| Literature DB >> 32363545 |
Claudia Jäggi-Wickes1, Pascale Brasier-Lutz1, Sabine Schaedelin2, Rosemarie Burian1, Cora-Ann Schoenenberger3,4, Rosanna Zanetti-Dällenbach5.
Abstract
PURPOSE: To prospectively compare the diagnostic accuracy of radial breast ultrasound (r-US) to that of conventional meander-like breast ultrasound (m-US), patients of a consecutive, unselected, mixed collective were examined by both scanning methods.Entities:
Keywords: Diagnostic accuracy; Ductosonography; Examination time; False negative; Meander-like breast ultrasound; Radial breast ultrasound
Mesh:
Year: 2020 PMID: 32363545 PMCID: PMC7246244 DOI: 10.1007/s00404-020-05554-x
Source DB: PubMed Journal: Arch Gynecol Obstet ISSN: 0932-0067 Impact factor: 2.344
Fig. 1Radial and meander-like ultrasound. a Scheme of radial scanning movement (top) and corresponding ultrasound image (bottom). Left panel: radial movement and radial movement for the axillary tail; right panel: anti-radial movement. b Scheme of meander-like scanning movement in two orthogonal planes (top) and corresponding ultrasound image (bottom). Both ultrasound images are from the same 59-year-old patient diagnosed with invasive ductal breast cancer
Patient characteristics
| All study subjects | Patients with suspicious breast lesions ( | |||||
|---|---|---|---|---|---|---|
| No or no suspicious breast lesions | Suspicious breast lesions | Benign histology | Malignant histology | |||
| Number of patients (%) | 1251 (100) | 148 (100) | 115 (100) | 33 (100) | ||
| Mean age in years (min, max) [± SD] | 48.7 (18, 88) [± 14.3] | 47.1 (19, 86) [± 14.7] | 0.229 | 44.1 (19, 86) [± 14.0] | 57.8 (30, 79) [± 12.1] | < 0.01 |
| Personal history of breast cancer (%) | 47 (3.8) | 3 (2.0) | 0.402 | 2 (1.7) | 1 (3.0) | 0.520 |
| Negative family history (%) | 894 (71.5) | 95 (64.2) | 0.081 | 74 (64.3) | 21 (63.6) | 0.485 |
| Positive family history (%) | 357 (28.5) | 53 (35.8) | 41 (35.7) | 12 (36.4) | ||
| Breast cancer | 319 | 43 | 32 | 11 | ||
| Ovarian cancer | 1 | 2 | 2 | 0 | ||
| Breast and ovarian cancer | 14 | 2 | 1 | 1 | ||
| Endometrial cancer | 10 | 6 | 6 | 0 | ||
| Breast and endometrial cancer | 13 | 0 | 0 | 0 | ||
Findings of clinical palpation and mammography and final BI-RADS assessment for benign and malignant breast lesions
| Malignant lesions | Benign lesions | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| Malignant lesions | Invasive ductal | Invasive lobular | DCIS | Benign lesions | B2 lesions | Fibrosis/sclerosis | Fibroadenoma | B3 lesions | |
| 36 (100) | 31 (86.1) | 3 (8.3) | 2 (5.6) | 132 (100) | 35 (26.5) | 41 (31.1) | 50 (37.9) | 6 (4.5) | |
| Meander-like ultrasound | |||||||||
| BI-RADS 5 | 15 (41.7) | 12 (80.0) | 3 (20.0) | 1 (0.8)+ | 1 (100) | ||||
| BI-RADS 4 | 17 (47.1) | 16 (94.1) | 1 (5.9) | 17 (12.9)+ | 5 (29.4) | 7 (41.2) | 4 (23.5) | 1 (5.9) | |
| BI-RADS 3 | 2 (5.6)* | 1 (50.0) | 1 (50.0) | 104 (78.7) | 29 (27.9) | 30 (28.8) | 41 (39.4) | 4 (3.9) | |
| BI-RADS 2 | 0 (0.0) | 9 (6.8) | 1 (11.1) | 3 (33.3) | 5 (55.6) | ||||
| Missing | 2 (5.6)҂ | 2 (100) | 1 (0.8)‡ | 1 (100) | |||||
| Radial ultrasound | |||||||||
| BI-RADS 5 | 15 (41.7) | 13 (86.7) | 2 (13.3) | 2 (1.5)+ | 2 (100) | ||||
