Literature DB >> 32267891

Reliability and agreement of a novel portable laser height metre.

Gustav Valentin Blichfeldt Sørensen1,2, Johannes Riis1,2, Mathias Brix Danielsen1,2, Jesper Ryg3,4, Tahir Masud5, Stig Andersen1,2, Martin Gronbech Jorgensen1.   

Abstract

BACKGROUND: Human height is a simple measure with great applicability. Usually, stadiometers are used to measure height accurately. However, these may be impractical to transport and expensive. Therefore, we developed a portable and low-cost laser height metre (LHM).
OBJECTIVE: We aimed to (1) determine intrarater and interrater reliability of our LHM and compare it to a wall-fixed stadiometer, (2) examine its agreement with the same stadiometer, and (3) determine the minimum number of recordings needed to obtain an accurate and reliable height measurement using the LHM.
METHODS: We recruited 32 participants (18+ years)-both men and women. Two raters performed assessments on the same day blinded to each other and their reference standard measurements. We calculated intraclass correlation coefficient (ICC), coefficient of variation (CV), standard error of measurement (SEM), and Bland-Altman plots with limits of agreement (LOA).
RESULTS: For both the LHM and stadiometer, we found ICC values of 0.99-1.00 (95% CI: 0.997-1.000) for both intrarater and interrater reliability. Regarding LHM intrarater reliability, SEM, CV, and LOA were 0.34 cm, 0.16%, and -1.07 to 0.73 cm, respectively. In terms of LHM interrater reliability, SEM, CV, and LOA were 0.27 cm, 0.12%, and -0.32 to 0.84 cm, respectively. As to agreement with stadiometers using one measurement, the mean difference was -0.14 cm and LOA ranged from -0.81 to 0.77 cm.
CONCLUSION: A portable and low-cost LHM, for measuring body height once, showed an excellent reproducibility within and between raters along with an acceptable agreement with a stadiometer thereby representing a suitable alternative.

Entities:  

Year:  2020        PMID: 32267891      PMCID: PMC7141692          DOI: 10.1371/journal.pone.0231449

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


Introduction

Body height is used in many different contexts, e.g. when calculating body mass index [1] or reference intervals for normal lung function [2]. Also, height measures may be used when screening for vertebral fractures in patients with osteoporosis [3] and growth retardation in children [4] together with being used for baseline measurements in clinical studies. In a clinical setting, height is typically measured using a wall-fixed stadiometer, which serves as a costly golden standard [2,5,6]. However, newer studies have investigated alternative ways of measuring height using portable measuring devices with laser distance metres [7-9]. The advantage of these devices is the ability to perform reliable measures in alternating settings. Recently, researchers developed such a measuring device that required manual adjustments of measuring axes, which showed a systematic bias of 0.45 cm when compared to a stadiometer [7]. To optimise this, the authors suggested that future measuring devices using laser distance metres should be fixed in one or more measurement axes. Other researchers have developed such an instrument where the laser distance metre was fixated in two out of three measurement axes, which reduced the bias to 0.07 cm [9]. However, to measure the height using this fixated device, the distance from the skull to the ceiling has to be subtracted from the floor-to-ceiling distance, which is a more time-consuming procedure that potentially can introduce calculation errors. Therefore, we developed a portable laser height metre (LHM) using a laser distance metre fixated in two axes, which can produce height estimates without having to do post-measurement calculations. Thus, we aimed to determine intrarater and interrater reliability of the LHM along with comparing it to a stadiometer, examine its agreement with the stadiometer, and finally to determine the minimum number of recordings needed when performing a height measurement.

Materials and methods

This paper is reported according to the Guidelines for Reporting Reliability and Agreements Studies (GRRAS) [10]. A protocol was registered a priori at Clinicaltrials.gov: NCT04070638.

Participants

We recruited 32 participants (59% women) and performed measurements over two consecutive days, with 15 participants on the first day and 17 on the second day. The only inclusion criterium for the study was a minimum age of 18 years. Participants were students and colleagues at Aalborg University Hospital and Aalborg University recruited through convenience sampling, i.e. a non-consecutive manner, to maximise recruitment efforts during the time of the study. We collected data on gender, age, and height. No statistical sample size calculations were performed, but we decided to aim for the inclusion of 30 participants or above based on previous recommendations [11].

Ethics

Written informed consent was obtained from all participants before participation in the study. The local ethics committee of Region Nordjylland, Denmark, was consulted, and approval of the study was not required for this study according to the Danish Act on the Scientific Ethical Committee System (Act no. 593, section 14, subsection 2). Approval was obtained from the Danish Data Protection Agency (record number 2019–100).

Measuring devices

Laser height metre

A laser distance metre (Bosch Zamo, Robert Bosch GmbH, Gerlingen-Schillerhöhe, Germany) was mounted perpendicularly to the end of a wooden lath (3×4×50 cm) (Fig 1). The other end of the lath was mounted perpendicularly to a T-shaped metal plate thereby fixating the x- and z-axis of the LHM. To adjust the y-axis, a bubble level was mounted on the end of the lath to ensure that the LHM measured vertically down in its z-axis from the participants’ vertex to the floor in front of the participants. The LHM weighed 800 g and was produced for 70 €. The device reports measures in meters to the thousandth’s decimal equivalent to centimetres to the tenth’s decimal.
Fig 1

The laser height metre.

Stadiometer (reference standard)

Wall-fixed stadiometers were used as a reference (Harpenden Stadiometer, Holtain Limited, Crosswell, UK [6]). The stadiometers are calibrated once a week by a biomedical laboratory technician. The device reports measures in centimetres to the tenth’s decimal.

