Literature DB >> 32309731

Comparison of Colorimetric Analyses to Determine Cortisol in Human Sweat.

Ethan Tu1,2, Paul Pearlmutter1, Michelle Tiangco1, Gia Derose3, Lina Begdache3, Ahyeon Koh1.   

Abstract

Colorimetric analysis, which relies on a chemical reaction to facilitate a change in visible color, is a great strategy for detecting cortisol, which is necessary to diagnose and manage the wide variety of diseases related to the hormone, because it is simple in design, inexpensive, and reliable as a standard cortisol analysis technique. In this study, four different colorimetric cortisol analyses that use various chromogens, which include sulfuric acid, Porter-Silber reagent, Prussian blue, and blue tetrazolium, are studied. Modifications to the classic Porter-Silber method are made by increasing the carbon content of the alcohol and adding gold nanoparticles, which result in a twofold increase in reaction rate and a slight decrease in the limit of detection (LoD). After a comparison of the reaction rate, LoD, dynamic range, characteristic peaks, and color stability of all methods, blue tetrazolium demonstrates a low LoD (97 ng/mL), broad dynamic range (0.05-2 μg/mL), and quick reaction rate (color development as fast as 10 min), which are well within the requirements for human biofluids. Cortisol in artificial saliva and sweat and in human sweat was determined while confirming that no excipients or other biomarkers interfered with the reactions. Twenty-one human sweat samples were tested using blue tetrazolium and revealed a significant difference between male and female apocrine cortisol concentrations and showed a highly significant difference between apocrine and eccrine cortisol concentrations. Colorimetric methods of cortisol can compete with existing electrochemical sensors because of their similar accuracy and detection range in certain wearable biosensor applications. The simplicity of colorimetric methods advances potential applications in skin-interfaced bio-electronics and point-of-care devices.
Copyright © 2020 American Chemical Society.

Entities:  

Year:  2020        PMID: 32309731      PMCID: PMC7161047          DOI: 10.1021/acsomega.0c00498

Source DB:  PubMed          Journal:  ACS Omega        ISSN: 2470-1343


Introduction

Cortisol is an important biomarker that is secreted by the adrenal cortex to regulate blood pressure, glucose levels, and metabolism.[1] Although cortisol is always present in human systems, its secretion outside of circadian dependencies is induced by psychological and emotional stress, which is why monitoring cortisol concentrations in the body is crucial in understanding stress-related diseases.[2,3] Normal levels of cortisol range from 8 to 142 ng/mL in sweat, 40 to 250 ng/mL in plasma, and 1 to 11 ng/mL in saliva, with the highest concentrations occurring during the morning and slowly decreasing throughout the day.[4−6] Using any biofluid to identify and track abnormal cortisol concentration can help patients and caregivers manage a variety of diseases such as Cushing’s syndrome and Addison’s disease, which cause characteristically high or low levels of cortisol, respectively.[7,8] The current gold standard in cortisol analysis uses an enzyme-linked immunosorbent assay (ELISA). However, colorimetric reactions also have been studied historically to determine cortisol concentrations. Early studies report fluorescent reactions among a number of steroids with concentrated sulfuric acid.[9−11] Each steroid–acid mixture has similar but unique absorption spectra and obeys Beer’s law. Zimmermann’s reagent was adopted later on as a spot test for steroids and replaced pure sulfuric acid.[12] In addition to greater accuracy and specificity, the absorption spectra were more easily differentiated.[13] Corticosteroid specific tests first appeared with the Porter–Silber reagent in 1950, where a pale yellow hue was reported when hydrocortisone and cortisone were added to a solution that contains dilute sulfuric acid, methanol, and phenylhydrazine.[14] The Porter–Silber reagent has been improved upon throughout the years, which results in faster reaction rates and a more robust color.[15−17] Concurrent with the Porter–Silber reagent, Clark developed another corticosteroid specific reagent, which consists of dilute sulfuric acid and diphenylamine.[18] Later on, reagents that used blue tetrazolium provided an even greater steroid assay, which lowered the reaction time to 10 min compared to the 90 min of Porter–Silber.[19,20] Singh and Verma recently reported another method using cortisol’s reduction properties to form Prussian blue from iron(III) and hexacyanoferrate(III), which is comparable to past formulations in terms of sensitivity, precision, and speed.[21] With the rise of competitive protein binding immunoassay techniques, colorimetric reactions have been largely made obsolete, with little research published since the 1970s. However, with increased interest in cortisol monitoring and detection in sweat and saliva with disposable biosensors, colorimetric analysis has become a prime candidate for such tasks that use simple and easy analytical analysis. The in situ analysis of biofluids, such as saliva, urine, and sweat, has drawn significant attention over the years because of its noninvasive collection methods. In skin-interfaced and wearable devices, electrochemical techniques are preferred over competitive protein binding techniques because of their portability and potential point-of-care applications.[22] Additionally, ELISA techniques and other immunoassay-based tests are expensive, are chemically complex, require a lengthy amount of time to complete, are difficult to scale up, and require cold-chain managements.[23−26] Multiple research groups have worked to improve analyte stability,[27] flexibility of electronic components,[28,29] wearability,[30−32] and sensitivity of electrochemical sensors.[27,33,34] However, short shelf life, battery dependence, inflexible batteries, and relatively high cost for complete electronic systems currently prevent mainstream success.[22] Colorimetric-based patch-type microfluidic devices have seen success with other biomarkers such as glucose, pH, chloride, lactate, creatine, ammonia, and ethanol.[31,35,36] Such techniques could provide a solution to the drawbacks of electrochemical systems, while retaining many benefits. The lack of electronic components can lower manufacturing cost and is capable of becoming disposable and offers more straightforward skin integration.[32] Microfluidic channel design can also be tailored to accommodate continuous time analysis or single time analysis.[37] Herein, we revisit and study the various colorimetric analyses of cortisol and improve its analytical performances that enable its use for wearable devices. Reactions reported by Zimmerman and Clark, as previously mentioned, were not studied because of the highly toxic components, which are not suitable for biosensor applications. Traditionally, both the Porter–Silber and Prussian blue methods have been slow to develop color, with reaction times that reach as long as 4 h at room temperature. The average reaction rate of the Porter–Silber was accelerated by substituting ethanol for isopropanol and adding gold nanoparticles (AuNPs) as a catalyst. In addition, the reaction rate was enhanced by completing the Prussian blue reaction with cortisol in ethanol as opposed to deionized (DI) water. Overall, the blue tetrazolium method was found to be the most robust, which excels in both speed and accuracy when compared to other methods. The small-n human pilot study confirms that the cortisol in biofluid can be determined.

