Ultrasound Standing Wave-Based Cell-to-liquid Separation for Measuring Viscosity and Aggregation of Blood Sample.

Gwangho Kim1, Sanghwa Jeong2, Yang Jun Kang2.   

Abstract

When quantifying mechanical properties of blood samples flowing in closed fluidic circuits, blood samples are collected at specific intervals. Centrifugal separation is considered as a required procedure for preparing blood samples. However, the use of centrifuge is associated with several issues, including the potential for red blood cell (RBC) lysis, clotting activation, and RBC adhesions in the tube. In this study, an ultrasonic transducer is employed to separate RBCs or diluent from blood sample. The ultrasonic radiation force is much smaller than the centrifugal force acting in centrifuge, it can avoid critical issues occurring under centrifuge. Then, the RBC aggregation and blood viscosity of the blood sample are obtained using the microfluidic technique. According to the numerical results, ultrasonic transducers exhibited a maximum quality factor at an excitation frequency of 2.1 MHz. Periodic pattern of acoustic pressure fields were visualized experimentally as a column mode. The half wavelength obtained was as 0.5 λ = 0.378 ± 0.07 mm. The experimental results agreed with the analytical estimation sufficiently. An acoustic power of 2 W was selected carefully for separating RBCs or diluent from various blood samples (i.e., Hct = 20% ~ 50%; diluent: plasma, 1x phosphate-buffered saline (PBS), and dextran solution). The present method was employed to separate fixed blood samples which tended to stack inside the tube while using the centrifuge. Fixed RBCs were collected easily with an ultrasonic transducer. After various fixed blood samples with different base solutions (i.e., glutaraldehyde solution, 1x PBS, and dextran solution) were prepared using the present method, RBC aggregation and the viscosity of the blood sample are successfully obtained. In the near future, the present method will be integrated into ex vivo or in vitro fluidic circuit for measuring multiple mechanical properties of blood samples for a certain longer period.

Entities:  

Keywords:  RBCs aggregation; RBCs separation; blood viscosity; microfluidic device; ultrasonic transducer

Year:  2020        PMID: 32316446      PMCID: PMC7219585          DOI: 10.3390/s20082284

Source DB:  PubMed          Journal:  Sensors (Basel)        ISSN: 1424-8220            Impact factor:   3.576


1. Introduction

To monitor the mechanical properties of blood samples flowing under an extracorporeal rat bypass loop [1,2], blood samples are collected at specific intervals. The repetitive collection of blood sample results in decreasing the volume of blood flowing in the circuit substantially. To maintain a fixed blood volume, 1x phosphate-buffered saline (PBS) is infused. The hemodilution procedure leads to significantly change mechanical properties of blood samples (i.e., blood viscosity, hematocrit, and pressure) [3]. In addition, experimental tests have been limited to durations less than 2 h. The use of different blood storage techniques have led to large variations in the mechanical properties of the blood [4,5]. Among these properties, blood viscosity has been used extensively to evaluate the variations in blood samples, and it has been influenced by several factors, including hematocrit, red blood cells (RBCs) aggregation, RBC deformability, and plasma viscosity. As an important step, a centrifuge is used to separate components (i.e., cells or liquid) from the blood sample. After setting the hematocrit [Hct] to a specific level [6], the blood viscosity is measured with a viscometer. During the centrifuge operation, a strong centrifugal force acts on the blood cells. Fixed RBCs stack and adhere to the surface of the tube because of the strong centrifugal forces. The use of a centrifuge has led to several issues, including RBC lysis, clotting activation, and inflammatory cascades [7]. For the reason, it is difficult to collect fixed RBCs from the tube. Thus, it hinders in adding fixed RBCs into specific diluent (i.e., 1x PBS, plasma, and dextran solution). Our previous studies do not study the effect of diluents on fixed RBCs [8,9,10]. To resolve critical issues (i.e., periodic blood collection and blood separation with a centrifuge), it is necessary to separate RBCs or plasma from blood samples without using a centrifuge. Recently, several separation systems that utilize acoustic standing waves have been proposed to separate serum, lipids, platelets, and exosome from blood samples [11,12,13,14,15,16]. According to previous studies, ultrasonic standing pressure fields are employed to separate RBCs flowing in microfluidic channels. Acoustic radial forces contribute to migrating RBCs laterally along the flow streams. The ultrasonic transducer has a much smaller scale of acoustic radiation force when compared with a centrifugal force. Thus, it has been considered as a promising method that can resolve critical issues related to the use of a centrifuge. However, the methods are not effective to separate large volume of blood sample (~ 5 mL). Additionally, most previous researches only focus on cell separation of blood samples. In this study, instead of a centrifuge, a bulk-sized ultrasonic transducer is suggested and designed to accelerate sedimentation of RBCs in stationary column chamber. To effectively separate RBCs or diluent from blood sample, the performance validation of ultrasonic transducer is validated by means of numerical analysis and experimental demonstration. First, numerical simulations are used to select optimum excitation frequency of acoustic standing wave by evaluating the resonance frequency, acoustic pressure fields, and quality factor. Second, after filling the ultrasonic transducer with the blood sample (Hct = 2%), focusing and sedimentation of RBCs inside the transducer are visualized experimentally with an elapse of time. Third, the contributions of the acoustic power and blood samples (i.e., hematocrit or base solution) to the separation performance are experimentally examined by measuring the separation height (Hs) over time. Fourth, variations of Hs are monitored by changing the degree of the RBC hardness or diluents. After separating RBCs from blood samples using an ultrasonic transducer, the blood sample is adjusted to a specific level of hematocrit by adding RBCs to a specific concentration of diluent. A microfluidic technique is employed to evaluate the mechanical properties of the blood samples (i.e., RBC aggregation and blood viscosity) with respect to the fixed RBCs or diluents of the blood sample.