Procedure

Measurements were performed at Aalborg University Hospital, Aalborg, Denmark, from the 26th to 27th of August 2019 in two separate rooms next to each other with stadiometers installed. Two raters were conveniently chosen; a medical doctor and a medical student who both were researchers at the Department of Geriatric Medicine at the same hospital, but had no prior experience or training using either of the devices. They independently, i.e. without communicating with each other, performed height measurements on all participants in the two separate rooms with each measuring device, i.e. the LHM and stadiometer. Independence between raters was ensured by the first author observing both raters’ behaviour simultaneously from outside the rooms. Each rater performed three measures per device. During measurements, participants were asked to stand flat on the floor–a hard surface–without shoes. Furthermore, participants were asked to stand with their heels positioned together and against the wall during both types of measurements. Each rater made sure that participants’ heads were positioned in the Frankfurt plane, defined as the horizontal “plane passing through the upper periphery of the external auricular canals and the lowest point of the left orbit [12], and encouraged them to stand with a straight back against the wall. If participants had piled-up hair, they were asked to smooth this out to minimise any potential overestimation of height. Then for the stadiometer, the measuring plate was pulled down to the skull. Afterwards, the participants were asked to take a deep breath and hold it after which the measurement was performed. For the LHM, the device was placed on top of each participant’s vertex and fixated in two axes by holding the T-shaped piece against the wall (Fig 2). The lath was rotated until the laser distance metre pointed vertically according to the bubble level at the end of the lath. The laser distance metre was activated by clicking on the “START” button. Participants were asked to take a deep breath and hold it, after which the measurement was locked and recorded. Following each measurement with both the stadiometer and LHM, the participant took a step away from the measuring area while the LHM was removed from the wall along with the position of the measuring plate of stadiometer was changed. Afterwards, the participant stepped back again into the measuring area to replicate the measurement procedure. Measurements were recorded in centimetres to tenth’s place since this was considered more clinically relevant than metres to the thousandth’s place.
Fig 2

Demonstration of how to use the laser height metre.

Blinding

The blinding process is illustrated in Fig 3. Thus, all participants were first measured by each rater using the LHM. When this stage was completed, measurements were archived and made inaccessible to blind the raters when performing stadiometer measurements again afterwards. All participants were measured in the same sequential order by both raters. The participants were not informed of their results of the height measurements and therefore blinded. We assumed that each rater was not able to remember each participant’s height when all participants were measured consecutively. After the stadiometer measurements were performed, data were also archived and made inaccessible to raters. After 30 minutes, this entire procedure was repeated once more. Measurements were collected using a data capturing tool [13]. To guide readers, we defined a session as the process where one rater performed three measurements using one device, i.e. either the stadiometer or LHM.
Fig 3

Overview of the measuring sessions.

Statistics

Histograms and QQ-plots were used to check for normal distributions. Age was reported in median and interquartile range due to a non-normal distribution that did not normalise when using logarithmic transformation, see S1 Appendix. Gender was reported in proportions. All heights measures were normally distributed and reported as mean (SD). For agreement measures, the mean of all three measurements for each participant per session was calculated. Afterwards, a group mean was calculated by taking the average of all participant means for each session. Subsequently, the difference between the group means was calculated by subtracting the LHM group mean from the stadiometer group mean for each rater. Furthermore, a two-tailed paired t-test was used for assessing systematic bias using the abovementioned difference. To evaluate whether using the average of three measures would have a similar agreement when using only two or one measurement, this process was also performed with the mean of the two first measurements, and finally exclusively the first measurement. We generated Bland-Altman plots with limits of agreement (LOA) for all the above-mentioned combinations. The same approach was used for intrarater and interrater reliability, where the difference in test-retest and between raters were investigated instead of between devices. LOA were calculated as mean difference ±1.96×SDdifference, where SDdifference is the standard deviation of differences between two measurements on a sample of participants [14]. For the mean difference, 95% confidence intervals were calculated as mean difference ±2.0395×SEbias, where 2.0395 is the t-value for 31 degrees of freedom (n-1) and a significance level of 0.05, and SEbias = SDdifference/√n where n is the sample size. For LOA, 95% confidence intervals were calculated as LOA ± 2.0395×SELOA, where SELOA = 1.71×SEbias [14]. For relative reliability measures, we calculated intraclass correlation coefficients (ICC) with 95% confidence intervals. ICCs for relative intrarater reliability were based on a single measurement, absolute agreement, two-way mixed effect model. For relative interrater reliability, we used a single measurement, absolute agreement, two-way random effect model (ICC 2,1) [11]. ICC estimates were interpreted as poor (<0.5), moderate (0.5–0.75), good (0.75–0.9), and excellent (>0.9) indicators of reliability [11]. For absolute reliability measures, we generated Bland-Altman plots, LOA, standard error of measurements (SEM), and coefficient of variation (CV) between raters on the same device (interrater) and as a test-retest setup (intrarater). SEM was calculated as the square root of the mean squared error term [15]. CV was calculated using the following formula CV(%) = ((Σ(SD/mean))/n)×100 [16]. Both reliability and agreement measures were analysed with a paired samples t-test to check for systematic bias and statistically significant differences, respectively [15]. Heteroscedasticity was inspected for using Bland-Altman plots. To be clinically applicable, we considered a difference in height measures less than 1 cm to be acceptable for both reliability and agreement measures. To avoid calculations based on typing errors, any identified observation on the measures differing more than 20 standard deviations from the remaining two measurements in that session were replaced with the mean of the same two measurements. The observation was replaced with the mean of the remaining two measurements. Findings in the results section were reported conservatively by only displaying the most deviating measures. Thus, for each device, we reported the largest mean difference, lowest ICC, along with the highest CV, SEM and LOA measures found. Statistical analyses were performed using Microsoft Excel for Mac version 16.28 (Microsoft Office, Microsoft Corporation, WA), IBM SPSS Statistics for MacOS version 25.0 (IBM Corporation, Armonk, NY), and STATA for MacOS (StataCorp. 2019. Stata Statistical Software: Release 16. College Station, TX: StataCorp LLC).

Results

Median age (IQR) was 23 (22–26) years. Mean (SD) height of participants was 175 (9) cm. Approximate time spent per participant was one minute for performing three measurements and recording them in the database. In the data set, one single outlier in the stadiometer measurements of 188.4 cm, equal 60 standard deviations from the remaining two measurements, was found. This was replaced with the mean of the two remaining measurements equal to 184.15cm. This has been highlighted in the complete and original data set that is fully available without any restrictions, see S2 Appendix. All analyses were performed on data from the 32 participants and reported in centimetres to the hundredth's place due to the statistical analyses summarising the measures originally reported in centimetres to the tenth’s place.

Intrarater reliability

Table 1 reports findings for intrarater reliability. For both the LHM and stadiometer, systematic bias for intrarater reliability was seen in both devices with -0.21 cm (95% CI: -0.33 cm to -0.07 cm; p = 0.003) and -0.14 cm (95% CI: -0.23 cm to -0.03 cm; p = 0.01), respectively. Both devices showed excellent relative intrarater reliability with ICC values of 0.99 (0.997–1.000). Also, absolute intrarater reliability was acceptable with a SEM, CV, and LOA of 0.34 cm, 0.16% and -1.07 (95% CI: -1.35 cm to -0.79 cm) to 0.73 cm (95% CI: 0.45 cm to 1.01 cm), respectively, for the LHM. This was 0.22 cm, 0.10%, and -0.69 (95% CI: -0.88 cm to -0.50 cm) to 0.53 cm (95% CI: 0.34 cm to 0.72 cm), respectively, for the stadiometer.
Table 1

Intrarater reliability for height measurements on each rater with the laser height metre and stadiometer (n = 32).