Results and Discussion

In general, cortisol is believed to be a critical biomarker for monitoring stress levels. Therefore, a simple and sensitive colorimetric analysis of cortisol holds immense potential for daily-monitoring wearable devices. Four colorimetric analyses of cortisol detections that include sulfuric acid, Porter–Silber reagent, Prussian blue, and blue tetrazolium were investigated and further developed in this study. The mechanism of each colorimetric reaction was experimentally validated with optical color changes that are capable of visual inspection of cortisol concentration in the system (Figures and S2). In Table , the characteristic absorption peak, limit of detection (LoD), time to color development, color change, and color stability are summarized for each method tested. Two types of LoDs are given: analytical and experimental, both of which are described in our Materials and Methods section. Characteristic absorption peaks from Prussian blue and blue tetrazolium methods differed slightly from previous reports. Singh and Verma reported a single absorbance peak at 780 nm using Prussian blue, while this study found two peaks, one at 300 nm and the other at 670 nm.[21] Multiple studies report a peak at 525 nm for blue tetrazolium; however, our results show a peak at 510 nm.[19,20] The reason for both discrepancies is the difference in the solvent system.[40] The blue tetrazolium method has the lowest LoD at 97 ng/mL. The AuNP Porter–Silber and blue tetrazolium methods are the only two capable of detecting cortisol within normal human sweat levels (8–142 ng/mL).[4] The time to color development listed in Table is an approximation of the time it would take for the reagent to form an absorption peak discernible by the spectrometer. The time to color development appears to be dependent on the amount of cortisol added to the solution. Higher concentrations of cortisol produce more chromogens within a given time and therefore decrease color development time. Color stability also varies greatly among the methods. Cortisol in sulfuric acid causes a rapid and progressive change to the absorption magnitude. In agreement with previous reports, the Porter–Silber reagent itself becomes cloudy and opaque after about 24 h, which interferes with spectroscopic analysis.[41] The cloudiness is most likely due to the deterioration of phenylhydrazine, as it is both light and air sensitive. Prussian blue was stable for a week with no change in absorption and may last much longer. Baseline solution of the blue tetrazolium reagent also turns magenta, where after 12 h, solutions with and without cortisol are nearly indistinguishable.
Figure 1

Optical images of color change with increasing concentration of cortisol: (A) sulfuric acid, (B) AuNP Silber–Porter, (C) Prussian blue, and (D) blue tetrazolium.