2. Materials and Methods

2.1. Cell-to-Liquid Separation with Ultrasonic Transducer

As shown in Figure 1A, the experimental setup was composed of an ultrasonic system, a light illumination, and an image acquisition system. An ultrasonic system consisted of an ultrasonic transducer, and a controller (APS 990, AppliSens, Netherlands). The controller generated an electric signal with an excitation frequency of 2.1~2.15 MHz and transmitted it to the ultrasonic transducer. Right side panel showed piezo resonator in ultrasonic transducer. The acoustic power (AP) of the ultrasonic transducer was adjusted to AP = 0, 2, and 3 W by changing the voltage appropriately. A portable light-emitting diode (LED) light was employed to illuminate the RBCs present inside the ultrasonic transducer. A high-speed camera (FASTCAM MINI, Photron, Tokyo, Japan) was used to capture the cell-to-liquid separation (or separation height) inside the ultrasonic transducer. When a function generator (WF1944B, NF Corporation, Yokohama, Japan) triggered the high-speed camera at an interval of 1 s, the snapshot was captured sequentially at a frame rate of 250 Hz. The camera offered a spatial resolution of 1280 × 1000 pixels, and each pixel corresponded to 10 μm.
Figure 1

Proposed method for separating RBCs from blood sample with ultrasonic transducer. (A) Ultrasonic experimental setup, including an ultrasonic transducer and a controller. Right side panel showed piezo resonator in ultrasonic transducer. (B) The working principle of RBC-to-liquid separation with the ultrasonic standing wave. (a) Separation sequences including disaggregation, focusing, aggregation, and sedimentation. (b) Schematic of an ultrasonic transducer and square blood container (each width [L] = 12.3 mm). The ultrasonic transducer consists of a piezo resonator, carrier, and reflector. An ultrasonic standing wave of 2.1 MHz existed inside the blood container, and it remained perpendicular to the gravitational direction. (C) As a preliminary demonstration, the blood sample (normal RBCs in plasma, Hct = 50%, blood volume = 5 mL) was infused into the blood container. (a) Snapshots captured at a specific time (t) (t = 0, 300, 600, 900, and 1000 s) at AP = 0. (b) Snapshots captured at a specific time (t) (t = 0, 300, 600, 900, and 1000 s) at AP = 2 W. Hs represented the liquid height, and was used to quantify the RBC-to-liquid separation. (c) Temporal variations of Hs with respect to AP = 0 and 2 W.

2.2. Design and Analysis of the Ultrasonic Transducer

As shown in Figure 1B(a), the PRF caused the disaggregated cells to rapidly move to either of the nodes or anti-nodes of pressure fields. It contributed to decreasing the average distance between the RBCs. The SRF was induced to aggregate the RBCs continuously until the gravity force of the aggregated RBCs overcame buoyancy and the sedimentation of RBCs began. In the transverse direction, the Bernoulli force (BF) played a significant role in packing the RBCs closely, as well as maintaining their positions. The aggregated RBCs tended to decrease [17]. The PRF in the axis direction (F) was derived as follows: and In Equations (1) and (2), , , and λ represented the pressure amplitude, RBCs volume, and wavelength, respectively. The compressibility (β) was expressed as β = 1/(ρ c2). ϕ (β, ρ) referred to the acoustic contrast factor, which was determined by the density (ρ, ρ) and compressibility (β, β) of cells and fluid, respectively. A positive value of the acoustic contrast factor indicated that cells tended to move toward the node of pressure fields, whereas a negative value of the acoustic contrast factor indicated that the cell tended to move toward the anti-node of pressure fields. According to Equation (1), the PRF was proportional to 2 and . As fluidic flow is considered as negligible in ultrasonic container, the motion of RBCs is determined by three forces (buoyant force, drag force, and gravity force) and size of the aggregated RBCs. Gravity force forces RBCs to move downward. However, as drag force and buoyant force act in upward direction, they retard RBCs sedimentation. When the gravity force of the aggregated RBCs overcomes buoyancy and the drag force, the sedimentation of RBCs begins. For the reason, sedimentation procedure of aggregated RBCs takes a longer time in an ultrasonic container. A previous study reported that standing wave frequency with much lower than 1 MHz resulted in acoustic cavitation or shock waves. However, the pressure field with a standing wave frequency of 1~3 MHz did not impact the cell viability in mammalian cells and RBCs. For this reason, a standing wave frequency with f > 1 MHz was proposed in order to separate micron-sized cells effectively [18]. As shown in Figure 1B(b), the ultrasonic transducer consisted of a piezo resonator, carrier, blood container, and reflector. The blood container had a square shape (each length (L) = 12.3 mm). It was fabricated with transparent Pyrex material for visualizing RBC sedimentation. The resonator was bonded strongly to the carrier with an adhesive. Acoustic energy was transmitted sufficiently to the blood container because of the impedance matching functionality. The size and material property of the ultrasonic transducer are summarized as Table 1.
Table 1

Summary of size and material property of an ultrasonic transducer.

LayerPiezoelectricCarrierBlood ContainerReflector
ℓ (mm)0.55.1412.35.14
MaterialPIC181PyrexPlasmaRBCsPyrex
ρ (kg/m3)78502200102510932200
v (m/s)44605400157016455400
RemarkAs acoustic contrast factor took positive value (ϕ = 0.21), RBCs tended to focus and aggregate at the nodes of pressure fields.
In this study, the excitation frequency of the ultrasonic standing wave was kept constant at f = 2.1 MHz. When the blood container was filled with pure plasma or pure RBCs, the value of wavelength (λ) was in the range 0.75 ~ 0.78 mm. In addition, the acoustic contrast factor was calculated as a positive value (i.e., ϕ = 0.21 > 0). Because the wavelength was much greater than the size of RBCs, the acoustic standing wave led to acoustic streaming flow. Thus, the RBCs were focused and aggregated toward the nodes of acoustic pressure fields.