DeviceRaterCombination of measurementsTest mean, cmRetest mean, cmMean difference, cm (95% CI)ICC (95% CI)CV, %Lower LOA, cm (95% CI)Upper LOA, cm (95% CI)SEM, cm
Laser height metreRater 1First measurement174.85175.02-0.17 (-0.33; 0.00)*0.99 (0.997;0.999)0.16-1.04 (-1.32; -0.77)0.71 (0.43; 0.99)0.33
Mean of first two measurements174.85175.02-0.17 (-0.30; -0.03)*0.99 (0.998;1.000)0.14-0.90 (-1.14; -0.67)0.57 (0.34; 0.81)0.29
Mean of all three measurements174.86175.04-0.18 (-0.32; -0.04)*0.99 (0.997;1.000)0.15-0.93 (-1.17; -0.69)0.57 (0.34; 0.81)0.30
Rater 2First measurement174.77174.94-0.17 (-0.33; 0.00)*0.99 (0.997;0.999)0.15-1.07 (-1.35; -0.79)0.73 (0.45; 1.01)0.34
Mean of first two measurements174.77174.96-0.19 (-0.33; -0.05)*0.99 (0.997;0.999)0.13-0.96 (-1.20; -0.72)0.57 (0.33; 0.81)0.30
Mean of all three measurements174.76174.97-0.21 (-0.33; -0.07)*0.99 (0.997;1.000)0.12-0.90 (-1.12; -0.68)0.50 (0.28; 0.72)0.29
StadiometerRater 1First measurement174.74174.88-0.14 (-0.23; -0.03)*0.99 (0.999;1.000)0.09-0.67 (-0.84; -0.50)0.41 (0.24; 0.58)0.21
Mean of first two measurements174.79174.90-0.11 (-0.19; -0.02)*1.00 (0.999;1.000)0.08-0.58 (-0.73; -0.43)0.37 (0.22; 0.52)0.18
Mean of all three measurements174.79174.89-0.10 (-0.19; -0.01)*1.00 (0.999;1.000)0.09-0.60 (-0.76; -0.44)0.40 (0.25; 0.56)0.19
Rater 2First measurement174.88174.96-0.08 (-0.19; 0.03)0.99 (0.999;1.000)0.10-0.69 (-0.88; -0.50)0.53 (0.34; 0.72)0.22
Mean of first two measurements174.88174.94-0.06 (-0.14; 0.03)1.00 (0.999;1.000)0.07-0.50 (-0.64; -0.36)0.39 (0.25; 0.53)0.16
Mean of all three measurements174.88174.94-0.06 (-0.13; 0.01)1.00 (0.999;1.000)0.07-0.45 (-0.57; -0.33)0.32 (0.20; 0.44)0.14

* = significant difference in a paired samples t-test (p < 0.05), cm = centimetres, ICC = Intraclass Correlation Coefficient based on a single measurement, absolute agreement two-way mixed effect model, CI = Confidence Interval, CV = Coefficient of Variation, SEM = Standard Error of Measurements, LOA = Limit of Agreement

* = significant difference in a paired samples t-test (p < 0.05), cm = centimetres, ICC = Intraclass Correlation Coefficient based on a single measurement, absolute agreement two-way mixed effect model, CI = Confidence Interval, CV = Coefficient of Variation, SEM = Standard Error of Measurements, LOA = Limit of Agreement

Interrater reliability

Table 2 displays results for interrater reliability. No statistically significant systematic bias was seen in the LHM with a difference of 0.1 cm (95% CI: -0.04 cm to 0.24 cm; p = 0.15) between raters. However, a systematic bias of -0.14 cm (95% CI: -0.22 cm to -0.05 cm; p = 0.0026) was present for the stadiometer. For the LHM, SEM, CV, and LOA were 0.27 cm, 0.12%, and -0.64 (95% CI: -0.88 cm to -0.41 cm) to 0.84 cm (95% CI: 0.61 cm to 1.08 cm), respectively. Regarding the stadiometer, the same measures were 0.20 cm, 0.09%, and -0.63 (95% CI: -0.81 cm to -0.46 cm) to 0.47 cm (95% CI: 0.29 cm to 0.64 cm), respectively. Relative reliability was excellent with ICC values for both devices of 0.99 (95% CI: 0.998 to 1.000). When using the average of several LHM measurements, LOA narrowed and SEM along with CV decreased up to 0.08 cm and 0.03%, respectively, for both intrarater and interrater reliability.
Table 2

Interrater reliability for height measurements on each test session with the laser height metre and stadiometer (n = 32).

DeviceSessionCombination of measurementsRater 1 mean, cmRater 2 mean, cmMean difference, cm (95% CI)ICC (95% CI)CV, %Lower LOA, cm (95% CI)Upper LOA, cm (95% CI)SEM, cm
Laser height metreTestFirst measurement174.85174.770.08 (-0.05; 0.21)0.99 (0.998;1.000)0.11-0.64 (-0.86; -0.41)0.80 (0.57; 1.03)0.26
Mean of first two measurements174.85174.770.08 (-0.05; 0.22)0.99 (0.998;1.000)0.11-0.64 (-0.86; -0.41)0.80 (0.58; 1.03)0.26
Mean of all three measurements174.86174.760.10 (-0.04; 0.24)0.99 (0.998;1.000)0.12-0.64 (-0.88; -0.41)0.84 (0.61; 1.08)0.27
RetestFirst measurement175.02174.940.08 (-0.04; 0.20)0.99 (0.999;1.000)0.10-0.56 (-0.76; -0.36)0.72 (0.52; 0.92)0.23
Mean of first two measurements175.02174.960.06 (-0.03; 0.15)1.00 (0.999;1.000)0.08-0.43 (-0.58; -0.28)0.54 (0.39; 0.70)0.18
Mean of all three measurements175.04174.970.07 (0.00; 0.15)*1.00 (0.999;1.000)0.07-0.32 (-0.45; -0.20)0.47 (0.35; 0.60)0.15
StadiometerTestFirst measurement174.74174.88-0.14 (-0.22; -0.05)*1.00 (0.999;1.000)0.09-0.59 (-0.73; -0.45)0.32 (0.18; 0.46)0.19
Mean of first two measurements174.79174.88-0.09 (-0.16; -0.02)*1.00 (0.999;1.000)0.07-0.45 (-0.57; -0.34)0.28 (0.16; 0.39)0.14
Mean of all three measurements174.79174.88-0.09 (-0.15; -0.02)*1.00 (0.999;1.000)0.07-0.45 (-0.56; -0.34)0.28 (0.16; 0.39)0.14
RetestFirst measurement174.88174.96-0.08 (-0.18; 0.02)0.99 (0.999;1.000)0.08-0.63 (-0.81; -0.46)0.47 (0.29; 0.64)0.20
Mean of first two measurements174.90174.94-0.04 (-0.13; 0.05)0.99 (0.999;1.000)0.07-0.54 (-0.69; -0.38)0.46 (0.30; 0.61)0.18
Mean of all three measurements174.89174.94-0.05 (-0.13; 0.03)1.00 (0.999;1.000)0.07-0.50 (-0.64; -0.36)0.39 (0.25; 0.53)0.16