Table 1

Comparison of Different Methods of Cortisol Detection

method of detectioncharacteristic absorption peak (nm)analytical LoD (μg/mL)experimental LoD (μg/mL)dynamic range (μg/mL)time to color development (min)visual color changecolor stability
pure concentrated sulfuric acid290, 394, and 4803.001.030–100immediateclear to florescent yellowunstable
Porter–Silber reagent with AuNPs4100.1450.4010–7060clear to viscous yellowstable up to 24 h
potassium hexacyanoferrate(III) and iron(III) reagent300 and 6690.7310.320undetermined45pale yellow to blue-greenstable for at least 1 week
blue tetrazolium reagent5100.0970.2140–1.55–10pale yellow to magentastable for at least 1 week
Optical images of color change with increasing concentration of cortisol: (A) sulfuric acid, (B) AuNP Silber–Porter, (C) Prussian blue, and (D) blue tetrazolium. The characteristic peaks and the resulting concentration curve of various colorimetric analyses are shown in Figure for further quantitative spectroscopic determination. For sulfuric acidcortisol experiments, the color change is immediate upon adding cortisol to sulfuric acid and continues to increase absorbance indefinitely. The full absorbance spectrum of this reaction at 5 min post mixing presents peaks at 290, 394, and 480 nm while with increased intensity with respect to increased cortisol concentration (Figure A). A calibration curve was obtained at 394 nm, and the LoD was calculated to be 3.00 μg/mL with a sensitivity of 0.0098 Abs (μg/mL)−1 (Figure E). Color stability of the sulfuric acid method is highly unstable and never reaches a plateau even after 100 min (Figure S3A). Overall, the sulfuric acid method is the simplest and quickest test. However, because of its LoD being in the upper range of human levels and spontaneous gradual color changes, this method functions better as a qualitative test rather than a quantitative test.
Figure 2

Representative absorbance spectrum of cortisol in various reagents and their concentration curves: (A,E) sulfuric acid reagent with characteristic peaks at 290, 394, and 470 nm. The concentration curve plotted at 394 nm. (B,F) AuNP Porter–Silber reagent with the characteristic peak at 410 nm. (C,G) Prussian blue reagent with characteristic peaks at 300, 410, and 669 nm. The concentration curve plotted at 669 nm. (D,H) Blue tetrazolium reagent with the characteristic peak at 510 nm.

Representative absorbance spectrum of cortisol in various reagents and their concentration curves: (A,E) sulfuric acid reagent with characteristic peaks at 290, 394, and 470 nm. The concentration curve plotted at 394 nm. (B,F) AuNP Porter–Silber reagent with the characteristic peak at 410 nm. (C,G) Prussian blue reagent with characteristic peaks at 300, 410, and 669 nm. The concentration curve plotted at 669 nm. (D,H) Blue tetrazolium reagent with the characteristic peak at 510 nm. The traditional Porter–Silber reagent requires a minimum of 2 h to complete the reaction at room temperature. The previously reported strategy to increase the reaction rate was to increase the temperature at which the reaction took place to 70 °C.[15] However, for potential wearable biosensor applications, simply increasing the temperature to 70 °C would be inapplicable. Investigations into the mechanism of the Porter–Silber reaction replaced ethanol with methanol to slow down the reaction, which shows that the carbon content of the alcohol would have a large effect on the reaction rate.[42]Figure A and Table S1 show a comparison of different alcohols as a solvent system and their effect on the reaction rate. Lower carbon content alcohols, such as methanol and ethanol, require a longer time to complete the reaction as opposed to higher carbon content alcohols. Octanol exhibits a 33% decrease in reaction time, while butanol exhibits a 17% decrease. Isomers of alcohols also increase the reaction rate. Both isopropanol and isobutanol outperform their primary alcohol counterparts, nearly halving the total time to full color development where absorbance maxima began to plateau. Isopropanol yields optimum results for bioanalysis because of the solubility of isopropanol in dilute sulfuric acid, which also requires for Porter–Silber reaction. Octanol and isobutanol are immiscible with dilute sulfuric acid and produce a distinct phase separation (Figure S4).
Figure 3

Effect of carbon content of alcohol and AuNPs on the color development time of Porter–Silber reaction: (A) time to initial color development at 23 and 70 °C with respect to increasing carbon content of the alcohol component of the Porter–Silber reaction. (B) Increase of AuNPs causes a decrease in color development time of the Porter–Silber reagent until the AuNP concentration exceeded 112.5 pM. The absorbance was determined at 410 nm in 5 min intervals as a function of AuNP concentration. The time at which each reaction reached an absorbance of 0.70 was recorded.