2.3. RBC Separation and Sedimentation Using an Ultrasonic Transducer

As a preliminary demonstration, the blood sample (normal RBCs suspended in plasma, Hct = 50%, blood volume = 5 mL) was infused into the blood container. By setting the acoustic power as AP = 0 and 2 W, the RBCs-to-plasma separation was monitored over time. Figure 1C(a) shows snapshots captured at a specific time (t = 0, 300, 600, 900, and 1000 s) at AP = 0 (i.e., no ultrasonic standing wave). The experimental result indicates that the RBCs were not separated from the blood samples under a gravitational force. In other words, under the condition of no ultrasonic standing wave, the gravitational force alone did not contribute to the separation of RBCs from the blood sample (Hct = 50%). Figure 1C(b) shows the snapshots obtained at a specific time (t = 0, 300, 600, 900, and 1000 s) at AP = 2 W. The experimental result indicated that the acoustic radiation force contributed significantly to the separation of RBCs from the blood sample (Hct = 50%) when compared with the experimental results under pure gravitational force. To evaluate the RBC-to-plasma separation over time, the separation height (Hs) was defined as the liquid height. It was quantified with a digital image processing with Matlab (Ver. 2018a, Mathworks, USA). After a gray-scale image was converted into a binary-scale image using Otsu’s method [19], Hs was obtained by taking the arithmetic average of image intensity over the specific ROI (300 × 820 pixels) inside the blood container. As shown in Figure 1C(c), temporal variations of Hs were obtained with respect to AP = 0 and 2 W. Under no acoustic force (i.e., AP = 0), Hs was estimated as Hs = 1 mm after an elapse of 1100 s. However, when setting the acoustic power as AP = 2 W, Hs tended to increase significantly. The RBCs were separated largely from plasma (i.e., Hs = 7 mm at t = 1100 s). From the preliminary experiment, the acoustic radiation force contributed to the separation of RBCs or plasma from the blood sample sufficiently when compared with pure gravitational force.

3. Results and Discussion

3.1. Numerical Simulation for Estimating Acoustic Pressure Fields Inside Ultrasonic Transducer

When an AC potential was applied to the piezo resonator, the dynamic behavior of the ultrasonic transducer system was estimated as the electrical admittance with respect to frequency (f) [20]. Using the commercial software (COMSOL Ver. 5.4, COMSOL Inc., USA), variations in electric admittance were obtained by varying the frequency from f = 1.5 MHz to f = 2.5 MHz. Figure 2A shows the electrical admittance spectrum of the ultrasonic transducer (i.e., multi-layer resonator) filled with a blood sample. The spectrum represented that several resonant frequencies were present between f = 1.76 MHz and f = 2.212 MHz. The resonant frequency of 2.212 MHz was located near the fundamental resonant frequency of the piezo resonator (i.e., 2.2 MHz) [17]. According to the simulation results, overtone resonances with different amplitudes existed within fundamental frequencies of lead zirconate titanate PZT because the acoustic wavelength in the blood container was much smaller than the thickness of the reflector. Figure 2B shows variations in the acoustic pressure fields with respect to the resonant frequency ((a) f = 1.916 MHz, (b) f = 1.978 MHz, (c) f = 2.1 MHz, (d) f = 2.154 MHz, (e) f = 2.188 MHz, and (f) f = 2.212 MHz). This result indicated that three resonant frequencies (i.e., f = 1.916, 1.978, and 2.1 MHz) exhibited column mode in the perpendicular direction of acoustic wave propagation. Among them, a resonant frequency of 2.1 MHz had a higher amplitude of acoustic pressure fields. Thus, in this study, the excitation frequency of the controller was fixed at 2.1 MHz. The ultrasonic standing wave present in the blood container played a critical role in focusing and aggregating the RBCs at the node of acoustic pressure fields. Figure 2C(a) shows the averaged value of the acoustic pressure (

) for three resonant frequencies (i.e., f = 1.916, 1.978, and 2.1 MHz). As expected, the

exhibited a maximum value at f = 2.1 MHz. In addition, to induce a large displacement, it is required fundamentally to have a higher value of quality factor. Variations of quality factor were obtained by varying the frequency from 2 MHz to 2.2 MHz. As shown in Figure 2C(b), the simulation results indicated that the quality factor exhibited a maximum value at f = 2.098 MHz. From the numerical simulation, it was confirmed that

and quality factor had a maximum value near f = 2.1 MHz. Thus, the ultrasonic transducer showed maximum performance at an excitation frequency of 2.1 MHz, and yielded satisfactory performance in the range of resonant frequencies [21].

Figure 2

Numerical simulation of resonance frequency and acoustic pressure fields with respect to frequency. (A) Variations of admittance with respect to frequency for multilayer piezoelectric resonator. (B) Acoustic pressure fields at resonance frequency (f) ((a) f = 1.916 MHz, (b) f = 1.978 MHz, (c) f = 2.1 MHZ, (d) f = 2.154 MHz, (e) f = 2.188 MHz, and (g) f = 2.212 MHz). (C) Variations of averaged acoustic pressure (

) and quality factor with respect to excitation frequency values ranging from f = 2 MHz to f = 2.2 MHZ. (a) Variation of

with respect to L and f = 1.916, 1.978, and 2.1 MHz. (b) Variations of quality factor with respect to frequency.

3.2. Experimental Visualization of RBCs Suspended in Ultrasonic Transducer

To visualize the behaviors of RBCs suspended in the ultrasonic transducer, the blood sample was diluted extremely to Hct = 2% by adding normal RBCs to 1x PBS. The acoustic power of the controller was kept constant at AP = 2 W. Snapshots were then captured sequentially at intervals of 0.5 s. Figure 3A shows a snapshots captured at a specific time ((a) t = 0, (b) t = 30 s, (c) t = 60 s, and (d) t = 150 s).
Figure 3

Experimental visualization of focusing and sedimentation of RBCs in ultrasonic transducer. (A) Snapshot captured at a specific time (t) ((a) t = 0, (b) t = 30 s, (c) t = 60 s, and (d) t = 150 s). (B) Variations of image intensity obtained along a straight line with respect to time (t) ((a) t = 0, (b) t = 30 s, (c) t = 60 s, and (d) t = 150 s). Inset of Figure 3B(c) shows periodic patterns of image intensity along the width direction.

