Can Combined Screening of Ultrasound and Elastography Improve Breast Cancer Identification Compared with MRI in Women with Dense Breasts-a Multicenter Prospective Study.

Lu-Ying Gao1, Yang Gu1, Wen Xu1, Jia-Wei Tian2, Li-Xue Yin3, Hai-Tao Ran4, Wei-Dong Ren5, Yu-Ming Mu6, Jie-Ying Zhang7, Cai Chang8, Jian-Jun Yuan9, Chun-Song Kang10, You-Bin Deng11, Hui Wang12, Xiao-Yan Xie13, Bao-Ming Luo14, Sheng-Lan Guo15, Qi Zhou16, En-Sheng Xue17, Wei-Wei Zhan18, Tong Jiao19, Qing Zhou20, Jie Li21, Ping Zhou22, Pin-Tong Huang23, Hong-Yuan Xue24, Chun-Quan Zhang25, Man Chen26, Xiang-Xiang Jing27, Ying Gu28, Jian-Feng Guo29, Hong-Yu Ding30, Jin-Feng Xu31, Wu Chen32, Li Liu33, Yu-Hong Zhang34, Hong-Qiao Wang35, Zhong-Ping Mu36, Jian-Chu Li1, Hong-Yan Wang1, Yu-Xin Jiang1.   

Abstract

Objectives: To assess the performance of elastography (ES) and ultrasound (US) in predicting the malignancy of breast lesions and to compare their combined diagnostic value with that of magnetic resonance imaging (MRI). Materials and
Methods: The study prospectively enrolled 242 female patients with dense breasts treated in 35 heath care facilities in China between November 2018 and October 2019. Based on conventional US and elastography, radiologists classified the degree of suspicion of breast lesions according to the US Breast Imaging Reporting and Data System (BI-RADS) criteria. The diagnostic value was compared between US BI-RADS and MRI BI-RADS, with pathological results used as the reference standard.
Results: The results demonstrated that irregular tumor shape, a nonparallel growth orientation, indistinct margins, angular contours, microcalcifications, color Doppler flow and ES score on US imaging were significantly related to breast cancer in dense breasts (P=0.001; P=0.001; P=0.008; P<0.001; P=0.019; P=0.008; P=0.002, respectively). The sensitivity, specificity, PPV, NPV, accuracy and AUC of US BI-RADS category were 94.7%, 90.7%, 95.8%, 88.0%, 93.4% and 0.93 (95%CI, 0.88-0.97), respectively, while those of MRI BI-RADS category were 98.2%, 57.5%, 84.3%, 83.3%, 86.0% and 0.78 (95%CI, 0.71-0.85), respectively. MRI BI-RADS showed a significantly higher sensitivity than US BI-RADS (98.2% vs 94.7%, P=0.043), whereas US BI-RADS showed significantly higher specificity (90.7% vs 57.5%, P<0.001). US BI-RADS showed better diagnostic efficiency in differentiating nodules in dense breasts than MRI BI-RADS (AUC 0.93 vs 0.78, P<0.001).
Conclusion: By combining the use of ES and conventional US, US BI-RADS had better diagnostic efficiency in differentiating nodules in dense breasts than MRI. For the diagnosis of malignant tumors in patients with dense breasts, MRI and US BI-RADS can be used as supplemental diagnostic tools to detect lesions, with US BI-RADS considered the preferred adjunctive resource. © The author(s).

Entities:  

Keywords:  Breast Imaging Reporting and Data System (BI-RADS); Breast cancer; Dense breast; Elastography; Magnetic resonance imaging (MRI); Ultrasound (US)

Year:  2020        PMID: 32328194      PMCID: PMC7171498          DOI: 10.7150/jca.43326

Source DB:  PubMed          Journal:  J Cancer        ISSN: 1837-9664            Impact factor:   4.207


Introduction

Mammography is a standard screening test that has been proven to reduce breast cancer-related mortality 1,2. However, dense glandular structures reduce the sensitivity of mammography, causing delayed diagnosis and worse outcomes 3. Ultrasound and magnetic resonance imaging (MRI) as adjuncts to mammography can aid in assessing breast lesions in dense breasts. US is a commonly used modality for detecting early breast cancers in dense breasts 4,5. However, the sonographic appearance of benign and malignant nodules overlaps to some extent, causing overtreatment biopsies that are a major limitation of US. US elastography (ES) is a new technique that improves the diagnostic value of US 6,7. ES has been incorporated into the fifth edition of the Breast Imaging Report and Data System (BI-RADS) of the American College of Radiology (ACR), which suggests the risk stratification of breast lesions based on suspicious conventional US features and elastography 8. MRI has been considered the most sensitive screening modality for breast cancer. However, its average specificity is relatively low and varies according to the tumor indications 9. A previous study performed in Korea showed that the addition of elastography and color Doppler US to B-mode US can increase the PPV of screening US in women with dense breasts 10. However, they did not include MRI information and failed to compare the diagnostic value of US and MRI for lesions in dense breasts. The objective of this study was to explore the potential of combining US and US elastography and compare the diagnostic value of US and MRI for predicting breast cancer in dense breasts.