* = significant difference in a paired samples t-test (p < 0.05), cm = centimetres, ICC = Intraclass Correlation Coefficient based on a single measurement, absolute agreement two-way random effects model, CI = Confidence Interval, CV = Coefficient of Variation, SEM = Standard Error of Measurements, LOA = Limit of Agreement

* = significant difference in a paired samples t-test (p < 0.05), cm = centimetres, ICC = Intraclass Correlation Coefficient based on a single measurement, absolute agreement two-way random effects model, CI = Confidence Interval, CV = Coefficient of Variation, SEM = Standard Error of Measurements, LOA = Limit of Agreement

Agreement

Table 3 displays method comparison results. We found acceptable agreement regardless of the number of measurements used. When only measuring once, the mean difference was -0.14 cm (95% CI: -0.23 cm to -0.06 cm; p = 0.002) and LOA ranged from -0.81 (95% CI: -1.03 cm to -0.59 cm) to 0.77 cm (95% CI: 0.56 cm to 0.98 cm). Also, when using the average of three measurements, the mean difference was -0.15 cm (95% CI: -0.22 cm to -0.08 cm; p < 0.001) and LOA ranged from -0.71 (95% CI: -0.92 cm to -0.51 cm) to 0.76 cm (95% CI: 0.56 cm to 0.97 cm). To aid clinicians in interpreting these results, Table 4 gives examples of how to apply systematic bias, SEM, CV, and ICC in clinical practice. For Bland-Altman plots along with histograms and QQ-plots, see S3 Appendix. In summary, none of the Bland-Altman plots showed signs of heteroscedasticity.
Table 3

Agreement between laser height metre and stadiometer on each rater and test session (n = 32).

SessionRaterCombination of measurementsStadiometer mean, cmLaser height metre mean, cmMean difference, cm (95% CI)Lower LOA, cm(95% CI)Upper LOA, cm(95% CI)
TestRater 1First measurement174.74174.85-0.11 (-0.24; 0.02)-0.81 (-1.03; -0.59)0.60 (0.38; 0.82)
Mean of first two measurements174.79174.85-0.06 (-0.17; 0.04)-0.64 (-0.83; -0.46)0.53 (0.34; 0.71)
Mean of all three measurements174.79174.86-0.07 (-0.19; 0.05)-0.71 (-0.92; -0.51)0.57 (0.37; 0.78)
Rater 2First measurement174.88174.770.11 (-0.01; 0.23)-0.55 (-0.76; -0.34)0.77 (0.56; 0.98)
Mean of first two measurements174.88174.770.11 (-0.01; 0.23)-0.54 (-0.54; -0.33)0.76 (0.56; 0.97)
Mean of all three measurements174.88174.760.11 (-0.01; 0.23)-0.53 (-0.74; -0.33)0.76 (0.56; 0.97)
RetestRater 1First measurement174.88175.02-0.14 (-0.23; -0.06)*-0.60 (-0.75; -0.46)0.32 (0.18; 0.47)
Mean of first two measurements174.90175.02-0.12 (-0.20; -0.04)*-0.54 (-0.68; -0.41)0.30 (0.17; 0.43)
Mean of all three measurements174.89175.04-0.15 (-0.22; -0.08)*-0.55 (-0.68; -0.42)0.25 (0.12; 0.38)
Rater 2First measurement174.96174.940.02 (-0.08; 0.12)-0.54 (-0.71; -0.36)0.58 (0.40; 0.75)
Mean of first two measurements174.94174.96-0.02 (-0.11; 0.06)-0.47 (-0.61; -0.33)0.43 (0.28; 0.57)
Mean of all three measurements174.94174.97-0.02 (-0.10; 0.05)-0.44 (-0.57; -0.31)0.39 (0.26; 0.52)

* = significant difference in a paired samples t-test (p < 0.05), cm = centimetres, LOA = Limits of Agreement, CI = Confidence Interval

Table 4

How should clinicians interpret our results?.

Parameter and estimateExamples and explanations
Systematic biasExample: 0.2 cmIntrarater: If a subject is measured two times by the same rater over a period of time, a change of ≤ 0.2 cm can be due to bias and can be adjusted for.Interrater: If a subject is measured one time by two raters, a difference of ≤ 0.2 cm can be due to bias and can be adjusted for.
Limits of agreement (LOA)Intrarater and interrater reliabilityExample: -0.32 to 0.84 cmMethod comparisonExample: -0.81 to 0.77 cmIntrarater and interrater reliability: If we measure height once, we can expect that the next value will lie within the LOA with 95% certainty. Thus, for a LOA of -0.32 to 0.84 cm, if one rater measures a participant with the height 180 cm, we would assume the height found by the second rater, or the same rater in a retest session, will be somewhere between 179.68 cm (180–0.32 cm) to 180.84 cm (180+0.84 cm) with 95% certainty.Method comparison: If we measure the height of a participant with a new method and compare it to a reference standard, we can expect that 95% of the differences between the methods lies from -0.81 to 0.77 cm. Thus, a participant with a height of 180 cm measured on the reference standard would lie between 179.19cm (180–0.81 cm) and 180.77cm (180+0.77 cm) on the new method.
Standard Error of Measurement (SEM)Example: 0.30 cmA SEM of 0.30 cm means that for a given height measure, the true height of the individual will lie within ± 1.96×0.3 cm = ±0.588 cm of the obtained height from the measurement. Thus, for a person with a height of 180 cm measured once, a second measurement will have to be above 180.588 cm (180+0.588cm) or below 179.412 cm (180–0.588 cm) to be a real change in height.
Coefficient of variation (CV)CV is a unitless indicator on how much the measuring device varies from the first to the second measurement. Thus, a lower CV equals a more reliable method. Since the indicator is unitless, comparison can be made between measuring devices using different measuring scales, e.g. centimetres and inches.
Intraclass correlation coefficient (ICC)A high ICC value means that there is a low random and systematic measurement error, and thus high relative reliability, when measuring on the same subject several times (intrarater) or by two raters (interrater).
* = significant difference in a paired samples t-test (p < 0.05), cm = centimetres, LOA = Limits of Agreement, CI = Confidence Interval

Discussion

This study aimed to determine intrarater and interrater reliability of the LHM along with comparing it to a stadiometer, examine its agreement with the stadiometer, and finally to determine the minimum number of recordings needed when performing a height measurement. We found excellent relative intrarater and interrater reliability for LHM along with an acceptable systematic bias, absolute reliability, and agreement. Furthermore, measuring the height only once was found adequate, as outcomes did not change substantially with more measures used. We consider the LHM a suitable alternative to stadiometers as differences are very small between devices. Also, the LHM has other advantages in terms of price and portability.