At an initial state (t = 0), there was no distinctive pattern with respect to RBCs. After t = 30 s, the RBCs were focused at several nodes of acoustic pressure fields. Several column modes are shown in the perpendicular direction of sound propagation. At t = 60 s, the RBCs were aggregated and formed clusters. After t =150 s, the RBC count tended to decrease significantly over time. To quantify the column mode of RBCs in the blood container, the image intensity was obtained along a straight line, as shown in each figure of Figure 3A. Figure 3B shows variations of the image intensity obtained along the width with respect to a specific time ((a) t = 0, (b) t = 30 s, (c) t = 60 s, and (d) t = 150 s). The RBC aggregation contributed to the varying image intensity values at each node. As shown in the inset of Figure 3B(c), the image intensity exhibited periodic patterns with respect to the width direction. This indicated that RBCs were aggregated at each node. To observe the variations in the image intensity, temporal variations in the image intensity were redrawn between 5.2 mm to 6.8 mm. As shown in the inset of Figure 3B(c), the periodic pattern was clearly illustrated to be along the direction of width. Subsequently, the half wavelength (i.e., 0.5 λ) was obtained as 0.5 λ = 0.378 ± 0.07 mm (N = 5). When compared with the half wavelength estimated from the acoustic model (i.e., 0.5 λ = 0.375 ~ 0.39 mm), there was little difference between the theoretical estimation and experimental result. Thus, it was estimated that the dynamic behavior of the ultrasonic transducer could be sufficiently estimated using the acoustic model.

3.3. Contributions of Acoustic Power to RBC-to-Liquid Separation

In the study, concentrated RBCs were purchased from the blood bank. The concentrated RBCs and 1x PBS were mixed and separated with centrifuge. Pure RBCs were then collected by removing buffy coat which included white blood cells or platelet. Then, blood sample was prepared by adding RBCs into various diluents (i.e., 1x PBS, plasma, and dextran solution). Thus, ultrasonic standing waves had an influence on RBCs inside ultrasonic transducer. It was necessary to evaluate the effect of the acoustic power on the RBC-to-liquid separation. A blood sample (Hct = 50%) was prepared by adding normal RBCs to the plasma, and blood sample was infused into the ultrasonic transducer. The acoustic power was set to AP = 0, 2, and 3 W. Figure 4A(a) shows temporal variations of Hs with respect to the acoustic power. In addition, Figure 4A(b) shows the snapshots captured at t = 1080 s with respect to the acoustic power. Under the condition of no acoustic power (i.e., AP = 0), no distinctive change in Hs was observed over time. However, when the acoustic power is set to AP = 2 W, Hs tended to increase substantially over time. Furthermore, when the acoustic power is from 2 W to 3 W, Hs achieved saturation rapidly. The results indicated that the acoustic power contributed to accelerating RBC-to-liquid separation significantly. When the RBCs were exposed to a higher value of acoustic power (i.e., AP = 3 W) for a longer duration of 1000 s, the heat generated by the piezo resonator contributed to adhering RBCs to the surface of the blood container. As shown in Figure 4A(b), the RBCs adhered to the surface of the blood container. The plasma was not shown transparently as a liquid layer. To mitigate this issue (i.e., thermal aggregation of RBCs), it was necessary to restrict the acoustic power below AP = 3 W, especially for a longer duration of operation (~1000 s). The power was therefore fixed at AP = 2 W, except when evaluating the effect of acoustic power on the separation efficiency. In addition, according to previous study [22], cavitation bubble was nucleated at low frequency of 200 kHz. Bubble size was estimated at 14 μm. However, at higher frequency with over 1 MHz, bubble size deceased to 1~3 μm significantly. However, throughout experiments, cavitation bubbles inside ultrasonic transducer was not shown clearly. There was no issue on RBCs damage resulting from bubble collapse.
Figure 4

Contributions of acoustic power to RBC-to-liquid separation. (A) Variations of Hs for blood sample (Normal RBCs in plasma, Hct = 50%) with respect to the acoustic power (AP). (a) Temporal variations of Hs with respect to AP. (b) Snapshot captured at t = 1080 s with respect to AP r. (B) Variations of Hs for blood sample (Normal RBCs in dextran solution (C = 15 mg/mL), Hct = 50%) with respect to AP. (a) Temporal variations of Hs with respect to AP. (b) Snapshot captured at t = 200 s with respect to AP.

As RBCs were suspended in a higher concentration of dextran solution (C = 15 mg/mL], the sedimentation rate of RBCs tended to increase significantly [23]. The blood sample was prepared once again by varying diluent from plasma to the dextran solution. Figure 4B(a) shows variations of Hs with respect to AP = 0, 2, and 3 W. Figure 4B(b) illustrates a snapshot captured at t = 200 s with respect to AP. Under no acoustic power, Hs tended to increase gradually because the dextran solution contributed to an increasing sedimentation rate of RBCs. When the acoustic power is set to AP = 2 or 3 W, Hs achieved saturation rapidly. From the results, the acoustic power contributed to increasing RBC-to-liquid separation significantly.

