Improved image quality of temporal bone CT with an ultrahigh-resolution CT scanner: clinical pilot studies.

Arisa Ohara1, Haruhiko Machida2, Hisae Shiga1, Wataru Yamamura3, Kenichi Yokoyama1.   

Abstract

PURPOSE: Ultrahigh-resolution CT (UHRCT) with slice collimation of 0.25 mm × 160 and matrix size of 1024 × 1024 has become clinically available. We compared the image quality of temporal bone CT (TBCT) between UHRCT and conventional multidetector CT (MDCT).
MATERIALS AND METHODS: We retrospectively enrolled 20 patients who underwent TBCT by MDCT (matrix size, 512 × 512) and subsequently by UHRCT (matrix size, 1024 × 1024). Two independent reviewers subjectively graded delineation of normal stapes, oval window, facial nerve canal, incudostapedial joint, and tympanic tegmen. We also quantified image noise in the cerebellar hemisphere. Between MDCT and UHRCT, we compared mean subjective grades using the Wilcoxon signed-rank test and the image noise using paired t test.
RESULTS: Grades were significantly higher with UHRCT than with MDCT for all the anatomies (P < 0.001), whereas noise was significantly higher with UHRCT than with MDCT (P = 0.002).
CONCLUSION: For TBCT, UHRCT shows better delineation of the fine anatomical structures compared with MDCT.

Entities:  

Keywords:  Image quality; Temporal bone CT; Ultrahigh-resolution CT

Mesh:

Year:  2020        PMID: 32394364      PMCID: PMC7452920          DOI: 10.1007/s11604-020-00987-5

Source DB:  PubMed          Journal:  Jpn J Radiol        ISSN: 1867-1071            Impact factor:   2.374


Introduction

Multidetector row CT (MDCT) is a major diagnostic tool in temporal bone imaging but is sometimes limited in the delineation of fine and complex anatomical structures of the middle and inner ears [1-4]. Since March 2017, a state-of-the-art ultrahigh-resolution CT (UHRCT) scanner has been clinically available to improve in- and through-plane spatial resolution of CT images. Major features of this CT scanner include an improved detector system (minimal slice thickness, 0.25 mm; maximal channel number, 1792) and x-ray focus (smallest size, 0.4 × 0.5 mm) compared with conventional MDCT scanners. Whereas UHRCT has been reported to be clinically useful in various CT examinations, including chest CT [5, 6], coronary CT angiography [7-9], and CT of the artery of Adamkiewicz [10], few reports appear to have examined temporal bone CT [11]. Yamashita et al. [11] previously compared the delineation of fine anatomical structures on temporal bone CT between UHRCT with a 512-matrix size and conventional MDCT. A greater matrix size can improve microstructure delineation. We thus compared image qualities of temporal bone CT between UHRCT with a 1024-matrix size and conventional MDCT.

Materials and methods

Patient population

We retrospectively enrolled 132 consecutive patients who underwent routine non-contrast temporal bone CT with the UHRCT scanner between April 1, 2017 and October 31, 2017 as a follow-up study, and who had undergone CT of the same region using a conventional 64-detector CT scanner previously at our institution without any significant interval change in clinical status based on their medical records. When two or more examinations were performed with the conventional MDCT scanner, only those CT images acquired at the most recent examination were used for this study. An experienced head-and-neck radiologist with 15 years of clinical experience reviewed the axial and MPR coronal temporal bone CT images on both sides and finally included 20 non-traumatic patients (12 men, 8 women; mean age, 62 ± 23 years; age range, 11–85 years) who showed normal temporal bone by both UHRCT and conventional MDCT without intervening change on at least one side. Median interval between the two CT examinations was 13.5 months (range, 5–40 months). This study was approved by the institutional review board at our institution. Informed consent was obtained from every patient.