Effect of carbon content of alcohol and AuNPs on the color development time of Porter–Silber reaction: (A) time to initial color development at 23 and 70 °C with respect to increasing carbon content of the alcohol component of the Porter–Silber reaction. (B) Increase of AuNPs causes a decrease in color development time of the Porter–Silber reagent until the AuNP concentration exceeded 112.5 pM. The absorbance was determined at 410 nm in 5 min intervals as a function of AuNP concentration. The time at which each reaction reached an absorbance of 0.70 was recorded. Introduction of AuNPs improves the reaction rate of the Porter–Silber reaction assisted by catalytic reaction.[43] However, an excess amount of AuNPs (concentrations > 112.5 pM) decreases the reaction time further without inhibiting the LoD or distorting the characteristic peak. With optimal AuNPs, the time to initial color development has been reduced nearly by half, from 120 to 60 min at room temperature. At 70 °C, the reduction is from 60 to 30 min. The AuNP solution itself exhibits a red-salmon color, which has an absorption peak at 560 nm. However, low concentration of the AuNP (<100 pM) exhibited no interference in the final color development and the AuNP characteristic peak was nearly undetectable. Simply increasing the concentration of AuNPs does not necessarily mean faster color development. In concentrations larger than 150 pM, the AuNPs hinder the reaction, which results in a lengthier color development time than the traditional Porter–Silber solution (Figure B). Figure B,F shows the results of the AuNP Porter–Silber reagent experiments that use isopropanol as a solvent, which is an optimized, improved, and modified Porter–Silber analysis. The reagent turns from clear to a translucent yellow and exhibits a maximum absorbance peak at 410 nm, which increases in amplitude with respect to increased concentrations of cortisol. Plotting the absorbance peak against cortisol concentration at 30 min yields the calibration curve of y = 0.0252x + 0.126 (R2 = 0.988). The 410 nm peak reaches a stable maximum and maintained color presentation after approximately 2 h. As a nontoxic dye commonly used in paints, Prussian blue poses a promising alternative, whereas previous methods require highly acidic and cytotoxic solvents. The solution begins at a pale yellow and changes to a green-blue. The absorption spectrum of the solution yields two peaks at 300 and 669 nm. The absorption maxima of both peaks increase with respect to increasing concentrations of cortisol, as shown in Figure C,G. Although the relationship between the absorption value and cortisol concentration is linear, the standard deviation of the curve is quite high. The unreliability of this method arises because of the difference in particle size of the resulting colloidal solution. The oxidative state of iron (Fe2+ vs Fe3+) drastically changes the particle size of the colloid, about 40–100 nm in diameter, and it therefore changes the absorption peaks.[34] Nevertheless, the Prussian blue method is still capable of determining cortisol in various concentrations (1.0–15 μg/mL). Blue tetrazolium in conjunction with tetramethylammonium hydroxide yields a magenta solution. In this study, we used methanol as a primary solvent, while previous studies used either ethanol or dichloromethane.[19,20] The use of methanol decreased the reaction time (approximately 25% faster than ethanol) with no changes in sensitivity. Figure D,H shows that the peak absorption increases with respect to increasing concentrations of cortisol. The calibration curve shows linear relationship with a sensitivity of 0.038 Abs (μg/mL)−1 and dynamic range from 0.2 to 1.6 μg/mL. Both the blank and cortisol solution turn magenta because the reaction depends on the hydrolysis rate of the cyclic diacetyl in blue tetrazolium.[44] Although absorbance saturates over time, the time until the full color development varies significantly as a function of cortisol concentration. To stabilize the reaction and to slow the redox process, the reaction can be carried out in a solvent with a large dielectric constant, such as water. The largest advantage of the blue tetrazolium method is its quick color development, which is an advancement for an industry pushing for rapid result diagnostics. All four methods of determining cortisol in the complex biofluid matrix, such as saliva and sweat, resulted in a similar analytical response. Quantitative analytical analysis in sensitivity, LoD, and dynamic ranges demonstrate insignificant changes with artificial sweat and saliva (Figure S5). Overall, blue tetrazolium is the best overall method of detecting cortisol because of rapid reaction rate, low toxicity, low LoD within physiologically relevant levels, and a single peak for simple analysis, which are all beneficial for wearable optoelectronic applications and stress monitoring. Blue tetrazolium methods may be comparable to the gold-standard ELISA technique for certain biosensor applications. It is noteworthy that the blue tetrazolium method’s greatest asset is the simplicity of its one-step reaction, which completes in under 10 min; it is far easier and faster than ELISA methods. Additionally, blue tetrazolium has a significantly greater range of detection, which is useful for diagnosing those with disorders such as Cushing’s syndrome, which produces characteristically high levels of cortisol in patients. In Figure , both blue tetrazolium and ELISA concentration curves made using artificial sweat are shown. For blue tetrazolium in artificial sweat, the LoD increases slightly from 97 to 146 ng/mL, which is expected to be because of the sweat’s composition comprising mostly water. ELISA is still the gold standard in laboratory settings; however, we believe it is not as appealing as blue tetrazolium in point-of-care applications.
Figure 4

