Dentinal Microcrack Formation after Root Canal Instrumentation by XP-Endo Shaper and ProTaper Universal: A Microcomputed Tomography Evaluation.

Sarah M Alkahtany1, Ebtissam M Al-Madi1.   

Abstract

AIM: To evaluate dentinal microcrack formation on root canals instrumented, continuously in the body temperature, with XP-endo shaper (XPES) and ProTaper Universal (PTU), by means of microcomputed tomographic (micro-CT) analysis. Methodology. Nineteen mesial roots with two separate canals (Vertucci Type IV) of extracted mandibular molars were used in this study. The root canals (N = 38) were divided into 2 groups. Group 1 (n = 19): all MB canals were instrumented with XPES. Group 2 (n = 19): all ML canals were instrumented with PTU. All roots were scanned with micro-CT before and after instrumentation. Two precalibrated examiners evaluated the cross-sectional images of each sample with DataViewer program. The dentinal microcracks (complete and incomplete) were counted in each third of the root for the preinstrumentation and the postinstrumentation images. Wilcoxin signed-rank and Mann-Whitney U tests were used for statistical analysis at a significance level of P < 0.05.
RESULTS: The number of microcracks increased significantly (P < 0.05) after instrumentation with XPES in the middle and cervical thirds. The number of microcracks increased significantly (P < 0.05) after instrumentation with PTU in the cervical third only. There was no significant difference between the groups in the cervical and apical thirds. In the middle third, the XPES induced more incomplete microcracks than PTU (P < 0.05).
CONCLUSION: Within the limitations of this study, there was no significant difference in the dentinal microcrack formation between XPES and PTU in the apical and cervical thirds of the root. However, XPES instrumentation induced more incomplete microcracks than PTU in the middle third of human roots.
Copyright © 2020 Sarah M. Alkahtany and Ebtissam M. Al-Madi.

Entities:  

Year:  2020        PMID: 32322272      PMCID: PMC7168705          DOI: 10.1155/2020/4030194

Source DB:  PubMed          Journal:  Int J Dent        ISSN: 1687-8728


1. Introduction

Disinfection of the root canal system is essential for successful root canal treatment [1]. The antibacterial effect of sodium hypochlorite cannot reach all bacteria in the dentinal tubules [2]. Therefore, root canal mechanical enlargement is required to ensure the removal of infected dentine [3]. Cleaning and shaping procedures have significantly improved with the use of NiTi rotary instruments [4]. However, NiTi files might lead to dentinal defects and can induce microcracks in the dentinal walls of the root canal [5]. Microcomputed tomography (micro-CT) is the method of choice to evaluate and assess dentinal defects and microcracks induced by root canal instrumentation with different systems. It allows the investigators to evaluate hundreds of axial sections per tooth to accurately detect the exact location of a microcrack [6]. Furthermore, micro-CT is a nondestructive and noninvasive technique to obtain two-dimensional and three-dimensional images of any tooth [7]. It enables scanning of the same sample for multiple tests without damage allowing each sample to be used as its own control [8]. The ProTaper Universal (PTU) (Dentsply Tulsa Dental Specialties, Tulsa, OK) Ni–Ti rotary system is one of the most commonly used files. It is machined from conventional superelastic (SE) austenitic Ni-Ti wire. It features a variable taper over the entire cutting blade length with convex triangular cross sections. This file design can help clinicians properly instrument and flare canals with anatomical difficulties [9]. However, PTU has been reported to be associated with a high incidence of microcrack formation [5, 10–15]. The XP-endo shaper (XPES) (FKG Dentaire, Switzerland) is made of MaxWire alloy, a martensite-austenite-electropolish thermomechanically treated NiTi alloy. This file will curve on exposure to body temperature, due to the phase transformation from the M-phase (martensitic state) to the A-phase (austenitic state) [16]. The manufacturer claims that the flexibility and preset shape enable the XPES to contract and expand within the canal itself and to reach areas that conventional files cannot access. Furthermore, XPES has an ISO size 30 diameter and 0.01 taper, which could minimize the physical stresses on the canal dentinal wall. Recent publications reported that the XPES system performed well in root canal instrumentation including severely curved canals but leaves untouched dentinal wall areas [17, 18]. Previous studies reported that XPES instrumentation will cause no or few dentinal microcracks compared with other NiTi rotary systems [19-21]. None of these studies exposed the XPES files to body temperature during the instrumentation. Therefore, in the present study, we aim to evaluate dentinal microcrack formation on root canals instrumented, continuously in the body temperature, with the XP-endo shaper (XPES) and ProTaper Universal (PTU), by means of microcomputed tomographic (micro-CT) analysis.

2. Materials and Methods

This research was conducted in King Saud University, Riyadh, Saudi Arabia. The research protocol was approved by the Institutional Review Board (E-17-2646).