Intrarater and interrater reliability

Three previous studies have examined height measuring devices with laser distance metres [7-9], all of which investigated intrarater reliability. Only one study [7] reported an interrater reliability ICC value. This was 0.991 (0.988–0.994) which was lower compared to our findings. The other two studies reported findings on intrarater reliability using technical error of measurement [8,9]. This measure is not included in the reporting recommended according to GRASS guidelines and we did not include this [10]. Furthermore, none of the studies provided LOA, CV, or SEM values. This hampers comparison with our findings. Only one study investigated interrater reliability on LHMs [8] and they reported only technical error of measurement. In our study, relative reliability was excellent with ICC values similar to the reference standard for both intrarater and interrater reliability. Also, the absolute reliability was not as low as for the stadiometer, but we consider it clinically acceptable since SEM differed less than 1 cm between raters and sessions. Differences between the LHM and stadiometer may be due to the former being slightly more unstable in terms of fixation compared to the latter. Also, the y-axis of the LHM was adjusted using a bubble level which required subjective judgements when interpreted. This could have been optimised using a digital level. Our findings are in line with prior studies with laser distance metres with mean differences of 0.07 to 0.45 cm, compared to our findings of 0.15 cm, when using the average of three measurements [7-9]. Furthermore, LOA ranged from -0.49 to 0.63 cm [9] to -3.3 to 2.8 cm [8] compared with our findings ranging from -0.71 to 0.76 cm. Reason for discrepancies may be due to inclusion of children by others [7,8], different reference standards used for comparison [7-9], and the lack of blinding [7-9]. Interestingly, the laser measuring device with fixated axes [9] differed less from the reference standard and produced clinically acceptable LOA compared to the two devices without fixation [7,8]. This emphasises the importance of stabilising the measuring device. We thus suggest future studies to investigate ways to enhance the fixation of the x-, z-, and y-axis of the LHM more accurately without compromising portability of the tool. This could be done by adding an adhesive material to the T-shaped base to enhance fixation to the wall.

Portability, practicality, and price

Our LHM shows excellent reliability and acceptable agreement. Additional advantages include the device being portable in contrast to the wall-fixed stadiometer. This may be an advantage to field studies in alternating settings, e.g. participants’ own homes. Also, it is possible to obtain a height measure within a few seconds which may ease implementation of LHM. Time spent was not reported by the previous studies [7-9]. This would have been relevant to compare since time spent may have differed between devices. For example, to generate a height measure in one of the studies, the distance from the skull to the ceiling had to be subtracted from the floor-to-ceiling distance [9]. Furthermore, this procedure had to be performed three times to finally obtain the average height of the participant. Thus, the feasibility in clinical practice and research of previous LHMs remains to be settled. Besides, our LHM is only a prototype. However, the production cost for the prototype was low at around 70 € compared to the sales price of a wall-fixed stadiometer of 1,269 € [6]. This may support the implementation of LHMs. Finally, commercialised portable stadiometers of similar price ranges as the LHM are available [17]. However, during our literature search in MEDLINE, we did not find other validation or reliability studies on portable stadiometers not already mentioned in this paper. Thus, even though alternative products are available, the scientific transparency may seem limited on these devices thereby hindering device comparison. Thus, we consider our LHM a portable, practical, and low-cost solution.

Strengths and limitations

First, we examined both reproducibility and agreement along with reporting our study according to guidelines [10]. Second, we used a blinded setup with both raters blinded to each other and their measurement on the reference standard. We cannot rule out that a Hawthorne effect affected the raters, i.e. they may have tried to perfect their measurements due to an awareness of being observed since both knew their ratings would be compared to each other [18]. However, the time spent of approximately one minute for performing and recording three measurements may have hindered the raters from accomplishing this. Thus, we believe such an effect to be limited. Future studies could overcome this by blinding the raters to the display on the measuring devices, after which the reading may be seen by a second person who records it in a database. Also, this may reduce the risk of typing errors. As mentioned earlier, a single observation had a difference of 4 cm from the other two observations within that session. When correcting this, the reliability and agreement improved substantially for the stadiometer, but not the LHM. Thus, even though correction of the data set was performed, we consider this as not having influenced our results of the LHM. Third, we did not perform sample size calculations. However, in retrospect, this is less relevant since the ability to detect a clinically significant change, i.e. less than 1 cm, was seen. Fourth, given the participants’ age and height ranges, the findings of this study may have limited generalisability to populations outside of these ranges, e.g. a paediatric population. Future studies could overcome this by including a larger and more diverse sample of participants. Fifth, all participants were measured in the same sequential order by both raters, rather than a randomised order, thereby potentially introducing a risk of systematic bias in the measurements. However, we found a clinically irrelevant difference between devices of 0.14 centimetres for which reason the participants or raters had to systematically adjust their height or measuring devices with this length for each session. Even though in theory, this may be possible, raters and participants were blinded to prior measurements thereby compromising the opportunity to systematically adjust these measures. Thus, we consider this potential bias as only being able to influence our results to a minor degree. A final limitation is the need for a vertical wall for obtaining similar results found in this study. This could be overcome by mounting a second bubble level along the side of the lath of the LHM.

Conclusion

Our findings combined with prior studies on height measuring devices with laser distance metres have shown great potential for accurate and reliable measures of height. Previous studies used technologies with different practical and methodological limitations. This study improves on these limitations by having developed a portable, quick and low-cost measure for human height that can provide reliable and accurate readings using only one measurement with a performance that compared to a stadiometer. The ability of our device, and others, to fixate two out of three measuring axes may have improved the reliability and agreement. This suggests a need for further research in ways to stabilise the devices in all three measuring axes.