CT data acquisition

We performed temporal bone CT with both the UHRCT (Aquilion Precision; Canon Medical Systems, Otawara, Tochigi, Japan) and conventional MDCT scanners (Aquilion 64; Canon Medical Systems) using a standard helical scan method, as described elsewhere [12-14]. Scan parameters for UHRCT were: tube voltage, 120 kVp; noise index, 6.5 Hounsfield Units for 2-mm thickness with filtered back-projection by automatic exposure control; number of channels, 1792; slice collimation, 0.25 mm × 160 rows; pitch factor, 0.569; rotation time, 0.75 s; and x-ray focus size, 0.4 × 0.5 mm. Scan parameters for conventional MDCT were: tube voltage, 120 kVp; tube current, 250 mA; number of channels, 896; slice collimation, 0.5 mm × 4 rows; pitch factor, 0.625; rotation time, 0.75 s; and x-ray focus size, 0.9 × 0.8 mm. With this MDCT scanner, we fixed the tube current because automatic exposure control was unavailable; and used only the central 4 rows of detectors to reduce the effect of cone angle and thus applied quarter-offset reconstruction to improve in-plane spatial resolution. We recorded the volume CT dose index (measured in mGy) displayed on the dose report on the CT scanner for each patient and calculated the mean volume CT dose index in our patients.

Image reconstruction and display

For each patient, we used filtered back-projection to reconstruct temporal bone CT axial and MPR coronal images by both UHRCT and conventional MDCT using the following parameters: FOV, 10 cm; matrix size, 1024 × 1024; slice thickness and interval, 0.6 mm; and bone kernel, FC 80 for the UHRCT; FOV, 10 cm; matrix size, 512 × 512; slice thickness and interval, 0.6 mm; and bone kernel, FC 80 for the conventional MDCT. All CT images were transferred to a PACS (SYNAPSE; FUJIFILM Medical, Tokyo, Japan) and displayed with a window width of 4000 Hounsfield Units and window level of 600 Hounsfield Units.

Image quality assessment

Two independent radiologists with 15 (Reviewer 1) and 5 (Reviewer 2) years of clinical experience freely scrolled the temporal bone CT images on the PACS using a paging method and subjectively graded the delineation of the following high-contrast structures only on the normal side in each patient: the stapes, oval window, and tympanic segment of the facial nerve canal on axial images, and the incudostapedial joint, oval window, and tympanic tegmen on MPR coronal images. We used a 5-point scale for grading: one (poor), not visible; two (fair), partially visible; three (moderate), generally visible with slight blurring or discontinuity; four (good), clearly visible with complete continuity; and five (excellent), very clearly visible with complete continuity. These anatomical structures were selected as clinically critical, but often difficult to completely and clearly delineate with a conventional MDCT scanner. Dehiscence is known to be identified in the tympanic segment of the facial nerve canal (mostly around the oval window) and the tympanic tegmen even in normal cases [15-17]. We thus defined a subjective grade of three points or less as the presence of dehiscence in these two structures. For quantitative assessment of image noise, we placed a circular region of interest of approximately 10 mm2 in the cerebellar hemisphere on the same side to measure the standard deviation of the CT value as the background noise on temporal bone CT images by UHRCT and conventional MDCT in each patient.

Statistical analysis

All continuous variables are expressed as mean ± standard deviation. We performed statistical analyses using commercially available software (IBM SPSS Statistics version 23; IBM, Armonk, NY). We used the Wilcoxon signed-rank test to compare subjective image quality graded for each anatomy on the temporal bone CT axial and MPR coronal images, the McNemar test to compare the incidence of dehiscence in the tympanic segment of the facial nerve canal and the tympanic tegmen between UHRCT and conventional MDCT. We used paired t-test to compare quantitative image noise after confirming the normal distribution and the Wilcoxon signed-rank test to compare the mean volume CT dose index between UHRCT and conventional MDCT. Values of P < 0.05 were considered statistically significant. We also used weighted kappa (κ) statistics to quantify inter-reviewer agreement on subjective image quality grade. Values of κ were interpreted as follows: poor, 0.00–0.20; fair, 0.21–0.40; moderate, 0.41–0.60; good, 0.61–0.80; and excellent, 0.81–1.00.