3.4. Contributions of Hematocrit and Dextran Solution to RBC-to-Liquid Separation

First, as hematocrit hindered the sedimentation rate of RBCs in the stationary flow condition, it was necessary to evaluate the effect of hematocrit on the RBC-to-liquid separation under acoustic power condition. A blood sample with Hct = 20%, 30%, 40%, and 50% was prepared by adding normal RBCs to diluent (i.e., 1x PBS, and plasma). To eliminate the effect of plasma protein on the sedimentation rate of RBCs, a blood sample was prepared by adding normal RBCs to 1x PBS. Figure 5A(a) shows temporal variations of Hs with respect to the acoustic power and hematocrit. Under no acoustic power, Hs changed slightly over time. Variations of Hs tended to decrease at higher values of hematocrit. In other words, a blood sample with Hct = 20% exhibited approximately Hs = 3 mm after an elapse of 3600 s. Furthermore, a blood sample with Hct = 50% exhibited an Hs value of approximately 1 mm after a duration of 3600 s. However, under an acoustic power of AP = 2 W, Hs achieved a saturation height (i.e., Hs = 12 mm) within a short time. The saturation time (T) of the corresponding hematocrit was estimated to be T = 200 s (Hct = 20%), T = 600 s (Hct = 30%), and T = 2100 s (Hct = 40%). For the blood sample with Hct = 50%, Hs tended to increase greatly. However, it did not achieve the saturation height after an elapse of 4100 s. The results indicated that the acoustic power was considered as effective for separating the RBCs from the blood sample. The value of Hs was therefore inversely proportional to the hematocrit level. According to a previous study [21], when RBCs were suspended into plasma rather than 1x PBS, the sedimentation rate of RBCs increased substantially. To verify the variation of the sedimentation rate, the blood sample was prepared by adding normal RBCs to the plasma. Figure 5A(b) shows temporal variations of Hs with respect to the acoustic power and hematocrit. When compared with the results shown in Figure 5A(a), plasma contributed to increasing Hs significantly. The saturation time tended to decrease significantly with respect to Hct = 20%, 30%, and 40%. Interestingly, the blood sample (Hct = 50%) did not show a significant difference depending on diluent. As the higher value of hematocrit hindered the sedimentation rate of RBCs significantly, plasma did not have a strong influence on variations of Hs when compared with 1x PBS. Based on the results, it was found that variations of Hs tended to decrease at a higher level of hematocrit. The acoustic power played a significant role in accelerating Hs over time. Under an acoustic power of 2 W, when compared with 1x PBS, plasma contributed to increasing variations of Hs, especially below Hct = 50%.
Figure 5

Effect of hematocrit and dextran solution on RBC-to-liquid separation. (A) Variations of Hs with respect to hematocrit (Hct = 20%, 30%, 40%, and 50%) and base solution (1x PBS, plasma). (a) Temporal variations of Hs of blood sample (normal RBCs in 1x PBS) with respect to acoustic power and hematocrit. (b) Temporal variations of Hs of blood sample (normal RBCs in plasma, Hct = 50%) with respect to acoustic power and hematocrit. (B) Variations of Hs with respect to concentration of dextran solution (C = 0, 5, 10, 15, and 20 mg/mL) and acoustic power (0, 2 W). (a) Temporal variations of Hs with respect to C and acoustic power = 0. (b) Temporal variations of Hs with respect to C and acoustic power = 2 W.

Second, to evaluate the effect of acoustic power on the blood sample (Hct = 50%), dextran solution was employed to increase the sedimentation rate of RBCs instead of 1x PBS or plasma. To significantly enhance the sedimentation of RBCs in the blood sample (Hct = 50%), various concentrations of dextran solution (C = 0, 5, 10, 15, and 20 mg/mL) were prepared by adding dextran powder to 1x PBS. Here, C = 0 meant 1x PBS. Hematocrit of blood sample was adjusted to 50% by adding normal RBCs into each dextran solution. Figure 5B(a) shows temporal variations of Hs with respect to C under no acoustic power (AP = 0). Above C = 10 mg/mL, Hs increased significantly over time. It achieved saturation height rapidly at a higher concentration of dextran solution. The blood sample (C = 5 mg/mL) contributed to a slight increase in Hs when compared with the blood sample (1x PBS). Under an acoustic power of 2 W, variations of Hs were obtained to evaluate the contributions of the acoustic power with respect to C. Figure 5B(b) represents temporal variations of Hs with respect to C under AP = 2 W. As a result, the acoustic power contributed to increasing Hs substantially over time. Thus, with respect to dextran solution of C = 10, 15, and 20 mg/mL, Hs arrived at a saturation height within 700 s. In addition, a lower concentration of dextran solution (C = 5 mg/mL) contributed to a slight increase in Hs when compared with the control blood sample (1x PBS). Based on the result, dextran solution contributed to increasing RBC-to-liquid separation significantly, even for a blood sample with Hct = 50%. An acoustic power of 2 W played a significant role in accelerating the sedimentation of RBCs substantially when compared to the case with no acoustic power.

3.5. Hardened Blood Sample Preparation with Ultrasonic Transducer

When preparing fixed RBCs with glutaraldehyde (GA) solution, normal RBCs were added and mixed to a specific concentration of GA solution. After an elapse of specific time, normal RBCs were fixed. To collect fixed RBCs, a centrifugal separator was usually employed to separate fixed RBCs and liquid from the blood sample. During the centrifugal operation, RBCs were stacked and adhered to the surface of the tube because of the strong centrifugal forces in outward direction. Thus, it was difficult to collect fixed RBCs from the tube. As shown in Figure A2A (Appendix A), the blood sample (Hct = 50%) was prepared by adding normal RBCs to a specific concentration of GA solution (C = 0, 5, 10, 15, and 20 μL/mL). After fitting tubes into a centrifuge, it rotated at 4000 rpm for 10 min. Figure A2B (Appendix A) shows tubes filled with blood samples after operating the centrifuge. Owing to the centrifugal force, RBCs were positioned diagonally inside a tube. To evaluate the RBC adhesion simply, a tube was inverted at about 180°. As shown in Figure A2C (Appendix A), the blood sample (i.e., fixed RBCs with C <= 10 μL/mL) flowed downward owing to the gravitational force. The fixed RBCs were then collected after removing liquid as the upper layer. However, the blood sample (i.e., fixed RBCs with C values above 15 μL/mL) did not flow downward because RBCs adhered to the inner surfaces of the tube. For the reason, it was difficult to collect fixed RBCs from the tube.
Figure A2

RBCs stacked and adhered to surface of tube after operating centrifuge. Blood sample (Hct = 50%) was prepared by adding normal RBCs into a specific concentration of GA solution (C = 0, 5, 10, 15, and 20 μL/mL). (A) Tubes filled with blood samples before operating centrifuge. (B) Tubes filled with blood samples after operating centrifuge (4000 rpm x 10 min). (C) Inverted tubes filled with hardened blood samples with respect to CGA. Above C =15 μL/mL, hardened RBCs adhered to the surface of tubes, and it was difficult to collect hardened RBCs from the tube.