Histograms and QQ-plots to check for normality.

(PDF) Click here for additional data file.

Complete and original data set.

(XLSX) Click here for additional data file.

Bland-Altman plots for intrarater and interrater reliability along with method comparison.

(PDF) Click here for additional data file. 26 Feb 2020 PONE-D-20-02226 Intrarater and interrater reliability and agreement of a novel portable laser height metre PLOS ONE Dear Dr. Sørensen, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. We would appreciate receiving your revised manuscript by Apr 11 2020 11:59PM. When you are ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. To enhance the reproducibility of your results, we recommend that if applicable you deposit your laboratory protocols in protocols.io, where a protocol can be assigned its own identifier (DOI) such that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols Please include the following items when submitting your revised manuscript: A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). This letter should be uploaded as separate file and labeled 'Response to Reviewers'. A marked-up copy of your manuscript that highlights changes made to the original version. This file should be uploaded as separate file and labeled 'Revised Manuscript with Track Changes'. An unmarked version of your revised paper without tracked changes. This file should be uploaded as separate file and labeled 'Manuscript'. Please note while forming your response, if your article is accepted, you may have the opportunity to make the peer review history publicly available. The record will include editor decision letters (with reviews) and your responses to reviewer comments. If eligible, we will contact you to opt in or out. We look forward to receiving your revised manuscript. Kind regards, Jason S Ng, OD, PhD Academic Editor PLOS ONE Additional Editor Comments (if provided): Both reviewers found the study useful and interesting. The reviewers did not find significant faults with the study and have found minor revisions that should be investigated and responses given. The study is overall well planned with important issues being addressed a priori and the manuscript is well written with good attention to detail as well as very useful tables and figures. Further comments include: 1. One important limitation of this study is that all subjects were young adults and with a limited age range (22-26) and a limited height range (175cm/SD = 9cm). Thus, the generalizability to a pediatric population or otherwise is lacking. This should be addressed. Given how seemingly simple the data collection was, the sample size could have been much larger and a more diverse sample of subjects recruited. 2. Portable stadiometers of similar price are actually available, though they are not quite as portable as the LHM – however, some of these do not even require a wall: Hopkins road rod stadiometer. The authors should further address that portable/affordable stadiometers do exist and comment on the place of the LHM with respect to this idea. 3. The resolution of the LHM is not clear. Figure 1 shows the meter reading to the thousandths place. All reported data is to the hundredth. Please discuss. 4. It is stated that ‘all participants were first measured…using the LHM” by both raters. Were these participants presented to both raters in the same sequential order each time? Or was there sequence of data collection randomized? This is not addressed and could introduce other effects of bias. 5. Eliminating data based on outliers (and replacing it with better data), while it is stated in the discussion that it likely had not effect, in order to ‘avoid calculations based on typing errors’ does not seem like good science and can potentially be a source of bias. Data on the repeatability of LHM device on known measured sources would be helpful here. Data entry into the REDCap system should have been simple enough and gives readers a sense of doubt if such simple data cannot be collected and recorded perfectly the first time. 6. The data cited in the text from table 1, does not seem perfectly accurate. For example, the -0.21 and -0.14 are from different categories so this is confusing. Did the authors select the largest mean difference per device? But isn’t this for intrarater reliability and not agreement between devices (Table 3). Further, the data discussing on lines 188-190 have similar questions. SEM seems to align with 0.15% and not 0.16%. Again the reason why those particular data points are chosen is not clear to the reader. The data cited in the text on lines 192-193 from Table 2 also suffers from this lack of clarity on why/how that particular data point was chosen among others to make the point. The data cited on line 196 does not align with the cited CV. The same issue arises from data cited on lines 202-205 and table 3. Journal Requirements: When submitting your revision, we need you to address these additional requirements: 1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at http://www.plosone.org/attachments/PLOSOne_formatting_sample_main_body.pdf and http://www.plosone.org/attachments/PLOSOne_formatting_sample_title_authors_affiliations.pdf 2. Please include a caption for Figure 1. [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes Reviewer #2: Yes ********** 2. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: Yes Reviewer #2: Yes ********** 3. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes Reviewer #2: Yes ********** 4. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #2: Yes ********** 5. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: Overall evaluation This is an interesting article describing the development of a laser height meter and tested the validity and reliability of the device. Usage of the device would enable accurate anthropometric measurement of the height. Authors may consider some minor changes as listed below. Title Authors may think making the title concise by omitting “intrarater and interarter”. Abstract Objective – in the second aim, the term “same stadiometer” is not clear. You may please revise it. Main Text Result section may be divided into different paragraph for a better presentation. Conclusion may be paraphrased for better clarity. Reference Avoid old references wherever possible. Reviewer #2: Dear Authors, Thank you for your contribution to this field of study. Your article is well written and your data supports your conclusions. I am suggesting a minor edit: Line 180 - Please edit "(9 cm)" it is unclear what this is associated with. ********** 6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: Yes: Ashley Bauman [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files to be viewed.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email us at figures@plos.org. Please note that Supporting Information files do not need this step. 16 Mar 2020 Response to reviewers: PONE-D-20-02226 Additional Editor Comments: 1. One important limitation of this study is that all subjects were young adults and with a limited age range (22-26) and a limited height range (175cm/SD = 9cm). Thus, the generalizability to a pediatric population or otherwise is lacking. This should be addressed. Given how seemingly simple the data collection was, the sample size could have been much larger, and a more diverse sample of subjects recruited. a. Answer: We acknowledge this, and it has now been emphasised in the “Strengths and limitations” section (line 300). It now specifically states: “Fourth, given the participants’ age and height ranges, the findings of this study may have limited generalisability to populations outside of these ranges, e.g. a paediatric population. Future studies could overcome this by including a larger and more diverse sample of participants.”. 2. Portable stadiometers of similar price are actually available, though they are not quite as portable as the LHM – however, some of these do not even require a wall: Hopkins road rod stadiometer. The authors should further address that portable/affordable stadiometers do exist and comment on the place of the LHM with respect to this idea. a. Answer: We agree that portable stadiometers of similar price are available. However, during our literature search on scientifically validated instruments, we did not find further studies to support this point. Thus, the products are available but the validity of these has not been reported, and comparison between devices may not be possible. This information has been added to the “Portability, practicality, and price” section (line 279) which now specifically states: “Finally, commercialised portable stadiometers of similar price ranges as the LHM are available [17]. However, during our literature search in MEDLINE, we did not find other validation or reliability studies on portable stadiometers not already mentioned in this paper. Thus, even though alternative products are available, the scientific transparency may seem limited on these devices thereby hindering device comparison.”. 3. The resolution of the LHM is not clear. Figure 1 shows the meter reading to the thousandths place. All reported data is to the hundredth. Please discuss. a. Answer: We apologise if this was unclear. The LHM reports measures in metres to the thousandths place, which is displayed in Figure 1, but not mentioned in the manuscript. In the S3 Appendix, the complete data set is reported in centimetres to the tenths place which was judged as more clinically relevant than meters to the thousandth's place. However, all tables report measures to the hundredth's place due to the statistical analyses summarising the measures. This has been added to the manuscript: i. First, the “Laser height metre” section (line 95) now reads: “The device reports measures in meters to the thousandth’s decimal equivalent to centimetres to the tenth’s decimal”. ii. Second, the “Stadiometer (reference standard)” section (line 100) now reads: “The device reports measures in centimetres to the tenth’s decimal.”. iii. Third, the “Procedure” section (line 127) now reads: “Measurements were recorded in centimetres to tenth’s place since this was considered more clinically relevant than metres to the thousandth’s place.”. iv. Fourth, the “Results” section (line 193) now reads: "All analyses were performed on data from the 32 participants and reported in centimetres to the hundredth's place due to the statistical analyses summarising the measures originally reported in centimetres to the tenth’s place.”. 4. It is stated that ‘all participants were first measured…using the LHM” by both raters. Were these participants presented to both raters in the same sequential order each time? Or was there sequence of data collection randomized? This is not addressed and could introduce other effects of bias. a. Answer: We acknowledge this, and we cannot completely rule out a potential risk of systematic bias since participants were presented to both raters in the same sequential order each time. However, we consider this as insignificant due to the small difference found between measuring devices and the blinding of both participants and raters towards the measurements. This has been added to the manuscript: i. First, the “Blinding” section (line 133) now reads: “All participants were measured in the same sequential order by both raters. The participants were not informed of their results of the height measurements and therefore blinded.”