Materials and Methods

Patients

This was a multicenter study conducted at regional medical centers in China, including 35 hospitals from 23 different provinces. All hospitals completed real-name registration on the website (www.nuqcc.cn) and a data survey after approval. The study prospectively enrolled 3292 patients with breast lesions who underwent biopsy or surgery at the 35 hospitals between November 2018 and October 2019. The following inclusion criteria were applied: (1) female patients older than 18 years of age; (2) patients assessed as having dense breast by mammography; and (3) patients for whom conventional US, elastography and MRI screening were performed. Patients who received treatment before surgery were excluded. A total of 242 patients were finally included [Figure 1]. All of these patients underwent breast ultrasound examination prior to core needle biopsy or surgical pathology. The final pathologic results were considered the diagnostic gold standard. The clinical features of the patients were recorded.
Figure 1

Flow chart of patient selection

Breast examination and prospective evaluation

All US examinations were performed with Resona7 or 8 devices (Mindray Medical, Shenzhen, China) equipped with 5-14 MHz linear-array transducers. US images were prospectively evaluated by 35 radiologists who were experienced in breast US and were blinded to the patient clinical data. First, conventional US images of the lesions were obtained, including B-mode US and color Doppler images. The tumor size, shape, echogenicity, growth orientation, margin, and contour, the presence of architectural distortion, the presence of duct ectasia, acoustic shadowing and microcalcifications were evaluated by B-mode US. Vascularity was classified into 4 patterns (no flow, minimal, moderate, or marked) by color Doppler flow 11. After conventional US, elastography images were generated by the same radiologists. Each lesion was assigned an elasticity score according to a 5-point scoring system 12. Based on B-mode US, color Doppler and elastography, the radiologists classified the degree of suspicion of breast lesions according to the BI-RADS criteria 8. MRI images were acquired on different scanners at the 35 referring region medical centers. All examination protocols included a T2-weighted and/or STIR sequence as well as T1-weighted contrast-enhanced dynamic images before and after single-dose Gd-based contrast media injection at 1.5 or 3 T, all in line with EUSOBI and EUSOMA recommendations 13,14. Based on the MRI, the radiologists classified the degree of suspicion of breast lesions according to the BI-RADS criteria 8. All radiologists completed real-name registration on the website (www.nuqcc.cn), and all the patients' images were uploaded. All the data and images from the website were separately reviewed by three experienced radiologists in our hospitals. In cases involving a discrepancy a consensus was reached after discussion.

Statistical analysis

Quantitative data are presented as the means ± standard deviations (SDs). Qualitative data are presented as frequencies. To assess the correlations between features and axillary lymph node metastasis, the χ2-test was used. The sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV) and accuracy were calculated through a comparison with the pathological findings. Receiver operating characteristic (ROC) curve analysis was performed. A P value <0.05 was considered statistically significant. Statistical analyses were performed with SPSS software (Version 19.0, SPSS Chicago, IL, USA) and MedCalc 11.4.2.0 software (MedCalc Software, Ostend, Belgium).

Results

Clinical and pathological status of the patients

The age range of the included patients was 20 to 77 years (median, 48 years). Based on pathology, 169 cases (69.8%) had breast cancer, and 73 (30.2%) did not. Among the patients with breast cancer, 99 had invasive carcinoma, 39 had intraductal carcinoma, one had mucinous carcinoma, 2 had adenocarcinoma, 3 had medullary carcinoma, 3 had a malignant phyllodes tumor, 5 had lobular carcinoma in situ, 3 had ductal carcinoma in situ, 3 had carcinoma with neuroendocrine features, 4 had micropapillary carcinoma, 2 had tubular carcinoma, 4 had papillary carcinoma and one had lymphoma. Among the patients without breast cancer, 27 had fibrocystic disease and adenosis, 26 had fibroadenoma, 10 had intraductal papilloma, 2 had inflammatory and related lesions, one had papilloma associated with fibroadenoma, a benign phyllodes tumor, 2 had a cyst, one case of fat necrosis, one case of sclerosing adenosis, one case of benign breast tissue, and one case of stromal fibrosis [Table 1].
Table 1