Results

Table 1 shows average image quality grades for each anatomy by UHRCT and conventional MDCT by the two Reviewers. The grades on both axial and MPR coronal images were significantly better by UHRCT than by conventional MDCT for all the anatomies (P < 0.001) (Figs. 1, 2, 3, 4).
Table 1

Average image quality grades by Reviewers 1 and 2

UHRCTMDCTP value
Axial
 Stapes4.8 ± 0.43.6 ± 0.5 < 0.001
 Oval window4.7 ± 0.53.6 ± 0.6 < 0.001
 Facial nerve canal4.0 ± 0.83.2 ± 0.6 < 0.001
Coronal
 Incudostapedial joint4.8 ± 0.43.6 ± 0.5 < 0.001
 Oval window4.8 ± 0.43.2 ± 0.4 < 0.001
 Tympanic tegmen4.9 ± 0.33.9 ± 0.7 < 0.001

MDCT multidetector row CT, UHRCT ultrahigh-resolution CT

Fig. 1

Axial images of the left temporal bone CT from the same case. Image quality scores for the tympanic segment of the facial nerve canal (arrow) by Reviewers 1 and 2 are five and four, respectively, on UHRCT (a), and three on MDCT (b) by both Reviewers. By both Reviewers, dehiscence in this anatomy was considered to be present on MDCT but not on UHRCT

Fig. 2

Axial images of the right temporal bone CT from the same case. Image quality scores for the tympanic segment of the facial nerve canal (arrow) by Reviewers 1 and 2 are three and four on UHRCT (a), respectively, and two and three on MDCT (b). Dehiscence in this anatomy was considered to be present by both Reviewers on MDCT; only by Reviewer 1 but not by Reviewer 2 on UHRCT

Fig. 3

Coronal images of the right temporal bone CT from the same case. Image quality scores for the tympanic tegmen (arrow) are five on UHRCT (a), and three on MDCT (b) by both Reviewers. By both Reviewers, dehiscence in this anatomy was considered to be present on MDCT but not on UHRCT

Fig. 4

Coronal images of the left temporal bone CT from the same case. Image quality scores of the incudostapedial joint (arrow) are five on UHRCT (a), and three on MDCT (b) by both Reviewers

Average image quality grades by Reviewers 1 and 2 MDCT multidetector row CT, UHRCT ultrahigh-resolution CT Axial images of the left temporal bone CT from the same case. Image quality scores for the tympanic segment of the facial nerve canal (arrow) by Reviewers 1 and 2 are five and four, respectively, on UHRCT (a), and three on MDCT (b) by both Reviewers. By both Reviewers, dehiscence in this anatomy was considered to be present on MDCT but not on UHRCT Axial images of the right temporal bone CT from the same case. Image quality scores for the tympanic segment of the facial nerve canal (arrow) by Reviewers 1 and 2 are three and four on UHRCT (a), respectively, and two and three on MDCT (b). Dehiscence in this anatomy was considered to be present by both Reviewers on MDCT; only by Reviewer 1 but not by Reviewer 2 on UHRCT Coronal images of the right temporal bone CT from the same case. Image quality scores for the tympanic tegmen (arrow) are five on UHRCT (a), and three on MDCT (b) by both Reviewers. By both Reviewers, dehiscence in this anatomy was considered to be present on MDCT but not on UHRCT Coronal images of the left temporal bone CT from the same case. Image quality scores of the incudostapedial joint (arrow) are five on UHRCT (a), and three on MDCT (b) by both Reviewers The incidence of dehiscence in the tympanic segment of the facial nerve canal and tympanic tegmen was comparable between UHRCT and conventional MDCT for Reviewer 1 (P = 0.182 and 0.133, respectively), although that in both anatomies was significantly lower by UHRCT than by conventional MDCT for Reviewer 2 (P < 0.001 and = 0.023, respectively) (Table 2; Figs. 1, 2, 3).
Table 2

Presence of dehiscence on CT

Reviewer 1UHRCTMDCTP value
Facial nerve canal7/20 (35%)12/20 (60%)0.182
Tympanic tegmen0/20 (0%)4/20 (20%)0.133

MDCT multidetector row CT, UHRCT ultrahigh-resolution CT

Presence of dehiscence on CT MDCT multidetector row CT, UHRCT ultrahigh-resolution CT Inter-reviewer agreement of all subjective image quality scores was good (κ = 0.69). Mean volume CT dose index was significantly lower with UHRCT (45.3 ± 2.1 mGy) than with conventional MDCT (128.2 ± 9.9 mGy; P < 0.001). Quantitative noise was significantly higher by UHRCT (151.7 ± 26.2 Hounsfield Units) than by conventional MDCT (126.9 ± 17.6 Hounsfield Units; P = 0.002).