. ii. Second, the “Discussion” section (line 303) now reads: “Fifth, all participants were measured in the same sequential order by both raters, rather than a randomised order, thereby potentially introducing a risk of systematic bias in the measurements. However, we found a clinically irrelevant difference between devices of 0.14 centimetres for which reason the participants or raters had to systematically adjust their height or measuring devices with this length for each session. Even though in theory, this may be possible, raters and participants were blinded to prior measurements thereby compromising the opportunity to systematically adjust the measures. Thus, we consider this potential bias as only being able to influence our results to a minor degree.”. 5. Eliminating data based on outliers (and replacing it with better data), while it is stated in the discussion that it likely had not effect, in order to ‘avoid calculations based on typing errors’ does not seem like good science and can potentially be a source of bias. Data on the repeatability of LHM device on known measured sources would be helpful here. Data entry into the REDCap system should have been simple enough and gives readers a sense of doubt if such simple data cannot be collected and recorded perfectly the first time. a. Answer: We agree that eliminating data based on outliers can be suboptimal in terms of good scientific practice and may be a source of bias. However, not being transparent about such a post-measurement correction may be even worse. Thus, we further highlighted the change in “Red” in the appendix with the original data set to increase transparency. We still believe this was a data entry error. Even though REDCap can be set up using a simple design, we decided not to put any minimum or maximum boundaries in the system when entering data, in order to detect any actual differences on several, i.e. in this case - more than one, measurements between devices. Finally, the repeatability results on the LHM remained the same, as the error only pertained to a single data entry on the stadiometer. The “Results” section (line 189) now reads: “In the data set, one single outlier in the stadiometer measurements of 188.4 cm, equal 60 standard deviations from the remaining two measurements, was found. This was replaced with the mean of the two remaining measurements equal to 184.15cm. This has been highlighted in the complete and original data set that is fully available without any restrictions, see S2 Appendix.” 6. The data cited in the text from table 1, does not seem perfectly accurate. For example, the -0.21 and -0.14 are from different categories so this is confusing. Did the authors select the largest mean difference per device? But isn’t this for intrarater reliability and not agreement between devices (Table 3). a. Answer: We apologise if this was unclear. As mentioned in the “Statistics” section (line 178), “findings in the results section were reported conservatively by only displaying the most deviating measures”. Instead of reporting all 36 mean differences for intra- and interrater reliability along with agreement measures, we chose to report only the largest mean difference per device in the manuscript. Also, to increase transparency, we reported all mean differences in the tables. The same approach was used for the SEM, CV, LOA and ICC measures. Thus, for the systematic bias in intrarater reliability measures (Table 1), mean differences ranged between -0.17 cm to -0.21 cm for the LHM and -0.06 cm to -0.14 cm for the stadiometer. Therefore, the most conservative measures reported in the manuscript (line 198) in this example would be -0.21 cm for the LHM and -0.14 cm for the stadiometer. For agreement between devices (Table 3), measures ranged between 0.02 cm to -0.14 cm when performing one measurement only. Thus, a difference of -0.14 cm was found to be the most conservative measure and was therefore reported in the manuscript (line 220). This has been clarified in the “Statistics” section (lines 178-180) which now reads: “Findings in the results section were reported conservatively by only displaying the most deviating measures. Thus, for each device, we reported the largest mean difference, lowest ICC, along with the highest CV, SEM and LOA measures found.” 7. Further, the data discussing on lines 188-190 have similar questions. SEM seems to align with 0.15% and not 0.16%. Again, the reason why those particular data points are chosen is not clear to the reader. a. Answer: We apologise if this was unclear. Line 200-202 (previously 188-190) reads: “Also, absolute intrarater reliability was acceptable with a SEM, CV, and LOA of 0.34 cm, 0.16% and -1.07 (95% CI: -1.35 cm to -0.79 cm) to 0.73 cm (95% CI: 0.45 cm to 1.01 cm), respectively, for the LHM“. From Table 1 it can be seen that the CV for the LHM ranged between 0.12% to 0.16%. The SEM ranged between 0.29 cm to 0.34 cm for which reason. Therefore, we chose to report the latter CV and SEM due to the same reason as mentioned in Item 6. 8. The data cited in the text on lines 192-193 from Table 2 also suffers from this lack of clarity on why/how that particular data point was chosen among others to make the point. The data cited in line 196 does not align with the cited CV. The same issue arises from data cited on lines 202-205 and table 3. a. Answer: We apologise if this was unclear. Lines 207-208 (previously lines 192-193) read: “No statistically significant systematic bias was seen in the LHM with a difference of 0.1 cm (95% CI: -0.04 cm to 0.24 cm; p = 0.15) between raters.”. Line 211 (previously line 196) reads: “Regarding the stadiometer, the same measures were 0.20 cm, 0.09%, and -0.63 (95% CI: -0.81 cm to -0.46 cm) to 0.47 cm (95% CI: 0.29 cm to 0.64 cm), respectively.”. Lines 220-222 (previously line 202-205) read: “When only measuring once, the mean difference was -0.14 cm (95% CI: -0.23 cm to -0.06 cm; p = 0.002) and LOA ranged from -0.81 (95% CI: -1.03 cm to -0.59 cm) to 0.77 cm (95% CI: 0.56 cm to 0.98 cm)”. These values correspond to the most conservative results displayed in Table 2 and Table 3 within each statistical measure, i.e. mean difference, ICC, CV, Lower LOA, Upper LOA, and SEM. The reasoning for this choice of reporting has been given in Item 6. Reviewers’ comments: 1. Title: Authors may think of making the title concise by omitting “intrarater and interrater”. a. Answer: We appreciate this and have now changed the title to “Reliability and agreement of a novel portable laser height metre”. 2. Main Text: Result section may be divided into different paragraph for a better presentation. a. Answer: We acknowledge this and have now made a division in the results section that now includes the following subheadings: “Intrarater reliability”, “Interrater reliability”, and “Agreement”. 3. Conclusion may be paraphrased for better clarity. a. Answer: We recognise this and have now paraphrased the conclusion section (line 315) which now reads: “Our findings combined with prior studies on height measuring devices with laser distance metres have shown great potential for accurate and reliable measures of height. Previous studies used technologies with different practical and methodological limitations. This study improves on these limitations by having developed a portable, quick and low-cost measure for human height that can provide reliable and accurate readings using only one measurement with a performance that compared to a stadiometer. The ability of our device, and others, to fixate two out of three measuring axes may have improved the reliability and agreement. This suggests a need for further research in ways to stabilise the devices in all three measuring axes.” 4. Reference: Avoid old references wherever possible. a. Answer: We agree that references should be based on the latest evidence. However, we find our references essential even though their publication date seems older. 5. Line 180 – Please edit “(9 cm)” it is unclear what this is associated with. a. Answer: We apologise if this was unclear. Line 187 (previously line 180) reads: “Mean (SD) height of participants was 175 cm. (9 cm).”. The “(9 cm)” is the standard deviation on the mean height of the participants. This has been edited and the sentence now reads: “Mean (SD) height of participants was 175 (9) cm”. This was reported according to the “Statistical Analyses and Methods in the Published Literature (SAMPL) Guidelines” recommended by the PLOS ONE submission guidelines. Submitted filename: Response to reviewers.docx Click here for additional data file. 25 Mar 2020 Reliability and agreement of a novel portable laser height metre PONE-D-20-02226R1 Dear Dr. Sørensen, We are pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it complies with all outstanding technical requirements. Within one week, you will receive an e-mail containing information on the amendments required prior to publication. When all required modifications have been addressed, you will receive a formal acceptance letter and your manuscript will proceed to our production department and be scheduled for publication. Shortly after the formal acceptance letter is sent, an invoice for payment will follow. To ensure an efficient production and billing process, please log into Editorial Manager at https://www.editorialmanager.com/pone/, click the "Update My Information" link at the top of the page, and update your user information. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org. If your institution or institutions have a press office, please notify them about your upcoming paper to enable them to help maximize its impact. If they will be preparing press materials for this manuscript, you must inform our press team as soon as possible and no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org. With kind regards, Jason S Ng, OD, PhD Academic Editor PLOS ONE Additional Editor Comments: The authors have addressed the editors and authors comments adequately. 27 Mar 2020 PONE-D-20-02226R1 Reliability and agreement of a novel portable laser height metre Dear Dr. Sørensen: I am pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department. If your institution or institutions have a press office, please notify them about your upcoming paper at this point, to enable them to help maximize its impact. If they will be preparing press materials for this manuscript, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org. For any other questions or concerns, please email plosone@plos.org. Thank you for submitting your work to PLOS ONE. With kind regards, PLOS ONE Editorial Office Staff on behalf of Dr. Jason S Ng Academic Editor PLOS ONE
  14 in total

