Ethics Committee of Sir Run Run Shaw Hospital, and the research was performed in accordance with relevant guidelines. Informed consent for study participation and publication of identifying information/images in an online open-access publication was obtained from all participants. Eligible participants were consecutive patients or healthy volunteers who were referred to sleep labs in Qingchun campus and Xiasha campus of Sir Run Run Shaw Hospital, and Zhejiang Hospital. Patients with unstable medical conditions, such as severe chronic pulmonary disease, severe heart disease, severe obesity (BMI >45 kg/m2), neuromuscular disorder, inability to apply the diagnostic device or other factors that result in incompletion of the sleep detection, patients unable to sign the study consent were excluded. In total, 176 participants in three hospital sleep centers from September 2015 to February 2017 were recruited with written consent. All the participants were assessed by physicians and detected with PSG and UWB radar simultaneously in the hospitals. There were 125 male and 51 female patients, with a mean age of 38.07±14.99, BMI 24.7±4.24, and Epworth sleepiness scale of 8.8±1.4 ().
Table 1

Demographics of participants

ItemTotal
Subjects number176
Male sex125
Age38.1±15.0
BMI (kg/m2)24.7±4.2
Epworth sleepiness scale8.8±1.4
Total sleep time (min)376.7±89.8
Oxygen desaturation index of 4% (/h)32.8±33.8
Sleep latency (min)27.8±39.0
Sleep efficiency index (%)79.0±15.1
Arousal index (/h)16.3±15.1

Polysomnography and UWB radar sleep screening device

All participants underwent PSG (Alice5, Respironics Inc, USA) and wireless UWB radar sleep screening device (ZG-S01A, Hangzhou Zhaoguan Sensor Technology) detection simultaneously throughout the whole night. PSG channels included four-channel electroencephalogram (EEG), electro-oculogram (EOG), air-flow at the nose and mouth, submental and leg electromyogram (EMG), electrocardiogram (ECG), chest and abdominal respiratory movement, oxygen saturation (pulse oximetry), a body position sensor, and a snoring microphone. All results were manually interpreted by a sleep technician according to the standard criteria of the ‘2012 AASM Manual for the Scoring of Sleep and Associated Events’ and were subsequently reviewed by certified physicians. The UWB wireless radar sleep screening device (ZG-S01A), a lightweight ball with a diameter of approximately 15 centimeters, was placed on the side table beside the bed 15–25 centimeters above the mattress and about 1.5 meters away from the patient, with its antenna directed towards the patient’s chest and abdomen. The patient is free to turn body position during sleep. ZG-S01A has a detection chip inside and connects to an Android app via Bluetooth. The UWB radar, which is within FCC spectrum limitations (6G–8G Hz), can detect both central and obstructive sleep apnea patterns. It begins to work once powered on, and ends once powered off. TST (total sleep time) was calculated in the same fashion as the portable PSG, adjusted based on participant’s body motion, respiratory rate and its variation. The device can tell if the patient is in the bed, or body position is changing. The wake time is removed from the total sleep time of the patient during AHI calculation. Radar AHI results were calculated blindly to PSG technician.

Statistical analysis

Continuous variables have been summarized as mean ± standard deviation (SD). Intraclass correlation coefficient and Bland-Altman plots were used to assess the correlation and agreement between the PSG AHI and Radar AHI. Kendall tau-b was used to evaluate the concordance of the severity of AHI between PSG and radar results. ROC analysis was adopted to determine the sensitivity and specificity of the radar device. Statistical analysis and graphs were performed with STATA 14.0, SPSS 23.0, and PRISM 7.0. A P value of <0.05 was considered to be statistically significant.