Analytical performance comparison of the (A) blue tetrazolium reagent and (B) ELISA for monitoring cortisol in a different range of concentrations: (A) blue tetrazolium capable of detecting μg/mL of ranges and an experimental LoD of 0.045 μg/mL. (B) ELISA method offers more sensitivity than the blue tetrazolium method. However, its linear dynamic range is significantly lower, which may fit for ng/mL determination.

Analytical performance comparison of the (A) blue tetrazolium reagent and (B) ELISA for monitoring cortisol in a different range of concentrations: (A) blue tetrazolium capable of detecting μg/mL of ranges and an experimental LoD of 0.045 μg/mL. (B) ELISA method offers more sensitivity than the blue tetrazolium method. However, its linear dynamic range is significantly lower, which may fit for ng/mL determination. To demonstrate the viability of blue tetrazolium as a diagnostic tool, we analyzed apocrine and eccrine sweat samples from different subjects (n = 21) and correlated their sweat cortisol concentration to their psychological stress levels. Subjects self-reported their psychological distress using the Kessler Psychological Distress Scale (K10) with 40 min of exercising. K10 scores range from 10 to 50, with scores under 20, 20–24, 25–29, and over 30 correlating with low, mild, moderate, and severe anxiety and stress, respectively.[45] Overall, both men and women qualitatively exhibited a positive correlation between their K10 score and absorbance (cortisol concentration). The resulting sweat cortisol concentration is higher than normal cortisol concentration, which is attributed to the physical stress subjects had to undergo to produce sweat. Normal pituitary response to psychological or physical stress is to release cortisol. Therefore, the greater the stress in aptitude, the higher the basal cortisol concentration in the body. The results show a significant difference in cortisol concentration in apocrine sweat produced by men versus women. As shown in Figure , male participants (n = 6) had an average apocrine cortisol absorbance of 0.136 ± 0.028, whereas female participants (n = 15) had an average of 0.106 ± 0.031. Statistical analysis uses a two-tailed t-test, which reveals a p-value of 0.042 that indicates a significant difference between the two groups in the 95% confidence interval. Likewise, there is a highly significant difference between apocrine and eccrine cortisol concentrations. Apocrine sweat samples have an average of 0.12 ± 0.032, while eccrine samples have an average of 0.072 ± 0.012 (Figure ). A two-tailed paired t-test results in a p-value of 0.000042. Based on our calibration curves, the basal cortisol concentration varies greatly between subjects, with a range of 13.3–443 μg/mL. Further studies are needed with more subjects to confirm our results and to compare them to existing methods, such as ELISA, or more sophisticated instrumental analysis, such as high-performance liquid chromatography–mass spectrometry.
Figure 5

Apocrine and eccrine cortisol concentration of male and female participants: boxplot analysis shows a significant difference (p-value = 0.042) between male and female cortisol levels and a highly significant difference (p-value = 0.0000042) between apocrine and eccrine sweat in both males and females.