2.1. Specimen Selection

Thirty-six extracted mandibular molars were collected, sterilized in 10% buffered formalin. All teeth were decoronated, and the lengths of roots were standardized to 16 mm. The roots were split at the furcation area by using ISOMET 2000 PRECISION SAW (Buehler, USA). Straight and angulated conventional radiographs were taken for all mesial roots to verify the inclusion criteria. The inclusion criteria were as follows: mesial roots with two separate canals (Vertucci Type IV), free from calcifications and pulp stones, and a root curvature between 10° and 30° as verified by measurement of Schneider [22]. Finally, nineteen mesial roots with thirty-eight root canals were selected and included in this study (N = 38). Each root was mounted in clear acrylic block with a mark on the buccal side. This mark will assist the positioning of each sample in the same orientation for pre- and postinstrumentation micro-CT scan.

2.2. Preinstrumentation Micro-CT Scan

The roots were scanned before instrumentation (preinstrumentation scan) with Skyscan1172 (Bruker, USA) 100 kV/98 μA with a Hamamatsu 10-MP camera. The camera pixel size was 11.40 μm with median filtering and flat-field correction.

2.3. Sample Preparation

The working length of all the canals was determined and confirmed by radiographs. Coronal flaring was performed for all canals by using a Gates-Glidden size 2, followed by preparation of a glide path for all canals with hand files (K file) up to size 15. RC-Prep® (Premier Dental, USA) was used as the lubricant. Canals were irrigated with 1 mL of 5% sodium hypochlorite (NaOCl) before and after each file. The mesial root canals (N = 38) were divided into two groups: XPES group (n = 19): mesiobuccal canals were instrumented with the XPES system. The files were mounted on the X-smart handpiece (Dentsply Tulsa Dental Specialties, Tulsa, OK) and used at a speed of 800 rpm and 1 N·cm, according to the manufacturer's instructions. Each file was used in 3 gentle strokes to the full working length. RC-Prep® was used as the lubricant, and 1 mL of 5% NaOCl was used for irrigation after instrumentation. PTU group (n = 19): mesiolingual canals were instrumented with the PTU system. The files were mounted on the X-smart handpiece and used at a speed of 300 rpm and 1 N·cm, according to the manufacturer's instructions. The canals were instrumented with S1, S2, F1, F2, and F3. Each file was lubricated with RC-Prep®, and 1 mL of 5% NaOCl was used for irrigation after each file. All the roots were submerged in a water bath at 37°C during instrumentation to mimic the body temperature. All files were used 3 times and then discarded to prevent separation.

2.4. Postinstrumentation Micro-CT Scan

After instrumentation, all samples were scanned again with Skyscan1172 (100 kV/98 μA) with the Hamamatsu 10-MP camera. The camera pixel size was 11.40 μm with median filtering and flat-field correction.

2.5. Dentinal Microcrack Evaluation

Two precalibrated examiners evaluated cross-sectional images of each sample with the DataViewer program (version 1.5.2.4, Bruker, USA). Each reconstructed root image was divided into thirds (cervical, middle, and apical). Next, dentinal microcracks (complete and incomplete) were counted in each third of the root in the preinstrumentation and postinstrumentation images.

2.6. Statistical Analysis

Inter-rater reliability was assessed by calculating the percentage agreement between the two examiners. The data were analyzed statistically and summarized using the chi-squared test to calculate the percentage of microcracks in each group. A comparison between the number of microcracks before and after instrumentation was performed with the Wilcoxon signed-rank test. The Mann–Whitney U test was used to compare the differences between the XPES and PTU groups at a significance level of P < 0.05 by using IBM SPSS® Statistics.

3. Results

A total of 17,430 cross sections were evaluated by two examiners. The inter-rater percentage agreement was excellent (90%). The percentages of complete microcracks in each group are illustrated in Figure 1, while the percentages of incomplete microcracks in each group are illustrated in Figure 2. The total number of complete and incomplete microcracks in each group before and after instrumentation is shown in Table 1.
Figure 1

Percentage of complete microcracks before and after instrumentation in the XPES and PTU groups.

Figure 2

Percentage of incomplete microcracks before and after instrumentation in the XPES and PTU groups.

Table 1

The number of complete and incomplete microcracks before and after instrumentation with XPES and PTU.

Type of microcrackGroupsNumber of microcracks
ApicalMiddleCervical
CompleteXPESBefore018
After0319
P value1.0000.1570.0001∗
PTUBefore1211
After1520
P value1.0000.0830.011∗
XPES vs. PTU P value1.0000.6360.685

IncompleteXPESBefore32341
After34145
P value1.0000.0001∗0.465
PTUBefore82735
After103737
P value0.3170.3170.627
XPES vs. PTU P value0.3170.025∗0.662

∗ Statistically significant difference.