Discussion

UHRCT can improve in- and through-plane spatial resolution compared with conventional MDCT and achieve the maximal spatial resolution of approximately 0.15 mm, mainly due to the improved detector system and x-ray focus [5, 6]. In the present study, use of UHRCT significantly improved delineation of all high-contrast, fine, and critical anatomical structures on temporal bone CT compared with conventional MDCT. Yamashita et al. [11] demonstrated that only the stapedius tendon was more clearly visualized on temporal bone CT by UHRCT than by 320-detector CT. However, all other anatomical structures were similarly visualized between both, possibly due to the use of a 512-matrix size also for UHRCT, different from the situation in our study (i.e., 1024-matrix size). Dehiscence is known to be identified in the facial nerve canal and tympanic tegmen even in normal cases [15-17]. The incidence of dehiscence of both the facial nerve canal and tympanic tegmen was higher by conventional MDCT than by UHRCT, although not significantly for Reviewer 1, possibly resulting from a greater partial volume effect due to lower spatial resolution in conventional MDCT. According to published data from cadaveric and surgical studies [15, 18–21], the incidence of facial nerve canal dehiscence ranged from 25 to 74%. According to published data from cadaveric studies, the incidence of tympanic tegmen dehiscence was reported as 20% [16, 17]. Thus, dehiscence of both the facial nerve canal and tympanic tegmen might have been overestimated by Reviewer 2 using conventional MDCT alone, as that Reviewer had less experience with temporal bone CT interpretation. We performed helical scanning and reconstructed 0.6-mm-slice images by filtered back-projection as per our clinical routine examinations. Use of axial or volume scanning for decreasing artifacts related to helical scanning and a thinner slice thickness for improving spatial resolution, if necessary, without increasing image noise by hybrid or model-based iterative reconstruction can not only further improve delineation of the aforementioned structures and diagnostic accuracy of the dehiscence of both the anatomies, but can also achieve visualization of more subtle structures such as mucosal folds of the middle ear. Mean volume CT dose index by conventional MDCT without using automatic exposure control (tube current-exposure time product: 187.5 mAs) was lower than that in previous studies [1, 11], but significantly higher than that by UHRCT in our study. Mean volume CT dose index by conventional MDCT was higher, but that by UHRCT was lower than the diagnostic reference levels for head CT (56–85 mGy) in the United States, European countries, and Japan [22-24]. Use of axial or volume scanning and hybrid or model-based iterative reconstruction may further reasonably reduce radiation dose by UHRCT. Whereas quantitative image noise was greater by UHRCT with better spatial resolution than by conventional MDCT, the use of the iterative reconstruction can also reduce the noise. Our study was limited in the retrospective design and inclusion of only a small study population from a single institution. We only used the single scan and reconstruction protocol for temporal bone CT. Use of different scan and reconstruction protocols may affect the study results. The diagnosis of dehiscence of the facial nerve canal and the tympanic tegmen by temporal bone CT was not surgically confirmed. Intervals between conventional MDCT and UHRCT examinations were long (median, 13.5 months) due to the retrospective nature of the study, although shorter-interval examinations would not be practical due to the increased radiation exposure. Although we only evaluated the delineation of normal anatomical structures on temporal bone CT, further studies are needed to assess diagnostic performance for middle ear diseases such as cholesteatoma and otosclerosis. However, a comparison of diagnostic performance between conventional MDCT and UHRCT in the same patients is challenging, as these diseases are likely to change over time. In conclusion, the latest UHRCT increases image noise but appears useful by offering sharper, clearer, more continuous delineation of fine anatomical structures in routine temporal bone CT scanning with higher spatial resolution compared with conventional MDCT. These benefits are mainly due to the improved detector system and x-ray focus.
  22 in total

Review 1.  Middle and inner ear: improved depiction with multiplanar reconstruction of volumetric CT data.

Authors:  John I Lane; E Paul Lindell; Robert J Witte; David R DeLone; Colin L W Driscoll
Journal:  Radiographics       Date:  2006 Jan-Feb       Impact factor: 5.333

2.  Anatomical variations of the tympanic and mastoid portions of the facial nerve canal. A radioanatomical investigation.

