Literature DB >> 3228767

Clinical performance of Bjork-Shiley mechanical heart valves: a perspective on outlet strut fractures in the 60 degrees and 70 degrees convexo-concave disc models.

H E Scully1, A Damle, B S Goldman, C Tong, J Azuma, L L Mickleborough, L Schwartz, R J Baird.   

Abstract

The convexo-concave disc model of Bjork-Shiley mechanical heart valve is a significant improvement over the spherical disc model and compares favourably to other valves, including bioprostheses, with respect to hemodynamics, thromboembolism, anticoagulant related hemorrhage and prosthetic endocarditis. A small number of the 60 degrees and 70 degrees welded outlet strut Bjork-Shiley convexo-concave valves are at risk of sudden structural failure. The risk is constant or decreasing with time. Elective explanation is not recommended. The integral outlet monostrut 70 degrees Bjork-Shiley convexo-concave valve demonstrates further improvements in design and durability, and is recommended for use in patients who are appropriate candidates for mechanical cardiac prostheses.

Entities:  

Mesh:

Year:  1988        PMID: 3228767

Source DB:  PubMed          Journal:  Can J Cardiol        ISSN: 0828-282X            Impact factor:   5.223


  2 in total

1.  Publicity about defects in artificial heart valves caused needless panic, doctor contends.

Authors:  D Spurgeon
Journal:  CMAJ       Date:  1991-11-15       Impact factor: 8.262

2.  Discrimination in vitro between the acoustic emissions from Bjork-Shiley convexo-concave valves with and without a broken minor strut.

Authors:  D K Walker; L N Scotten
Journal:  Med Biol Eng Comput       Date:  1991-09       Impact factor: 2.602