Results

Respiratory movements and body motion signals were collected by radar device simultaneously with PSG monitoring (). The radar device was developed and further optimized to improve consistency, and the different types the sleep apnea were detected () (17). In total, 176 participants were recruited to validate this device; all participants successfully completed simultaneous detection through PSG and wireless UWB radar detection. According to AHI criteria, PSG detection revealed 113 sleep apnea syndrome patients, including 26 mild cases, 32 moderate cases, 55 severe cases, and 63 normal cases. The mean PSG AHI was 23.69±24.63 events/hr. The wireless radar device revealed 113 sleep apnea syndrome patients, including 25 mild cases, 34 moderate cases, 54 severe cases, and 63 normal cases. The mean radar AHI was 22.66±24.63 events/hr ().
Figure 1

Simultaneous monitoring with the UWB radar and PSG (enlarged UWB radar photo on the left top of the figure). UWB, ultra-wideband; PSG, polysomnography.

Figure 2

Segments of the radar signal illustrating the scoring of respiratory events. (A) Normal breathing signal; (B) hypopnea showing a slight reduction in the excursion; (C) central apnea showing a complete absence of excursion; (D) obstructive apnea showing a major reduction in the excursion, some excursion appears to still exist, suggesting breathing effort.

Table 2

Intraclass correlation coefficient of AHI results between PSG and radar

PSG severityPSG AHI (mean)Radar AHI (mean)PSG group (n)Radar group (n)Intraclass correlation95% confidence intervalP
Normal2.322.7563630.520.21–0.710.002
Mild11.349.5026250.49−0.12–0.770.047
Moderate22.1120.832340.700.40–0.85<0.001
Severe54.9353.7555540.910.84–0.95<0.001
Total23.6922.661761760.980.98–0.99<0.001

AHI, apnea-hypopnea index; PSG, polysomnography.

Simultaneous monitoring with the UWB radar and PSG (enlarged UWB radar photo on the left top of the figure). UWB, ultra-wideband; PSG, polysomnography. Segments of the radar signal illustrating the scoring of respiratory events. (A) Normal breathing signal; (B) hypopnea showing a slight reduction in the excursion; (C) central apnea showing a complete absence of excursion; (D) obstructive apnea showing a major reduction in the excursion, some excursion appears to still exist, suggesting breathing effort. AHI, apnea-hypopnea index; PSG, polysomnography. With a cut-off value of AHI =5, the radar device demonstrated a sensitivity and specificity of 100% (). ROC analysis () showed the area under the curve was 1.00±0.00 (95% confidence interval 1.00–1.00), while the positive predictive value and negative predictive value were both 100%, indicating the high accuracy of the radar device in sleep apnea syndrome screening.
Table 3

Sensitivity and specificity of radar AHI compared with PSG AHI

AHI cut-off valueSensitivitySpecificityPPV (%)NPV (%)
511100100
150.970.9698.9100
300.950.9998.2198.37

AHI, apnea-hypopnea index; PSG, polysomnography; PPV, positive predictive value; NPV, negative predictive value.

Figure 3

Correlation of wireless UWB radar-AHI and PSG-AHI. (A) ROC curve analysis with a cut-off value of AHI =5, area under curve 1.00±0.00, 95% confidence interval: 1.00–1.00; (B) ROC curve analysis with cut-off value of AHI =15, area under curve 0.99±0.005, 95% confidence interval: 0.98–1.00; (C) ROC curve analysis with cut-off value of AHI =30, area under curve 0.99±0.004, 95% confidence interval: 0.99–1.00; (D) significant correlation was detected between wireless UWB radar sleep screening device-AHI and PSG-AHI (Pearson’s r =0.97, 95% confidence interval: 0.96–0.98, P<0.0001, n=176). UWB, ultra-wideband; AHI, apnea-hypopnea index; PSG, polysomnography.