Apocrine and eccrine cortisol concentration of male and female participants: boxplot analysis shows a significant difference (p-value = 0.042) between male and female cortisol levels and a highly significant difference (p-value = 0.0000042) between apocrine and eccrine sweat in both males and females. The differences between men and women and apocrine and eccrine could be explained by the pituitary gland’s regulation of cortisol. The main regulatory organ of cortisol is the pituitary gland, which diverges in development when males and females go through puberty.[46] The pituitary gland also has a role in regulating sweat. Adenylate cyclase, a second messenger released by the pituitary gland, promotes different levels of eccrine sweat secretion in men versus women.[47] It is also possible that cortisone and corticosterone, both metabolites of cortisol, occur in higher concentrations in men. However, any interference due to each steroid would be minimal because of its inherently lower concentration in the body and delayed because of the metabolizing process.[48] The exact cause of the difference between groups needs to be further investigated. Overall, this pilot study shows the great potential that such methods have for monitoring and diagnosing cortisol-related illnesses.

Conclusions

All the methods presented in this study can be used as a fast and simple qualitative and quantitative diagnostic tool for the detection of cortisol concentration in biofluids such as saliva and sweat. For qualitative assessment, the pure sulfuric reaction can be used like a pH litmus test and provide a good approximation of cortisol concentration in a short amount of time. For quantitative measurements, the AuNP Porter–Silber or the blue tetrazolium reaction with methanol grants in situ analysis with less concern of cytotoxicity. Such cortisol colorimetric methods have the benefit over electrochemical sensors by not relying on electronic components and can be easily analyzed by the naked eye, which provides intuitive understanding of the cortisol level. These colorimetric analyses are especially beneficial in analytical settings with limited access to state-of-the-art instrumentations. Tests performed on common noninvasive biofluids, such as sweat and saliva, disclosed that no excipients or other biomarkers interfere with these reactions. The pilot study revealed a previously unreported difference in cortisol levels between apocrine and eccrine sweat glands and between men and women (p-value = 0.000042 and 0.042, respectively). Although other steroid hormones, such as testosterone and estrogen, do not affect the reaction, it is unknown whether or not more closely related steroids, such as cortisone or corticosterone, may influence spectrophotometric results. Therefore, research into the selectivity of each method will be further studied. In our future work, we will incorporate colorimetric methods into a single, skin-interfaced microfluidics biosensor to provide both qualitative and quantitative measurements to monitor stress and stress-related illnesses.[31,49] The development of a smartphone application that enables the user’s phone camera in place of a spectrometer and reads pixel values in lieu of absorbance is underway.[50,51] The simplicity of such low-cost methods facilitates the widespread use of such devices of disposable, low-cost wearable devices.

Materials and Methods

For a detailed list of the materials and their manufacturer, please refer to the Supporting Information section of this paper.

Preparation of Cortisol Stock

The cortisol stock solutions were prepared fresh every week. For analytical assessments in general, the solution contained 20 ± 0.1 mg of hydrocortisone dissolved in 10 mL of absolute ethanol (5.5 mM). This solution was primarily used to obtain calibration curves and absorbance spectra of the different methods. To determine cortisol contents in artificial biofluids, the calibration curves were constructed with solutions that contain 4.0 ± 0.1 mg (0.55 mM) and 2.0 ± 0.1 mg (0.28 mM) of hydrocortisone in 20 mL of artificial saliva and artificial sweat, respectively.

Acid-Induced Fluorescence Method

Appropriate amounts of cortisol stock were added directly to concentrated sulfuric acid into quartz cuvettes. The solution was then well mixed with pipettes. Measurements were taken 5 min after mixing.

Porter–Silber Reagent Method with AuNPs

The traditional Porter–Silber reagent was prepared according to the original article’s specifications.[14] Furthermore, we studied various solvent systems while replacing ethanol in the original reagent to improve analytical performances. These alcohol solvents include methanol, isopropanol, butanol, isobutanol, and octanol. These Porter–Silber reagents, which contain 4 mL of concentrated sulfuric acid, 2.5 mL of distilled water, 3.5 mL of alcohol, and 20 mg of phenylhydrazine, were prepared fresh daily. Additional tests were carried out with solutions that contain AuNPs. In this solution, 2.4 mL of distilled water with 100 μL of AuNPs was used instead of 2.5 mL of DI water. To create a calibration curve, appropriate amounts of cortisol stock were added to the Porter–Silber reagents at room temperature (23 °C) for approximately 1 h until color formation. Measurements were taken every 15 min after the addition of cortisol.