Most micro-CT images showed microcracks in the cervical and middle thirds before the instrumentation (52%–79% incomplete and 5%–37% complete microcracks). In general, the number of microcracks in the cervical and middle thirds increased after instrumentation (Figures 3 and 4). The number of complete microcracks in the cervical third increased significantly (P < 0.05) after instrumentation in both XPES and PTU groups, while the number of incomplete microcracks in the middle third increased significantly (P < 0.05) after instrumentation only in the XPES group (Figure 4). The apical third in all groups did not show any complete microcracks and only a few incomplete microcracks (Figure 5). There was no significant increase in the number of microcracks in the apical third after instrumentation.
Figure 3

Micro-CT cross-sectional image of the cervical third of mandibular molar mesial root: (a) Preinstrumentation; (b) postinstrumentation. The MB canal (left) instrumented with XPES and the ML canal (right) instrumented with PTU. The microcracks that appeared in the postinstrumentation image for both canals were the propagation of previous microcracks observed in the preinstrumentation image.

Figure 4

Micro-CT cross-sectional image of the middle third of mandibular molar mesial root: (a) preinstrumentation; (b) postinstrumentation. The MB canal instrumented with XPES (left) and the ML canal instrumented with PTU (right). One microcrack that appeared in the postinstrumentation image in the ML canal (PTU group) was the same as the preexisting microcrack observed in the preinstrumentation image. However, in the XPES group, two new incomplete microcracks were observed in the postinstrumentation image.

Figure 5

Micro-CT cross-sectional image of the apical third of mandibular molar mesial root. (a) Preinstrumentation; (b) postinstrumentation. The MB canal instrumented with XPES (left) and ML canal instrumented with PTU (right). No microcracks could be detected in the pre- or postinstrumentation images.

There were no significant intergroup differences in the number of incomplete microcracks in the cervical and apical thirds. However, in the middle third, the XPES induced significantly more incomplete microcracks than PTU.

4. Discussion

Endodontic practice aims to restore and preserve remaining natural dentition using safe instruments and techniques. However, mechanical root canal preparation might induce microcracks that could propagate to root fractures, leading to poor prognosis [11, 23]. Therefore, it is essential to assess the safety of any new file including the incidence of dentinal microcrack formation. The present study used micro-CT evaluation to compare the dentinal microcrack formation on root canals instrumented with PTU and XPES. The results showed that most of the microcracks seen in postinstrumentation images were present in the preinstrumentation images. However, many postinstrumentation complete microcracks were incomplete microcracks in the preinstrumentation images. According to Stringhet et al., this change in the type of microcracks was due to root canal lumen enlargement rather than a true propagation of the previous incomplete microcrack [6]. No attempt was made in this study to measure the length of the microcracks. Only the number and the type of microcracks (complete or incomplete) were evaluated. Our results demonstrated that instrumentation with both tested files induced a few new microcracks in root canal walls. However, the increase in the number of microcracks was statistically significant only in the middle and cervical thirds. There was a significant increase in the number of complete microcracks after instrumentation in the cervical third in both groups. This could be due to the coronal flaring with Gates–Glidden drills or due to rotary file movement. Furthermore, the results showed a significant increase in the number of incomplete microcracks in the middle third in the XPES group in comparison with the PTU group. This might be attributed to the high-speed rotation (800 rpm) of the file and/or the nature of the movement of XPES. When the file is exposed to body temperature, it can contract and expand while rotating inside the canal due to its flexibility and preset shape. During our experiment, the operator experienced more vibration while using XPES compared with PTU. Our results disagree with the findings of the previous studies. Bayram et al. compared the number of microcracks induced by XPES and ProTaper Gold (PTG), and their results showed that the PTG system significantly increased the incidence of microcracks, while the XPES system did not induce any new dentinal microcracks [19]. Ugur Aydin et al. compared the percentages of new microcracks formed after instrumentation with Reciproc Blue, XPES, and WaveOne Gold. Their investigation concluded that none of the used rotary systems caused new microcracks formation or propagation of existing microcracks [20]. Furthermore, our results contradict the findings of Aksoy et al.; their study concluded that PTU induced more microcracks than XPES [21]. This disagreement might be attributed to differences in methodology; in our study, the file was tested and used at body temperature (37°C) throughout the instrumentation. However, in the previous studies, the file was exposed to body temperature only once before the instrumentation. In the present investigation, we did not use the fresh cadaveric model suggested by De-Deus et al. because this model was not readily available in most institutes, inluding ours [24]. Therefore, human extracted teeth were used, although a previous investigation reported that extracted teeth showed some microcracks induced by the extraction procedure itself [25]. In our study, all preexisted microcracks were recorded in the preinstrumentation micro-CT images, and they were not related to the root canal preparation. The results of this study should be interpreted with caution due to some limitations. All extracted teeth were decoronated before the instrumentation; this was performed to limit the variation between the root lengths. Furthermore, the sectioned roots were mounted directly in hard acrylic blocks. However, these conditions do not represent a real clinical scenario. Moreover, the root canals were not randomly distributed between the groups; this might cause sampling bias. Finally, our results might be affected by the use of Gates-Glidden for preflaring of root canals. The solo use of an endodontic file inside the root canal is recommended. We recommend future investigators to use soft material around the teeth before mounting them in acrylic blocks and to complete the root canal instrumentations through the crowns without any sectioning to have more realistic results.