AHI, apnea-hypopnea index; PSG, polysomnography; PPV, positive predictive value; NPV, negative predictive value. Correlation of wireless UWB radar-AHI and PSG-AHI. (A) ROC curve analysis with a cut-off value of AHI =5, area under curve 1.00±0.00, 95% confidence interval: 1.00–1.00; (B) ROC curve analysis with cut-off value of AHI =15, area under curve 0.99±0.005, 95% confidence interval: 0.98–1.00; (C) ROC curve analysis with cut-off value of AHI =30, area under curve 0.99±0.004, 95% confidence interval: 0.99–1.00; (D) significant correlation was detected between wireless UWB radar sleep screening device-AHI and PSG-AHI (Pearson’s r =0.97, 95% confidence interval: 0.96–0.98, P<0.0001, n=176). UWB, ultra-wideband; AHI, apnea-hypopnea index; PSG, polysomnography. With a cut-off value of AHI =15, the radar device demonstrated a sensitivity of 96.55% and specificity of 95.51% (). ROC analysis () showed the area under the curve was 0.99±0.005 (95% confidence interval: 0.98–1.00), with a positive predictive value of 98.9% and a negative predictive value of 100%. With a cut-off value of AHI =30, the radar device showed a sensitivity of 94.55% and specificity of 99.17% (). ROC analysis () showed the area under the curve was 0.99±0.004 (95% confidence interval: 0.99–1.00), with a positive predictive value of 98.21% and a negative predictive value of 98.37%. Pearson’s correlation, Intraclass correlation, and Bland-Altman plots each revealed significantly high correlation and good agreement between the radar AHI and PSG AHI (Pearson’s r =0.97, 95% confidence interval: 0.96–0.98, P<0.0001, n=176; Intraclass correlation average measure 0.98, 95% confidence interval: 0.98–0.99, P<0.001; Bland-Altman AHI bias −1.03, 95% limits of agreement −13.04–10.99) (, ). The radar AHI and PSG AHI agreement was not affected by body mass index (). Both radar AHI and PSG AHI increased consistently with increasing age (), but radar AHI tended to non-significantly underestimate the severity of OSA with increased age, or PSG AHI (). Subgroup AHI correlation analysis also revealed that both moderate and severe OSA patients had high correlation between radar AHI and PSG AHI, while mild OSA patients did not (). There was a significantly high concordance of AHI severity (Kendall tau-b =0.834, P<0.0001, n=176) between PSG and radar results ().
Figure 4

Radar AHI and PSG AHI showed high agreement despite severity and age variations. (A) Radar AHI/PSG AHI agreement was estimated according to Bland-Altman comparison; (B) correlation between BMI and Radar AHI/PSG AHI agreement; (C) correlation between age and Radar AHI/PSG AHI values; (D) correlation between age and Radar AHI/PSG AHI agreement. AHI, apnea-hypopnea index; PSG, polysomnography.

Table 4

Severity of AHI concordance between PSG and radar results (n=176)

Severity according to PSGSeverity according to radar
Normal (AHI <5)Mild (5≤ AHI <15)Moderate (AHI ≥30)Severe (AHI ≥30)Total
Normal (AHI <5)6300063
Mild (5≤ AHI <15)0251026
Moderate (15≤ AHI <30)0031132
Severe (AHI ≥30)0025355
Total63253454176

AHI, apnea-hypopnea index; PSG, polysomnography.

Radar AHI and PSG AHI showed high agreement despite severity and age variations. (A) Radar AHI/PSG AHI agreement was estimated according to Bland-Altman comparison; (B) correlation between BMI and Radar AHI/PSG AHI agreement; (C) correlation between age and Radar AHI/PSG AHI values; (D) correlation between age and Radar AHI/PSG AHI agreement. AHI, apnea-hypopnea index; PSG, polysomnography. AHI, apnea-hypopnea index; PSG, polysomnography. Bland-Altman plot revealed significantly high correlation and good agreement between radar AHI and PSG AHI both in male and female groups (). PSG AHI was higher in the male group (mean difference −25.93, 95% confidence interval: −30.83–−21.03, P<0.0001) (), but radar AHI/PSG AHI agreement between male and female participants showed no statistical difference (mean difference −0.64, 95% confidence interval: −2.12–0.84, P=0.39) (). In hypertensive group and normal blood pressure group, Bland-Altman plot showed no significant AHI difference between the two groups, the AHI bias is −0.81±9.42 with 95% limits of agreement (−19.26–17.64).
Figure 5

