Literature DB >> 32356946

Guided-motion hinged knee replacement prosthesis: early survival rate and postoperative patient function and satisfaction.

David L Perrin1, Thomas R Turgeon1.   

Abstract

Background: Literature on the survival rates and function of hinged total knee replacement (HTKR) prostheses is scarce, and to our knowledge there is not yet any published literature on the Legion HK Hinge Knee Replacement prosthesis (Smith & Nephew) with guided-motion articulation. The objective of this study was to establish the early survival rate of this modern HTKR at a single institution and to investigate postoperative patient function and satisfaction.
Methods: This retrospective study included patients who received the Legion HTKR prosthesis with guided-motion inserts as a primary or revision implant between October 11 and March 2016 at a tertiary care centre in Manitoba, Canada. Preoperative and postoperative functional scores on the 12-item Oxford Knee Score and postoperative patient satisfaction were assessed.
Results: Thirty-nine HTKR implantations (38 patients) were included in this study: 12 primary cases and 27 revision cases. Three revision surgeries and 4 perioperative complications were noted at a mean follow-up of 29.1 months. The 2-year survivorship of the HTKR system was 90.7%. Postoperative functional scores improved significantly and the majority of patients were satisfied or very satisfied at all follow-up time points.
Conclusion: The early survival rate of a modern guided-motion HTKR prosthesis is similar to the survival rates of other hinged knee prostheses published in the literature. The prosthesis demonstrated substantial postoperative functional improvement when used in the setting of complex primary or revision total knee replacement.
© 2020 Joule Inc. or its licensors.

Entities:  

Mesh:

Year:  2020        PMID: 32356946      PMCID: PMC7829011          DOI: 10.1503/cjs.002419

Source DB:  PubMed          Journal:  Can J Surg        ISSN: 0008-428X            Impact factor:   2.089


  20 in total

1.  Constraint in total knee arthroplasty: when and what?

Authors:  James P McAuley; Gerard A Engh
Journal:  J Arthroplasty       Date:  2003-04       Impact factor: 4.757

2.  Rotating-hinge total knee for revision total knee arthroplasty.

Authors:  Alessandro Bistolfi; Giuseppe Massazza; Federica Rosso; Maurizio Crova
Journal:  Orthopedics       Date:  2012-03-07       Impact factor: 1.390

3.  Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030.

Authors:  Steven Kurtz; Kevin Ong; Edmund Lau; Fionna Mowat; Michael Halpern
Journal:  J Bone Joint Surg Am       Date:  2007-04       Impact factor: 5.284

4.  Repeated early failure of a newly designed hinged knee system.

Authors:  Joshua K L Lee; Vikram Chatrath; Paul R Kim
Journal:  J Arthroplasty       Date:  2012-06-28       Impact factor: 4.757

5.  Complications following rotating hinge Endo-Modell (Link) knee arthroplasty.

Authors:  B Guenoun; L Latargez; M Freslon; G Defossez; N Salas; L-E Gayet
Journal:  Orthop Traumatol Surg Res       Date:  2009-10-17       Impact factor: 2.256

6.  Implant survival and outcome after rotating-hinge total knee revision arthroplasty: a minimum 6-year follow-up.

Authors:  Asgeir Gudnason; Jan Milbrink; Nils P Hailer
Journal:  Arch Orthop Trauma Surg       Date:  2011-06-09       Impact factor: 3.067

7.  Mid-term survival following primary hinged total knee replacement is good irrespective of the indication for surgery.

Authors:  Paul Baker; Rebecca Critchley; Andrew Gray; Simon Jameson; Paul Gregg; Andrew Port; David Deehan
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2012-12-14       Impact factor: 4.342

8.  The GSB total knee arthroplasty. A medium- and long-term follow-up and survival analysis.

Authors:  C J van Loon; C Pluk; M C de Waal Malefijt; M de Kock; R P Veth
Journal:  Arch Orthop Trauma Surg       Date:  2001       Impact factor: 3.067

9.  Mid-term results after implantation of rotating-hinge knee prostheses: primary versus revision.

Authors:  Turgay Efe; Philip P Roessler; Thomas J Heyse; Carsten Hauk; Caroline Pahrmann; Alan Getgood; Jan Schmitt
Journal:  Orthop Rev (Pavia)       Date:  2012-12-11

10.  Early results of a new rotating hinge knee implant.

Authors:  Alexander Giurea; Hans-Joachim Neuhaus; Rolf Miehlke; Reinhard Schuh; Richard Lass; Bernd Kubista; Reinhard Windhager
Journal:  Biomed Res Int       Date:  2014-06-25       Impact factor: 3.411

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.