Clinical and imaging characteristics of benign and malignant breast lesions

Benign, n (%)Malignant, n (%)TotalP
Age44.5±9.449.3±10.60.001
Age<0.001
<5055(39.0)86(61.0)141
≥5018(17.8)83(82.2)101
First-degree relatives with breast cancer0.70
073(30.3)168(69.7)241
10(0.0)1(100.0)1
Tumor size (cm)<0.001
<268(57.1)51(42.9)119
≥215(12.2)108(87.8)123
Tumor shape0.012
Round, oval73(31.7)157(68.3)230
Irregular0(0.0)12(100.0)12
Growth orientation<0.001
Parallel68(37.0)116(63.0)184
Nonparallel5(8.6)53(91.4)58
Margin<0.001
Circumscribed22(12.0)162(88.0)184
Indistinct51(87.9)7(12.1)58
Contour<0.001
Smooth, lobulated66(45.8)78(54.2)144
Angular7(7.1)91(92.9)98
Acoustic shadowing0.51
Yes11(28.9)27(71.1)38
NO62(30.4)142(69.6)204
Microcalcification<0.001
Yes11(28.9)27(71.1)38
NO62(30.4)142(69.6)204
Echogenicity0.15
Hypo-echoic63(28.9)155(71.1)218
Complex10(41.7)14(58.3)24
Presence of architectural distortion<0.001
Yes0(0.0)33(100.0)33
No73(34.9)136(65.1)209
Presence of duct ectasia0.46
Yes5(26.3)14(73.7)19
No68(30.5)155(69.5)223
CDFI level<0.001
0-154(57.4)40(42.6)94
2-319(12.8)129(87.2)148
ES score<0.001
15(71.4)2(28.6)7
222(75.9)7(24.1)29
336(48.6)38(51.4)74
410(11.2)79(88.8)89
50(0.0)43(100.0)43
ES score<0.001
1-363(57.3)47(42.7)110
4-510(7.6)122(92.4)132
US BI-RADS category<0.001
21(100.0)0(0.0)1
326(96.3)1(3.7)27
4a39(83.0)8(17.0)47
4b5(16.1)26(83.9)31
4c2(2.6)76(97.4)78
50(0.0)58(100.0)58
US BI-RADS category<0.001
3-4b66(88.0)9(12.0)75
4b-57(4.2)160(95.8)167
MRI BI-RADS category<0.001
14(80.0)1(20.0)5
28(100.0)0(0.0)8
330(93.8)2(6.3)32
431(20.8)118(79.2)149
50(0.0)48(100.0)48
MRI BI-RADS category<0.001
0-342(93.3)3(6.7)45
4-531(15.7)166(84.3)197

Abbreviations: BI-RADS: Breast Imaging-Reporting and Data System; US: ultrasound; MRI: magnetic resonance imaging; ES: elastography; CDFI: color doppler flow imaging

Clinical and imaging characteristics of benign and malignant breast lesions

Comparisons of US and clinical features between malignant and benign lesions are shown in Table 1. There were significant differences in age (P<0.001), tumor size (P<0.001), tumor shape (P=0.012), growth orientation (P<0.001), margin (P<0.001), contour (P<0.001), microcalcifications (P<0.001), presence of architectural distortions (P<0.001), color Doppler flow (P<0.001), elastography (ES) score (P<0.001) and US BI-RADS category (P<0.001) between the malignant and benign groups. However, echogenicity (P=0.15), acoustic shadowing (P=0.51) and the presence of duct ectasia (P=0.46) were not associated with lesion malignancy [Table 1]. The overall MRI BI-RADS categories were also significantly different (P<0.001) between the malignant and benign groups.