Radar AHI and PSG AHI exhibited high agreement for both male and female participants. (A) Radar AHI/PSG AHI agreement in male participants was estimated according to Bland-Altman plot; (B) radar AHI/PSG AHI agreement in female participants was estimated according to Bland-Altman plot; (C) PSG AHI was higher in male participants than in female counterparts (mean difference −25.93, 95% confidence interval: −30.83–−21.03, P<0.0001); (D) radar AHI/PSG AHI agreement between male and female participants showed no statistical difference (mean difference −0.64, 95% confidence interval: −2.12–0.84, P=0.39). AHI, apnea-hypopnea index; PSG, polysomnography.

Radar AHI and PSG AHI exhibited high agreement for both male and female participants. (A) Radar AHI/PSG AHI agreement in male participants was estimated according to Bland-Altman plot; (B) radar AHI/PSG AHI agreement in female participants was estimated according to Bland-Altman plot; (C) PSG AHI was higher in male participants than in female counterparts (mean difference −25.93, 95% confidence interval: −30.83–−21.03, P<0.0001); (D) radar AHI/PSG AHI agreement between male and female participants showed no statistical difference (mean difference −0.64, 95% confidence interval: −2.12–0.84, P=0.39). AHI, apnea-hypopnea index; PSG, polysomnography.