| BI-RADS 4 | 17 (47.1) | 15 (88.2) | 1 (5.9) | 1 (5.9) | 12 (9.1)+ | 2 (16.7) | 6 (49.9) | 2 (16.7) | 2 (16.7) |
| BI-RADS 3 | 2 (5.6)* | 1 (50.0) | 1 (50.0) | 100 (75.7) | 27 (27.0) | 27 (27.0) | 42 (42.0) | 4 (4.0) | |
| BI-RADS 2 | 0 (0.0) | 15 (11.4) | 6 (40.0) | 5 (33.3) | 4 (26.7) | ||||
| Missing | 2 (5.6)҂ | 2 (100) | 3 (2.3)‡ | 1 (33.3) | 2 (66.7) | ||||
| Palpation | |||||||||
| Lesion palpable | 23 (63.9) | 21 (91.2) | 1 (4.4) | 1 (4.4) | 50 (37.9) | 16 (32.0) | 9 (18.0) | 24 (48.0) | 1 (2.0) |
| Lesion not palpable | 8 (22.2) | 6 (75.0) | 1 (12.5) | 1 (12.5) | 73 (55.3) | 18 (24.7) | 29 (39.7) | 22 (30.1) | 4 (5.5) |
| Missing | 5 (13.9) | 4 (80.0) | 1 (20.0) | 9 (6.8) | 1 (11.1) | 3 (33.3) | 4 (44.5) | 1 (11.1) | |
| Mammography | |||||||||
| Lesion detected | 30 (83.3) | 25 (83.3) | 3 (10.0) | 2 (6.7) | 39 (29.5) | 12 (30.8) | 10 (25.6) | 13 (33.3) | 4 (10.3) |
| Lesion not detected | 5 (13.9) | 5 (100) | 24 (18.2) | 11 (45.8) | 10 (41.7) | 3 (12.5) | |||
| Not performed ≤ 40 years | 0 (0.0) | 40 (30.3) | 7 (17.5) | 6 (15.0) | 26 (65.0) | 1 (2.5) | |||
| Missing | 1 (2.8) | 1 (100) | 29 (22.0) | 5 (17.2) | 15 (51.8) | 8 (27.6) | 1 (3.4) | ||
*False negative, +False positive, ҂Missed cancers, ‡Benign lesion not revealed
False-negative lesions and lesions missed by either scanning method
| Histology | Meander-like ultrasound | Radial ultrasound | Lesion palpable | Lesion detected by mammography |
|---|---|---|---|---|
| False negative lesions | ||||
| DCIS | BI-RADS 3 | BI-RADS 3 | No | Yes |
| Inv. ductal | BI-RADS 3 | BI-RADS 3 | Yes | Yes |
| Cancers which were missed by one method | ||||
| Inv. ductal | Missed | BI-RADS 4 | No | Yes |
| Inv. ductal | Missed | BI-RADS 4 | No | Yes |
| Inv. ductal | BI-RADS 4 | Missed | No | No |
| Inv. ductal | BI-RADS 4 | Missed | No | No |
| Benign lesions which were missed by one method | ||||
| B3-Lesion | Missed | BI-RADS 3 | No | Yes |
| Fibrosis | BI-RADS 4 | Missed | No | Yes |
| Fibroadenoma | BI-RADS 3 | Missed | No | No mammography |
| Fibroadenoma | BI-RADS 3 | Missed | No | No mammography |
Diagnostic accuracy of conventional meander-like ultrasound and radial ultrasound
| Meander-like ultrasound | Radial ultrasound | CI | ||||||
|---|---|---|---|---|---|---|---|---|
| % | CI | % | CI | |||||
| Malignant lesions | 36 | 100 | 36 | 100 | ||||
| Cancers identified | 34 | 94.4 | [81.4; 99.0] | 34 | 94.4 | [81.4; 99.0] | 1 | [0.1; 13.8] |
| Cancers missed | 2 | 5.6 | [1.0; 18.6] | 2 | 5.6 | [1.0; 18.6] | 1 | [0.1; 13.8] |
| True positive | 32 | 88.9 | [74.2; 96.1] | 32 | 88.9 | [74.2; 96.1] | 1 | [0.1; 13.8] |
| False negativea (BI-RADS 3) | 2 | 5.6 | [1.0; 18.6] | 2 | 5.6 | [1.0; 18.6] | 1 | [0.0; 13.8] |
| False negativeb (BI-RADS 3 and missed cancersc) | 4 | 11.1 | [3.9; 25.8] | 4 | 11.1 | [3.9; 25.8] | 1 | [0.1; 13.8] |
| Benign lesions | 132 | 100 | 132 | 100 | ||||
| Benign lesions identified | 131 | 99.2 | [96.1; 100.0] | 129 | 97.7 | [93.7; 99.4] | 0.63 | [0.0; 4.2] |
| Benign lesions missed | 1 | 0.8 | [0.0; 3.9] | 3 | 2.3 | [0.6; 6.3] | 0.63 | [0.0; 4.2] |
| True negative | 113 | 85.6 | [78.6; 90.8] | 115 | 87.1 | [80.5; 92.1] | 0.79 | [0.2; 2.5] |