Diagnostic performance of US and MRI BI-RADS by category

In terms of the US BI-RADS category, the ROC curves demonstrated that the best cut-off value was US BI-RADS 4b. The sensitivity, specificity, PPV, NPV, accuracy and AUC were 94.7%, 90.7%, 95.8%, 88.0%, 93.4% and 0.93 (95%CI, 0.88-0.97), respectively. In terms of the MRI BI-RADS category, the ROC curves demonstrated that the best cut-off value was MRI BI-RADS 4. The sensitivity, specificity, PPV, NPV, accuracy and AUC were 98.2%, 57.5%, 84.3%, 83.3%, 86.0% and 0.78 (95%CI, 0.71-0.85), respectively. When combining the optimal US BI-RADS and MRI BI-RADS categories, the sensitivity, specificity, PPV, NPV, accuracy, and AUC were 92.9%, 91.8%, 96.3%, 84.8%, 92.6% and 0.93 (95%CI, 0.88-0.97), respectively. The diagnostic value of the US BI-RADS category was better than that of the MRI BI-RADS category alone (AUC 0.93 vs 0.78, P<0.001). The US BI-RADS category showed a significantly higher specificity than the MRI BI-RADS category (90.7% vs 57.5%, P<0.001), whereas the MRI BI-RADS category yielded a higher sensitivity (98.2% vs 94.7%, P=0.043). The diagnostic value of combining the US BI-RADS with the MRI BI-RADS category is equal to the US BI-RADS stratification alone (AUC 0.93 vs 0.93, P=0.81) [Table 2].
Table 2

Diagnostic efficiency of US and MRI BI-RADS for predicting breast cancer in dense breasts

Sensitivity (%)Specificity (%)PPV (%)NPV (%)Accuracy (%)AUC (95%CI)
MRI BI-RADS category98.257.584.383.386.00.78 (0.71-0.85)
US BI-RADS category94.790.795.888.093.40.93 (0.88-0.97)
MRI and US BI-RADS category92.991.896.384.892.60.93 (0.88-0.97)

Abbreviations: BI-RADS: Breast Imaging-Reporting and Data System; US: ultrasound; MRI: magnetic resonance imaging

Discussion

Dense breasts can hide nodules on mammography, and a negative result on mammography does not reliably exclude the presence of breast lesions. The most widely available supplemental screening options for patients with dense breast tissue are US and MRI, but there has been a lack of information to guide the decision to utilize one or the other versus both. The results of the present study showed that US showed a higher specificity for detecting malignant breast lesions, whereas MRI yielded a higher sensitivity. Overall, US showed better diagnostic efficiency in differentiating nodules in dense breasts than MRI. The results also showed that the diagnostic value of US and MRI combined was equal to the value of US alone. US has been applied to more than 200,000 women and is capable of increased lesion detection in addition to mammography for women with dense breasts, similar to the results of this study 15-16. Among the conventional US characteristics of breast lesions, this study showed that irregular tumor shape, nonparallel growth orientation, indistinct margin, angular contour, microcalcification and color Doppler flow on US imaging were significantly related to breast cancer in dense breasts. However, there are several barriers to implementing screening US in practice. One of these has been the high rates of false positives from US. A recent study showed a PPV of 48% for biopsies performed based on the results of screening US in women at average risk for breast cancer with dense breasts 17. The revisions to the BI-RADS Fifth Edition (2013) grant ES a complementary role in the ultrasonic diagnosis of breast nodules. A 5-point scale was adopted according to the hardness of the nodules in strain elastography 18. Previous studies have reported an increase in accuracy when combining B-mode US and ES 19-20. A previous study reported that the addition of elastography and color Doppler US to B-mode US in dense breasts increased the AUC to 0.96 and specificity to 76.4% without loss of sensitivity 10. In this study, similarly, the highest AUC and specificity of US were achieved when elastography was added to conventional US for diagnosing breast cancer in dense breasts. Moreover, the fifth BI-RADS category had a PPV of 95.8%, and the high PPV helped reduce the number of false-positive findings without missing cancers. A relatively high PPV helps reduce unnecessary biopsies. For high-risk women of any breast density, supplemental screening with annual MRI has been proven to reduce late-stage disease and increase metastasis-free survival 21-22. According to a systematic review, the use of MRI for high-risk women improved the sensitivity of lesion detection by mammography from 32 to 84% 23. Moreover, the use of supplemental MRI screening in women with extremely dense breast tissue and normal results on mammography resulted in the diagnosis of significantly fewer interval cancers than mammography alone 24-25. MRI has been considered the most sensitive screening modality for women with dense breasts 26. In this study, MRI also performed well in differentiating breast nodules in dense breasts, exhibiting a higher sensitivity (98.2%) than US. Previous studies have shown that conventional ultrasound did not improve detection over MRIs and that conventional US leads to a greater number of false-positive breast cancers compared to MRI 27. In this study, combined with ES, US showed higher specificity and had a better diagnostic efficiency in differentiating nodules in dense breasts than MRI. There are several limitations to the study. First, the 35 hospitals are referral cancer centers instead of community hospitals, which may have resulted in an increase in the malignancy ratio of the nodules in the patient population. Second, all of the patients had pathology results. Therefore, the study included more malignant breast nodules than benign nodules, which may have led to selection bias and resulted in the underestimation of the NPV and the overestimation of the PPV. Third, potential biases may have been present because the many lesions detected during the study period were not subjected to MRI examinations, as these were performed at the discretion of the patient. Fourth, because 35 radiologists performed the MRI and US examinations, there may have been interobserver differences. In conclusion, this study found that tumor shape, growth orientation, margin, microcalcification, color Doppler flow and ES score on US were independently associated with breast cancer in patients with dense breasts. US BI-RADS showed a higher specificity, and MRI yielded a higher sensitivity. US BI-RADS had a better diagnostic efficiency in differentiating nodules in dense breasts than MRI. For the diagnosis of malignancy in patients with dense breasts, MRI and US BI-RADS are supplemental diagnostic tools to detect lesions, with US BI-RADS being the preferred adjunctive resource.
  23 in total