Discussion

This study aimed to validate a novel, portable wireless UWB radar sleep screening device in obstructive sleep apnea. The UWB radar can detect signals indicating respiratory movement and body motion, count the patient’s sleep events and frequency, and adjust the total sleep time. Since no attached sensors on the patient are needed, the portable radar device can simultaneously work with PSG and is non-contact, regardless of the bed mattress and sleep posture. All those features combine to make the device suitable for sleep screening both in the hospital or at home. Given the high cost and long waitlist of classic standard polysomnography, many types of portable home sleep test devices have recently emerged. Evidence suggests that such devices are a reasonable substitute for in-laboratory polysomnography in most patients, except for some patients with severe cardiopulmonary or neuromuscular diseases (18). Sleep test devices were classified by the American Academy of Sleep Medicine into four levels based on the type of leads and settings. Advanced technology has resulted in a variety of portable monitoring devices that are easy to connect even without attendance of technicians, and can be used at home for obstructive sleep apnea screening or even diagnosis. Level 2 sleep test devices have more than seven channels and are used at home to increase accessibility and decrease costs. Nevertheless, the proper use of corresponding electrode cables and wires limits unattended home use. However, portable devices simplify these settings at the cost of amounts of data for accurate diagnosis. Thus, the paradox between channel numbers and diagnostic data for unattended devices urge creation of breakthrough technologies. Systemic review confirmed that level 3 portable devices showed good diagnostic performance in adult patients with a high pre-test probability of moderate to severe OSA with no unstable comorbidities (19). For most patients with suspected OSA, including mild OSA patients, home respiratory polygraphy is similarly effective to PSG (20). The accuracy of level 4 portable devices, on the other hand, merits more investigation. These devices adopt overnight ECG recordings, pulse oxygen saturation, or thoracoabdominal motion detection (21,22). Other devices with mattress- and sheet-like technology can diminish wires, increasing the speed of setup; however, size and cost limit widespread monitoring in homes (23,24). Most of these devices are contact in nature and may generate hygiene or comfort problems. Breath sound detection with microphones has been reported with acceptable accuracy and is easy to apply at home, but many such technologies are snore-dependent and may be susceptible to surrounding acoustic noise such as from other patients in the same general ward, limiting generalization (25). Remote sensing devices with digital camera- and vision-based technology are non-contact and unobtrusive, but are susceptible to movement artifacts and obscurity of patients with cover sheets (26,27). Radar-based devices are non-contact and patients can use bedding articles freely, but those are still in early experimental stages. In 2002, the Federal Communications Commission (FCC) granted permission for the marketing and operation of a new class of products incorporating ultra-wideband (UWB) technology. The main advantages of UWB radar for real-life implementation of home healthcare application include: high material-penetration capability, low electromagnetic interference and specific absorption rates, good immunity against multipath interference, thereby enhancing measurement reliability, low energy consumption, and diminished size of devices (28). Importantly, costs are dramatically reduced. In 2006, Chen first reported respiration rate estimation algorithms with the UWB radar (28); Baboli also proposed a new algorithm that detected heartbeat and respiration rate, while dramatically diminishing interfering environmental motions (29). The algorithm was also able to acquire respiratory features such as the inspiration and expiration speed, respiration intensity, and respiration holding ratio (30). In 2014, Lazaro first reported the application of UWB radar technology in sleep-related breathing disorders on one volunteer (31), and in 2016, Javaid combined the UWB radar and microphone data to detect sleep apnea events, comparing results with PSG data on three subjects, yielding a sensitivity and specificity of 80% and 70%, respectively (32). Our study aimed to validate the accuracy of UWB radar detection in sleep-related breathing disorders. We found a statistically high correlation between UWB radar AHI and PSG AHI.ROC analysis with three different cut-off values revealed satisfying diagnostic efficacy, and high sensitivity and specificity were reached in varying AHI severity groups. Respiratory movement varies with sex and age, and our device can diminish these differences to obtain high concordance with PSG AHI. In addition, UWB radar is non-contact and can be easily delivered via mail or express services, primed for easy installation at home for continuous monitoring without technician attendance. These home applications merit further research, as this unobtrusive device might also benefit pediatric patients or patients with burn injuries. Limitations of this study include the fact that fewer female participants were recruited, which is commonly seen in most sleep disorder studies. Since this device does not include electroencephalogram, it theoretically can’t record micro arousals. The thoracoabdominal movement was detected based on physical stretch of bands. It was reported that assessment of the thoracoabdominal movement can indirectly detect respiratory effort-related arousal (33), and the UWB radar, which is based on radar signal, is more sensitive to detect the thoracoabdominal movement, so it can more accurately record the respiratory effort-related arousal. Although the UWB radar does not need the measurement of SpO2, lack of oximetry is the second limitation. In this study, a small group of healthy participants were recruited to verified the cutoff value of AHI =5, and influenced the mean Epworth Sleepiness Scale. Further examinations outside the sleep center are required to validate the radar device, and adding additional channels such as pulse oximetry or heart rate may increase device accuracy and diagnostic value.

Conclusions

This study suggested that the UWB radar AHI demonstrated an outstanding consistency with PSG AHI. The technology's high sensitivity and specificity meet the competency of obstructive sleep apnea screening or following dynamic monitoring at home, and may be a portable, convenient, and reliable device to screen patients suspected of obstructive sleep apnea. Integrations of big data and artificial intelligence technologies might be promising for this automatic sleep screening device in the near future. The article’s supplementary files as
  33 in total

Review 1.  Portable Monitoring for the Diagnosis of OSA.

Authors:  Jessica A Cooksey; Jay S Balachandran
Journal:  Chest       Date:  2016-01-06       Impact factor: 9.410

2.  A UWB Radar Signal Processing Platform for Real-Time Human Respiratory Feature Extraction Based on Four-Segment Linear Waveform Model.

Authors:  Chi-Hsuan Hsieh; Yu-Fang Chiu; Yi-Hsiang Shen; Ta-Shun Chu; Yuan-Hao Huang
Journal:  IEEE Trans Biomed Circuits Syst       Date:  2015-02-03       Impact factor: 3.833

3.  A non-contact vision-based system for respiratory rate estimation.

Authors:  Michael H Li; Azadeh Yadollahi; Babak Taati
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2014

4.  Photoplethysmography as a single source for analysis of sleep-disordered breathing in patients with severe cardiovascular disease.