When compared with the centrifuge, the ultrasonic transducer did not apply a strong force to the RBCs. In other words, as the ultrasonic radiation force was much smaller than the centrifugal force, the contributions of ultrasonic radiation force to the RBC adhesion were sufficiently negligible. Inside the blood container, RBCs come down inversely in gravitational direction and stacked with the ultrasonic transducer. Using the fixed blood sample discussed in Figure A2 (Appendix A), the RBC-to-liquid separation was evaluated using the ultrasonic transducer. As shown in Figure 6A, blood sample preparation was divided into three stages: step (I), step (II), and step (III). Throughout the three stages, the diluent was changed only after operating the ultrasonic transducer. The hematocrit of the blood sample was fixed at Hct = 50%. First, Figure 6A(a) illustrates the procedure involved in the blood sample preparation in step (I). Normal RBCs (~3 mL) and a specific concentration of GA solution (~ 3 mL) were added and mixed in a blood container. To sufficiently fix the membrane of normal RBCs, blood sample had been remained for 10 min. After operating the ultrasonic transducer for 30 min, the blood sample was separated into RBCs (in the lower layer) and the GA solution (in the upper layer). Second, Figure 6A(b) shows the procedure of blood sample preparation in step (II). The GA solution (~1 mL) was removed from the upper layer. 1x PBS (~ 1 mL) was then added and mixed in a blood container. After operating the ultrasonic transducer for 30 min, the blood sample was separated into RBCs and GA solution diluted with 1x PBS. Third, Figure 6A(c) shows the procedure of blood sample preparation at step (III). The diluted GA solution (~1 mL) was removed from the top layer. A specific concentration of dextran solution (C = 15 mg/mL) (~ 1 mL) was then added and mixed in the blood container. After operating the ultrasonic transducer for 30 min, the blood sample was separated into RBCs and liquid.
Figure 6

Measurement of separation height (Hs) after adding normal RBCs into GA solution, 1x PBS, and dextran solution (C = 15 mg/mL). (A) Schematic of the three steps for preparing blood sample. (a) Step (I): normal RBCs (3 mL) and a specific concentration of GA solution (3 mL) were infused and mixed in blood container. (b) Step (II): after operating ultrasonic transducer for 30 min, GA solution (~1 mL) was removed from top layer. 1x PBS was then added and mixed in blood container. (c) Step (III): after operating ultrasonic transducer for 30 min, diluted GA solution (~1 mL) was removed from the top layer. A specific concentration of dextran solution (C = 15 mg/mL) was added and mixed in blood container. (B) Temporal variations of Hs for the blood sample (C = 5 μL/mL) following the sample preparation process [step (I), step (II), and step (III)] for the blood sample mixed with normal RBCs and C: (a) CGA = 5 μL/mL, (b) CGA = 10 μL/mL, (c) CGA = 15 μL/mL, and (d) CGA = 20 μL/mL.

With respect to the three stages, the temporal variations of Hs for the blood sample were obtained by varying the concentration of the GA solution. Figure 6B(a) shows the temporal variations of Hs with respect to each step. Here, at step (I), the GA solution of 5 μL/mL was used to fix the membrane of normal RBCs. The result indicated that Hs tended to increase gradually by changing the diluent (i.e., GA solution → 1x PBS dextran solution). Figure 6B(b) shows temporal variations of Hs for the hardened blood sample with CGA=10 μL/mL. Step (I) and step (II) did not show substantial differences with respect to the variation of Hs over time. Compared with step (I) and step (II), step (III) (i.e., dextran solution as diluent) exhibited a slight increase of Hs over time. However, as shown in Figure 6B(c,d), with respect to the higher concentration of GA solution (C = 15, and 20 μL/mL), step (III) shows a decrease of Hs over time. According to the experimental results, Hs tended to decrease gradually at higher concentrations of the GA solution. Below CGA = 10 μL/mL, the dextran solution [step (III)] contributed to an increasing Hs. However, above C = 15 μL/mL, the dextran solution contributed to a gradual decrease in Hs. At a higher concentration of GA solution, fixed RBCs did not contribute to RBC aggregation. In other words, the dextran solution did not contribute to the acceleration of Hs over time. During the procedure of blood sample preparation, dextran solution was added to dilute GA solution at step (III). The suctioning or discharging of the blood sample with a pipette required a large force. For the reason, it was inferred that the mixed solution between GA solution and dextran solution may hinder significantly the RBC-to-liquid separation. From the experimental results, the ultrasonic transducer could be used effectively to separate the hardened blood sample without RBC adhesion or the occurrence of stacks while operating the centrifuge.