1.  American Cancer Society guidelines for breast screening with MRI as an adjunct to mammography.

Authors:  Debbie Saslow; Carla Boetes; Wylie Burke; Steven Harms; Martin O Leach; Constance D Lehman; Elizabeth Morris; Etta Pisano; Mitchell Schnall; Stephen Sener; Robert A Smith; Ellen Warner; Martin Yaffe; Kimberly S Andrews; Christy A Russell
Journal:  CA Cancer J Clin       Date:  2007 Mar-Apr       Impact factor: 508.702

2.  Dense Breast Ultrasound Screening After Digital Mammography Versus After Digital Breast Tomosynthesis.

Authors:  Elizabeth H Dibble; Tisha M Singer; Nneka Jimoh; Grayson L Baird; Ana P Lourenco
Journal:  AJR Am J Roentgenol       Date:  2019-09-25       Impact factor: 3.959

3.  Prospective study of breast cancer incidence in women with a BRCA1 or BRCA2 mutation under surveillance with and without magnetic resonance imaging.

Authors:  Ellen Warner; Kimberley Hill; Petrina Causer; Donald Plewes; Roberta Jong; Martin Yaffe; William D Foulkes; Parviz Ghadirian; Henry Lynch; Fergus Couch; John Wong; Frances Wright; Ping Sun; Steven A Narod
Journal:  J Clin Oncol       Date:  2011-03-28       Impact factor: 44.544

4.  Detecting Breast Cancer in Women with Dense Breasts.

Authors:  Dan L Longo
Journal:  N Engl J Med       Date:  2019-11-28       Impact factor: 91.245

5.  Swedish two-county trial: impact of mammographic screening on breast cancer mortality during 3 decades.

Authors:  László Tabár; Bedrich Vitak; Tony Hsiu-Hsi Chen; Amy Ming-Fang Yen; Anders Cohen; Tibor Tot; Sherry Yueh-Hsia Chiu; Sam Li-Sheng Chen; Jean Ching-Yuan Fann; Johan Rosell; Helena Fohlin; Robert A Smith; Stephen W Duffy
Journal:  Radiology       Date:  2011-06-28       Impact factor: 11.105

6.  Current Status of Supplemental Screening in Dense Breasts.

Authors:  Wendie A Berg
Journal:  J Clin Oncol       Date:  2016-03-09       Impact factor: 44.544

Review 7.  Systematic review: using magnetic resonance imaging to screen women at high risk for breast cancer.

Authors:  Ellen Warner; Hans Messersmith; Petrina Causer; Andrea Eisen; Rene Shumak; Donald Plewes
Journal:  Ann Intern Med       Date:  2008-05-06       Impact factor: 25.391

Review 8.  Ultrasound as an Adjunct to Mammography for Breast Cancer Screening: A Health Technology Assessment.

Authors: 
Journal:  Ont Health Technol Assess Ser       Date:  2016-07-01

9.  Combined screening with ultrasound and mammography vs mammography alone in women at elevated risk of breast cancer.

Authors:  Wendie A Berg; Jeffrey D Blume; Jean B Cormack; Ellen B Mendelson; Daniel Lehrer; Marcela Böhm-Vélez; Etta D Pisano; Roberta A Jong; W Phil Evans; Marilyn J Morton; Mary C Mahoney; Linda Hovanessian Larsen; Richard G Barr; Dione M Farria; Helga S Marques; Karan Boparai
Journal:  JAMA       Date:  2008-05-14       Impact factor: 56.272

10.  Breast MRI: guidelines from the European Society of Breast Imaging.

Authors:  R M Mann; C K Kuhl; K Kinkel; C Boetes
Journal:  Eur Radiol       Date:  2008-04-04       Impact factor: 5.315