Prussian Blue Methods

A modified method was used to form a light-toned Prussian blue (C18Fe7N18).[38] An iron(III) solution was prepared by adding 10 mg of iron(III) chloride to 20 mL of ethanol. Potassium hexacyanoferrate(III) solutions were similarly prepared by adding 13.5 mg of potassium hexacyanoferrate(III) to 20 mL of ethanol. Cortisol stock was first mixed with equal parts of sulfuric acid and iron(III) solution (2 mL each) to oxidize the corticosteroid. After 5 min of mixing, 0.5 mL of potassium hexacyanoferrate solution was added. The solution was then put in a water bath at 70 °C until color formation. Measurements were taken every 10 min after the addition of cortisol.

Blue Tetrazolium Methods

A 1% v/v tetramethylammonium hydroxide solution (N(CH3)4+ OH–) was prepared by diluting 5 mL of the aqueous solution in 45 mL of methanol. A second solution that contains 100 mg of blue tetrazolium dissolved in 50 mL of methanol was made. Equal parts of both solutions were then mixed along with appropriate amounts of cortisol stock at room temperature. Measurements were taken every 10 min after the addition of cortisol.

Spectroscopy Analysis

For each colorimetric method, a concentration curve was constructed by adding appropriate amounts of cortisol stock to the respective reagent. Measurements were taken at predetermined intervals in 1 cm2 quartz cuvettes (VWR) and using a UV–Vis Spectrometer (Cary 60, Agilent Technologies) from 300 to 800 nm. All solutions were measured against a baseline solution, which includes all components of the respective reagent without the addition of cortisol.

Small-n Human Pilot Study

The research protocol was approved by Binghamton University’s Institutional Review Board (MOD00000592). Participants were healthy subjects aged between 19 and 30, of which seven are males and 14 are females. Sweat samples collected from 21 subjects were analyzed using the blue tetrazolium method. Measurements were taken at 10 min at 510 nm wavelength. Subjects completed the Kessler Psychological Distress Scale (K10) prior to exercise. Cortisol concentrations from both armpits and lower back were then compared to the scores obtained through the K10 test. The full sweat study design is outlined in the Supporting Information, Figure S1.
  7 in total

1.  Nano gold-doped molecularly imprinted electrochemical sensor for rapid and ultrasensitive cortisol detection.

Authors:  Sanjida Yeasmin; Bo Wu; Ye Liu; Ahasan Ullah; Li-Jing Cheng
Journal:  Biosens Bioelectron       Date:  2022-03-03       Impact factor: 12.545

2.  Recent progress, challenges, and opportunities for wearable biochemical sensors for sweat analysis.

Authors:  Roozbeh Ghaffari; John A Rogers; Tyler R Ray
Journal:  Sens Actuators B Chem       Date:  2021-01-07       Impact factor: 7.460

3.  All Silica Micro-Fluidic Flow Injection Sensor System for Colorimetric Chemical Sensing.

Authors:  Vedran Budinski; Denis Donlagic
Journal:  Sensors (Basel)       Date:  2021-06-14       Impact factor: 3.576

4.  State of Sweat: Emerging Wearable Systems for Real-Time, Noninvasive Sweat Sensing and Analytics.

Authors:  Roozbeh Ghaffari; Da Som Yang; Joohee Kim; Amer Mansour; John A Wright; Jeffrey B Model; Donald E Wright; John A Rogers; Tyler R Ray
Journal:  ACS Sens       Date:  2021-08-05       Impact factor: 9.618

Review 5.  Enzymatic Methods for Salivary Biomarkers Detection: Overview and Current Challenges.

Authors:  Alonso Ornelas-González; Margarita Ortiz-Martínez; Mirna González-González; Marco Rito-Palomares
Journal:  Molecules       Date:  2021-11-20       Impact factor: 4.411

6.  Sweat and saliva cortisol response to stress and nutrition factors.

Authors:  Paul Pearlmutter; Gia DeRose; Cheyenne Samson; Nicholas Linehan; Yuqiao Cen; Lina Begdache; Daehan Won; Ahyeon Koh
Journal:  Sci Rep       Date:  2020-11-04       Impact factor: 4.379

7.  Development of a Prototype Lateral Flow Immunoassay of Cortisol in Saliva for Daily Monitoring of Stress.

Authors:  Elizaveta Panfilova
Journal:  Biosensors (Basel)       Date:  2021-05-07
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.