Review 1.  Measuring agreement in method comparison studies.

Authors:  J M Bland; D G Altman
Journal:  Stat Methods Med Res       Date:  1999-06       Impact factor: 3.021

2.  The "Hawthorne effect"--what did the original Hawthorne studies actually show?

Authors:  G Wickström; T Bendix
Journal:  Scand J Work Environ Health       Date:  2000-08       Impact factor: 5.024

3.  Correct measurement of height is important when assessing lung function values.

Authors:  Sidsel Bendixen Holm Renstrøm; Christian Skjoldvang Andersen; Camilla Helene Borup Pedersen; Flemming F Madsen
Journal:  Dan Med J       Date:  2012-02       Impact factor: 1.240

4.  Body mass index as a measure of body fatness: age- and sex-specific prediction formulas.

Authors:  P Deurenberg; J A Weststrate; J C Seidell
Journal:  Br J Nutr       Date:  1991-03       Impact factor: 3.718

5.  Research electronic data capture (REDCap)--a metadata-driven methodology and workflow process for providing translational research informatics support.

Authors:  Paul A Harris; Robert Taylor; Robert Thielke; Jonathon Payne; Nathaniel Gonzalez; Jose G Conde
Journal:  J Biomed Inform       Date:  2008-09-30       Impact factor: 6.317

Review 6.  Statistical methods for assessing measurement error (reliability) in variables relevant to sports medicine.

Authors:  G Atkinson; A M Nevill
Journal:  Sports Med       Date:  1998-10       Impact factor: 11.136

7.  A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability Research.

Authors:  Terry K Koo; Mae Y Li
Journal:  J Chiropr Med       Date:  2016-03-31

Review 8.  Anthropometry as a tool for measuring malnutrition: impact of the new WHO growth standards and reference.

Authors:  M B Duggan
Journal:  Ann Trop Paediatr       Date:  2010

9.  The reliability of height measurement (the Wessex Growth Study).

Authors:  L D Voss; B J Bailey; K Cumming; T J Wilkin; P R Betts
Journal:  Arch Dis Child       Date:  1990-12       Impact factor: 3.791

10.  Examining the utility of a laser device for measuring height in free-living adults and children.

Authors:  Sandra N Mayol-Kreiser; Vanessa M Garcia-Turner; Carol S Johnston
Journal:  Nutr J       Date:  2015-09-08       Impact factor: 3.271

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.