3.6. Measurement of RBC Aggregation and Blood Viscosity for Hardened Blood Sample

As discussed with respect to Figure 6, a blood sample was collected at each step. For each step, the blood sample was collected as fixed RBCs suspended in GA solution (step [I]), fixed RBCs suspended in 1x PBS (step [II]), and fixed RBCs suspended in dextran solution (step [III]). In other words, RBCs and diluent were separated by operating ultrasonic transducer. Blood sample was prepared by adding RBCs into specific diluents (i.e., GA solution, 1x PBS, and dextran solution). Blood sample and 1x PBS were infused simultaneously by controlling flow rate of both fluids with syringe pumps. While varying diluent at each step, RBC aggregation and blood viscosity were measured simultaneously. Figure 7A(a) shows the experimental setup for measuring RBC aggregation and blood viscosity. The experimental setup was composed of microfluidic device, and two syringe pumps. As shown in Figure 7A(b), the image intensity () in the blood channel (BC) was obtained to monitor RBC aggregation with respect to the blood flow rate. In addition, the blood viscosity in the coflowing channel (CC) was obtained by evaluating the interface. The blood sample (i.e., test fluid) and 30% glycerin solution (i.e., reference fluid) were injected by controlling two syringe pumps at the same flow rate (i.e., Q = Q = Q, Q: flow rate of blood sample, Q: flow rate of glycerin solution, and Q: flow rate of syringe pump). The flow rate of syringe pump increased gradually from Q = 0.1 mL/h to Q = 3.1 mL/h at an interval of 0.2 mL/h [24]. Hematocrit of blood sample was adjusted to 50% by adding normal RBCs to 1x PBS. As shown in Figure 7A(c), a simple mathematical model was constructed with discrete fluidic circuit elements (i.e., flow rate (Q, Q), and resistance element (R, R). Based on the sample pressure of both fluids in the coflowing channel, the blood viscosity (μ) was expressed as . Here, correction factor (CF) was included to compensate the boundary difference between real physical model and simple mathematical model. As shown in Figure 7B(a), variations of , <α>, and Q were obtained with respect to time. As 1x PBS did not include plasma protein, remained constant under a varying flow rate. <α> tended to decrease at a higher flow rate. According to a previous study [24], AI and μ were obtained from temporal variations of and <α> respectively. Figure 7B(b) shows variations of AI and μ with respect to shear rate (). As expected, AI remained constant with respect to . In addition, μ tended to decrease gradually with respect to because the blood sample behaved as a non-Newtonian fluid.
Figure 7

Quantification of RBC aggregation (AI) and blood viscosity (μ) in microfluidic environment. (A) Simultaneous measurement of AI and μ with microfluidic device. (a) Schematic of experimental setup for measuring AI and μ. (b) Quantification of microscopic image intensity () in blood channel and interface (α) in coflowing channel. (c) Discrete fluidic circuit model for blood viscosity in coflowing channel. (B) Quantification of AI and μ for normal blood sample (normal RBCs in 1x PBS, Hct = 50%). (a) Variations of , α, and Qsp with an elapse of time. (b) Variations of AI and μ over shear rate.

First, to evaluate the effects of RBCs or diluent on RBC aggregation, RBC aggregation were obtained by changing diluents (i.e., step (I), step (II), and step (III)) and GA solution (i.e., C = 5, 10, 15, and 20 μL/mL). As shown in Figure 8A, step (I) and step (II) did not exhibit variations of AI with respect to shear rate. However, step (III) shows that AI tended to decrease largely for up to C =10 μL/mL. Above C = 15 μL/mL, there was little variation of AI with respect to C. From this result, fixed RBCs with GA solution of C = 15~20 μL/mL did not contribute to the RBC aggregation.
Figure 8

Quantitative evaluations of AI and blood viscosity for fixed blood sample with different diluents with respect to step (I, II, and III). (A) Variations of AI with respect to C ((a) C = 5 μL/mL, (b) C = 10 μL/mL, (c) C = 15 μL/mL, and (d) C = 20 μL/mL) and shear rate. (B) Variations of μ with respect to C ((a) C = 5 μL/mL, (b) C = 10 μL/mL, (c) C = 15 μL/mL, and (d) C = 20 μL/mL) and shear rate.

Second, to evaluate the contribution of RBCs or diluent to the blood viscosity, variations in the blood viscosity were obtained by varying the diluent (i.e., step (I), step (II), and step (III)) and C. As shown in Figure 8B, variations of μ were obtained with respect to C and shear rate. At lower shear rates (i.e., < 100 , μ showed large variations with respect to shear rate. RBC aggregation posed difficulties when evaluating the interface between the blood sample and 30% glycerin solution. Thus, it led to reduced measurement accuracy. However, at a higher shear rate, μ showed smaller variations over the shear rate. As shown in Figure 8B(a), step (III) shows an increase of μ when compared with step (I) and step (II). As the syringe pump was installed vertically, the erythrocyte sedimentation rate (ESR) occurred in the driving syringe. Because of the continuous ESR in the driving syringe, the hematocrit of the blood sample infused in the microfluidic channel increased over time. In other words, dextran solution contributed to increasing RBC aggregation, which resulted in an increase in the hematocrit and blood viscosity. As shown in Figure 8B(b), the dextran solution resulted in an increased blood viscosity. However, the variation of the blood viscosity was relatively reduced when compared with Figure 8B(a). This result indicated that the effect of dextran solution on blood viscosity tended to decrease at higher concentrations of GA solution (i.e., fixed RBCs). As shown in Figure 8B(c,d), dextran solution did not contribute to the increase in blood viscosity. In addition, the blood viscosity remained constant over the shear rate. In other words, as fixed RBCs did not include flexibility with respect to the shear rate, the blood viscosity remained constant over the shear rate. From the aforementioned experimental results, it lead to the conclusion that RBC aggregation and blood viscosity measured by the microfluidic technique could be employed to monitor fixed blood samples. In addition, the dextran solution contributes to increasing RBC aggregation and blood viscosity below certain concentrations of GA solution (i.e., C = 5 and 10 μL/mL).

