| Literature DB >> 32341937 |
Lisa Heck1, Elena Eggl1, Susanne Grandl2, Martin Dierolf1, Christoph Jud1, Benedikt Günther1, Klaus Achterhold1, Doris Mayr3, Bernhard Gleich1, Karin Hellerhoff2, Franz Pfeiffer1,4, Julia Herzen1.
Abstract
Purpose: Although the mortality rate of breast cancer was reduced with the introduction of screening mammography, many women undergo unnecessary subsequent examinations due to inconclusive diagnoses. Superposition of anatomical structures especially within dense breasts in conjunction with the inherently low soft tissue contrast of absorption images compromises image quality. This can be overcome by phase-contrast imaging. Approach: We analyze the spatial resolution of grating-based multimodal mammography using a mammographic phantom and one freshly dissected mastectomy specimen at an inverse Compton x-ray source. Here, the focus was on estimating the spatial resolution with the sample in the beam path and discussing benefits and drawbacks of the method used and the estimation of the mean glandular dose. Finally, the possibility of improving the spatial resolution is investigated by comparing monochromatic grating-based mammography with the standard one.Entities:
Keywords: inverse Compton x-rays; mammography; phase-contrast imaging; radiation; spatial resolution
Year: 2020 PMID: 32341937 PMCID: PMC7175026 DOI: 10.1117/1.JMI.7.2.023505
Source DB: PubMed Journal: J Med Imaging (Bellingham) ISSN: 2329-4302
Fig. 1Schematic drawing of the experimental setup at the MuCLS (not to scale): the CLS is a storage ring-based ICS. Therefore, x-rays are generated by the collision of electrons accelerated to relativistic energies with a counterpropagating laser pulse stored in an enhancement cavity. The measurements were performed at an experimental station that is located about 15 m away from the interaction point.
Acquisition parameters whereby the exposure times always refer to the MGD in the column to the left. The exposure times are given for the whole acquisition including all steps and all stitching scans.
| Sample | Energy (keV) | MGD mAC-Mx (mGy) | Exposure time (s) | MGD mgb-Mx (mGy) | Total exposure time (s) | Number of steps | Stitching |
|---|---|---|---|---|---|---|---|
| I | 25 | 0.3 | 75 | 0.9 | 275 | 11 | |
| Phantom | 25 | 1.0 to 2.0 | 40 to 80 | 0.7 to 1.8 | 28 to 72 | 7 or 9 |
Acquisition parameters of the clinical in vivo and ex vivo measurements.
| Sample | X-ray tube settings | MGD civAC-Mx (mGy) | MGD cevAC-Mx (mGy) | Compressed thickness (cm) |
|---|---|---|---|---|
| I | 30 kVp (W/Rh), 100 mAs | 2.9 | 1.4 (AP) | 4.5 |
| Phantom | 28 kVp (W/Rh), 200 mAs | — | 2.0 | 4.5 |
Fig. 2Spectral power analysis for the determination of spatial frequency: (a) Visualization of the resolution criterion. The maximal spatial frequency is defined as the frequency where the spectral power of the signal equals the spectral power of the noise . In experimental data, only the total signal, including noise, is measured. The maximal spatial frequency is then given by the intersection of the spectral power of the experimental signal and the double of the spectral power of the noise baseline. (b) Example for the determination of the maximal spatial frequency on the experimental data. The smallest and highest spatial frequencies where the noisy spectra power curve intersects the line is given by the black, vertical dotted lines. Then, the calculation of the mean value results in the actual spatial resolution of the image. Subfigure (a) adapted from Ref. 40.
Fig. 3Absorption-only radiographs of the mammographic accreditation phantom: (a) clinical mammography (cevAC-Mx) with an MGD of 2.0 mGy, (b) mAC-Mx with an MGD of 2.0 mGy, and (c) with an MGD of 1.6 mGy. All images were scaled for maximum detail visibility.
Fig. 4Monochromatic grating-based phase-contrast imaging showing (a) mgbAC-Mx, (b) mgbDPC-Mx, and (c) mgbDFC-Mx for an MGD of 1.8 mGy. All images were scaled for maximum detail visibility.
CNR calculated for a quantitative analysis with the mammographic accreditation phantom. The numbers have been chosen so that the small numbers refer to the largest structures and the large numbers to the smaller structures.
| Modality | MGD (mGy) | Fibers | Calcifications | Tumor masses | |||
|---|---|---|---|---|---|---|---|
| 1 | 4 | 1 | 3 | 1 | 5 | ||
| cevAC-Mx | 2.0 | 2.51 | 1.99 | 36.84 | 13.89 | 5.92 | 0.53 |
| mAC-Mx | 1.0 | 3.16 | 0.11 | 30.86 | 11.94 | 8.64 | 1.50 |
| mAC-Mx | 1.6 | 3.51 | 1.61 | 38.47 | 15.60 | 10.70 | 2.73 |
| mAC-Mx | 2.0 | 4.71 | 1.39 | 44.11 | 16.85 | 12.19 | 2.67 |
| mgbAC-Mx | 1.8 | 4.46 | 0.15 | 25.32 | 13.25 | 7.14 | 0.89 |
| mgbDFC-Mx | 1.8 | 0.65 | 0.16 | 6.50 | 10.42 | 15.00 | 9.59 |
Spatial resolution calculated with an analysis of the noise power spectrum.
| Sample | cevAC-Mx (lp/mm) | mAC-Mx (lp/mm) | mgbAC-Mx (lp/mm) | mgbDFC-Mx (lp/mm) |
|---|---|---|---|---|
| I | ||||
| Phantom |
Fig. 5Advanced detection of tumor lesions. (a) The cevAC-Mx image in AP position, (b) civAC-Mx image in crandio-caudal position, and (c) the histopathological analysis of the sample are depicted in the first row. The second row presents the results acquired with grating interferometry at the MuCLS including (d) the mgbAC-Mx, (e) the mgbDPC-Mx, and (f) the mgbDFC-Mx images. All images in the second row are taken in AP position The red arrow depicts the mamilla, whereas the orange arrows in the radiographs and the black arrows in the histopathology image depict the cancerous lesions. All images were scaled for maximum detail visibility.