| False positived (BI-RADS 4 or 5) | 18 | 13.6 | [8.4; 20.6] | 14 | 10.6 | [6.2; 16.8] | 0.39 | [0.5; 9.1] |
| False positived (BI-RADS 4 or 5 and missed lesionse) | 19 | 14.4 | [9.2; 21.4] | 17 | 12.9 | [7.9; 19.5] | 0.79 | [0.2; 2.5] |
| Diagnostic accuracy | ||||||||
| Sensitivity | 88.9 | [74.2; 96.1] | 88.9 | [74.2; 96.1] | 1 | [0.1;13.8] | ||
| Specificity | 86.4 | [79.4; 91.6] | 89.4 | [83.2; 93.8] | 0.39 | [0.1; 1.9] | ||
| Accuracy | 86.9 | [81.1; 91.6] | 89.3 | [83.8; 93.4] | 0.45 | [0.2; 1.8] | ||
| PPVf | 64.0 | [50.0; 76.4] | 69.6 | [54.4; 82.2] | 0.90 | [0.25; 3.38] | ||
| NPVg (excluding missed cancers) | 98.3 | [94.2; 99.7] | 98.3 | [94.4; 99.7] | 0.45 | [1.0; 1.0] | ||
| NPVg (including missed cancers) | 96.6 | [91.8; 98.8] | 96.7 | [92.1; 98.9] | 0.95 | [0.97; 1.03] | ||
aThe same two breast lesions were classified as BI-RADS 3 by both methods
bFalse negative and NPV were calculated excluding and including missed cancers and dfalse positive was calculated excluding and including missed lesionse. In general, studies do not include a second US examination and thus, the cancer miss rate and the number of missed benign lesions remain unknown. Consequently, missed cancers/missed benign lesions are not represented in the calculation of false negative/false positive values and NPV
cCancers missed by one method but correctly identified by the other
eBenign lesions not revealed by one of the two methods
fPPV, positive predictive value; gNPV, negative predictive value
Duration of bilateral meander-like ultrasound versus radial ultrasound examination
| Meander-like ultrasound | Radial ultrasound | ||
|---|---|---|---|
| Mean examination duration revealing suspicious lesions requiring breast biopsy (min) | 22.6 | 14.8 | < 0.01 |
| (min, max) [± SD] | (4.7, 74.0) [± 12.8] | (3.9, 47.0) [± 7.6] | |
| Malignant lesions | 26.6 | 14.8 | < 0.01 |
| (min, max) [± SD] | (4.7, 69.0) [± 18.0] | (3.9, 38.5) [± 7.1] | |
| Benign lesions | 21.7 | 14.8 | < 0.01 |
| (min, max) [± SD] | (7.1, 74.0) [± 11.2] | (5.2, 47.0) [± 7.8] | |
| Mean examination duration revealing no or no suspicious lesions (min) | 13.3 | 7.6 | < 0.01 |
| (min, max) [± SD] | (2.0, 58.2) [± 6.2] | (2.3, 64.2) [± 5.8] | |
| No lesion detected | 9.7 | 4.0 | < 0.01 |
| (min, max) [± SD] | (4.2, 22.2) [± 3.8] | (2.3, 9.5) [± 1.3] | |
| 1 lesion detected | 13.2 | 6.6 | < 0.01 |
| (min, max) [± SD] | (4.7, 32.3) [± 5.9] | (2.4, 24.5) [± 3.3] | |
| > 1 lesion detected | 14.6 | 9.3 | < 0.01 |
| (min, max) [± SD] | (2.0, 58.2) [± 6.6] | (2.7, 64.2) [± 6.8] | |
| Mean examination duration of all examinations (min) | 15.0 | 8.9 | < 0.01 |
| (min, max) [± SD] | (2.0, 74.0) [± 8.6] | (2.3, 64.2) [± 6.7] |
Comparison of the first and last 50 radial ultrasound examinations revealing suspicious lesions
| First 50 radial | Last 50 radial | ||
|---|---|---|---|
| Duration of radial ultrasound examination | |||
| Mean time (min) | 19.8 | 9.9 | < 0.01 |
| (min, max) [± SD] | (5.2, 47.0) [± 8.6] | (3.9, 17.6) [± 3.4] | |
| Diagnostic accuracy | |||
| Sensitivity [CI] | 70.0 [38.1; 91.3] | 91.7 [63.4; 99.6] | 0.45 |
| Specificity [CI] | 90.0 [76.8; 96.5] | 86.8 [72.8;94.7] | 1 |
| Accuracy [CI] | 86.0 [73.4; 93.7] | 88.0 [76.4; 94.6] | 1 |