4. Conclusions

In this study, to separate RBCs and liquid from the blood sample, an ultrasonic transducer was proposed instead of a centrifuge. Using the microfluidic technique, RBC aggregation and blood viscosity were obtained with respect to each blood sample. First, a numerical simulation was conducted to evaluate the characteristics of the ultrasonic transducer (i.e., natural frequency, acoustic pressure fields, and quality factor). Based on the variations in the acoustic power (

) and quality factor with respect to frequency, the ultrasonic transducer showed a maximum performance at an excitation frequency of 2.1 MHz. Second, the periodic pattern was visualized as a column mode experimentally. The half wavelength was found to be 0.5 λ = 0.378 ± 0.07 mm. The experimental result was in good agreement with the analytical estimation. Third, the acoustic power effectively separated the RBCs or base solution from various blood samples (i.e., Hct = 20% ~ 50%, base solution: plasma, 1x PBS, and dextran solution). The experimental results indicated that RBCs with hematocrit of 50% were separated substantially within 10 min under acoustic power of 2 W. Additionally, ESR (erythrocyte sedimentation rate) was highly accelerated under acoustic pressure fields. Fourth, the present method was employed to separate fixed RBCs from fixed blood samples composed of fixed RBCs suspended in GA solution. According to experimental results, the ultrasonic transducer did not induce RBC adhesion or the occurrence of stacks. The present method separated fixed RBCs from fixed blood sample easily. Instead of the centrifuge, the proposed method was employed to prepare various fixed blood samples while changing diluents (i.e., GA solution, 1x PBS, and dextran solution). Then, RBC aggregation and the blood viscosity of the blood sample were obtained successfully by varying the blood samples. In conclusion, the present method showed promising potentials for blood sample preparation with ultrasonic-based separation, and quantification of mechanical properties of blood samples in microfluidic device. In the near future, if a microfluidic platform and ultrasonic transducer will be integrated into a closed fluidic circuit, multiple mechanical properties of blood samples could be monitored simultaneously for a certain longer period.

  17 in total

1.  Clarification of plasma from whole human blood using ultrasound.

Authors:  C M Cousins; P Holownia; J J Hawkes; C P Price; P Keay; W T Coakley
Journal:  Ultrasonics       Date:  2000-03       Impact factor: 2.890

2.  Simultaneous measurement of blood pressure and RBC aggregation by monitoring on-off blood flows supplied from a disposable air-compressed pump.

Authors:  Yang Jun Kang
Journal:  Analyst       Date:  2019-05-03       Impact factor: 4.616

3.  Multiphysics modelling of the separation of suspended particles via frequency ramping of ultrasonic standing waves.

Authors:  Francisco J Trujillo; Sebastian Eberhardt; Dirk Möller; Jurg Dual; Kai Knoerzer
Journal:  Ultrason Sonochem       Date:  2012-09-13       Impact factor: 7.491

4.  Label-free viscosity measurement of complex fluids using reversal flow switching manipulation in a microfluidic channel.

Authors:  Yang Jun Kang; Jeongeun Ryu; Sang-Joon Lee
Journal:  Biomicrofluidics       Date:  2013-07-26       Impact factor: 2.800

5.  High-throughput acoustic separation of platelets from whole blood.

Authors:  Yuchao Chen; Mengxi Wu; Liqiang Ren; Jiayang Liu; Pamela H Whitley; Lin Wang; Tony Jun Huang
Journal:  Lab Chip       Date:  2016-08-01       Impact factor: 6.799

6.  Isolation of exosomes from whole blood by integrating acoustics and microfluidics.

Authors:  Mengxi Wu; Yingshi Ouyang; Zeyu Wang; Rui Zhang; Po-Hsun Huang; Chuyi Chen; Hui Li; Peng Li; David Quinn; Ming Dao; Subra Suresh; Yoel Sadovsky; Tony Jun Huang
Journal:  Proc Natl Acad Sci U S A       Date:  2017-09-18       Impact factor: 11.205

7.  Particle separation using ultrasound can be used with human shed mediastinal blood.

Authors:  Henrik Jönsson; Andreas Nilsson; Filip Petersson; Mats Allers; Thomas Laurell
Journal:  Perfusion       Date:  2005-01       Impact factor: 1.972

8.  Microfluidic biosensor for monitoring temporal variations of hemorheological and hemodynamic properties using an extracorporeal rat bypass loop.

Authors:  Yang Jun Kang; Eunseop Yeom; Sang-Joon Lee
Journal:  Anal Chem       Date:  2013-10-14       Impact factor: 6.986

9.  Hybrid System for Ex Vivo Hemorheological and Hemodynamic Analysis: A Feasibility Study.

Authors:  Eunseop Yeom; Yang Jun Kang; Sang Joon Lee
Journal:  Sci Rep       Date:  2015-06-19       Impact factor: 4.379

10.  The Effect of Pulsatile Versus Nonpulsatile Blood Flow on Viscoelasticity and Red Blood Cell Aggregation in Extracorporeal Circulation.

Authors:  Chi Bum Ahn; Yang Jun Kang; Myoung Gon Kim; Sung Yang; Choon Hak Lim; Ho Sung Son; Ji Sung Kim; So Young Lee; Kuk Hui Son; Kyung Sun
Journal:  Korean J Thorac Cardiovasc Surg       Date:  2016-06-05
View more
  3 in total

1.  Comparing of Frequency Shift and Impedance Analysis Method Based on QCM Sensor for Measuring the Blood Viscosity.

Authors:  Shuang Liao; Peng Ye; Cheng Chen; Jie Zhang; Lin Xu; Feng Tan
Journal:  Sensors (Basel)       Date:  2022-05-17       Impact factor: 3.847

2.  Red Blood Cell Sedimentation Index Using Shear Stress of Blood Flow in Microfluidic Channel.

Authors:  Yang Jun Kang
Journal:  Biosensors (Basel)       Date:  2022-07-21

3.  Contributions of Red Blood Cell Sedimentation in a Driving Syringe to Blood Flow in Capillary Channels.

Authors:  Yang Jun Kang
Journal:  Micromachines (Basel)       Date:  2022-06-08       Impact factor: 3.523

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.