Authors:  P A Dimopoulos; C Muren; O Smedby; K Wadin
Journal:  Acta Radiol Suppl       Date:  1996

3.  Atrial Fibrillation: Diagnostic Accuracy of Coronary CT Angiography Performed with a Whole-Heart 230-µm Spatial Resolution CT Scanner.

Authors:  Daniele Andreini; Gianluca Pontone; Saima Mushtaq; Edoardo Conte; Marco Perchinunno; Marco Guglielmo; Valentina Volpato; Andrea Annoni; Andrea Baggiano; Alberto Formenti; Maria Elisabetta Mancini; Virginia Beltrama; Valentina Ditali; Alessandro Campari; Cesare Fiorentini; Antonio L Bartorelli; Mauro Pepi
Journal:  Radiology       Date:  2017-04-26       Impact factor: 11.105

4.  Temporal Bone CT: Improved Image Quality and Potential for Decreased Radiation Dose Using an Ultra-High-Resolution Scan Mode with an Iterative Reconstruction Algorithm.

Authors:  S Leng; F E Diehn; J I Lane; K K Koeller; R J Witte; R E Carter; C H McCollough
Journal:  AJNR Am J Neuroradiol       Date:  2015-05-21       Impact factor: 3.825

5.  Ultra-high-resolution CT angiography of the artery of Adamkiewicz: a feasibility study.

Authors:  Kunihiro Yoshioka; Ryoichi Tanaka; Hidenobu Takagi; Yuta Ueyama; Kei Kikuchi; Takuya Chiba; Kazumasa Arakita; Joanne D Schuijf; Yasuo Saito
Journal:  Neuroradiology       Date:  2017-10-28       Impact factor: 2.804

6.  Effect of Matrix Size on the Image Quality of Ultra-high-resolution CT of the Lung: Comparison of 512 × 512, 1024 × 1024, and 2048 × 2048.

Authors:  Akinori Hata; Masahiro Yanagawa; Osamu Honda; Noriko Kikuchi; Tomo Miyata; Shinsuke Tsukagoshi; Ayumi Uranishi; Noriyuki Tomiyama
Journal:  Acad Radiol       Date:  2018-01-17       Impact factor: 3.173

7.  Macroscopic bony deficiency of the tegmen tympani in adult temporal bones.

Authors:  D V Lang
Journal:  J Laryngol Otol       Date:  1983-08       Impact factor: 1.469

8.  Coronary artery disease: diagnostic accuracy of CT coronary angiography--a comparison of high and standard spatial resolution scanning.

Authors:  Gianluca Pontone; Erika Bertella; Saima Mushtaq; Monica Loguercio; Sarah Cortinovis; Andrea Baggiano; Edoardo Conte; Andrea Annoni; Alberto Formenti; Virginia Beltrama; Andrea Igoren Guaricci; Daniele Andreini
Journal:  Radiology       Date:  2014-02-08       Impact factor: 11.105

9.  Tegmental and petromastoid defects in the temporal bone.

Authors:  T R Kapur; W Bangash
Journal:  J Laryngol Otol       Date:  1986-10       Impact factor: 1.469

10.  Ultra-High-Resolution Computed Tomography of the Lung: Image Quality of a Prototype Scanner.

Authors:  Ryutaro Kakinuma; Noriyuki Moriyama; Yukio Muramatsu; Shiho Gomi; Masahiro Suzuki; Hirobumi Nagasawa; Masahiko Kusumoto; Tomohiko Aso; Yoshihisa Muramatsu; Takaaki Tsuchida; Koji Tsuta; Akiko Miyagi Maeshima; Naobumi Tochigi; Shun-Ichi Watanabe; Naoki Sugihara; Shinsuke Tsukagoshi; Yasuo Saito; Masahiro Kazama; Kazuto Ashizawa; Kazuo Awai; Osamu Honda; Hiroyuki Ishikawa; Naoya Koizumi; Daisuke Komoto; Hiroshi Moriya; Seitaro Oda; Yasuji Oshiro; Masahiro Yanagawa; Noriyuki Tomiyama; Hisao Asamura
Journal:  PLoS One       Date:  2015-09-09       Impact factor: 3.240