Authors:  Offer Amir; Deganit Barak-Shinar; Antonietta Henry; Frank W Smart
Journal:  J Sleep Res       Date:  2011-06-14       Impact factor: 3.981

Review 5.  The Last 25 Years of Obstructive Sleep Apnea Epidemiology-and the Next 25?

Authors:  Paul E Peppard; Erika W Hagen
Journal:  Am J Respir Crit Care Med       Date:  2018-02-01       Impact factor: 21.405

Review 6.  Obstructive sleep apnea is a common disorder in the population-a review on the epidemiology of sleep apnea.

Authors:  Karl A Franklin; Eva Lindberg
Journal:  J Thorac Dis       Date:  2015-08       Impact factor: 2.895

7.  The validity of the static charge sensitive bed in detecting obstructive sleep apnoeas.

Authors:  O Polo; L Brissaud; B Sales; A Besset; M Billiard
Journal:  Eur Respir J       Date:  1988-04       Impact factor: 16.671

8.  Assessment of thoracoabdominal bands to detect respiratory effort-related arousal.

Authors:  J F Masa; J Corral; M J Martín; J A Riesco; A Sojo; M Hernández; N J Douglas
Journal:  Eur Respir J       Date:  2003-10       Impact factor: 16.671

9.  Blood pressure variability and neurocognitive functioning.

Authors:  Simona Lattanzi; Fabrizio Vernieri; Mauro Silvestrini
Journal:  J Clin Hypertens (Greenwich)       Date:  2018-02-21       Impact factor: 3.738

10.  Obstructive sleep apnea syndrome and the nocturnal blood pressure profile.

Authors:  Simona Lattanzi; Francesco Brigo; Mauro Silvestrini
Journal:  J Clin Hypertens (Greenwich)       Date:  2018-05-30       Impact factor: 3.738

View more
  5 in total

1.  Sleep Position Detection with a Wireless Audio-Motion Sensor-A Validation Study.

Authors:  Wojciech Kukwa; Tomasz Lis; Jonasz Łaba; Ron B Mitchell; Marcel Młyńczak
Journal:  Diagnostics (Basel)       Date:  2022-05-11

2.  Toward standardizing the clinical testing protocols of point-of-care devices for obstructive sleep apnea diagnosis.

Authors:  Vivek Tangudu; Kahkashan Afrin; Sandy Reddy; Nicolaas E P Deutz; Steven Woltering; Satish T S Bukkapatnam
Journal:  Sleep Breath       Date:  2020-08-31       Impact factor: 2.816

3.  Automated Detection of Sleep Apnea-Hypopnea Events Based on 60 GHz Frequency-Modulated Continuous-Wave Radar Using Convolutional Recurrent Neural Networks: A Preliminary Report of a Prospective Cohort Study.

Authors:  Jae Won Choi; Dong Hyun Kim; Dae Lim Koo; Yangmi Park; Hyunwoo Nam; Ji Hyun Lee; Hyo Jin Kim; Seung-No Hong; Gwangsoo Jang; Sungmook Lim; Baekhyun Kim
Journal:  Sensors (Basel)       Date:  2022-09-21       Impact factor: 3.847

4.  Evaluation of a non-contact ultra-wideband bio-radar sleep monitoring device for screening of sleep breathing disease.

Authors:  Zhijing Wei; Jiahuan Xu; WenYang Li; Xingjian Wang; Zheng Qin; Jiawei Zhou; Wei Wang
Journal:  Sleep Breath       Date:  2021-07-24       Impact factor: 2.816

5.  Apnea Detection in Polysomnographic Recordings Using Machine Learning Techniques.

Authors:  Marek Piorecky; Martin Bartoň; Vlastimil Koudelka; Jitka Buskova; Jana Koprivova; Martin Brunovsky; Vaclava Piorecka
Journal:  Diagnostics (Basel)       Date:  2021-12-08
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.