5. Conclusion

Within the limitations of this study, there was no significant difference in the dentinal microcrack formation between XP-endo shaper (XPES) and ProTaper Universal (PTU) in the apical and cervical thirds of the root. However, XPES instrumentation induced more incomplete microcracks than PTU in the middle third of human roots.
  25 in total

1.  Extension of bactericidal effect of sodium hypochlorite into dentinal tubules.

Authors:  Debbie T S Wong; Gary S P Cheung
Journal:  J Endod       Date:  2013-11-09       Impact factor: 4.171

2.  A comparison of canal preparations in straight and curved root canals.

Authors:  S W Schneider
Journal:  Oral Surg Oral Med Oral Pathol       Date:  1971-08

Review 3.  New thermomechanically treated NiTi alloys - a review.

Authors:  J Zupanc; N Vahdat-Pajouh; E Schäfer
Journal:  Int Endod J       Date:  2018-04-19       Impact factor: 5.264

4.  Comparison of dentinal and apical crack formation caused by four different nickel-titanium rotary and reciprocating systems in large and small canals.

Authors:  Xin Zhou; Shan Jiang; Xiao Wang; Shuai Wang; Xiaofei Zhu; Chengfei Zhang
Journal:  Dent Mater J       Date:  2015       Impact factor: 2.102

Review 5.  State of the art of Micro-CT applications in dental research.

Authors:  Michael V Swain; Jing Xue
Journal:  Int J Oral Sci       Date:  2009-12       Impact factor: 6.344

6.  Stereomicroscopic evaluation of dentinal defects induced by new rotary system: "ProTaper NEXT".

Authors:  Deepa Deepak Shori; Pratima Ramakrishna Shenoi; Arshia R Baig; Rajesh Kubde; Chetana Makade; Swapnil Pandey
Journal:  J Conserv Dent       Date:  2015 May-Jun

Review 7.  Microcomputed Tomography Studies of the Effectiveness of XP-endo Shaper in Root Canal Preparation: A Review of the Literature.

Authors:  Christianne Velozo; Diana Albuquerque
Journal:  ScientificWorldJournal       Date:  2019-08-19

8.  A Stereomicroscopic Evaluation of Dentinal Cracks at Different Instrumentation Lengths by Using Different Rotary Files (ProTaper Universal, ProTaper Next, and HyFlex CM): An Ex Vivo Study.

Authors:  Fauzia Ashraf; Pushpa Shankarappa; Abhinav Misra; Asheesh Sawhney; Nandamuri Sridevi; Anu Singh
Journal:  Scientifica (Cairo)       Date:  2016-06-29

9.  Comparative Analysis of Apical Root Crack Propagation after Root Canal Preparation at Different Instrumentation Lengths Using ProTaper Universal, ProTaper Next and ProTaper Gold Rotary Files: An In vitro Study.

Authors:  Sonalkumari Vakilram Nishad; G B Shivamurthy
Journal:  Contemp Clin Dent       Date:  2018-06

10.  Shaping Ability of XP Endo Shaper File in Curved Root Canal Models.

Authors:  Abdulmohsen Alfadley; Abdalrhman Alrajhi; Hamad Alissa; Faisal Alzeghaibi; Lubna Hamadah; Khalid Alfouzan; Ahmed Jamleh
Journal:  Int J Dent       Date:  2020-02-17
View more
  2 in total

1.  Dentinal microcracks induced by endodontic procedures: A scientometric and bibliometric analysis.

Authors:  Stéphanie Quadros Tonelli; Kênia Soares de Toubes; André Maués Brabo Pereira; Antônio Carlos de Oliveira Miranda; Letícia Ladeia Fernandes; Eduardo Nunes; Frank Ferreira Silveira
Journal:  J Conserv Dent       Date:  2022-05-02

2.  Influence of obturating techniques on root dentin crack propagation: A micro-computed tomography assessment.

Authors:  Praveen Kumar Chellapilla; Mohan Rao Boddeda; Mandava Jyothi; Lakshman Varma Uppalapati; Ravi Kumar Konagala; Lohita Dasari
Journal:  J Conserv Dent       Date:  2021-07-05
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.